Skip to main content

ORIGINAL RESEARCH article

Front. Physiol., 03 June 2021
Sec. Exercise Physiology

Comparison of Two Metabolic Simulators Used for Gas Exchange Verification in Cardiopulmonary Exercise Test Carts

\r\nTjeu SourenTjeu Souren1Edward Rose*Edward Rose2*Herman GroepenhoffHerman Groepenhoff3
  • 1Independent Consultant, Utrecht, Netherlands
  • 2Independent Consultant, Novi, MI, United States
  • 3Vyaire Medical, Mettawa, IL, United States

Introduction: Metabolic simulators (MS) produce simulated human breaths for the purpose of verification of cardiopulmonary exercise test (CPET) equipment. MS should produce consistent identical breaths with known CO2 and O2 gas concentrations over a range of breath rates and tidal volumes. Reliability of a CPET metabolic cart depends on ongoing quality control and maintenance of the device, including intermittent verification with a MS. We compared two MS devices against two standard CPET systems.

Methods: The Vacumed 17056 (Vacumetrics, Ventura, CA) and Relitech (Relitech Systems BV, Nijkerk, The Netherlands) were used with two standard metabolic carts (Vyntus CPX and Vyntus ONE, both Vyaire Medical, Mettawa, IL, United States). Tidal volume (VT) was set at 2 and 3 L and breathing frequency ranged from 20 to 80 breaths per minute for each MS. At each set point, we measured three sets of 40 breaths. Primary outcome parameters collected were VT, oxygen consumption (v.O2), carbon dioxide production (v.CO2), and respiratory exchange ratio (RER).

Results: VT, RER, v.O2, and v.CO2 results as obtained from both MS were all within the limits of acceptability, at both tidal volume settings, and all ventilatory rates. No significant trends were identified for either MS device. The Relitech MS produced tidal volumes that were closer to the target VT for both CPET carts at both VT and all rates, but the results of both MS were within acceptable ranges.

Conclusion: Verification of CPET equipment using either the VM or RT metabolic simulator, producing highly accurate and predictable simulated breaths of known composition, enabling CPET laboratory managers to rely on subject test data obtained during cardiopulmonary exercise testing.

Introduction

Cardiopulmonary exercise testing (CPET) is increasingly used for diagnosing complex medical conditions covering a broad spectrum of conditions from cardiopulmonary diseases to metabolic and even mitochondrial disorders (Muraresku et al., 2018). CPET is helpful for identifying the presence and severity of heart failure, atherosclerotic heart disease, peripheral vascular disease, pulmonary hypertension, deconditioning, restrictive, or obstructive pulmonary disease, alveolar diffusion disorders, and cardiomyopathy (American Thoracic Society, 2003). CPET is also widely used for tracking progress of performance testing, rehabilitation and training, response to treatment, risk assessment for major surgery, and eligibility for transplant (Levett et al., 2018). Results from CPET can guide nutrition and parenteral replacement therapy (West et al., 2017).

The data from CPET provide a highly accurate evaluation of linked physiological systems—ventilation, pulmonary gas exchange, circulation, and metabolic processes—against a known workload (American Thoracic Society, 2003). The ability to identify subtle changes in metabolic rate and metabolic efficiency at a tissue, cellular, or even organelle level requires highly sensitive and accurate equipment producing reliable data. To ensure reliability of CPET equipment, quality control should be performed on a routine basis. Society guidelines from the American College of Chest Physicians (ACCP), American Thoracic Society (ATS), and European Respiratory Society (ERS) as well as the manufacturers of the CPET metabolic carts recommend specific calibration procedures, combined with periodic external maintenance, calibration, and verification against known standards (American Thoracic Society, 2003). CPET laboratories must have confidence that the external systems used for calibration or verification are certified, accurate, and reliable “gold standards.” To date, research often compares CPET machines against a metabolic simulator (Casaburi et al., 1997; Porszasz et al., 2007; Beijst et al., 2013; Van Iterson et al., 2016; Levett et al., 2018; Radtke et al., 2019), although evidence that the metabolic simulators themselves are accurate and reproducible has been lacking (American Thoracic Society, 2003).

Automated breath and metabolism simulators, used for the validation of CPET systems, have been used for over 30 years. Simulators provide reproducible simulations of human breathing at various breathing frequencies and tidal breath volumes. Ideally, the respiratory and metabolic variables of tidal volume (VT), respiratory rate (RR), minute ventilation (v.E), oxygen consumption (v.O2), and carbon dioxide production (v.CO2) generated by a metabolic simulator should be indistinguishable from the patterns produced by humans. The simulator produces and controls those variables with high precision and consistency so that every breath, at every breath rate, is identical. Consequently, one can test any metabolic cart with high precision, reproducibility, and accuracy. The validity of VT, respiratory exchange ratio (RER), v.O2, and v.CO2 produced by a simulator is crucial (Casaburi et al., 1997; Gore et al., 1997).

The respiratory exchange ratio (RER) is one metabolic parameter reported by CPET carts. RER is defined as the ratio between the amount of exhaled v.CO2 from metabolism and v.O2 uptake, resulting in a dimensionless value that correlates with the metabolic rate (respiratory quotient) and the fuel being used by the subject (fats vs. carbohydrates). If the CPET laboratory staff follows the manufacturer’s recommended onboard calibration procedures and external validation, it is assumed that the metabolic analyzer will produce precise and valid test results (within the specifications of the manufacturer of the device) from which clinical decisions may be made. However, this may not always be true if there is discrepancy between metabolic simulators (Myers, 1996).

The operation of the metabolic simulator utilizes a high precision motor-driven cylindrical piston pump similar to a syringe that creates very precise stroke volumes of air (VT) at adjustable stroke rates (breathing frequency, BF). By injecting known amounts of CO2 and N2 into the stroke volume, an exact volume of CO2 and of O2 per breath is created. Nitrogen is infused to reduce the O2 concentration of the ambient air to simulate oxygen consumption, and CO2 is added to the ambient air to simulate CO2 production. The gases, CO2 and N2, are added to the simulator’s tidal strokes through highly precise mass flow controllers (Huszczuk et al., 1990). With the injection of known masses of CO2 and N2, in combination with adjustments for pump stroke frequency and tidal volume (VT), accurate v.O2, v.CO2, and v.E can be created.

Both Relitech and Vacumed metabolic simulators share the same design principles. The user interface allows the operator to set different stroke volumes (1–3 L, in steps of 0.5 L) and adjustable stroke frequency (from 10 to 80 strokesmin–1), thus enabling simulated ventilation ranges from 10 Lmin–1 up to 240 Lmin–1 at a simulated breathing rate up to 80min–1. The metabolic gases for both systems are mixed by the metabolic simulator using room air that is pumped back and forth, injecting amounts of pure CO2 and N2. The 100% CO2 creates a gas that simulates a precise CO2 production rate at the different stroke rates (ventilation), while 100% N2 dilutes the ambient air O2 to a specific O2 concentration to simulate pre-set oxygen consumption rates. The amount of injected CO2 and N2 during each breath exhaled by the metabolic simulator is regulated by high precision mass flow controllers (MFC’s), resulting in a level of precision of simulated metabolic rate for O2 and CO2 of <0.2%. Certification of the metabolic simulator piston pump and MFC’s are performed initially, with re-certification every 2 years by the manufacturer of the MFCs. Together, this level of volume and MFC precision enable the metabolic simulators to create v.O2 and v.CO2 with an accuracy of <0.5% even at high ventilation ranges.

International guidelines recommend the use of a metabolic simulator to perform a thorough systematic check on the overall performance of a CPET metabolic system (Casaburi et al., 1997; Porszasz et al., 2007; Radtke et al., 2019). However, the reliability and comparability of different brands of commercially available metabolic simulators have not been discussed in the literature and may not necessarily be consistent across CPET carts (American Thoracic Society, 2003; Bunn et al., 2011). Therefore, we conducted a study to verify the uniformity of outcome response of two types of metabolic simulators against two standard CPET metabolic carts.

Materials and Methods

Two metabolic simulators equipped with mass flow controllers (Vacumed 17056, Vacumetrics, Ventura, CA, United States and Relitech, Relitech Systems BV, Nijkerk, The Netherlands) were used with two standard metabolic carts (Vyntus CPX and Vyntus ONE, both Vyaire Medical equipped with Sentry Suite Software version 3.10, Mettawa, IL, United States). The Vacumed (VM) metabolic simulator used for this investigation was modified to include two mass flow controllers (MFC), one for 100% oxygen and one for 100% nitrogen, instead of its typical one MFC for a gas mixture, to more closely mimic the double MFC setup in the Relitech (RT) system (see Figure 1).

FIGURE 1
www.frontiersin.org

Figure 1. Schematic view of metabolic simulator with sample output (original artwork).

The Vyntus CPX uses a calibration and verification procedure that is fully automatic using a built-in flow generator, producing precise constant airflows. This sets a flow calibration factor using a calibration flow at 2 Ls–1 followed by a low flow verification at 0.2 Ls–1 with limits of acceptability of ±1% of the expected flow. The 2 Ls–1 flow is used to set the gain: the measured flow at 2 Ls–1 should be within 1% of the expected flow, and a gain is set to adjust. The lower flow rate of 0.2 Ls–1 is for verification of the digital (rotating vane) volume sensor for undisturbed rotations at flow initiation.

In contrast, although the Vyntus ONE uses the identical digital volume sensor, volume is manually calibrated using a certified 3 L calibration syringe. A manual verification procedure is performed, discharging the 3 L syringe at least three times to give volume strokes of 3 L at flow ranges from 0.5 up to 12 Ls–1 (Porszasz et al., 2007). The limits of acceptability are 3 L ± 3.5% for each flow range, according to international standards (Miller, 2005).

Both Vyntus models use the same gas sensors: a high speed digital O2 sensor based on an electrochemical principle; and a fast response digital CO2 sensor based on the principle of infrared absorption. For both Vyntus models, the gas calibration procedures are the same, consisting of determining the O2 and CO2 signals, initially analyzing room ambient air and subsequently a highly accurate calibration gas mixture consisting of CO2, O2, and balanced nitrogen (nominal 5% CO2 and 15% O2 in N2, with 0.05% accuracy). The calibration certificate values as defined by the calibration gas supplier are set as reference values. The gas calibration procedure determines gain factors, response time, and delay time for the O2 as well as the CO2 sensor.

Both metabolic simulators were used in random sequence for both CPET devices. The primary CPET parameters (VT, RER, v.O2, and v.CO2) were verified using a standardized protocol: the testing sequence was first performed with a tidal volume of 2 L and subsequently 3 L at different frequencies: 20min–1, 40min–1, 60min–1, and 80min–1. Every frequency level consisted of three sets of at least 40 “breath” cycles. Vyntus CPX and the Vyntus ONE were both set to breath by breath (BxB) mode.

Metabolic simulators operate in ambient conditions, producing the pre-set breath cycles using ambient temperature and pressure conditions (ATP). Within the Vyntus system application software (SentrySuite®), a setting was used (SeS QM) which stops the conversion of “breaths” from ATP to body temperature, pressure and water vapor saturated (BTPS) for VT and standard temperature, and pressure, dry (STPD) for v.O2 and v.CO2 calculations.

Statistics

Application of the Bland–Altman Plot was used for the interpretation of comparisons between both Vyntus metabolic carts and both metabolic simulators. Differences between the VM and RT simulators, for different levels of BF for VT and RER and different levels of gas flow outcome values for VO2 and VCO2, were assessed using two-way analysis of variance (ANOVA). For analysis of bias, differences between metabolic simulator (VM, RT) vs. Vyntus (CPX, ONE) was tested by paired t-test. All data were analyzed using GraphPad Prism 5 (GraphPad Software, San Diego, CA 92108). Non-parametric Kendall’s Tau was used to check for correlation between proportional error of measured gas exchange estimates between metabolic simulators and Vyntus CPET systems (IBM Corp. Released 2020. IBM SPSS Statistics for Windows, Version 27.0. Armonk, NY: IBM Corp.). Significance was set at p < 0.05. Pre-defined acceptability limits (American Thoracic Society, 2003) are shown in Table 1.

TABLE 1
www.frontiersin.org

Table 1. Tolerance of differences between metabolic simulator and metabolic cart.

Results

Tidal Volume (VT)

Tidal volume results from the two simulators as measured by the Vyntus ONE and Vyntus CPX metabolic carts are shown in Figure 2. Mean VT results were significantly closer to the set point for the Relitech than for the Vacumed, but all results for both simulators were within acceptability ranges.

FIGURE 2
www.frontiersin.org

Figure 2. Tidal volume of Vacumed vs. Relitech simulators as measured by Vyntus CPX at 2 L (A) and 3 L (C) and Vyntus ONE at 2 L (B) and 3 L (D). Each data point represents the mean ± SD of three sets of 40 breaths. VM, Vacumed; RT, Relitech. All data points p < 0.001; VM vs. RT.

Gas Exchange: Oxygen Uptake (v.O2), Carbon Dioxide Output (v.CO2)

Oxygen uptake and carbon dioxide output flow rates as measured by Vyntus CPX and Vyntus ONE are shown in the Bland-Altman plots in Figures 3, 4. Although not all individual data points fell between the tolerance ranges for both simulators using both CPET devices, the mean bias for v.O2 and v.CO2 was less than 50 mlmin–1 and the upper and lower limits of agreement are within the acceptability limits identified in Table 1. Mean and SD data supporting these plots is given in Table 2.

TABLE 2
www.frontiersin.org

Table 2. Absolute gas (v.O2 and v.CO2) flow results for VT 2,000–3,000 mL.

FIGURE 3
www.frontiersin.org

Figure 3. Bland and Altmann plots showing the difference between oxygen delivery (VO2) from metabolic simulator as measured by Vyntus devices. CPX, Vyntus CPX; VM, Vacumed metabolic simulator; RT, Relitech metabolic simulator; ONE, Vyntus ONE; ULA, Upper limit of agreement; LLA, Lower limit of agreement.

FIGURE 4
www.frontiersin.org

Figure 4. Bland and Altmann plots showing the difference between carbon dioxide delivery (VCO2) from metabolic simulator as measured by Vyntus devices. CPX, Vyntus CPX; VM, Vacumed metabolic simulator; RT, Relitech metabolic simulator; ONE, Vyntus ONE; ULA, Upper limit of agreement, LLA, Lower limit of agreement.

From the v.O2 and v.CO2 results (Table 2) by both metabolic simulators it is shown that, using the Vyntus ONE at a VT of 2 L (both for v.O2 and v.CO2), a small but significant difference was found at 4,000 mlmin–1. The Vyntus CPX only showed a small but significant difference at a VT of 3 L at 3,000 mlmin–1 and 4,000 mlmin–1 for v.CO2.

Bias with 95% confidence intervals between Vyntus CPET systems and metabolic simulators is shown in Table 3. Kendall’s Tau correlation coefficient to check for proportional error of gas exchange estimates by Vyntus systems is shown in Table 4.

TABLE 3
www.frontiersin.org

Table 3. Bias with 95% confidence intervals between Vyntus CPET systems and metabolic simulators.

TABLE 4
www.frontiersin.org

Table 4. Kendall’s Tau correlation coefficient to check for proportional error of gas exchange estimates by Vyntus systems.

Respiratory Exchange Ratio (RER)

Results of the respiratory exchange ratio (v.CO2/v.O2) from the two simulators as measured by Vyntus ONE and Vyntus CPX metabolic carts are shown in Figure 5. The results show that RER at all breathing frequency levels generated by both metabolic simulators were within acceptable ranges. Graphs indicate the mean ± SD for three sets of 40 breaths.

FIGURE 5
www.frontiersin.org

Figure 5. (A,B) Respiratory exchange ratio (RER) from metabolic simulators as measured by Vyntus CPX (A) and Vyntus ONE (B) at 2 L. (C,D) Respiratory exchange ratio (RER) from metabolic simulators as measured by Vyntus CPX (C) and Vyntus ONE (D) at 3 L. Data points represent the mean value ± SD of three sets of 40 breaths (simulator strokes). RT, Relitech; VM, Vacumed; RR, respiratory rate. All not significant.

Discussion

We compared the results of two state-of-the-art metabolic simulators using two different metabolic carts. We measured tidal volumes, respiratory exchange ratios, and oxygen and carbon dioxide mixtures at both 2 and 3 L stroke volumes of the metabolic simulators, over a range of respiratory rates from 20 to 80min–1.

Both the metabolic simulator and the metabolic cart have error tolerances within which they must operate. Metabolic cart tolerances are defined by the manufacturer-based design standards set for the device. In this study, tolerances for the Vyntus CPX and Vyntus ONE, as specified by the manufacturer, are VT ± 2% or 50 mL, v.O2 and v.CO2 ± 3% or 50 mL/min (whichever is greater), and RER difference = 0.04 absolute. Both metabolic simulators, based on the manufacturers’ calibration and certification reports, generate VT with a precision of 0.2%. v.O2 and v.CO2 were generated using high precision mass flow controllers by injection of pure (100%) N2 and CO2, resulting in a level of precision of <0.5% for v.O2 and v.CO2. Using these levels of tolerance, the values for acceptability ranges in Table 1 were derived.

Vyntus calibration procedures were followed as recommended by the manufacturer consisting of a volume and flow calibration for the digital volume transducer (DVT) and gas calibration procedure for O2 and CO2 gas analyzers. There are important differences in the calibration procedures for the two systems, a significant impetus for us to use two different metabolic simulators for each CPET cart in this study. For the Vyntus CPX, the DVT sensor is calibrated and verified by a flow signal, while the Vyntus ONE is manually calibrated and verified by an absolute volume.

The dynamic calibration procedure is designed to determine the gas sample delay and response time to align both gas sensor signals (O2 and CO2) with the corresponding ventilation as measured directly at the outlet of the mouthpiece/facemask by the DVT sensor. Gain is the ratio between actual and measured gas concentrations. Delay time is defined as the time needed for the gas to reach the analyzer via the sample line. Response time is the time needed between reaching the analyzer to showing the measured value (Porszasz et al., 2007). Knowing the total sum of delay and response time dynamically enables the system to set the calibration gas gain in relation to the O2 and CO2 sensor response. In a traditional non-dynamic gas calibration, without the exact estimate of the total of response and delay time, a difference between O2 and CO2 sensor response can lead to underestimation of gas concentrations at medium to high ventilation (breathing frequency) during exercise. At high ventilation, the end value of the sampled gas is typically not 100%. This concept also ensures the compensation of small fluctuations such as changes in the sample line performance over time. For our experiments, prior to each test, a duplicate gas/delay time procedure ensured correct and repeatable reading of the O2 and CO2 sensor aligned to breath-by-breath ventilation.

Through precise and accurate timing of the switching between room air and the calibration gas, values for gain, response time, and delay time for CO2 and O2 are determined. The switching frequency is equivalent to breathing rate (frequency) during CPET breath-by-breath measurements. Since CPET is a dynamic, breathing frequency-dependent procedure, gain factors determined in relation to breathing frequency will give the most accurate readings for both CO2 and O2 at any breathing frequency during metabolic exercise testing.

The SentrySuite Quality Management (SeSQM) software system provided the benefit of reading the CPET gas exchange outcome values under ambient environmental conditions as generated by the metabolic simulators. Normally gas exchange (v.O2/v.CO2) within a patient record is presented under standard temperature, pressure, and dry (STPD) conditions while ventilation values such as tidal volume are presented under body temperature, pressure, and water vapor saturated (BTPS). By using SESQM, no manual corrections of Vyntus data results were needed, enabling direct comparison of metabolic simulator results for this study.

As shown in Figure 2 and Tables 1, 2, VT was set at either 2 or 3 L with breathing frequencies of 20–80 breaths per minute. This corresponds to ventilation ranging from 40 to 240 Lmin–1. Both metabolic simulators generated stroke volumes within the predefined limits of acceptability. The VM model generated significant slightly lower volumes over the full range of frequencies compared to the RT simulator, but both were within acceptable ranges. No significant breathing frequency dependency was observed for a VT of 2 L. During the VT 3 L experiments, a minor frequency dependency could be observed by both metabolic simulators due to a low breathing frequency for CPET.

In Figures 3, 4, we have presented the measured valued generated by each metabolic simulator (RT and VM) in a Bland-Altman plot (Bland and Altman, 1986) presentation. The X-axes define the mean values from the simulators and the metabolic carts, while the Y-axes show the differences between the Vyntus measured results minus the metabolic simulator-generated output. A positive difference indicates a value measured by Vyntus that is higher than the value generated by the simulator and a negative value on the Y-axes indicates Vyntus result that is lower than that generated by the simulator.

Bland and Altmann graphs for both gas exchange parameters v.O2 and v.CO2 demonstrate that the variance of the difference between metabolic simulator-produced values increases with the increase of scale of the values. This is in accordance with the pre-set range of acceptability criteria varying from 50 mlmin–1 for the lower values to 150 mlmin–1 for the higher gas exchange values as shown in Table 1.

As shown by the Bland and Altman plots and the data in Table 2, repeated single results, even within a single metabolic simulator, show a range of variance. Based on these study results, it can be suggested that when a metabolic simulator is used, multiple measurements should be obtained for clinical quality control and decision-making. In our example, we used the mean of three sets of measurements.

Both Vyntus devices showed small but significant differences for v.O2 and v.CO2 as delivered by both simulators, except for v.O2 as delivered by the Vacumed simulator. However, all 95% confidence intervals stayed within the level of clinical acceptance. In addition, we found a relation between proportional error and measured estimates (in the range 1,000–4,000 mLmin –1) for Vyntus ONE for v.O2 and v.CO2 and for Vyntus CPX for VO2 when each device was connected to the Vacumed simulator. This is consistent with the expectations as reflected by an increased range of acceptance at higher measured gas flow estimates.

Because the RER is calculated from v.O2 and v.CO2, the negative difference of RER trend results from these combined effects (Figure 5). Although this tendency is apparent, the differences are still well within the defined range of acceptability.

The four chosen metabolic rates produced by the metabolic simulators, 1, 2, 3, and 4 liters per minute for v.O2 and v.CO2, corresponded to increasing the breathing frequency through 20, 40, 60, and 80min–1, respectively. In Figure 2, with the simulator’s VT set to 3 L, we can see that there is a minor difference between the simulator and Vyntus ONE at lower breathing rates (20 and 40min–1), and the difference increases with higher breathing rates (at 60 and 80min–1), proportional to the frequency. This can be observed in both v.O2 and v.CO2 results from Vyntus ONE.

The respiratory exchange ratio (RER) is an important parameter used in the interpretation of exercise testing results. For example, an RER over 1.05 indicates that the subject is close to peak exercise and the highest obtained RER during CPET is used in interpretation of exercise performance. Since RER is a derived parameter from v.O2 and v.CO2, the acceptability range is set at an absolute difference of 0.04 from the expected generated value of 1.00. As shown in Figure 5, the differences as measured by Vyntus CPX and Vyntus ONE were within the 0.04 limit of acceptability for the total range of frequency and metabolic states (breathing frequency, v.O2, and v.CO2). Strict criteria for the RER help to understand the balance between generated v.O2 and v.CO2. If v.CO2 is high (within the acceptable ranges) but v.O2 is low (within the acceptable ranges), a falsely high RER could be seen, thus potentially leading to misinterpretation of the CPET results of the subject tested.

Of note, RER is not influenced by variations in VT, as VT equally affects v.O2 and v.CO2. Because of this, RER relative to the absolute v.O2, v.CO2, and VT as generated by the simulator is a very powerful quality check of CPET analyzers.

Conclusion

In our study, both standard metabolic simulators created comparable gas mixtures for O2 and CO2, to simulate highly accurate metabolic breath-by-breath parameters, within the ranges of pre-defined acceptability limits. The simulators are reliable across the spectrum of respiratory rates at tidal volumes of 2 and 3 L. Use of either of these metabolic simulators for standardized verification of metabolic cart equipment ensures accurate physiologic measurement of subjects undergoing cardiopulmonary exercise testing. No systematic differences were identified when comparing the two types of simulators.

Despite the use of a metabolic simulator, CPET labs must ensure the proper performance of a metabolic cart using both biological control (repeated measurements of a human subject over time) and expert clinical quality control with clinical interpretation of normal human subject testing.

Data Availability Statement

The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author/s.

Author Contributions

TS and HG were involved equally in the study design, data acquisition, interpretation of data, and review of the manuscript. ER was involved in the interpretation of data and wrote the manuscript. All authors contributed to the article and approved the submitted version.

Conflict of Interest

TS and ER are paid consultants for Vyaire Medical, Mettawa, IL, United States. HG is an employee of Vyaire Medical, Mettawa, IL, United States. Financial support for this project came from Vyaire Medical, Mettawa, IL, United States.

Acknowledgments

The results of our study are presented clearly, honestly, and without fabrication, falsification, or inappropriate data manipulation. The results of the present study do not constitute endorsement by ACSM.

References

American Thoracic Society (2003). ATS/ACCP statement on cardiopulmonary exercise testing. Am. J. Res. Crit. Med. 167:211.

Google Scholar

Beijst, C., Schep, G., Breda, E., Wijn, P. F., and Pul, C. (2013). Accuracy and precision of CPET equipment: a comparison of breath-by-breath and mixing chamber systems. J. Med. Eng. Technol. 37, 35–42. doi: 10.3109/03091902.2012.733057

PubMed Abstract | CrossRef Full Text | Google Scholar

Bland, J. M., and Altman, D. (1986). Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 327, 307–310.

Google Scholar

Bunn, J. A., Pittsley, J. L., Baker, S. V., and Yates, J. (2011). Assessment of Accuracy of the Vacu-Med 17053 Calibrator for Ventilation, Oxygen Uptake (v.O2), and Carbon Dioxide Production (v.CO2). Respirat. Care 56, 472–476. doi: 10.4187/respcare.00951

PubMed Abstract | CrossRef Full Text | Google Scholar

Casaburi, R., Prefaut, C., and Cotes, J. (1997). Equipment, measurements and quality control in clinical exercise testing. Monograph 6, 72–87.

Google Scholar

Gore, C. J., Catcheside, P. G., French, S. N., Bennett, J. M., and Laforgia, J. (1997). Automated VO2max calibrator for open-circuit indirect calorimetry systems. Med. Sci. Sports Exerc. 29, 1095–1103. doi: 10.1097/00005768-199708000-00016

PubMed Abstract | CrossRef Full Text | Google Scholar

Huszczuk, A., Whipp, B. J., and Wasserman, K. A. (1990). respiratory gas exchange simulator for routine calibration in metabolic studies. Eur. Respir. J. 3, 465–468.

Google Scholar

Levett, D. Z. H., Jack, S., Swart, M., Carlisle, J., Wilson, J., Snowden, C., et al. (2018). Perioperative cardiopulmonary exercise testing (CPET): consensus clinical guidelines on indications, organization, conduct, and physiological interpretation. Br. J. Anaesth. 120, 484–500. doi: 10.1016/j.bja.2017.10.020

PubMed Abstract | CrossRef Full Text | Google Scholar

Miller, M. A. T. S. (2005). /ERS task force: standardisation of spirometry. Eur. Respir. J. 26, 319–338.

Google Scholar

Muraresku, C. C., McCormick, E. M., and Falk, M. J. (2018). Mitochondrial Disease: Advances in clinical diagnosis, management, therapeutic development, and preventative strategies. Curr. Genet. Med. Rep. 6, 62–72.

Google Scholar

Myers, J. (1996). Essentials of cardiopulmonary exercise testing. Champaign, IL: Human Kinetics Publishers.

Google Scholar

Porszasz, J., Stringer, W., and Casaburi, R. (2007). Equipment, measurements and quality control in clinical exercise testing. Eur. Respirat. Monograph 40:108.

Google Scholar

Radtke, T., Crook, S., Kaltsakas, G., Louvaris, Z., Berton, D., Urquhart, D. S., et al. (2019). ERS statement on standardisation of cardiopulmonary exercise testing in chronic lung diseases. Eur. Respir. Rev. 28:154. doi: 10.1183/16000617.0101-2018

PubMed Abstract | CrossRef Full Text | Google Scholar

Van Iterson, E. H., Borlaug, B. A., Johnson, B. D., and Olson, T. P. (2016). The Calibration of Oxygen Uptake, Delivery, and Utilization is Influenced by Heart Failure With Preserved versus Reduced Ejection Fraction During Exercise. Circulation 134(Suppl._1), A16287–A.

Google Scholar

West, M. A., Wischmeyer, P. E., and Grocott, M. P. W. (2017). Prehabilitation and Nutritional Support to Improve Perioperative Outcomes. Curr. Anesthesiol. Rep. 7, 340–349.

Google Scholar

Keywords: cardiopulmonary exercise test, metabolic simulator, respiratory exchange ratio, calibration, metabolic cart, Vyntus ONE, Vyntus CPX

Citation: Souren T, Rose E and Groepenhoff H (2021) Comparison of Two Metabolic Simulators Used for Gas Exchange Verification in Cardiopulmonary Exercise Test Carts. Front. Physiol. 12:667386. doi: 10.3389/fphys.2021.667386

Received: 12 February 2021; Accepted: 07 May 2021;
Published: 03 June 2021.

Edited by:

Emiliano Cè, University of Milan, Italy

Reviewed by:

José Ramón Alvero Cruz, University of Malaga, Spain
Susanna Rampichini, University of Milan, Italy

Copyright © 2021 Souren, Rose and Groepenhoff. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Edward Rose, ZXJvc2VAMjBwb21zLmNvbQ==

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.