Influence of catheter thickness on respiratory physiology during less invasive surfactant administration in extremely preterm infants

\r\nChamindu C. Gunatilaka,,&#x;Chamindu C. Gunatilaka1,2,†Qiwei Xiao,,&#x;Qiwei Xiao1,2,†Alister J. Bates,,,&#x;Alister J. Bates1,2,3,†Axel R. Franz,,&#x;Axel R. Franz4,5,†Christian F. Poets,&#x;Christian F. Poets4,†Christian A. Maiwald,
&#x;\r\nChristian A. Maiwald4,5*
  • 1Center for Pulmonary Imaging Research, Cincinnati Children’s Hospital Medical Center, Cincinnati, OH, United States
  • 2Division of Pulmonary Medicine, Cincinnati Children’s Hospital Medical Center, Cincinnati, OH, United States
  • 3Department of Pediatrics, University of Cincinnati, Cincinnati, OH, United States
  • 4Department of Neonatology, University Children’s Hospital Tübingen, Tübingen, Germany
  • 5Center for Pediatric Clinical Studies (CPCS), University Hospital Tübingen, Tübingen, Germany

Introduction: Delivering surfactant via thin catheters (minimal-invasive surfactant therapy (MIST); less invasive surfactant administration (LISA)) has become a common procedure. However, the effect of tracheal obstruction caused by catheters of different sizes on tracheal resistance in extremely low gestational age newborns (ELGANs) is unknown.

Methods: To investigate the effect of catheters size 3.5, 5 and 6 French on airway resistance in ELGANs of 23–28 weeks gestational age during LISA, we performed calculations based on Hagen-Poiseuille's law and compared these with a clinically and physically more accurate method: computational fluid dynamics (CFD) simulations of respiratory airflow, performed in 3D virtual airway models derived from MRI.

Results: The presence of the above catheters decreased the cross-sectional area of the infants' tracheal entrance (the cricoid ring) by 13–53%. Hagen-Poiseuille's law predicted an increase in resistance by 1.5–4.5 times and 1.3–2.6 times in ELGANs born at 23 and 28 weeks, respectively. However, CFD simulations demonstrated an even higher increase in resistance of 3.4–85.1 and 1.1–3.5 times, respectively. The higher calculated resistances were due to the extremely narrow remaining lumen at the glottis and cricoid with the catheter inserted, resulting in a stronger glottal jet and turbulent airflow, which was not predicted by Hagen-Poiseuille.

Conclusion: Catheter thickness can greatly increase tracheal resistance during LISA-procedures in ELGANs. Based on these models, it is recommended to use the thinnest catheter possible during LISA in ELGANs to avoid unnecessary increases in airway resistance in infants already experiencing dyspnea due to respiratory distress syndrome.

Introduction

Surfactant therapy via less invasive surfactant administration (LISA) or minimally-invasive surfactant therapy (MIST) in neonates is increasingly used and its introduction into routine care in neonatal intensive care units in Europe occurred quite rapidly (14).

However, there is also critique concerning this method (5, 6). Compared to more invasive methods, the distribution of surfactant seems worse (7) and there is a higher risk of failure particularly in extremely low gestational age neonates (ELGANs) (8). A potential reason for this could be an impaired lung ventilation during LISA. In their critique of the LISA-method, De Luca et al. also referred to the influence of catheter thickness and argued that this can be calculated by using Hagen-Pouiselle's law (5). However, because Hagen-Pouiselle's law requires a circular cross-sectional area and a laminar flow and both conditions are not fulfilled during LISA, we performed a more detailed investigation on this topic.

In Germany, various approved and specifically designed LISA/MIST-catheters (e.g., Neofact®—Lyomark Pharma, Oberhaching, Germany and Surfcath™—Vygon, Aachen, Germany) and non-approved catheters intended for other purposes (e.g., gastric tubes, umbilical vein catheters, etc.) are being used for LISA (9, 10). While outer diameters of the CE-marked, approved catheters are 3.5 (1.2 mm; Neofact®) or 6 French (2.0 mm; Surfcath™) respectively, non-approved catheters reportedly vary between 3.5 and 5 French (1.2 and 1.7 mm).

The narrowest part of the airway is the cricoid ring which has an inner diameter that varies between 2.7–2.8 mm in infants with a gestational age (GA) of 23 weeks and 3.2–3.3 mm in infants with a GA of 28 weeks, respectively (11, 12).

This is not much greater than catheter thickness and could therefore represent a relevant obstacle to the infants’ spontaneous respiration, since tracheal obstruction by the catheter will likely increase airway resistance, resulting in an increased work of breathing (WOB), decreased tidal volume (Vt) and minute ventilation (MV). Considering the law of Hagen-Poiseuille, a reduction of the inner diameter of the airway increases resistance by the fourth power (13) “or the fifth power, in the case of non-laminar flow” … “because of the non-perfectly circular shape” of the tracheal cross-sectional area (5). It can therefore be assumed that the airway obstruction induced by the catheter during LISA/MIST will have the greatest impact on breathing in the smallest infants. And it may also explain why GA <28 weeks is an independent risk factor for LISA-failure (8).

However, in addition to the obstruction itself, the neonatal airway has some unique features that makes it much more difficult for the infant to compensate such a loss in Vt.

(1) WOB in preterm infants is mainly done by the diaphragm (14). However, stronger contractions of the diaphragm (to overcome increased resistance) would be partially off-set by the high elasticity of the chest wall (15).

(2) The horizontal position of the ribs hardly allows any support from thoracic muscles (14).

(3) The respiratory muscles mainly consist of quickly exhaustible type I muscle fibers (16).

(4) Pharynx, larynx and trachea are softer than in older infants and tend to collapse more quickly (17), which additionally raises resistance. This must be particularly true, when the flow velocity will be raised to maintain Vt/MV during an obstruction.


One should also consider the altered postnatal adaptation of ELGANs, where lung fluid clearance of the airways by aquaporin-4 channels (AQP4) is less effective (18), and the additional impact of the respiratory distress syndrome (RDS), which is the reason these infants should be treated via LISA/MIST. RDS decreases the functional residual capacity (19) and accelerates respiratory rate (20). The latter factor limits the ability to compensate a loss of Vt by the catheter-obstruction.

The points discussed above triggered this investigation into the potential influence of catheter thickness during LISA/MIST on respiratory physiology, as we suspected that the use of thicker catheters could affect spontaneous breathing during LISA (and thus possibly also the homogeneous distribution of surfactant), particularly in ELGANs, and could therefore represent a potential risk factor for treatment failure.

Material and methods

Method selection

We calculated the catheter-induced change in airway resistance using two methods. First, we calculated the pressure drop (and therefore resistance) of a laminar flow within a tube using Hagen-Poiseuille's law to estimate changes in the airway resistance, as discussed by specialists in this field (5). The rationale for using Hagen-Poiseuille's law is that it is a relatively simple calculation and can thus be performed quickly for any combination of patient and catheter size. However, the theory behind this law makes several assumptions about flow and geometry (listed below) which may not apply to tracheal airflow. Therefore, computational fluid dynamics (CFD) simulations, which are not based on the same assumptions, and are therefore capable of producing more accurate results, were also performed. The downside to CFD simulations is that they are computationally expensive to perform, and patient-specific simulations require medical imaging of the airway to create its geometry.

Theoretical calculations

The Hagen-Poiseuille model of airway resistance was made by first calculating the cross-sectional area (CSA) of the associated tracheal lumen (CSAlumen) at the narrowest part (cricoid ring) in infants with an GA of 23–28 weeks [diameters were calculated according to (11, 12)]. The CSA of the catheters were also calculated using 3.5, 5, and 6 French tubes (CSAcatheter). For each combination of patient and catheter size, the remaining CSA of the lumen with the catheter in place was calculated (CSAlumenCSAcatheter), and the radius of a circle with this remaining area (rremaining) calculated as follows:

rremaining=CSAlumenCSAcatheterπ

After that, the percentage of the remaining radius (rratio=rremaining/rtrachea, where rtrachea is the radius of the trachea) was calculated to determine the increase in resistance (ΔRes) by application of Hagen-Poiseuille's law (ΔRes=1/(rratio4)). This law estimates the resistance along a tube due to flow based on the following assumptions: laminar flow; constant circular CSA; no flow acceleration; incompressible liquid; diameter much smaller than the length of the tube. Since several of these assumptions are not true for respiratory airflow, a second method was also used to calculate the change in resistance (21).

All calculations were performed using Microsoft Excel 2010.

Computational fluid dynamics (CFD)

A neonatal intensive care unit patient at a post-menstrual age of 40 weeks was enrolled with approval from the Institutional Review Board and parental consent. The patient had no known airway or lung disease, was considered as a control for the purpose of this study and was imaged using ultrashort echo time magnetic resonance imaging (MRI) technique. Next, a three-dimensional airway model was created via segmentation of the MR images as previously described (2229). To create airway models at sizes equivalent to different ages, the three dimensional airway model of the 40 week old subject was downscaled uniformly until an average tracheal diameter of 3.0 and 3.8 mm was reached, which was equivalent to be the average diameter of an infant with a GA of 23 weeks and 28 weeks, respectively (11, 12). Breathing rate was set to 80 breaths/minute and inspiration time to 1/3 of a respiratory cycle. Tidal volume was calculated as 5.75 ml/kg bodyweight based on weights of 600 g for the 23-week model and 1,050 g for the 28-week model [50th percentile of female infants with 23 and 28 weeks of gestation (30)]. PEEP was set to 6 cmH2O, as this is the standard procedure at the neonatal intensive care unit in Tuebingen. CFD simulations were performed using Simcenter STAR-CCM+ 14.04.011-R8 (Siemens Digital Industries Software) commercial software package. These respiratory airflow simulations have been validated previously (31). CFD simulations model the airflow for a single respiratory cycle without catheters and subsequently with nasally guided catheters of 3.5, 5, and 6 French diameters were inserted 1.5 cm below the vocal cords into the trachea of the infant with a GA of 23 weeks and 2.0 cm in the infant with a GA of 28 weeks (32). For each simulation, airway resistance in the trachea at peak inspiration and the total pressure drop between the nostrils and the end of the main bronchi at peak expiration were calculated.

Results

Theoretical calculations

Depending on the diameter of the LISA-catheter and CSA of the trachea, the “remaining CSA” was reduced by 13%–53%. This resulted in a theoretical increase in resistance by a factor of ×1.3–4.5. The relevance of the differences increased exponentially with immaturity (i.e., lower tracheal CSA) of the infant. For details, see Figure 1 and calculations, see Table 1.

Figure 1
www.frontiersin.org

Figure 1. Overview of theoretical calculations based on the Law of Hagen–Pouiselle.

Table 1
www.frontiersin.org

Table 1. Calculations for the use of catheters of different thickness during LISA, based on Hagen–Pouiselle's law.

CFD

CFD simulations demonstrated a clear increase in airflow velocity at peak inspiration in the infant's trachea as the catheter increased in size (Figure 2). There was a high velocity jet from the vocal cords into the trachea when a 6 French catheter was inserted into the airway. However, the increase in airflow velocity was minimal when a 3.5 French catheter was inserted, although the airflow pattern was also changed. When 5 French and 6 French catheters were placed, airflow was shifted to the anterior aspect of the tracheal lumen, with little flow in the posterior part of the airway.

Figure 2
www.frontiersin.org

Figure 2. Airflow velocity at peak inspiration for an infant of 23 weeks’ gestation. Left to right: No tube, 3.5 French tube, 5 French tube, and 6 French tube. The black arrow on each airway represents the airflow direction.

We then calculated the pressure drop between nostrils and the carina at peak expiration expected from the increase in airflow resistance due to catheter obstruction. Figure 3 demonstrates the total pressure drop in a preterm infant of 23 weeks gestation.

Figure 3
www.frontiersin.org

Figure 3. Total pressure drop between nostrils and carina at peak expiration for an infant at a GA of 23 weeks. The total pressure drop between nostrils to carina is further divided into upper airway and trachea. The insertion of a 6 French catheter into the airway results in a 48-times increase in total pressure drop between the nostrils and carina compared to the unobstructed airway without catheter.

The tracheal resistance at peak in- and expiration significantly increased with immaturity and thickness of the tracheal catheter. Simulations in an infant at 23 weeks’ gestation demonstrated a threefold (3.5 Fr.), 17-fold (5 Fr.), and 85-fold (6 Fr.), respectively, increase in airway resistance at peak inspiration (Figure 4).

Figure 4
www.frontiersin.org

Figure 4. Increase in tracheal resistance at peak inspiration (black) and peak expiration (red) caused by a tracheal catheter in infants at 23- and 28-weeks’ GA. Tracheal resistance was raised by more than 85 times when a 6F catheter was inserted into the airway of an infant at 23 weeks’ GA.

Discussion

The clinical relevance of our findings arises from the purely logical consideration that in a critically ill infant with respiratory distress undergoing LISA, any additional, procedure-related WOB should be reduced to a minimum. The reported calculations serve to illustrate the potential magnitude of catheter-related factors during LISA.

Although clinical trials have shown the benefit of LISA in ELGANs, the pertinent Cochrane review (33) included only limited data on extremely preterm infants. Only Kribs et al. (34) reported a subgroup analysis of infants at 23–24 weeks gestation and found no benefit for LISA. Additionally, GA <28 weeks is an independent risk factor for LISA-failure (8) and surfactant distribution was shown to be worse in LISA compared to invasive surfactant application methods in lambs (7). Therefore, an improvement of the ventilatory situation during LISA (e.g., by a reduction in catheter size) might be helpful in these infants.

In-silico calculations were chosen for this study since in vivo measurements of airway resistance in spontaneously breathing infants on non-invasive respiratory support during LISA procedures are impractical. However, measurements of flow or electrical impedance tomography (35) might help to better understand these physiological aspects in the future.

Hagen-Poiseuille's law assumes a laminar flow in an ideal tube. Such theoretical assumptions are common in physical calculations (36) and were also mentioned in relation to LISA (5), but only rarely reflect clinical reality: as shown, CFD airflow becomes turbulent due to an increase in airflow velocity caused by catheter-related airway obstruction, and the resulting crescent-shaped CSA increases airway resistance substantially. Even assuming a fifth power relationship as suggested by De Luca et al. (corresponding to a 1.6–6.6-fold increase), one would strongly underestimate the increase in resistance.

Due to the above-mentioned features of the neonatal airway, ELGANs only have two options to compensate catheter-mediated airway obstruction. They can increase their respiratory rate or the pressure gradient (ΔP) across the trachea, which would have to be accomplished by an increase in diaphragmatic contractility. However, a higher velocity of tracheal gas flow leads to a reduced distending pressure according to the Bernoullie principle during inspiration and thus to an increased risk of airway collapse, a further increase in resistance and reduced effectiveness of the applied WOB. Taken together, the capacity for compensation of such a catheter-related partial airway obstruction appears to be limited in ELGANs suffering from RDS.

The role of PEEP during LISA with respect to airway resistance, effective Vt and WOB appears unpredictable with our model. On one hand, partial airway obstruction, particularly if combined with an increased respiratory rate, may lead to inadvertent PEEP, which might increase dead-space and therefore additionally reduces effective tidal volume. On the other hand, PEEP will be near zero during catheter insertion (due to an open mouth during laryngoscopy). Therefore, it remains unclear whether the catheter-related obstruction prevents any build-up of PEEP or will lead to an auto-PEEP.

Limitations of this study include that it relies on in-silico simulations and theoretical calculations based on physical principles rather than in vivo measurements. The calculations and simulations do not consider potential partial compensation by increased breathing efforts of the infant and the role of PEEP. In addition, our CFD calculations were based on a term infant whose proportions were uniformly scaled down to represent a premature infant. The anatomical proportions may vary in reality and therefore it is debatable whether the CFD results of one MRI can be considered representative of the target population.

However, in summary, we conclude that the use of thicker catheters in ELGANs may lead to decreased tidal volume due to increased resistance during the LISA procedure, potentially also affecting surfactant distribution (7, 37), and therefore it seems to be possible, that the use of thicker catheters increase failure rates of LISA in ELGANs.

Conclusion

The present calculations show that thickness of the selected catheter for LISA procedures matters, particularly in the most immature ELGANs with a small trachea. The increase in the infant's tracheal resistance and WOB can be relevant and may contribute to LISA failures. These findings may be relevant to the clinical setting, despite our results being only based on in silico modeling and theoretical considerations rather than in vivo measurements. We recommend that catheters selected for LISA in ELGANs should be as thin as possible—but the disadvantage of a potentially more difficult LISA procedure with a thinner and hence more flexible catheter have to be considered.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding author.

Ethics statement

The MRI data were collected prospectively under local IRB approval (Cincinnati Children’s Hospital Medical Center; IRB 2023-6101) from which an airway model was created. The de-identified airway model was then used retrospectively in this study.

Author contributions

CG: Conceptualization, Investigation, Writing – original draft, Writing – review & editing, Methodology. QX: Investigation, Writing – original draft, Writing – review & editing. AB: Conceptualization, Investigation, Methodology, Supervision, Validation, Writing – original draft, Writing – review & editing. AF: Conceptualization, Investigation, Methodology, Supervision, Writing – review & editing. CP: Conceptualization, Supervision, Writing – review & editing. CM: Conceptualization, Investigation, Methodology, Writing – original draft, Writing – review & editing.

Funding

The author(s) declare financial support was received for the research, authorship, and/or publication of this article.

Acknowledgments

We acknowledge support by Open Access Publishing Fund of University of Tübingen.

Conflict of interest

CA Maiwald has a financial relationship in the profits of the Neofact® product as inventor (Murrplastik Medizintechnik GmbH, Germany) and his research projects on LISA are, in part, supported by Lyomark Pharma GmbH, Germany.

The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Abbreviations

CFD, computational fluid dynamics; CSA, cross-sectional area; LISA, less invasive surfactant administration; MIST, minimal-invasive surfactant therapy; MV, minute ventilation; RDS, respiratory distress syndrome; Vt, tidal volume; WOB, work of breathing.

References

1. Klotz D, Porcaro U, Fleck T, Fuchs H. European perspective on less invasive surfactant administration-a survey. Eur J Pediatr. (2017) 176(2):147–54. doi: 10.1007/s00431-016-2812-9

PubMed Abstract | Crossref Full Text | Google Scholar

2. Fernandez C, Boix H, Camba F, Comuñas JJ, Castillo F. Less invasive surfactant administration in Spain: a survey regarding its practice, the target population, and premedication use. Am J Perinatol. (2020) 37(3):277–80. doi: 10.1055/s-0039-1678534

PubMed Abstract | Crossref Full Text | Google Scholar

3. Bhayat S, Kaur A, Premadeva I, Reynolds P, Gowda H. Survey of less invasive surfactant administration in England, slow adoption and variable practice. Acta Paediatr. (2020) 109(3):505–10. doi: 10.1111/apa.14995

PubMed Abstract | Crossref Full Text | Google Scholar

4. Heiring C, Jonsson B, Andersson S, Björklund LJ. Survey shows large differences between the Nordic countries in the use of less invasive surfactant administration. Acta Paediatr. (2017) 106(3):382–6. doi: 10.1111/apa.13694

PubMed Abstract | Crossref Full Text | Google Scholar

5. De Luca D, Shankar-Aguilera S, Centorrino R, Fortas F, Yousef N, Carnielli VP. Less invasive surfactant administration: a word of caution. Lancet Child Adolesc Health. (2020) 4(4):331–40. doi: 10.1016/S2352-4642(19)30405-5

PubMed Abstract | Crossref Full Text | Google Scholar

6. De Luca D, de Winter JP. Less invasive surfactant administration: all that glitters is not gold. Eur J Pediatr. (2020) 179(8):1295–6. doi: 10.1007/s00431-020-03688-3

PubMed Abstract | Crossref Full Text | Google Scholar

7. Niemarkt HJ, Kuypers E, Jellema R, Ophelders D, Hütten M, Nikiforou M, et al. Effects of less-invasive surfactant administration on oxygenation, pulmonary surfactant distribution, and lung compliance in spontaneously breathing preterm lambs. Pediatr Res. (2014) 76(2):166–70. doi: 10.1038/pr.2014.66

PubMed Abstract | Crossref Full Text | Google Scholar

8. Janssen LC, Van Der Spil J, van Kaam AH, Dieleman JP, Andriessen P, Onland W, et al. Minimally invasive surfactant therapy failure: risk factors and outcome. Arch Dis Child Fetal Neonatal Ed. (2019) 104(6):F636–42. doi: 10.1136/archdischild-2018-316258

PubMed Abstract | Crossref Full Text | Google Scholar

9. Herting E, Hartel C, Gopel W. Less invasive surfactant administration: best practices and unanswered questions. Curr Opin Pediatr. (2020) 32(2):228–34. doi: 10.1097/MOP.0000000000000878

PubMed Abstract | Crossref Full Text | Google Scholar

10. Bhattacharya S, Read B, Miller M, da Silva O. Impact of catheter choice on procedural success of minimally invasive surfactant therapy. Am J Perinatol. (2023) 40(11):1202–7. doi: 10.1055/s-0041-1733956

PubMed Abstract | Crossref Full Text | Google Scholar

11. Fishman RA, Pashley NR. A study of the premature neonatal airway. Otolaryngol Head Neck Surg. (1981) 89(4):604–7. doi: 10.1177/019459988108900419

PubMed Abstract | Crossref Full Text | Google Scholar

12. Fayoux P, Devisme L, Merrot O, Marciniak B. Determination of endotracheal tube size in a perinatal population: an anatomical and experimental study. Anesthesiology. (2006) 104(5):954–60. doi: 10.1097/00000542-200605000-00011

PubMed Abstract | Crossref Full Text | Google Scholar

13. Rocha G, Soares P, Gonçalves A, Silva AI, Almeida D, Figueiredo S, et al. Respiratory care for the ventilated neonate. Can Respir J. (2018) 2018:7472964. doi: 10.1155/2018/7472964

PubMed Abstract | Crossref Full Text | Google Scholar

14. Blackburn ST. Maternal, Fetal, & Neonatal Physiology: A Clinical Perspective. Amsterdam, Netherlands: Saunders Elsevier (2007).

Google Scholar

15. Papastamelos C, Panitch HB, England SE, Allen JL. Developmental changes in chest wall compliance in infancy and early childhood. J Appl Physiol (1985). (1995) 78(1):179–84. doi: 10.1152/jappl.1995.78.1.179

PubMed Abstract | Crossref Full Text | Google Scholar

16. Keens TG, Bryan AC, Levison H, Ianuzzo CD. Developmental pattern of muscle fiber types in human ventilatory muscles. J Appl Physiol Respir Environ Exerc Physiol. (1978) 44(6):909–13. doi: 10.1152/jappl.1978.44.6.909

PubMed Abstract | Crossref Full Text | Google Scholar

17. Neumann RP, von Ungern-Sternberg BS. The neonatal lung–physiology and ventilation. Paediatr Anaesth. (2014) 24(1):10–21. doi: 10.1111/pan.12280

PubMed Abstract | Crossref Full Text | Google Scholar

18. Walters DV, Ramsden CA, Olver RE. Dibutyryl cAMP induces a gestation-dependent absorption of fetal lung liquid. J Appl Physiol (1985). (1990) 68(5):2054–9. doi: 10.1152/jappl.1990.68.5.2054

PubMed Abstract | Crossref Full Text | Google Scholar

19. Goldsmith LS, Greenspan JS, Rubenstein SD, Wolfson MR, Shaffer TH. Immediate improvement in lung volume after exogenous surfactant: alveolar recruitment versus increased distention. J Pediatr. (1991) 119(3):424–8. doi: 10.1016/S0022-3476(05)82057-8

PubMed Abstract | Crossref Full Text | Google Scholar

20. Ainsworth SB. Pathophysiology of neonatal respiratory distress syndrome: implications for early treatment strategies. Treat Respir Med. (2005) 4(6):423–37. doi: 10.2165/00151829-200504060-00006

PubMed Abstract | Crossref Full Text | Google Scholar

21. Pisano A. From tubes and catheters to the basis of hemodynamics: viscosity and Hagen–Poiseuille equation. In: Physics for Anesthesiologists and Intensivists. Cham: Springer (2021). doi: 10.1007/978-3-030-72047-6_8

Crossref Full Text | Google Scholar

22. Yang MM, Higano NS, Gunatilaka CC, Hysinger EB, Amin RS, Woods JC, et al. Subglottic stenosis position affects work of breathing. Laryngoscope. (2021) 131(4):E1220–6. doi: 10.1002/lary.29169

PubMed Abstract | Crossref Full Text | Google Scholar

23. Gunatilaka CC, Higano NS, Hysinger EB, Gandhi DB, Fleck RJ, Hahn AD, et al. Increased work of breathing due to tracheomalacia in neonates. Ann Am Thorac Soc. (2020) 17(10):1247–56. doi: 10.1513/AnnalsATS.202002-162OC

PubMed Abstract | Crossref Full Text | Google Scholar

24. Gunatilaka CC, Schuh A, Higano NS, Woods JC, Bates AJ. The effect of airway motion and breathing phase during imaging on CFD simulations of respiratory airflow. Comput Biol Med. (2020) 127:104099. doi: 10.1016/j.compbiomed.2020.104099

PubMed Abstract | Crossref Full Text | Google Scholar

25. Gunatilaka CC, McKenzie C, Hysinger EB, Xiao Q, Higano NS, Woods JC, et al. Tracheomalacia reduces aerosolized drug delivery to the lung. J Aerosol Med Pulm Drug Deliv. (2024) 37(1):19–29. doi: 10.1089/jamp.2023.0023

PubMed Abstract | Crossref Full Text | Google Scholar

26. Xiao Q, Ignatiuk D, McConnell K, Gunatilaka C, Schuh A, Fleck R, et al. The interaction between neuromuscular forces, aerodynamic forces, and anatomical motion in the upper airway predicts the severity of pediatric OSA. J Appl Physiol (1985). (2024) 136(1):70–8. doi: 10.1152/japplphysiol.00071.2023

PubMed Abstract | Crossref Full Text | Google Scholar

27. Ignatiuk D, Xiao Q, McConnell K, Fleck R, Schuler C, Schuh A, et al. Computational assessment of upper airway muscular activity in obstructive sleep apnea—in vitro validation. J Biomech. (2022) 144:111304. doi: 10.1016/j.jbiomech.2022.111304

PubMed Abstract | Crossref Full Text | Google Scholar

28. Gunatilaka CC, Hysinger EB, Schuh A, Xiao Q, Gandhi DB, Higano NS, et al. Predicting tracheal work of breathing in neonates based on radiological and pulmonary measurements. J Appl Physiol (1985). (2022) 133(4):893–901. doi: 10.1152/japplphysiol.00399.2022

PubMed Abstract | Crossref Full Text | Google Scholar

29. Xiao Q, Cetto R, Doorly DJ, Bates AJ, Rose JN, McIntyre C, et al. Assessing changes in airflow and energy loss in a progressive tracheal compression before and after surgical correction. Ann Biomed Eng. (2020) 48(2):822–33. doi: 10.1007/s10439-019-02410-1

PubMed Abstract | Crossref Full Text | Google Scholar

30. Voigt M, Fusch C, Olbertz D, Hartmann K, Rochow N, Renken C, et al. Analyse des Neugeborenenkollektivs der Bundesrepublik Deutschland. Geburtshilfe Frauenheilkd. (2006) 66(10):956–70. doi: 10.1055/s-2006-924458

Crossref Full Text | Google Scholar

31. Xiao Q, Stewart NJ, Willmering MM, Gunatilaka CC, Thomen RP, Schuh A, et al. Human upper-airway respiratory airflow: in vivo comparison of computational fluid dynamics simulations and hyperpolarized 129Xe phase contrast MRI velocimetry. Plos One. (2021) 16(8):e0256460. doi: 10.1371/journal.pone.0256460

PubMed Abstract | Crossref Full Text | Google Scholar

32. Maiwald CA, Neuberger P, Franz AR, Engel C, Michel J, Esser M, et al. Catheter insertion depths in less-invasive surfactant administration. Arch Dis Child Fetal Neonatal Ed. (2022) 107(2):222–4. doi: 10.1136/archdischild-2021-321611

PubMed Abstract | Crossref Full Text | Google Scholar

33. Abdel-Latif ME, Davis PG, Wheeler KI, De Paoli AG, Dargaville PA. Surfactant therapy via thin catheter in preterm infants with or at risk of respiratory distress syndrome. Cochrane Database Syst Rev. (2021) 5(5):Cd011672. doi: 10.1002/14651858.CD011672.pub2

PubMed Abstract | Crossref Full Text | Google Scholar

34. Kribs A, Roll C, Göpel W, Wieg C, Groneck P, Laux R, et al. Nonintubated surfactant application vs conventional therapy in extremely preterm infants: a randomized clinical trial. JAMA Pediatr. (2015) 169(8):723–30. doi: 10.1001/jamapediatrics.2015.0504

PubMed Abstract | Crossref Full Text | Google Scholar

35. Rahtu M, Frerichs I, Waldmann AD, Strodthoff C, Becher T, Bayford R, et al. Early recognition of pneumothorax in neonatal respiratory distress syndrome with electrical impedance tomography. Am J Respir Crit Care Med. (2019) 200(8):1060–1. doi: 10.1164/rccm.201810-1999IM

PubMed Abstract | Crossref Full Text | Google Scholar

36. Vidal Melo MF. Clinical respiratory physiology of the neonate and infant with congenital heart disease. Int Anesthesiol Clin. (2004) 42(4):29–43. doi: 10.1097/00004311-200404240-00005

PubMed Abstract | Crossref Full Text | Google Scholar

37. Bohlin K, Bouhafs RKL, Jarstrand C, Curstedt T, Blennow M, Robertson B. Spontaneous breathing or mechanical ventilation alters lung compliance and tissue association of exogenous surfactant in preterm newborn rabbits. Pediatr Res. (2005) 57(5):624–30. doi: 10.1203/01.PDR.0000156502.84909.BC

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: less invasive surfactant administration (LISA), minimal invasive surfactant therapy (MIST), respiratory distress syndrome, ELGANS, respiratory physiology

Citation: Gunatilaka CC, Xiao Q, Bates AJ, Franz AR, Poets CF and Maiwald CA (2024) Influence of catheter thickness on respiratory physiology during less invasive surfactant administration in extremely preterm infants. Front. Pediatr. 12:1352784. doi: 10.3389/fped.2024.1352784

Received: 8 December 2023; Accepted: 2 September 2024;
Published: 17 September 2024.

Edited by:

Seung Han Shin, Seoul National University Children's Hospital, Republic of Korea

Reviewed by:

Scott Guthrie, Vanderbilt University Medical Center, United States
Theo Dassios, King's College Hospital NHS Foundation Trust, United Kingdom

Copyright: © 2024 Gunatilaka, Xiao, Bates, Franz, Poets and Maiwald. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Christian A. Maiwald, christian.maiwald@med.uni-tuebingen.de

ORCID:
Chamindu C. Gunatilaka
orcid.org/0000-0003-2713-6533
Qiwei Xiao
orcid.org/0000-0002-8276-1473
Alister J. Bates
orcid.org/0000-0002-8855-3448
Axel R. Franz
orcid.org/0000-0001-5066-0857
Christian F. Poets
orcid.org/0000-0002-1072-0066
Christian A. Maiwald
orcid.org/0000-0002-2240-6488

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.