Feasibility of fast, four-dimensional computed tomography-based O-ring LINAC plans for lung stereotactic body radiotherapy in patients with poor performance status

Eun Jeong Heo,Eun Jeong Heo1,2Minseok KimMinseok Kim3Chun Gun ParkChun Gun Park4Kyung Hwan ChangKyung Hwan Chang5Kwang Hyeon KimKwang Hyeon Kim6Jang Bo ShimJang Bo Shim7Young Je ParkYoung Je Park1Chul Yong KimChul Yong Kim1Nam Kwon Lee*Nam Kwon Lee1*Suk Lee*Suk Lee1*
  • 1Department of Radiation Oncology, College of Medicine, Korea University, Seoul, Republic of Korea
  • 2Department of Medical Physics, Graduate School of Korea University, Sejong, Republic of Korea
  • 3Department of Biostatistics and Computing, Yonsei University Graduate School, Seoul, Republic of Korea
  • 4Department of Mathematics, Kyonggi University, Suwon-si, Gyeonggi, Republic of Korea
  • 5Department of Radiologic Science, Far East University, Eumseong-gun, Chungcheongbuk-do, Republic of Korea
  • 6Department of Neurosurgery, Ilsan Paik Hospital, College of Medicine, Inje University, Goyang-si, Gyeonggi-do, Republic of Korea
  • 7Department of Radiation Oncology, Guro Hospital, Korea University Medical Center, Seoul, Republic of Korea

Purpose: We aimed to retrospectively analyzed the feasibility of fast four-dimensional computed tomography (4DCT)-based O-ring LINAC treatment for patients with an average respiratory amplitude was< 0.5 cm and who cannot endure long treatment times due to poor performance status in lung 4D-stereotactic body radiotherapy (SBRT).

Methods: This study included data of 38 patients who received lung 4D-SBRT and had average respiratory amplitude< 0.5 cm in the full phase. C-arm LINAC plans were based on 4DCT data obtained at phase values ranging from 20–70% using a C-arm LINAC. O-ring LINAC plans were retrospectively established based on 4DCT data obtained at phase values of 0–90% using an O-ring LINAC. The conformity index (CI), homogeneity index (HI), and gradient measurement of the planning target volumes (PTV) were analyzed to compare dosimetric data between C-arm LINAC and O-ring LINAC plans. Organs at risk were analyzed in accordance with the Radiation Therapy Oncology Group 0915 protocol. Treatment delivery time and total monitor units were analyzed to compare the efficiency of treatment delivery. Statistical comparisons were performed using the Wilcoxon signed-rank test (P< 0.05).

Results: For the PTV, there was no significant difference in the CI or HI between C-arm LINAC and O-ring LINAC plans. For organs-at-risk, all plans met the criteria for dose constraint. There was a significant difference between C-arm LINAC and O-ring LINAC plans except in the spinal cord. Treatment delivery time was 92% longer for C-arm LINAC plans than for O-ring LINAC plans. The total MU value for C-arm LINAC plans was 9.6% higher than that for O-ring LINAC plans.

Conclusion: We verified the feasibility of fast 4DCT-based O-ring LINAC treatment for patients with average respiratory amplitude< 0.5 cm and who cannot endure long treatment times due to poor performance status in lung 4D-SBRT.

1 Introduction

Stereotactic body radiotherapy (SBRT) is a targeted treatment for early-stage, medically inoperable non-small-cell lung cancer (stages I and II) (1). The Radiation Therapy Oncology Group (RTOG) reported a 3-year overall survival rate of 55% and a local control rate of over 90% for patients with medically inoperable stage I disease undergoing lung SBRT (2). Unlike conventional radiation treatment, SBRT can achieve a highly biologically effective dose by delivering large doses to well-defined targets in small fractions (3). Accordingly, SBRT has been reported to exhibit excellent biological effectiveness in terms of local tumor control and acceptable levels of late complications (4). To optimize outcomes, SBRT must be precisely localized to the target, the conformation of the target must be verified, and the dose fall-off outside the target region must be determined. In addition, calculations must accurately and reproducibly account for the movement of the target due to organ movement, such as those involved in respiration or related to patient positioning (5).

Some studies reported that when the dosimetric differences between non-gating and specific phase-based gating plans were analyzed after dividing according to the criteria of specific tumor motion, the larger the tumor motion, the more dosimetric benefits for organs-at-risk (OARs) (6, 7). Heo et al. reported that gating plans were dosimetric benefits compared with non-gating plans when the average respiratory amplitude in phase 20–70% was less than 0.5 cm at the time (8). However, Fox et al. reported that the treatment delivery efficiency of specific phase-based gating plans was lower than that of non-gating plans (9).

As patients with poor performance status cannot endure long treatment times, leading to unplanned and prolonged radiation treatment or discontinuation of treatment. Acute radiation-related toxicities such as dyspnea, cough, chest pain, and pneumonitis can occur in lung four-dimensional (4D)-SBRT. Acute radiation-related toxicity can also lead to unplanned treatment breaks. Moreover, prolonged radiation treatment appears to negatively affect survival for patients. The median overall survival rates were significantly worse for patients with prolonged radiation treatment time than in those with standard radiation treatment time (18.6 vs. 22.7 months, P< 0.0001) (10).

In the present study, we aimed to retrospectively verify the feasibility of fast four-dimensional computed tomography (4DCT)-based O-ring linear accelerator (LINAC) treatment in patients with average respiratory amplitude< 0.5 cm and who are unable to endure long treatment times due to poor performance status by comparing dosimetric differences and the efficiency of treatment delivery between C-arm LINAC and O-ring LINAC plans in lung 4D-SBRT.

2 Materials and methods

2.1 Patient characteristics

The current study was approved by the Institutional Review Board (IRB No. ED17317) at Korea University Anam Hospital in Seoul, Korea. According to Heo et al., we selected the patients if they presented an average respiratory amplitude of 0.5 cm or less (8). A total of 38 patients with lung cancer who completed C-arm LINAC-based lung 4D-SBRT were included in this study. The average respiratory amplitude was 0.5 cm (0.03–1.87 cm) at phase values of 0–90% and 0.34 cm (0.03–0.91 cm) at phase values of 20–70% when obtained from the respiratory amplitude distribution in the contouring mode of the Eclipse TPS (version 15.6; Varian Medical Systems, Palo Alto, CA, USA). Planning target volumes (PTVs) were obtained from the dose-volume histograms (DVHs) for each patient. The average PTV was 42.62 cc (11.10–115.10 cc) at phase values of 0–90% and 37.48 cc (9.00–103.00 cc) at phase values of 20–70%.

The mean patient age was 73 years (52–86); 25 were men (65.8%) and 13 were women (34.2%) (Table 1). The numbers of patients with American Joint Committee on Cancer (AJCC) stages IA, IB, IIB, and lung metastasis were 16 (42.1%), 2 (5.2%), 2 (5.3%), and 18 (47.4%), respectively. Tumor locations in each patient were noted as follows: 6 (15.8%) in the left lower lobe (LLL), 14 (36.8%) in the left upper lobe (LUL), 11 (28.9%) in the right lower lobe (RLL), 1 (2.8%) in the right middle lobe (RML), and 6 (15.8%) in the right upper lobe (RUL).

TABLE 1
www.frontiersin.org

Table 1 Patient characteristics.

2.2 4DCT acquisition and target definition

The 4DCT images were acquired using a Big Bore CT scanner (Philips Healthcare, Cleveland, OH, USA) and sorted into 10 respiratory phases labeled from 0 to 90%. For each patient, the target and organs (ipsilateral lung, contralateral lung, heart, spinal cord, and esophagus) were contoured in the 50% phase and deformably propagated across phases. Target organs and OARs were delineated in accordance with our standard protocol.

2.3 Planning conditions for C-arm LINAC plans

C-arm LINAC plans were based on 4DCT data obtained at phase values ranging from 20 to 70% using a C-arm LINAC (VitalBeam®, Varian Medical Systems, Palo Alto, CA, USA) in Eclipse TPS. C-arm LINAC plans relied on a Millennium 120 MLC with a leaf thickness of 5 mm and a 6 MV flattening-filter-free (FFF) beam. For the volumetric modulated arc therapy (VMAT) technique, the gantry was rotated at an angle of 0–180° by a 2-half arc. The prescription dose of 4,800 cGy was delivered in 4 fractions and optimized such that 95–105% of the prescription dose was delivered to the target. All plans were calculated using the advanced AurosXB system (Varian Eclipse TPS, version 15.6) with heterogeneity correction using the Photon Optimizer (PO) MLC algorithm. Dose constraints based on the RTOG 0915 protocol were applied for the ipsilateral lung, contralateral lung, heart, spinal cord, and esophagus for OARs (ipsilateral, contralateral lung: D1000cc< 12.4 Gy, D1500cc< 11.6 Gy, heart: Dmax< 34.0 Gy, V28Gy< 15 cc, spinal cord: Dmax< 26 Gy, V13.6Gy< 1.2 cc, V20.8Gy< 0.35 cc, esophagus: Dmax< 30 Gy, V18.8Gy< 5 cc) (1).

2.4 Planning conditions for O-ring LINAC plans

O-ring LINAC plans were retrospectively established based on 4DCT data for phase values of 0–90% using an O-ring LINAC. O-ring LINAC plans used a dual-layer MLC with a leaf thickness of 5 mm. O-ring LINAC has two times faster MLC speed (5.0 cm/s vs. 2.5 cm/s), and four times faster gantry speed (4 RPM vs. 1 RPM) compared with C-arm LINAC (Table 2).

TABLE 2
www.frontiersin.org

Table 2 Treatment delivery efficiency parameters of O-ring LINAC and C-arm LINAC.

2.5 Analysis of dosimetric differences

The dosimetric differences between C-arm LINAC and O-ring LINAC plans were analyzed using dosimetric parameters (Table 3). PTVs were compared using the conformity index (CI), homogeneity index (HI), and gradient measurement (GM) (1315). Notably, because the dose gradient should be sharp in all directions around the target, the dose fall-off must be evaluated to ensure precise delivery of the high dose to the tumor and minimize damage to OARs. The GM, one of the dosimetric parameters used to evaluate dose fall-off, was calculated as the difference in radius between the 50% prescription isodose volume and the 100% prescription isodose volume, which reflects intermediate dose spillage (15). OARs were analyzed by calculating D1000cc and D1500cc for the ipsilateral and contralateral lungs; Dmax and V28Gy for the heart; Dmax, V13.6Gy, and V20.8Gy for the spinal cord; and Dmax and V18.8Gy for the esophagus in accordance with RTOG 0915 methods (1).

TABLE 3
www.frontiersin.org

Table 3 Dosimetric parameters used in this study, with corresponding formula in terms of PTV.

The necessary parameters were extracted for dose analysis by creating a database using Python (version 3.9.6) after exporting DVH data for the 38 patients from the Eclipse TPS. The ipsilateral and contralateral lungs were defined based on the tumor location. When the tumor location was the right lung, the right lung was defined as the ipsilateral lung, while the left lung was defined as the contralateral lung. The mean ± standard deviation (SD) was used to analyze dosimetric differences between C-arm LINAC and O-ring LINAC plans.

2.6 Comparison of treatment delivery efficiency

Treatment delivery time and total monitor unit (MU) values obtained from the Eclipse TPS system QA summary were analyzed to compare the efficiency of treatment delivery between C-arm LINAC and O-ring LINAC plans. The treatment delivery time of O-ring LINAC plans was recorded as beam-on time (Figure 1A). The treatment delivery time of the C-arm LINAC plan was recorded as the sum of the gated treatment time, beam-on time, and beam-off time (Figure 1B). MU was defined as the sum of MU1 of Arc 1 and MU2 of Arc 2. The kV CBCT acquisition times were 16.6 s and 60 s when O-ring LINAC and C-arm LINAC were used, respectively. The kV CBCT acquisition time was excluded from the treatment delivery time.

FIGURE 1
www.frontiersin.org

Figure 1 Timeline for the acquisition of treatment delivery efficiency for 38 patients undergoing lung SBRT. Workflow of integrated kV CBCT acquisition and treatment delivery time for (A) C-arm LINAC and (B) O-ring LINAC plans. Gated treatment time: beam-off time to get a respiratory signal. Beam-on time: time reaching below the threshold of the predefined phase. Beam-off time: time reaching above the threshold of the predefined phase. SBRT, stereotactic body radiotherapy; CBCT, cone beam computed tomography.

2.6.1 Statistical analysis

Dosimetric differences and the efficiency of treatment delivery (mean ± standard deviation) for the O-ring LINAC and C-arm LINAC plans were compared using Wilcoxon signed-rank tests (P< 0.05). Statistical analyses were performed using Statistical Package for the Social Sciences software ver. 26.0 (IBM, Armonk, NY, USA).

3 Results

3.1 Analysis of dosimetric differences

3.1.1 Comparison of isodose distribution and average DVH

The isodose distributions in the axial, coronal, and sagittal planes were compared between the C-arm LINAC and O-ring LINAC plans (Figure 2). When the coronal and sagittal planes of the C-arm LINAC plans were compared with those of the O-ring LINAC plans, the 50% isodose line became smaller as the PTV volume decreased. The average DVH for target coverage and doses delivered to OARs were also compared between O-ring LINAC and C-arm LINAC plans (Figure 3). Both plans resulted in similar target coverage and met RTOG 0915-based dose constraints for OARs.

FIGURE 2
www.frontiersin.org

Figure 2 Dose distributions for a representative case (Patient 1) in the comparison between C-arm LINAC and O-ring LINAC plans for lung SBRT. Representative isodose distributions of the C-arm LINAC plan in the (A) axial, (B) coronal, and (C) sagittal planes. Representative isodose distributions of the O-ring LINAC plan in the (D) axial, (E) coronal (F) sagittal planes. Representative GM values are shown for the O-ring LINAC and C-arm LINAC plans. The isodose lines represent 105% (red), 100% (orange), 95% (cyan), 90% (blue), 50% (green), and 20% (dark blue). SBRT, stereotactic body radiotherapy; GM, gradient measurement.

FIGURE 3
www.frontiersin.org

Figure 3 Average dose-volume histogram (DVH) for the planning target volume (PTV) and organs-at-risk (OARs). The solid line and shaded bands represent the average and minimum-maximum range of the DVHs across all 38 patients who received SBRT.

3.1.2 Comparison of dosimetric parameters

In the analysis of PTV and OARs volumes, significant differences were noted between the C-arm LINAC and O-ring LINAC plans (all P< 0.05) (Table 4). The PTV for C-arm LINAC plans was 12.1% smaller than that for O-ring LINAC plans (37.484 ± 22.544 cc, 42.621 ± 24.867 cc, P< 0.001). The ipsilateral and contralateral lung volumes were 6.0% and 5.6% lower for C-arm LINAC plans than for O-ring LINAC plans, respectively (P< 0.001). The heart, esophagus, and spinal cord volumes were 4.6%, 8.9%, and 2.6% lower for C-arm LINAC plans than for O-ring LINAC plans, respectively.

TABLE 4
www.frontiersin.org

Table 4 Dosimetric differences between C-arm LINAC and O-ring LINAC plans.

In the analysis of PTVs, there were no statistically significant differences between the CI and HI (all P > 0.05). However, the difference in radius between the 50% isodose and 100% isodose lines for C-arm LINAC plans was reduced when compared with that for O-ring LINAC plans. In addition, the GM was 9.6% lower for C-arm LINAC plans than for O-ring LINAC plans (1.381 ± 0.222 cm and 1.528 ± 0.206 cm, P< 0.001) (Table 4, Figure 3). For OARs, all plans met the criteria for dose constraints based on the RTOG 0915 protocol (Table 4). Within the dose constraint criteria, there were no significant differences except in the spinal cord. The D1000cc and D1500cc values for the ipsilateral lung for C-arm LINAC plans were reduced by 31.6% and 59.9% when compared with those for O-ring LINAC plans, respectively (D1000cc: 34.755 ± 102.480 cGy, 50.780 ± 132.061 cGy, P< 0.001, D1500cc: 2.666 ± 7.926 cGy, 6.641 ± 14.598 cGy, P< 0.001). The D1000cc and D1500cc values for the contralateral lung for C-arm LINAC plans were reduced by 33.1% and 32.2%, respectively, when compared with those for O-ring LINAC plans (D1000cc: 20.671 ± 37.433 cGy, 30.902 ± 50.234 cGy, P< 0.001, D1500cc: 6.979 ± 13.507 cGy, 10.299 ± 18.258 cGy, P< 0.001). The Dmax and V20Gy values for the heart for C-arm LINAC plans were 7.1% and 28.5% lower than those for O-ring LINAC plans, respectively. The Dmax and V20Gy values for the esophagus for C-arm LINAC plans were reduced by 6.9% and 33.9%, respectively, when compared with those for O-ring LINAC plans. The Dmax, V13.6Gy, and V20.8Gy values for the spinal cord for C-arm LINAC plans were reduced by 2.5%, 31.3%, and 0.0%, respectively, when compared with those for the O-ring LINAC plans; however, these differences were not significant.

3.2 Comparison of treatment delivery efficiency

For treatment delivery efficiency, there were significant differences between the C-arm LINAC and O-ring LINAC plans (all P< 0.05) (Figure 4). Treatment delivery time was 92% longer for C-arm LINAC than for O-ring LINAC plans (9.6 ± 3.3 min, 5.0 ± 0.4 min, P = 0.043). The total MU value was 9.6% higher for C-arm LINAC plans than for O-ring LINAC plans (4016.5 ± 236.9, 3665.8 ± 230.7, P = 0.039).

FIGURE 4
www.frontiersin.org

Figure 4 Comparison of treatment delivery time between C-arm LINAC and O-ring LINAC plans.

4 Discussion

In this study, we tried to retrospectively verify that the feasibility of fast 4DCT-based O-ring LINAC treatment for patients with an average respiratory amplitude< 0.5 cm and who cannot endure long treatment times due to poor performance status in lung 4D-SBRT.

In our study, patients enrolled in this study were selected based on the average respiratory amplitude is< 0.5 cm in the full phase by Heo et al. They analyzed dosimetric differences between plan0-90%, plan20-70%, and plan40-60% in 40 patients treated with lung 4D-SBRT. Compared with plan0-90%, D1000cc of plan20-70%, and plan40-60% were reduced for ipsilateral lung (1.36 ± 1.58 Gy, 0.98 ± 0.96 Gy, 0.77 ± 0.79 Gy, P-value< 0.05). However, all plans met the criteria for dose constraints based on the RTOG protocol. The average respiratory amplitude in phase 20–70% was< 0.5 cm at the time (8). In addition, in our study, we included patients treated C-arm LINAC plans using C-arm LINAC, and we retrospectively established O-ring LINAC plans using a large-scale database (38 patients). A larger sample size is needed to verify statistically significant differences. Pokhrel et al. reported when compared dosimetric differences between non-gating plans in patients treated with lung SBRT, clinical follow-up results are considered essential (16).

Some studies compared dosimetric differences between specific phase-based gating and free-breathing-based non-gating plans that divided two groups as 0.5 cm< tumor motion< 1 cm and tumor motion > 1 cm. Prunaretty et al. reported that when the motion was > 1 cm, the HI of PTV was 6.67% higher for gating plans than for free-breathing plans. When the tumor motion was< 1 cm, the HI of PTV was 16.67% higher for gating plans than for free-breathing plans. In addition, when the tumor motion was > 1 cm, the V20Gy values of ipsilateral lung were 49.29% lower for gating plans than for free-breathing plans. When the tumor motion was< 1 cm, the V20Gy of ipsilateral lung were 21.18% lower than for free-breathing plans (7). All plans met the ROSEL constraints (17). Jang et al. reported that when the tumor motion was > 1 cm, the V20Gy values of the ipsilateral lung were 0.41% and 1.11% lower for gatingGW50% (5 phases) and gatingGW25% (3 phases) plans, respectively, when compared with non-gated plans. When the tumor motion was< 1 cm, the V20Gy values of the ipsilateral lung were 0.31% and 0.42% lower for gatingGW50% and gatingGW25% plans, respectively, when compared with non-gated plans. However, in our study, patients with average respiratory amplitude< 0.5 cm were selected. The average respiratory amplitude was 0.51 cm (0.03–1.87 cm) at phase values of 0–90% and 0.34 cm (0.03–0.91 cm) at phase values of 20–70%, respectively. For PTV, the CI and HI were 0.15% and 0% higher for C-arm LINAC plans than for O-ring LINAC plans, respectively. As in previous studies, there were no statistically significant differences between the CI and HI (all P > 0.05). For OARs, the D1000cc and D1500cc values of the ipsilateral lung were also 31.6% and 59.9% lower for C-arm LINAC plans than for O-ring LINAC plans, respectively. In addition, there were significant differences between the plans (all P< 0.001). However, all plans met the criteria for dose constraints based on the RTOG 0915 protocol.

Fox et al. analyzed the factors affecting the treatment delivery time during gating plans established based on specific phase in 15 patients treated with gating plans and 13 patients treated with non-gated plans. Thy reported that the beam-on time of gating plans increased by 5.5 times (range 1.2–12.2) than that of the non-gated treatment (9). Our study found similar results. Treatment delivery time was 92% longer for C-arm LINAC than for O-ring LINAC plans. We verified the feasibility of fast 4DCT-based O-ring LINAC treatment for patients with an average respiratory amplitude< 0.5 cm and who cannot endure long treatment times due to poor performance status in lung 4D-SBRT.

Wang et al. reported that tumor motion variation was higher in the upper lobe compared with the lower lobe (10 ± 4.8 mm vs. 2.3 ± 2.2 mm) (18). Therefore, considering the tumor location, the feasibility of a fast respiratory motion management decision support guideline for lung SBRT in patients with poor performance status must be considered in future studies.

5 Conclusion

We verified the feasibility of fast 4DCT-based O-ring LINAC treatment for patients with an average respiratory amplitude< 0.5 cm and who cannot endure long treatment times due to poor performance status in lung 4D-SBRT. Compared with OARs, both plans were acceptable according to the RTOG-0915 protocol, but no significant differences between O-ring LINAC and C-arm LINAC plans. However, treatment delivery time was 92% longer for C-arm LINAC plans than for O-ring LINAC plans.

Data availability statement

The original contributions presented in the study are included in the article/supplementary material. Further inquiries can be directed to the corresponding authors.

Ethics statement

The studies involving humans were approved by Korea University Anam Hospital Institutional Review Board. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.

Author contributions

EH: Conceptualization, Data curation, Formal Analysis, Investigation, Methodology, Software, Validation, Visualization, Writing – original draft. MK: Formal Analysis, Software, Validation, Visualization, Writing – review & editing. CP: Formal Analysis, Software, Validation, Visualization, Writing – review & editing. KC: Writing – review & editing. KK: Writing – review & editing. JS: Writing – review & editing. YP: Writing – review & editing. CK: Writing – review & editing. NL: Writing – review & editing. SL: Conceptualization, Investigation, Supervision, Writing – review & editing, Methodology.

Funding

The author(s) declare financial support was received for the research, authorship, and/or publication of this article. This work was supported by the National Research Foundation of Korea (NRF) grant funded by the Korea government (MSIT) (N0. 2021R1G1A1003584) and Korea University (No. K1925071). The funder was not involved in the study design, collection, analysis, interpretation of data, the writing of this article or the decision to submit it for publication.

Acknowledgments

We would like to thank Editage (www.editage.co.kr) for English language editing.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Videtic G, Hu C, Singh AK, Chang JY, Parker W, Olivier KR, et al. A Randomized Phase 2 Study Comparing 2 Stereotactic Body Radiation Therapy Schedules for Medically Inoperable Patients With Stage I Peripheral Non-Small Cell Lung Cancer: NRG Oncology RTOG 0915 (NCCTG N0927). Int J Radiat Oncol Biol Phys (2015) 93(4):757–64. doi: 10.1016/j.ijrobp.2015.07.2260

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Timmerman RD, Paulus R, Paulus HI, Gore EM, Edelman MJ, Galvin J, et al. Stereotactic body radiation therapy for operable early-stage lung cancer: findings from the NRG oncology RTOG 0618 trial. JAMA Oncol (2018) 4:1263–6. doi: 10.1001/jamaoncol.2018.1251

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Followill D, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, et al. Stereotactic body radiation therapy: the report of AAPM Task Group 101. Med Phys (2010) 37(8):4078–101. doi: 10.1118/1.3438081

PubMed Abstract | CrossRef Full Text | Google Scholar

4. i Garau MM. Radiobiology of stereotactic body radiation therapy (SBRT). Rep Pract Oncol Radiother (2017) 22:86–95. doi: 10.1016/j.rpor.2017.02.010

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Timmerman R, Papiez L, Suntharalingam M. Extracranial stereotactic radiation delivery: expansion of technology beyond the brain. Technol Cancer Res Treat (2003) 2(2):153–60. doi: 10.1177/153303460300200212

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Jang SS, Huh GJ, Park SY, Yang PS, Cho EY. The impact of respiratory gating on lung dosimetry in stereotactic body radiotherapy for lung cancer. Phys Med (2014) 30:682–9. doi: 10.1016/j.ejmp.2014.05.005

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Prunaretty J, Boisselier P, Aillères N, Riou O, Simeon S, Bedos L, et al. Tracking, gating, free-breathing, which technique to use for lung stereotactic treatments? A dosimetric comparison. Rep Pract Oncol Radiother (2019) 24:97–104. doi: 10.1016/j.rpor.2018.11.003

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Heo EJ, Kim MS, Park CG, Kim KH, Shim JB, Hwang H, et al. Determination of optimized phase through correlation between 4DCT analysis indices in lung 4D SBRT. Int J Radiat Oncol Biol Phys (2021) 111(3):e541. doi: 10.1016/j.ijrobp.2021.07.1472

CrossRef Full Text | Google Scholar

9. Fox T, Simon EL, Elder E, Riffenburgh RH, Johnstone PA, et al. Free breathing gated delivery (FBGD) of lung radiation therapy: analysis of factors affecting clinical patient throughput. Lung Cancer (2007) 56:69–75. doi: 10.1016/j.lungcan.2006.11.019

PubMed Abstract | CrossRef Full Text | Google Scholar

10. McMillan MT, Ojerholm E, Verma V, Higgins KA, Singhal S, Predina JD, et al. Radiation treatment time and overall survival in locally advanced non-small cell lung cancer. Int J Radiat Oncol Biol Phys (2017) 98:1142–52. doi: 10.1016/j.ijrobp.2017.04.004

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Varian Medical Systems. Halcyon physics, V.M. Systems. Palo Alto (2019).

Google Scholar

12. Varian Medical Systems. Specifications, V.M. Systems. Palo Alto (2015).

Google Scholar

13. Wilke L, Andratschke N, Blanck O, Brunner TB, Combs SE, Grosu AL, et al. ICRU report 91 on prescribing, recording, and reporting of stereotactic treatments with small photon beams. Strahlenther Onkol (2019) 195(3):193–8. doi: 10.1007/s00066-018-1416-x

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Hodapp N. The ICRU Report 83: prescribing, recording and reporting photon-beam intensity-modulated radiation therapy (IMRT). Strahlentherapie und Onkologie: Organ der Deutschen Rontgengesellschaft (2012) 188:97–9. doi: 10.1007/s00066-011-0015-x

CrossRef Full Text | Google Scholar

15. Duan Y, Lin Y, Wang H, Kang B, Feng A, Ma K, et al. How does the gradient measure of the lung SBRT treatment plan depend on the tumor volume and shape? Front Oncol (2021) 11. doi: 10.3389/fonc.2021.781302

CrossRef Full Text | Google Scholar

16. Pokhrel D, Visak J, Critchfield LC, Stephen J, Bernard ME, Randall M, et al. Clinical validation of ring-mounted halcyon linac for lung SBRT: comparison to SBRT-dedicated C-arm linac treatments. J Appl Clin Med Phys (2021) 22:261–70. doi: 10.1002/acm2.13146

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Hurkmans CW, Cuijpers JP, Lagerwaard FJ, Widder J, van der Heide UA, Schuring D, et al. Recommendations for implementing stereotactic radiotherapy in peripheral stage IA non-small cell lung cancer: report from the Quality Assurance Working Party of the randomised phase III ROSEL study. Radiat Oncol (2009) 4:1–14. doi: 10.1186/1748-717X-4-1

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Wang Y, Bao Y, Zhang L, Fan W, He H, Sun ZW, et al. Assessment of respiration-induced motion and its impact on treatment outcome for lung cancer. Biomed Res Int (2013) 2013:872739. doi: 10.1155/2013/872739

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: fast treatment, O-ring LINAC, poor performance status, 4D-SBRT, lung cancer

Citation: Heo EJ, Kim M, Park CG, Chang KH, Kim KH, Shim JB, Park YJ, Kim CY, Lee NK and Lee S (2023) Feasibility of fast, four-dimensional computed tomography-based O-ring LINAC plans for lung stereotactic body radiotherapy in patients with poor performance status. Front. Oncol. 13:1270677. doi: 10.3389/fonc.2023.1270677

Received: 01 August 2023; Accepted: 25 October 2023;
Published: 20 November 2023.

Edited by:

Guang Li, Memorial Sloan Kettering Cancer Center, United States

Reviewed by:

Vanessa Figlia, ARNAS Ospedali Civico Di Cristina Benfratelli, Italy
Ming Hsiang Liou, Yuan’s General Hospital, Taiwan

Copyright © 2023 Heo, Kim, Park, Chang, Kim, Shim, Park, Kim, Lee and Lee. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Nam Kwon Lee, nklee74@korea.ac.kr; Suk Lee, sukmp@korea.ac.kr

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.