ORIGINAL RESEARCH article

Front. Oncol., 28 April 2023
Sec. Radiation Oncology
This article is part of the Research Topic LINAC-Based Stereotactic Radiosurgery in Daily Practice View all 8 articles

Feasibility and dosimetric evaluation of single- and multi-isocentre stereotactic body radiation therapy for multiple liver metastases

Chunbo TangChunbo Tang1Changfei Gong*Changfei Gong2*Biaoshui LiuBiaoshui Liu3Hailiang GuoHailiang Guo1Zhongyang DaiZhongyang Dai1Jun YuanJun Yuan1Xiaoping WangXiaoping Wang2Yun Zhang*Yun Zhang2*
  • 1Department of Oncology, First Affiliated Hospital of Gannan Medical University, Ganzhou, China
  • 2Department of Radiation Oncology, Jiangxi Cancer Hospital, Nanchang, China
  • 3Department of Radiation Oncology, Sun Yat-sen University Cancer Center, Guangzhou, China

Objectives: Single-isocentre volumetric-modulated arc therapy (VMAT) stereotactic body radiation therapy (SBRT) improves treatment efficiency and patient compliance for patients with multiple liver metastases (MLM). However, the potential increase in dose spillage to normal liver tissue using a single-isocentre technique has not yet been studied. We comprehensively evaluated the quality of single- and multi-isocentre VMAT-SBRT for MLM and propose a RapidPlan-based automatic planning (AP) approach for MLM SBRT.

Methods: A total of 30 patients with MLM (two or three lesions) were selected for this retrospective study. We manually replanned all patients treated with MLM SBRT by using the single-isocentre (MUS) and multi-isocentre (MUM) techniques. Then, we randomly selected 20 MUS and MUM plans for training to generate the single-isocentre RapidPlan model (RPS) and the multi-isocentre RapidPlan model (RPM). Finally, we used data from the remaining 10 patients to validate RPS and RPM.

Results: Compared with MUS, MUM reduced the mean dose delivered to the right kidney by 0.3 Gy. The mean liver dose (MLD) was 2.3 Gy higher for MUS compared with MUM. However, the monitor units, delivery time, and V20Gy of normal liver (liver-gross tumour volume) for MUM were significantly higher than for MUS. Based on validation, RPS and RPM slightly improved the MLD, V20Gy, normal tissue complications, and dose sparing to the right and left kidneys and spinal cord compared with manual plans (MUS vs RPS and MUM vs RPM), but RPS and RPM significantly increased monitor units and delivery time.

Conclusions: The single-isocentre VMAT-SBRT approach could be used for MLM to reduce treatment time and patient comfort at the cost of a small increase in the MLD. Compared with the manual plans, RapidPlan-based plans, especially RPS, have slightly improved quality.

Introduction

Malignant tumors of the liver include primary hepatobiliary cancers and metastatic tumors. Liver metastases usually originate from colorectal cancer, lung cancer, pancreatic cancer, adenocarcinoma, breast cancer, and other malignant tumors (1). Surgical resection of liver metastases remains the primary treatment option. Unfortunately, only a small number of patients are suitable for surgery because of insufficient functional liver reserve and medical complications (2, 3). Alternative options include stereotactic body radiotherapy (SBRT), radiofrequency ablation, chemoembolization, and radioembolization. These techniques are promising in a considerable number of patients with liver metastases (4). SBRT is defined as an effective, non-invasive, and highly accurate hypo-fractionated radiotherapy technique that has been prospectively shown to provide a good local control rate (5). Several studies have studied the feasibility, safety, and clinical outcomes of SBRT for primary liver cancer and metastases (612).

Over the past few years, researchers have increasingly focused on the effectiveness of SBRT in treating systemic metastatic tumors and have analyzed the effect of dose distribution on tumor targets and organs at risk (OARs). Owing to the technical challenges inherent in treating multiple lesions as their spatial separation decreases, Hallaq et al. (13) investigated the rationale of technical requirements of SBRT for multiple metastases. Wang et al. (14) and Ruggieri et al. (15) conducted studies on brain metastasis with SBRT. They showed that extending the treatment time would increase the position error, resulting in a decrease in the accuracy of dose delivery. Based on a multi-institution study, Rusthoven et al. (16) demonstrated that high-dose liver SBRT is safe and effective for treating patients within three hepatic metastases. Clark et al. (17) evaluated the plan quality of single-isocentre versus multi-isocentre volumetric modulated arc therapy (VMAT) for multiple central nervous system metastases. The preliminary results indicated the delivery time of a single-isocentre was less than half of the multi-isocentre while maintaining comparable planning quality. Recently, with advances in optimization modalities and delivery techniques, dosimetric performance and therapeutic efficiency have been greatly improved for liver SBRT. By comparing the treatment plan quality, robustness, and plan complexity of robust optimization and the planning target volume (PTV)-based optimized plans, Miura et al. (18) found that robust optimization provides stable target coverage with shifted locations and helps to slightly reduce plan complexity for liver SBRT. Thaper et al. (19) studied whether the dynamic conformal arc integrated with the segment shape optimization and variable dose rate was superior to the classic VMAT. To harmonize liver SBRT practice and to fill the knowledge gap concerning the inter-system and inter-user differences for treatment techniques and treatment planning systems (TPS), Moustakis et al. (20) performed a meaningful study of planning benchmark for SBRT of liver metastases and provided the best practice guidelines for users. Although the aforementioned studies have studied the feasibility, safety and efficiency of liver SBRT, and have potential benefits in terms of solving some issues, to the best of our knowledge, none of the studies that investigated the dosimetric impact of single-isocentre SBRT for multiple liver metastases (MLM), and the potential increase in dose spillage to the normal liver tissue using a single-isocentre technique has not yet been studied. In this work, we aimed to investigate whether a single-isocentre technique leads to increased normal liver dose compared to a conventional multi-isocentre technique for MLM SBRT.

Technological advances in radiotherapy over the past decade have enabled the creation and delivery of smarter plans (21). Deliberately mimicking the behavior of experienced planners, automatic planning (AP) algorithms, including Auto-Planning (Philips Medical Systems, Best, The Netherlands) (22), RapidPlan (Varian Medical Systems, Palo Alto, CA) (23), and Multi-Criteria Optimization (RaySearch Laboratories, Stockholm, Sweden) (24), have been developed to accelerate the treatment planning process and drastically improve the planning efficiency. In particular, the knowledge‐based (KB) RapidPlan adopts machine learning or statistical methods to extract historical planning information and establishes a prediction model to estimate the expected dose-volume histograms (DVHs) for new patients (2527). Many studies have emerged on the application of RapidPlan to various anatomical locations (2830). Only two previous studies have evaluated the benefit of RapidPlan technology in the treatment of liver cancer. Gang et al. (31) constructed a special RapidPlan based on the distance between the right kidney and the PTV, as well as a general KB model and compared the prediction ability of the two models. Antonella Fogliata et al. (32) built another general KB model; the results showed the optimization engine can produce a clinically acceptable plan for liver cancer. However, there has yet to be a RapidPlan model to automate treatment planning for multi-lesion liver SBRT including deploying beam geometry and optimization. Based on these considerations, we also evaluated RapidPlan for MLM SBRT.

More specifically, our major contributions include the following. First, we verified the technical feasibility of single- and multi-isocentre for MLM SBRT and performed the relevant dosimetric evaluation. Second, we extended the RapidPlan model for MLM by incorporating SBRT and compared it with manual planning. Finally, we comprehensively evaluated these four approaches – that is, the manual single-isocentre technique (MUS), the manual multi-isocentre technique (MUM), the RapidPlan single-isocentre technique (RPS), and the RapidPlan multi-isocentre technique (RPM).

Materials and methods

Patients and target delineation

We selected 30 patients with MLM (two or three lesions) treated in our hospital between August 2020 and February 2022 for this retrospective study. All patients were immobilized using a stereotactic body frame with a vacuum fixation cushion to create a reproducible position. Four-dimensional computed tomography (4DCT) (CT0, CT1…CT9) was acquired with 2 mm slice thickness using Siemens Medical Systems and transferred to the Eclipse TPS (Version 15.5). The gross tumor volumes (GTVs) were contoured on 4DCT by one senior radiation oncologist according international guidelines (20, 33). The internal target volumes (ITVs) encompassing the whole respiratory tumor motion areas were generated after extension of GTVs and the final PTV was generated by uniformly expanding ITVs with a 5 mm margin. Critical structures including the normal liver, heart, spinal cord, and kidneys were also contoured. Table 1 lists the detailed clinical characteristics of these patients.

TABLE 1
www.frontiersin.org

Table 1 Characteristics of the 30 patients with MLM enrolled in this study.

Treatment planning

The prescribed dose was 56 Gy in 7 fractions for all patients. The machine optimization settings were as follows: 6 MV FFF beam with dose rate 1400 MU/min, AcurosXB with dose-to-water reporting mode and PO dose calculation algorithms with heterogeneity corrections with 2.0 × 2.0 × 2.0 mm calculation grid size, and the jaw tracking was activated during optimization to reduce MLC transmission or leakage (34). For MUM and RPM, multiple isocentres were placed at the center of each lesion, and 2-4 partial coplanar arcs with 30° collimator angles were utilized to reduce the MLC leakage dose. For MUS and RPS, all lesions were integrated into a single target area with a maximum inter-target distance of less than 15 cm, and then the single isocentre was placed in the geometric center of this target. The optimal setting of two partial coplanar arcs was adopted with a 30° collimator angle. All treatment plans were made by one senior medical physicist to avoid inter-planner variability.

For manual planning, all 30 patients treated with SBRT were retrospectively replanned using MUS and MUM. Based on clinical requirements, experience, and customization, we pre-defined the initial optimization objectives. To ensure appropriate dose coverage and homogeneity, we optimized all plans, giving great weight for the PTV and less weight for OARs based on the as low as reasonably achievable (ALARA) concept (20). Typically, to achieve acceptable SBRT plans, the planner needs to introduce auxiliary structures frequently and adjust the constraints repeatedly. We utilized dose-limiting shells and normal tissue objectives (NTOs) to limit hot spots or to control dose fall-off outside the PTV. Based on previous studies and SBRT/IGRT protocols (20, 33), at least 98% of the PTV received 100% of the prescribed dose, conformity index (CI) less than 1.2 is desirable, D2cm (maximum dose at any point 2 cm away from the PTV margin in any direction) has to be smaller than 50–70%, depending on the PTV size. The maximum dose to the PTV was planned fell inside the GTVs. We also considered vulnerable OARs constraints for evaluation of the mean liver dose (MLD); V20Gy of normal liver (liver-GTVs); dose to 0.5 cc (D0.5cc) of the stomach, esophagus, and small bowel; maximum dose (Dmax) to the planning risk volume (PRV) to the spinal cord (plus a 0.5 cm margin); and mean dose (Dmean) to the left and right kidneys. We performed multiple re-optimizations to meet the final clinical requirements. The settings of optimization goals and priority are in Table S1 (supplementary material).

For RapidPlan, we randomly selected 20 MUS and 20 MUM for training to generate RPS and RPM models, respectively. Figure 1 shows the clinical workflow of RapidPlan modelling and training steps. The RapidPlan optimization component consists of three main parts: the modelling and training engine, the automatic constraint prediction module, and new VMAT/IMRT optimization functions. The main steps for modelling and training were: (1) we selected the clinically accepted cases and created the unified RapidPlan model. (2) We defined the structures, optimization objectives, initial constraints, and priorities of the model. (3) We added each training case to the model and associated the model structure with the prescription and optimization goal of the plan. (4) We extracted the dose and geometric feature information by using the principal component analysis method (35, 36). (5) We trained and finalized the model. For testing, based on the published RPS and RPM models, we optimized 10 RPS and 10 RPM plans that were not included in the training dataset. The main steps of validation were: (1) we selected the corresponding RPS or RPM model. (2) We manually matched the validation plan structure to the model structure. (3) We generated the DVH estimation range of the relevant constraint structures by the sophisticated regression model and automatically generating the dose-volume constraints. (4) We optimized the verification plan.

FIGURE 1
www.frontiersin.org

Figure 1 Schematic of the workflow for RapidPlan modelling.

Plan evaluation and analysis

We calculated and compared the following dosimetric evaluation metrics: Dmean and Dmax (the mean dose and maximum dose to target volume), D2cm, D2, D95, D98, and V100% (Vx represents the volume receiving at least x% of the prescription dose) of the PTV. We evaluated the gradient index (GI) (V50%/V100%) and the CI, and performed DVH analysis for each plan (3740). The lower the GI value, the steeper the dose gradient. We defined the CI as the ratio between the PTV covered by the prescribed dose and the PTV. A CI value closer to 1 indicates better conformity.

We evaluated the MLD and V20Gy for normal liver (liver-GTVs) and analyzed the potential relationships between the MLD, the ITV, and the inter-target distance by linear regression analysis (34, 4144). We assessed the probability of normal tissue complications (NTCPs) defined as the occurrence of radiation-induced hepatitis with liver enzyme changes grade ≥ 2 based on the MLD (EQD2) (42, 45, 46) for all treatment plans. Dmax to the spinal cord and the Dmean to kidneys were also evaluated. Additionally, we calculated the planned monitor units (MUs) and delivery time (beam-on time) to assess efficiency. Wilcoxon’s signed-rank test was carried out to calculate the statistical significance at a 95% confidence interval to compare different techniques and Bonferroni correction was used in order to account for multiple hypothesis testing. A p < 0.05 was considered statistically significant.

Results

For the 30 cases, we generated 80 treatment plans (30 MUS, 30 MUM, 10 RPS and 10 RPM plans) for MLM SBRT. Table 2 shows the results of MUS and MUM for the 30 patients: the two approaches achieved good dosimetry goals and similar target performance. Compared with MUS, MUM had a lower D2cm (28.58 ± 3.85) and reduced the mean dose delivered to the right kidney by 0.3 Gy. The mean dose to the left kidney (1.2 Gy) and the maximum dose to the spinal cord (14.2 Gy) of MUM were similar to MUS. MUS increased the mean dose of MLD by 2.3 Gy compared with MUM. However, the MUs, delivery time, and V20Gy of MUM were significantly higher than those of MUS.

TABLE 2
www.frontiersin.org

Table 2 Quantitative evaluation of MUS and MUM for the 30 patients.

Figure 2 illustrates the transverse, coronal, and sagittal dose distributions for different regions of a representative patient and the corresponding cumulative DVHs for the PTV and OARs is shown in Figure 3. In the visual inspection, the isodose lines are more conformal and constricted to the PTV for RPS and MUS compared with RPM and MUM. Especially for the transverse and sagittal views, there is greater dose spill over as indicated by the red square in Figure 2, which shows a steeper dose gradient and improved conformity for single-isocentre plans (RPS and MUS). The detailed dosimetric differences of the target volume and individual organs are presented as DVHs for this patient in Figure 3.

FIGURE 2
www.frontiersin.org

Figure 2 Comparison of transverse, coronal, and sagittal dose distributions for one example case.

FIGURE 3
www.frontiersin.org

Figure 3 Comparison of DVHs for the PTV and OARs.

Table 3 shows an overview of the numerical findings of MUS, MUM, RPS, and RPM based on average DVH analysis of the PTV and OARs for the 10 patients. There were no significant differences for the PTV metrics except for D2 and Dmax, while the quality of D2cm and the GI for MUS (28.5 ± 2.1) and RPS (28.4± 3.2) was lower than that of MUM and RPM, and RPS (69.0 ± 1.1) had the best hotspot control. RPM had the lowest MLD and it was significantly better than those of RPS and MUS (p < 0.05). On the contrary, V20Gy of liver GTVs for RPM and MUM was larger than that of MUS and RPS. Overall, for the MLD, V20Gy, NTCPs, and dose sparing to the left and right kidneys and the spinal cord, AP plans showed slight improvements compared with manual plans (MUS vs RPS, MUM vs RPM), but the AP plans significantly increased MUs and delivery time. Linear regression showed no significant correlation between the MLD and the inter-target distance (p > 0.05) (Figure 4A). There was a significant correlation between the MLD and the ITV for all approaches (p < 0.05) (Figure 4B).

TABLE 3
www.frontiersin.org

Table 3 Quantitative evaluation on the average, standard deviation, and range of DVH metrics for the PTV and OARs among MUS, MUM, RPS, and RPM.

FIGURE 4
www.frontiersin.org

Figure 4 (A) Linear regression analysis between the MLD and the ITV. (B) linear regression analysis between the MLD and inter-target distance.

Discussion

Single-isocentre SBRT for MLM is an emerging technique for medically inoperable patients (69). However, there are little data about the dosimetric impact of single-isocentre SBRT for MLM compared with a conventional multi-isocentre SBRT technique, and the potential increase in dose spillage to the normal liver tissue using a single-isocentre technique. So, we conducted a comprehensive comparison between single- and multi-isocentre techniques and investigated whether a single-isocentre technique leads to increased normal liver dose for MLM SBRT. First, we manually designed the single- and multi-isocentre plans for all 30 patients. The two approaches achieved good dosimetry goals, as shown in Table 2. Compared with the single-isocentre technology, the multi-isocentre technology plans slightly improved the PTV, OARs sparing, and the MLD. On the other hand, the V20Gy, MUs, and delivery time of multi-isocentre plans were significantly higher than those of single-isocentre plans, and we found that these indicators were larger in patients with three lesions (5 patients) compared with patients with two lesions (25 patients). These results are consistent with recent studies that compared the dosimetric quality of single-isocentre multi-lesion lung SBRT with the multi-isocentre technique (47). Single-isocentre SBRT for MLM could be safely used to reduce position error and delivery time and to improve cost-effectiveness, even though we observed a relevant increase in the MLD.

Despite growing interest in single-isocentre/multiple-lesion VMAT SBRT treatments, difficulties due to complex multi-lesion SBRT planning have been described (13, 20, 4749). When treating multiple lesions with a single- or multi-isocentre VMAT plan, the beam fields in the target interval are likely to interact with each other, and it is difficult to control the dose spill over outside the target area. Currently, manual treatment planning is still the most utilized approach in clinical practice, whereas assigning parameters, tweaking constraints, and setting weights remains labor intensive. During the optimization process, dose-limiting shells, and regions of interest of hot/cold spots are frequently introduced to achieve a proper balance between target coverage and OARs sparing. Due to the complexity of MLM and variable skills among planners, one of the major challenges of VMAT-SBRT planning is the large variation in the quality of plans. With limited clinical resources and time, even senior physicists may not be able to get the expected MLM plan. Manual planning depends heavily on the experience of the planner. Moreover, it is impossible for physicians/dosimetrists to know whether optimization attempts have minimized the OARs doses to the lowest attainable level within an acceptable time frame. Trying to overcome the limitations of MLM SBRT planning, we have presented a novel implementation of the multi-isocentre AP approach for MLM SBRT alongside its operability and advantages with respect to previously manually designed plans in terms of planning quality and efficiency.

Based on the manually optimized single-/multi-isocentre plans and in line with the clinical requirements, we constructed the general single- and multi-isocentre AP models. For either single- or multi-isocentre plans, the AP plans are higher quality than the manual plans, although there were no significant differences in some PTV metrics. Based on OARs sparing and the probability of liver NTCPs, the AP plans, especially RPS, are also better than the manual plans. We showed that RPS and RPM improve sparing of the spinal cord and kidneys in the max and medium dose range compared with MUS and MUM (see Table 3). In our study, the MUs of single-isocentre plans decreased by 62% (MUS vs MUM) and 40% (RPS vs RPM) compared with multi-isocentre plans. Furthermore, the results of 30 cases of manual planning and 10 cases of AP showed a significant correlation between the MLD and the ITV for all techniques, but the MLD was not correlated with the distance between lesions. Many factors affect the quality of AP, but the main factors are the quality of the plan in the model library and the geometry of the target area. To ensure the consistency of the target structure and the efficiency of the model, we established two general training models. Our study can provide some guidance for the clinical application of multiple metastases of hepatocellular carcinoma.

There are some potential limitations to this work. First, although we enrolled 30 patients with MLM, due to the inherent requirements of the model training set, the number of test patients is relatively small. Since we only compared treatment plans and the single-isocentre technique has not yet been widely clinically applied for MLM SBRT, and our conclusions could be further supported by increasing the sample size in future studies. Second, dose/plan verification is not investigated in the current retrospective study, and this should be done before clinical radiotherapy. However, we consider these limitations to be of minimal importance and they have not significantly influenced the overall findings of the study.

Conclusion

We have shown the feasibility of single-/multi-isocentre SBRT using RapidPlan technique for MLM and have presented the resulting dosimetric impact for single- versus multi-isocentre MLM SBRT plans. One could use a single-isocentre approach for MLM to reduce treatment time and patient comfort at the cost of a small increase in the MLD. Compared with the manual plans, the RapidPlan-based plans slightly improve the PTV, OARs, and the MLD at the cost of increased MUs and delivery time.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material. Further inquiries can be directed to the corresponding authors.

Ethics statement

The studies involving human participants were reviewed and approved by the Institution Review Board and Ethics Committee of Jiangxi Cancer Hospital. Written informed consent to participate in this study was provided by the participants’ legal guardian/next of kin.

Author contributions

CT: data curation, methodology, project administration, writing – original draft. CG: conceptualization, formal analysis, writing – review, and editing. BL: data curation, formal analysis. HG: formal analysis, and investigation. ZD: investigation, project administration. JY: data curation, formal analysis. XW: data curation, and investigation. YZ: conceptualization, formal analysis, writing – review, and editing. All authors contributed to the article and approved the submitted version.

Funding

The authors acknowledge the support provided by Jiangxi Provincial Natural Science Foundation (No.20224BAB206070) and the Science and Technology Project of Jiangxi Provincial Health Commission of China (No.202310023) and ((No.202210054)).

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fonc.2023.1144784/full#supplementary-material

References

1. Tsilimigras DI, Brodt P, Clavien PA, Muschel RJ, D'Angelica MI. Liver metastases. Nat Rev Dis Primers (2021) 7(1):27. doi: 10.1038/s41572-021-00261-6

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Grundmann RT. Current state of surgical treatment of liver metastases from colorectal cancer. World J Gastrointest Surg (2011) 3(12):183–96. doi: 10.4240/wjgs.v3.i12.183

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Petrowsky H, Fritsch R, Guckenberger M, De Oliveira ML, Dutkowski P, Clavien PA. Modern therapeutic approaches for the treatment of malignant liver tumours. Nat Rev Gastroenterol Hepatol (2020) 17(12):755–72. doi: 10.1038/s41575-020-0314-8

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Lo SS, Moffatt-Bruce SD, Dawson LA, Schwarz RE, Teh BS, Mayr NA, et al. The role of local therapy in the management of lung and liver oligometastases. Nat Rev Clin Oncol (2011) 8(7):405–16. doi: 10.1038/nrclinonc.2011.75

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Lee MT, Kim JJ, Dinniwell R, Brierley J, Lockwood G, Wong R, et al. Phase I study of individualized stereotactic body radiotherapy of liver metastases. J Clin Oncol (2009) 27(10):1585–91. doi: 10.1200/JCO.2008.20.0600

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Rusthoven KE, Kavanagh BD, Burri SH, Chen C, Cardenes H, Chidel MA, et al. Multi-institutional phase I/II trial of stereotactic body radiation therapy for lung metastases. J Clin Oncol (2009) 27(10):1579–84. doi: 10.1200/JCO.2008.19.6386

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Scorsetti M, Comito T, Tozzi A, Navarria P, Fogliata A, Clerici E, et al. Final results of a phase II trial for stereotactic body radiation therapy for patients with inoperable liver metastases from colorectal cancer. J Cancer Res Clin Oncol (2015) 141(3):543–53. doi: 10.1007/s00432-014-1833-x

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Cacciola A, Parisi S, Tamburella C, Lillo S, Ferini G, Molino L, et al. Stereotactic body radiation therapy and radiofrequency ablation for the treatment of liver metastases: how and when? Rep Pract Oncol Radiother (2020) 25(3):299–306. doi: 10.1016/j.rpor.2020.02.010

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Comito T, Clerici E, Tozzi A, D'Agostino G. Liver metastases and SBRT: a new paradigm? Rep Pract Oncol Radiother (2015) 20(6):464–71. doi: 10.1016/j.rpor.2014.10.002

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Lo SS, Fakiris AJ, Chang EL, Mayr NA, Wang JZ, Papiez L, et al. Stereotactic body radiation therapy: a novel treatment modality. Nat Rev Clin Oncol (2010) 7(1):44–54. doi: 10.1038/nrclinonc.2009.188

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys (2008) 35(1):310–7. doi: 10.1118/1.2818738

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Bedford JL, Warrington AP. Commissioning of volumetric modulated arc therapy (VMAT). Int J Radiat oncology biology Phys (2009) 73(2):537–45. doi: 10.1016/j.ijrobp.2008.08.055

CrossRef Full Text | Google Scholar

13. Al-Hallaq HA, Chmura S, Salama JK, Winter KA, Robinson CG, Pisansky TM, et al. Rationale of technical requirements for NRG-BR001: the first NCI-sponsored trial of SBRT for the treatment of multiple metastases. Pract Radiat Oncol (2016) 6(6):e291–8. doi: 10.1016/j.prro.2016.05.004

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Wang CW, Lin YC, Tseng HM, Xiao F, Chen CM, Cheng WL, et al. Prolonged treatment time deteriorates positioning accuracy for stereotactic radiosurgery. PloS One (2015) 10(4):e0123359. doi: 10.1371/journal.pone.0123359

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Ruggieri R, Naccarato S, Mazzola R, Ricchetti F, Corradini S, Fiorentino A, et al. Linac-based VMAT radiosurgery for multiple brain lesions: comparison between a conventional multi-isocenter approach and a new dedicated mono-isocenter technique. Radiat Oncol (London England) (2018) 13(1):38. doi: 10.1186/s13014-018-0985-2

CrossRef Full Text | Google Scholar

16. Rusthoven KE, Kavanagh BD, Cardenes H, Stieber VW, Burri SH, Feigenberg SJ, et al. Multi-institutional phase I/II trial of stereotactic body radiation therapy for liver metastases. J Clin Oncol Off J Am Soc Clin Oncol (2009) 27(10):1572–8. doi: 10.1200/JCO.2008.19.6329

CrossRef Full Text | Google Scholar

17. Clark GM, Popple RA, Young PE, Fiveash JB. Feasibility of single-isocenter volumetric modulated arc radiosurgery for treatment of multiple brain metastases. Int J Radiat Oncol Biol Phys (2010) 76(1):296–302. doi: 10.1016/j.ijrobp.2009.05.029

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Miura H, Ozawa S, Kusaba H, Doi Y, Kenjo M, Yamada K, et al. Characterization of robust optimization for VMAT plan for liver cancer. Rep Pract Oncol Radiother (2020) 25(3):376–81. doi: 10.1016/j.rpor.2020.03.012

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Thaper D, Kamal R, Singh G, Oinam AS, Yadav HP, Kumar R, et al. Dosimetric comparison of dynamic conformal arc integrated with segment shape optimization and variable dose rate versus volumetric modulated arc therapy for liver SBRT. Rep Pract Oncol Radiother (2020) 25(4):667–77. doi: 10.1016/j.rpor.2020.04.017

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Moustakis C, Blanck O, Chan MKH, Boda-Heggemann J, Andratschke N, Duma MN, et al. Planning benchmark study for stereotactic body radiation therapy of liver metastases: results of the DEGRO/DGMP working group on stereotactic radiation therapy and radiosurgery. Int J Radiat Oncol Biol Phys (2022) 113(1):214–27. doi: 10.1016/j.ijrobp.2022.01.008

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Smith A, Granatowicz A, Stoltenberg C, Wang S, Liang X, Enke CA, et al. Can the student outperform the master? a plan comparison between pinnacle auto-planning and eclipse knowledge-based RapidPlan following a prostate-bed plan competition. Technol Cancer Res Treat (2019) 18:1533033819851763. doi: 10.1177/1533033819851763

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Hansen CR, Bertelsen A, Hazell I, Zukauskaite R, Gyldenkerne N, Johansen J, et al. Automatic treatment planning improves the clinical quality of head and neck cancer treatment plans. Clin Transl Radiat Oncol (2016) 1:2–8. doi: 10.1016/j.ctro.2016.08.001

PubMed Abstract | CrossRef Full Text | Google Scholar

23. Good D, Lo J, Lee WR, Wu QJ, Yin FF, Das SK, et al. A knowledge-based approach to improving and homogenizing intensity modulated radiation therapy planning quality among treatment centers: an example application to prostate cancer planning. Int J Radiat Oncol Biol Phys (2013) 87:176–81. doi: 10.1016/j.ijrobp.2013.03.015

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Breedveld S, Storchi PR, Voet PW, Heijmen BJ. iCycle: integrated, multicriterial beam angle, and profile optimization for generation of coplanar and noncoplanar IMRT plans. Med Phys (2012) 39:951–63. doi: 10.1118/1.3676689

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Wu B, Ricchetti F, Sanguineti G, Kazhdan M, Simari P, Jacques R, et al. Data-driven approach to generating achievable dose-volume histogram objectives in intensity-modulated radiotherapy planning. Int J Radiat Oncol Biol Phys (2011) 79(4):1241–7. doi: 10.1016/j.ijrobp.2010.05.026

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Scaggion A, Fusella M, Roggio A, Bacco S, Pivato N, Rossato MA, et al. Reducing inter- and intra-planner variability in radiotherapy plan output with a commercial knowledge-based planning solution. Physica Med PM an Int J devoted to Appl Phys to Med Biol Off J Ital Assoc Biomed Phys (AIFB) (2018) 53:86–93. doi: 10.1016/j.ejmp.2018.08.016

CrossRef Full Text | Google Scholar

27. Kubo K, Monzen H, Ishii K, Tamura M, Kawamorita R, Sumida I, et al. Dosimetric comparison of RapidPlan and manually optimized plans in volumetric modulated arc therapy for prostate cancer. Physica Med PM an Int J devoted to Appl Phys to Med Biol Off J Ital Assoc Biomed Phys (AIFB) (2017) 44:199–204. doi: 10.1016/j.ejmp.2017.06.026

CrossRef Full Text | Google Scholar

28. Palmiero AN, Fabian D, St Clair W, Randall M, Pokhrel D. Management of multiple brain metastases via dual-isocenter VMAT stereotactic radiosurgery. Med dosimetry Off J Am Assoc Med Dosimetrists vol (2021) 46:3. doi: 10.1016/j.meddos.2021.01.002

CrossRef Full Text | Google Scholar

29. Quan K, Xu KM, Lalonde R, Horne ZD, Bernard ME, McCoy C, et al. Treatment plan technique and quality for single-isocenter stereotactic ablative radiotherapy of multiple lung lesions with volumetric-modulated arc therapy or intensity-modulated radiosurgery. Front Oncol (2015) 5:213. doi: 10.3389/fonc.2015.00213

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Yin Y, Ma C, Gao M, Chen J, Ma Y, Liu T, et al. Dosimetric comparison of RapidArc with fixed gantry intensity-modulated radiotherapy treatment for multiple liver metastases radiotherapy. Med dosimetry Off J Am Assoc Med Dosimetrists (2011) 36(4):448–54. doi: 10.1016/j.meddos.2010.12.001

CrossRef Full Text | Google Scholar

31. Yu G, Li Y, Feng Z, Tao C, Yu Z, Li B, et al. Knowledge-based IMRT planning for individual liver cancer patients using a novel specific model. Radiat Oncol (London England) (2018) 13(1):52. doi: 10.1186/s13014-018-0996-z

CrossRef Full Text | Google Scholar

32. Fogliata A, Wang PM, Belosi F, Clivio A, Nicolini G, Vanetti E, et al. Assessment of a model based optimization engine for volumetric modulated arc therapy for patients with advanced hepatocellular cancer. Radiat Oncol (London England) (2014) 9:236. doi: 10.1186/s13014-014-0236-0

CrossRef Full Text | Google Scholar

33. Sterzing F, Brunner TB, Ernst I, Baus WW, Greve B, Herfarth K, et al. Stereotactic body radiotherapy for liver tumors: principles and practical guidelines of the DEGRO working group on stereotactic radiotherapy. Strahlenther Onkol (2014) 190:872–81. doi: 10.1007/s00066-014-0714-1

PubMed Abstract | CrossRef Full Text | Google Scholar

34. Son SH, Choi BO, Ryu MR, Kang YN, Jang JS, Bae SH, et al. Stereotactic body radiotherapy for patients with unresectable primary hepatocellular carcinoma: dose-volumetric parameters predicting the hepatic complication. Int J Radiat Oncol Biol Phys (2010) 78(4):1073–80. doi: 10.1016/j.ijrobp.2009.09.009

PubMed Abstract | CrossRef Full Text | Google Scholar

35. Söhn M, Alber M, Yan D. Principal component analysis-based pattern analysis of dose-volume histograms and influence on rectal toxicity. Int J Radiat Oncol Biol Phys (2007) 69(1):230–9. doi: 10.1016/j.ijrobp.2007.04.066

PubMed Abstract | CrossRef Full Text | Google Scholar

36. Zhu X, Ge Y, Li T, Thongphiew D, Yin FF, Wu QJ. A planning quality evaluation tool for prostate adaptive IMRT based on machine learning. Med Phys (2011) 38(2):719–26. doi: 10.1118/1.3539749

PubMed Abstract | CrossRef Full Text | Google Scholar

37. Shaw E, Kline R, Gillin M, Souhami L, Hirschfeld A, Dinapoli R, et al. Radiation therapy oncology group: radiosurgery quality assurance guidelines. Int J Radiat Oncol Biol Phys (1993) 27(5):1231–9. doi: 10.1016/0360-3016(93)90548-a

PubMed Abstract | CrossRef Full Text | Google Scholar

38. Paddick I, Bodo L. A simple dose gradient measurement tool to complement the conformity index. J Neurosurg (2006) 105Suppl:194–201. doi: 10.3171/sup.2006.105.7.194

CrossRef Full Text | Google Scholar

39. Huang Y, Chin K, Robbins JR, Kim J, Li H, Amro H, et al. Radiosurgery of multiple brain metastases with single-isocenter dynamic conformal arcs (SIDCA). Radiother Oncol J Eur Soc Ther Radiol Oncol (2014) 112(1):128–32. doi: 10.1016/j.radonc.2014.05.0090

CrossRef Full Text | Google Scholar

40. Wilke L, Andratschke N, Blanck O, Brunner TB, Combs SE, Grosu AL, et al. ICRU report 91 on prescribing, recording, and reporting of stereotactic treatments with small photon beams: statement from the DEGRO/DGMP working group stereotactic radiotherapy and radiosurgery. Strahlenther Onkol (2019) 195(3):193–8. doi: 10.1007/s00066-018-1416-x

PubMed Abstract | CrossRef Full Text | Google Scholar

41. Schefter TE, Kavanagh BD, Timmerman RD, Cardenes HR, Baron A, Gaspar LE. A phase I trial of stereotactic body radiation therapy (SBRT) for liver metastases. Int J Radiat Oncol Biol Phys (2005) 62(5):1371–8. doi: 10.1016/j.ijrobp.2005.01.002

PubMed Abstract | CrossRef Full Text | Google Scholar

42. Prayongrat A, Kobashi K, Ito YM, Katoh N, Tamura M, Dekura Y, et al. The normal tissue complication probability model-based approach considering uncertainties for the selective use of radiation modality in primary liver cancer patients. Radiother Oncol J Eur Soc Ther Radiol Oncol (2019) 135:100–6. doi: 10.1016/j.radonc.2019.03.003

CrossRef Full Text | Google Scholar

43. Miften M, Vinogradskiy Y, Moiseenko V, Grimm J, Yorke E, Jackson A, et al. Radiation dose-volume effects for liver SBRT. Int J Radiat Oncol Biol Phys (2021) 110(1):196–205. doi: 10.1016/j.ijrobp.2017.12.290

PubMed Abstract | CrossRef Full Text | Google Scholar

44. Grimm J, Marks LB, Jackson A, Kavanagh BD, Xue J, Yorke E. High dose per fraction, hypofractionated treatment effects in the clinic (HyTEC): an overview. Int J Radiat oncology biology Phys (2021) 110(1):1–10. doi: 10.1016/j.ijrobp.2020.10.039

CrossRef Full Text | Google Scholar

45. Pan CC, Kavanagh BD, Dawson LA, Li XA, Das SK, Miften M, et al. Radiation-associated liver injury. Int J Radiat Oncol Biol Phys (2010) 76(3 Suppl):S94–100. doi: 10.1016/j.ijrobp.2009.06.092

PubMed Abstract | CrossRef Full Text | Google Scholar

46. Dawson LA, Normolle D, Balter JM, McGinn CJ, Lawrence TS, Ten Haken RK. Analysis of radiation-induced liver disease using the Lyman NTCP model. Int J Radiat Oncol Biol Phys (2002) 53(4):810–21. doi: 10.1016/s0360-3016(02)02846-8

PubMed Abstract | CrossRef Full Text | Google Scholar

47. van Timmeren JE, Ehrbar S, Chamberlain M, Mayinger M, Hoogeman MS, Andratschke N, et al. Single-isocenter versus multiple-isocenters forr multiple lung metastases: evaluation of lung dose. Radiother Oncol (2022) 166:189–94. doi: 10.1016/j.radonc.2021.11.030

PubMed Abstract | CrossRef Full Text | Google Scholar

48. Sanford L, Molloy J, Kumar S, Randall M, McGarry R, Pokhrel D. Evaluation of plan quality and treatment efficiency for single-isocenter/two-lesion lung stereotactic body radiation therapy. J Appl Clin Med Phys (2019) 20:118–27. doi: 10.1002/acm2.12500

PubMed Abstract | CrossRef Full Text | Google Scholar

49. Trager M, Salama J, Yin FF, Adamson J. SBRT treatment of multiple extracranial oligometastases using a single isocenter with distinct optimizations. J Radiosurg SBRT (2017) 4:265–73.

PubMed Abstract | Google Scholar

Keywords: RapidPlan, multiple liver metastases, stereotactic body radiotherapy, single-isocentre, normal tissue complications

Citation: Tang C, Gong C, Liu B, Guo H, Dai Z, Yuan J, Wang X and Zhang Y (2023) Feasibility and dosimetric evaluation of single- and multi-isocentre stereotactic body radiation therapy for multiple liver metastases. Front. Oncol. 13:1144784. doi: 10.3389/fonc.2023.1144784

Received: 15 January 2023; Accepted: 13 April 2023;
Published: 28 April 2023.

Edited by:

Ferenc Lakosi, Somogy County Kaposi Mór Teaching Hospital, Hungary

Reviewed by:

Vanessa Panettieri, Peter MacCallum Cancer Centre, Australia
Yvonne Dzierma, Saarland University Hospital, Germany

Copyright © 2023 Tang, Gong, Liu, Guo, Dai, Yuan, Wang and Zhang. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Changfei Gong, Z29uZ2NmMTk5MEBvdXRsb29rLmNvbQ==; Yun Zhang, emhhbmd5dW5fMTk4M0Bzb2h1LmNvbQ==

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.