ORIGINAL RESEARCH article

Front. Oncol., 13 December 2022
Sec. Surgical Oncology

Evaluating breast ultrasound S-detect image analysis for small focal breast lesions

Boyuan Xing&#x;Boyuan Xing1†Xiangyi Chen&#x;Xiangyi Chen2†Yalin WangYalin Wang3Shuang LiShuang Li4Ying-Kui Liang*Ying-Kui Liang5*Dawei Wang,*Dawei Wang3,5*
  • 1Department of Ultrasound Imaging, The People’s Hospital of China Three Gorges University/the First People’s Hospital of Yichang, Yichang, Hubei, China
  • 2Department of Nuclear Medicine, First Affiliated Hospital of Guangxi Medical University, Nanning, China
  • 3Department of Medical Engineering, Medical Supplies Center of PLA General Hospital, Beijing, China
  • 4Department of Pathology, The People’s Hospital of China Three Gorges University/the First People’s Hospital of Yichang, Yichang, Hubei, China
  • 5Department of Nuclear Medicine, The Sixth Medical Center of People's Liberation Army General Hospital, Beijing, China

Background: S-Detect is a computer-assisted, artificial intelligence-based system of image analysis that has been integrated into the software of ultrasound (US) equipment and has the capacity to independently differentiate between benign and malignant focal breast lesions. Since the revision and upgrade in both the breast imaging-reporting and data system (BI-RADS) US lexicon and the S-Detect software in 2013, evidence that supports improved accuracy and specificity of radiologists’ assessment of breast lesions has accumulated. However, such assessment using S-Detect technology to distinguish malignant from breast lesions with a diameter no greater than 2 cm requires further investigation.

Methods: The US images of focal breast lesions from 295 patients in our hospital from January 2019 to June 2022 were collected. The BI-RADS data were evaluated by the embedded program and as manually modified prior to the determination of a pathological diagnosis. The receiver operator characteristic (ROC) curves were constructed to compare the diagnostic accuracy between the assessments of the conventional US images, the S-Detect classification, and the combination of the two.

Results: There were 326 lesions identified in 295 patients, of which pathological confirmation demonstrated that 239 were benign and 87 were malignant. The sensitivity, specificity, and accuracy of the conventional imaging group were 75.86%, 93.31%, and 88.65%. The sensitivity, specificity, and accuracy of the S-Detect classification group were 87.36%, 88.28%, and 88.04%, respectively. The assessment of the amended combination of S-Detect with US image analysis (Co-Detect group) was improved with a sensitivity, specificity, and accuracy of 90.80%, 94.56%, and 93.56%, respectively. The diagnostic accuracy of the conventional US group, the S-Detect group, and the Co-Detect group using area under curves was 0.85, 0.88 and 0.93, respectively. The Co-Detect group had a better diagnostic efficiency compared with the conventional US group (Z = 3.882, p = 0.0001) and the S-Detect group (Z = 3.861, p = 0.0001). There was no significant difference in distinguishing benign from malignant small breast lesions when comparing conventional US and S-Detect techniques.

Conclusions: The addition of S-Detect technology to conventional US imaging provided a novel and feasible method to differentiate benign from malignant small breast nodules.

Introduction

Breast cancer is the most diagnosed cancer and the leading cause of cancer mortality in women worldwide. Despite significant advances in screening, imaging, and surgical and adjunctive treatment, the management of breast cancer exacts a costly and psychological toll (1). Recently, breast cancer has become the most frequently diagnosed cancer in Chinese women (30.4 cases per 100,000; estimated incidence rate of 17.1%) (2). Ultrasound (US) examination is a valuable and commonly used technology for breast screening that has numerous advantages including relatively low cost, real-time imaging, no radiation exposure, millimeter-level resolution, and the opportunity for interventional applications. US evaluation is principally utilized to characterize two-dimensional lesion morphology, but can be extended to use elastography or acoustic radiation force impulse imaging (ARFI) to assess tissue stiffness, and three-dimensional US (3D-US) can yield more detailed imaging. As part of the breast cancer screening armamentarium, US has contributed to earlier diagnosis, which facilitates more conservative and more effective treatment with commensurate improved patient outcomes (2, 3). Tumor stage is a critical component of patient survival, as illustrated by the 10-year overall (OA) survival rate of 81%–93% in patients with breast cancers measuring ≤2 cm, compared to the OA rate of 60%–70% with cancers >2 cm (4, 5). Small breast nodules can have considerable variability in the shape, margin appearance, and echo pattern by conventional B-mode US that causes difficulty in distinguishing benign from malignant lesions (6, 7). Breast imaging-reporting and data system (BI-RADS) is the standard US lexicon aimed at standardizing image interpretation, reporting, and teaching breast imaging. S-Detect is a computer-assisted embedded diagnostic software program based on a deep-learning algorithm (without image postprocessing and therefore not influenced by human bias) that automatically analyzes the inner constructions, refracted echoes, interfaced regions, morphological manifestations, and other parameters of the specific breast nodule (8, 9). Hence, the intention of the present study was to investigate the statistical performance of conventional US, S-Detect technology, and the two techniques in combination (Co-Detect) in distinguishing benign from malignant small breast nodules (≤2 cm) as confirmed histologically.

Materials and methods

Study design

The study was carried out in accordance with The Code of Ethics of the World Medical Association (Declaration of Helsinki). Institutional review board approval and informed patient consent were obtained for this prospective study. Permission was obtained from the hospital for patient medical records review.

Between January 2019 and June 2022, 326 nodules (size range: 0.30–2.00 cm; mean, 1.19 cm ± 0.43 SD) were detected in 295 women (age range: 18–90 years; mean, 45.40 years ± 11.40 SD) from the Department of Ultrasound or the Department of Thyroid and Breast Surgery in the First People’s Hospital of Yichang city. Indicators for breast US and BI-RADS included the following: (1) a palpable lesion with a diameter ≤ 2 cm, (2) lesion tissue obtained by biopsy or conventional surgical excision, and (3) a definite pathological diagnosis. Exclusion criteria were as follows: (1) anamnesis and prior breast tissue biopsy or surgery, and (2) pregnant or breast-feeding patient.

Instruments and strategies

The US machine used was a type RS80A (Samsung) equipped with an L2-12A ultrasonic probe (5–13 MHz). The patient was positioned supine, the appropriate breast was fully inspected, and multi-slice and stereo scans from each quadrant were obtained. An independent experienced US radiologist/operator conducted the sonography, measured the maximal diameter of the index lesion(s), and assessed the adjacent tissue relationship of each lesion. Records of the size, morphology, orientation, inner-echo density, calcification(s), blood flow, homolateral lymph node metastases, and key sonographic features were kept. Then, US scan mode was switched to freeze and the built-in S-Detect mode was initiated. From the US BI-RADS classification, the lesions were separated into six categories (0–5), with Category 4 further subdivided into 4A–C. Ultimately, for each US group (conventional, S-Detect, and Co-Detect), the lesion was diagnosed as either benign or malignant.

Criteria of US image diagnoses

In the conventional US group, the two-dimensional gray-scale images were evaluated according to the BI-RADS lexicon, in which certain sonographic characteristics were suspicious for malignancy: irregular shape and edge, nonparallel expansion and growth, signal attenuation posteriorly, and microcalcifications. In the scoring system, lesions with one of the above characteristics were classified as 4a; lesions with two characteristics were classified as 4b, and lesions with ≥3 features were classified as 4c (1012). Scores of ≤4a were considered benign and scores of ≥4b were considered malignant. The S-Detect group diagnoses were assigned according to the pre-programmed definitions: Category 3 (probably benign, 98%); Category 4 [suspicious, 2%–95%; Category 4A (low suspicion, >2%–10%), Category 4B (moderate suspicion, >10%–50%), and Category 4C (high suspicion, >50%–95%)]; and Category 5 (highly suggestive, >95%). For the Co-Detect group, the conventional B-mode scan and S-Detect method were synchronized and the final scores reflected modification of the 2D-US score according to whether the S-Direct was assigned to “Malignant” or “Benign” as illustrated in the table below. (For example, if the 2D-US score was 3 and the S-Detect assignment was “Malignant”, the final score was changed to 4a. If the S-Detect assignment was “Benign”, the score would remain a 3. Similarly, if the 2D-US score was 4b, but the S-Detect assignment was “Malignant”, the score would be upgraded to 4c, whereas if the S-Detect was “Benign”, the score would be downgraded to 4a.)

www.frontiersin.org

Results acquired as followed: Scores <4b (benign), scores ≥4b (malignant).

The gray-scale two-dimensional US images of each lesion were performed and re-evaluated by two independently experienced radiologists (with more than 10 years’ working experiences), under blinded conditions. An associate or chief physician was consulted if the diagnosis was not consistent, and a final co-diagnosis was achieved to minimize personal biases.

Statistical analysis

Statistical analysis was performed through SPSS 22.0. Pathological diagnosis was regarded as the gold standard. MedCalc software was conducted to simulate the receiver operating characteristic (ROC) curve, and the area under curve (AUC) values of each curve was calculated. Z-test values were used to assess the differentiation of AUC values between groups (p < 0.05 means a significant difference).

Results

Pathological results

Of the 326 nodules, 87 were malignant (76 infiltrative ductal carcinoma, 6 intraductal carcinoma, 3 papillocarcinomas, and 1 Paget’s disease) and 239 were benign (106 fibroadenosis with fibroadenoma, 66 fibroadenosis, 38 fibroadenomas, 7 sclerosing adenosis, 11 intraductal papillomas, 5 granulomatous lobulitis, 5 fibrocystic diseases, and 1 adenomyoepithelioma) (Table 1).

TABLE 1
www.frontiersin.org

Table 1 Pathological results and BI-RADS scores of lesions detected.

Diagnostic results of conventional US, S-detect, and co-detect groups

Diagnoses of the conventional US group included 244 benign and 82 malignant nodules with sensitivity, specificity, and accuracy values of 75.86%, 93.31%, and 88.65%, respectively (Figure 1). The S-Detect system assigned 222 benign and 104 malignant lesions with sensitivity, specificity, and accuracy values of 87.36%, 88.28%, and 88.04%, respectively. The Co-Detect group had diagnoses of 234 benign and 82 malignant nodules with improved sensitivity, specificity, and accuracy values of 90.80%, 94.56%, and 93.56%, respectively. Representative sonographic images and corresponding histology pictures are shown in Figures 25. Correlation of pathological results and BI-RADS scores initially and in the Co-Detect group is depicted in Table 1, and diagnostic results with statistical comparisons are shown in Tables 2, 3.

FIGURE 1
www.frontiersin.org

Figure 1 Receiver operating characteristic curve (ROC) and area under curve (AUC) values of the conventional US group, the S-Detect group, and the Co-Detect group.

FIGURE 2
www.frontiersin.org

Figure 2 Non-specific invasive carcinoma from a 69-year-old patient (BI-RADS 4a). A hypoechoic nodule (0.57 × 0.51 cm) was found at the 2 o’clock point of her left breast; the ultrasound graphic showed an uneven edge, irregular shape, and uniform echo (A). S-Detect artificial intelligence recommendation was “possibly malignant” and the BI-RADS score receded from 4a to 4b (B). Pathologically confirmed as breast non-specific invasive carcinoma (C).

FIGURE 3
www.frontiersin.org

Figure 3 Fibroadenosis with fibroadenoma formation from a 35-year-old patient (BI-RADS 4a). A hypoechoic nodule (1.09 × 0.82 cm) was found at the 3 o’clock point of her left breast; ultrasound graphic showed glossy edge, irregular shape, partly angulated, and nonuniform internal echo (A). S-Detect artificial intelligence recommendation was “possibly malignant” and the BI-RADS score receded from 4a to 4b (B). Pathologically confirmed as breast fibroadenosis with fibroadenoma formation (C).

FIGURE 4
www.frontiersin.org

Figure 4 Non-specific invasive carcinoma from a 49-year-old patient (BI-RADS 4b). A hypoechoic nodule (0.86 × 0.68 cm) was found at the 10 o’clock point of her right breast with smooth and irregular edges; the inner echoes detected with nonuniform distribution came with spotted hyper-echoes (A). The S-Detect suggestion was “possibly benign”; the BI-RADS classification was adjusted from 4b to 4a (B). Pathological results are showed in (C) (HE, ×100). S-Detect gave a misdiagnosis due to the misrecognized microcalcifications that were found through operators.

FIGURE 5
www.frontiersin.org

Figure 5 Intraductal papilloma from a 54-year-old patient (BI-RADS 4a). A hypoechoic nodule (1.12 × 0.65 cm) was detected at the 10 o’clock point of her right breast with smooth and irregular/regular edges (S-Detect technology failed to draw the outline of this nodule), nonuniform distribution with some spotted hyper-echoes (A). The S-Detect suggestion was “possibly malignant” and BI-RADS score declined from 4a to 4b (B). Pathology result was intraductal papilloma shown in the picture (HE, ×100) (C). Inaccurate boundary layer distributed inappropriate malignant.

TABLE 2
www.frontiersin.org

Table 2 Ultrasonic diagnosis of different detected methods.

TABLE 3
www.frontiersin.org

Table 3 Efficacy comparison of different methods in diagnosis.

ROCs and AUC results of different methods

The Co-Detect group had better diagnostic efficiency with a higher AUC value than the conventional and S-Detect groups (Co-Detect vs. conventional US, Z = 3.882, p = 0.0001; Co-Detect vs. S-Detect, Z = 3.861, p = 0.0001), whereas there was no significant difference between the conventional US and S-detect group regarding diagnostic efficiency (Z = 1.294, p = 0.1956); the AUC values of the conventional US, S-Detect, and Co-Detect groups were 0.85, 0.88 and 0.93, respectively (Figure 1).

Discussion

Although ultrasonography is a commonly used technique to help differentiate benign and malignant breast nodules, not until BI-RADS was introduced by the American College of Radiology was there a standardized terminology to describe the characteristics and final assessment categories of breast lesions (13, 14). Nevertheless, the inevitable overlap of benign and suspicious characteristics together with inexperienced sonographers can still compromise the accuracy of US diagnostic accuracy (15). The S-Detect computer-assisted algorithm, based on AI and neural networks, “learning” on a large amount of training data, is a built-in program on US machines and, independent of human input, analyzes the morphological characteristics of the breast nodule according to the BI-RADS lexicon. Ultimately, it provides a dichotomous evaluation: “possibly benign” or “possibly malignant” (16, 17). In the present study, only small breast nodules (diameter ≤2 cm) were enrolled. According to the BI-RADS standard, the sensitivity, specificity, and accuracy of the diagnosis of malignant nodules were 75.86%, 93.31%, and 88.65%, respectively. Of the 87 malignant nodules, 19 were scored BI-RADS 4a: 10 nodules with a blurred edge and 9 nodules with an irregular shape; these lesions had only a single feature suspicious for malignancy (irregular shape and edges; non-parallel expansion and growth, signal attenuation posteriorly, and microcalcifications) and, therefore, scored as BI-RADS 4a. Of the 239 benign nodules, 14 had a BI-RADS 4b score and 2 had a BI-RADS 4c score; these BI-RADS classifications mismatched with pathological results. These misdiagnoses may be attributed to small size (diameter ≤1 cm, contour and inaccurate microcalcification programming observed), atypical echo features, and uncomprehensive morphological manifestations (irregular-shaped, parallel or unparallel orientation, micro-lobulated and irregular margins, and complex echogenicity in mass) that could be found in both benign and malignant lesions.

Several studies have confirmed that S-Detect technology improves the diagnostic accuracy of breast nodules (18, 19). However, there are very few studies focused on small breast nodules, especially with a maximum diameter limited to 2 cm. Briefly, in the present study, of 326 lesions in 295 patients, 87 were proven malignant and 239 were benign. The S-Detect group suggested 104 malignant and 222 benign nodules; 39 were misdiagnosed histologically. The sensitivity, specificity, and accuracy values of the S-Detect group were 87.36%, 88.28% and 88.04%, respectively. Of 11 malignancies misdiagnosed as benign, 5 were smaller than 1 cm with clear boundaries, and 1 nodule had inner microcalcification. Possible reasons for the misdiagnoses may be related to a size too small to automatically analyze, and the known limitation of the S-Detect technology is that it cannot analyze microcalcification within nodules (Figures 4A, B). Some microcalcification overlooked by S-Detect can easily be identified by senior radiologists, and the diagnosis was amended to reflect this finding. There were 28 nodules misdiagnosed as malignant: 10 with fibroadenosis and fibroadenoma, 9 with only fibroadenosis, 5 with intraductal papillomas, 2 cases of granulomatous lobular mastitis, 1 case of an adenomyoepithelioma, and 1 with sclerosing adenosis. In this study, we used a multipurpose ultrasonic probe (L2-12A, 5–13 MHz) to extend usage latitude. The possible reason for these inaccuracies may be related to the location and depth of masses; ultrasonic frequency is negatively correlated with detection range and imaging resolution (20, 21). In some cases, the increased divergence of interpretations of mammary gland images (5 cases were declared malignant in the S-Detect group) and shape irregularity (19 cases) was considered a higher risk in the S-Detect group (micro-lobulated, angular, and showing other suspicious marginal changes) that caused the S-Detect program to errantly define the borders of the nodule, potentially misrepresenting the edge of the nodules, which resulted in misdiagnosis as malignant (Figure 5).

In the Co-Detect group, S-Detect technology was used to adjust the BI-RADS scores assigned by conventional US. Markedly important, of 92 malignant and 234 benign nodules, 86.63% (149 nodules) of the 172 nodules with a BI-RADS 4a score by conventional US were downgraded to a BI-RADS 3 score due to the benign assignment by S-Detect. As a result, the clinical management of these patients changed from biopsy to clinical follow-up. Eventually, 95.97% (143 of 149) were confirmed as benign pathologically; groups of patients who had the same strategy of a corrected BI-RADS category may be able to avoid biopsy, thus avoiding unnecessary surgery and reducing costs. Although few in number, of concern were the 6 of 149 (4.03%) that were confirmed as malignant during follow up. Correctly downgraded from BI-RADS 4b to 4a were 9 of 77 (19.15%) nodules that were confirmed benign pathologically. Additionally, 38 nodules with a 4b score were upgraded to class 4c and subsequently confirmed as malignant. Clearly, although not perfect, both upgrade and downgrade score adjustments prompted by S-Detect improved the diagnostic accuracy of two-dimensional ultrasonography.

The Co-Detect group had impressive success in differentiating benign and malignant nodules with sensitivity, specificity, and accuracy values of 90.80%, 94.56%, and 93.56%, respectively. However, notably, there were 21 misdiagnoses: 8 pathologically malignant nodules were misdiagnosed as benign, including 5 infiltrating ductal carcinomas, 2 intraductal carcinomas, and 1 papillary carcinoma. The possible reasons for these “misses” might be attributed to an inappropriate BI-RADS 4a score in the traditional US group and unmatched BI-RADS 3 score in the Co-Detect group, optimistically regarded as advanced in the S-Detect group, with four of the nodules being quite small, <1 cm, edge sophisticated, irregular, with no circumscribed shape, irregular with angular margins, or lobulated. There were 13 benign nodules misdiagnosed as malignant, namely, 4 cases of fibroadenosis with fibroadenoma, 4 cases of intraductal papillomas, 2 cases of granulomatous mastitis, 2 cases of fibroadenoma, and 1 case of adenomyoepithelioma. Several of these had concerning characteristics based on the conventional US including serrated edge, shape irregularity, consistent echo, and parallel growth, which may cause incorrect BI-RADS 4a score in conventional US, incorrect classification in S-Detect, and incorrect diagnosis as BI-RADS 4b in the Co-Detect group (as malignant lesions). The AUC of the Co-Detect group relative to diagnostic efficiency was significantly higher than conventional US and S-Detect by statistical analysis (Z = 3.882, p = 0.0001; Z = 3.861, p = 0.0001).

Although we concluded that S-Detect clearly had definite clinical value for the diagnosis of small breast nodules, there were still some limitations. Specifically, S-Detect cannot identify microcalcifications within nodules, an important indicator suspicious for malignancy, whereas experienced radiologists could identify these ambiguities. S-Detect automatic nodule identification, particularly if small or if the edges are indistinct, might cause inappropriate classification; the addition of blood flow information, elastography, and other parameters could further enhance diagnostic characterization (22, 23). The present study demonstrated that S-Detect technology could provide accurate analysis, avoiding subjective physician bias. The combination of S-Detect and conventional ultrasonography could amplify the advantages of both, improving diagnostic efficiency, increasing diagnostic confidence, and significantly reducing the frequency of misdiagnosis. Furthermore, this feasible strategy for the evaluation of focal breast lesions we had conducted was applied to a single-center study; it should be further certified using a multi-center study, and the ability to reduce the margin of experience, which varies from resident doctors to top doctors, should be evaluated.

Conclusions

In conclusion, there was an unequivocal improvement in diagnostic accuracy of small breast nodules with the incorporation of S-Detect; its use in conjunction with conventional ultrasonography can be confidently recommended.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

(I) Conception and design: BX, XC, DW; (II) Administrative support: YW, Y-KL; (III) Provision of study materials or patients: BX, XC; (IV) Collection and assembly of data: BX, SL; (V) Data analysis and interpretation: BX, XC. All authors contributed to the article and approved the submitted version.

Funding

This work was supported by grants from the Yichang City’s Medical and Health Research Project (No. A21-2-029), the Scientific Research Program of Education Department of Hubei Province (No. B2022030), and the Innovation and Cultivation Fund of the Sixth Medical Center of PLA General Hospital (No. CXPY202005 to DW).

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Sung H, Ferlay J, Siegel RL, Laversanne M, Soerjomataram I, Jemal A, et al. Global cancer statistics 2020: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin (2021) 71(3):209–49. doi: 10.3322/caac.21660

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Cao M, Li H, Sun D, Chen W. Cancer burden of major cancers in China: A need for sustainable actions. Cancer Commun (Lond) (2020) 40(5):205–10. doi: 10.1002/cac2.12025

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Liu MJ, Men YM, Zhang YL, Zhang YX, Liu H. Improvement of diagnostic efficiency in distinguishing the benign and malignant thyroid nodules via conventional ultrasound combined with ultrasound contrast and elastography. Oncol Lett (2017) 14(1):867–71. doi: 10.3892/ol.2017.6245

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Fitzgibbons PL, Page DL, Weaver D, Thor AD, Allred DC, Clark GM, et al. Prognostic factors in breast cancer. Arch Pathol Lab Med (2000) 124(7):966–78. doi: 10.5858/2000-124-0966-PFIBC

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Narod SA. Age of diagnosis, tumor size, and survival after breast cancer: implications for mammographic screening. Breast Cancer Res Treat (2011) 128(1):259–66. doi: 10.1007/s10549-010-1318-9

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Uehiro N, Horii R, Iwase T, Tanabe M, Sakai T, Morizono H, et al. Validation study of the UICC TNM classification of malignant tumors, seventh edition, in breast cancer. Breast Cancer. (2014) 21(6):748–53. doi: 10.1007/s12282-013-0453-7

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Pistolese CA, Tosti D, Citraro D, Ricci F, Di Stefano C, Lamacchia F, et al. Probably benign breast nodular lesions(BI-RADS 3):correlation between ultrasound features and histologic findings. Ultrasound Med Biol (2019) 45(1):78–84. doi: 10.1016/j.ultrasmedbio.2018.09.004

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Segni DM, Soccio VD, Cantisani V, Bonito G, Rubini A, Di Segn, et al. Automated classification of focal breast lesions according to s-detect: validation and role as a clinical and teaching tool. J Ultrasound (2018) 21(2):105–18. doi: 10.1007/s40477-018-0297-2

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Wu J, Zhao Z, Zhang W, Liang M, Ou B, Yang HY, et al. Computer-aided diagnosis of solid breast lesions with ultrasound: Factors associated with false-negative and false-positive results. J Ultrasound Med (2019) 38(12):3193–202. doi: 10.1002/jum.15020

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Ma JF, Chen LY, Wu SL, Xu YY, Yao F, Jin F, et al. Clinical practice guidelines for ultrasound-guided breast lesions and lymph nodes biopsy: Chinese society of breast surgery (CSBrS) practice guidelines 2021. Chin Med J (Engl) (2021) 134(12):1393–5. doi: 10.1097/CM9.0000000000001549

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Bevers TB, Helvie M, Bonaccio E, Calhoun KE, Daly MB, Farrar WB, et al. Breast cancer screening and diagnosis, version 3.2018, NCCN clinical practice guidelines in oncology. J Natl Compr Canc Netw (2018) 16(11):1362–89. doi: 10.6004/jnccn.2018.0083

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Kapetas P, Clauser P, Woitek R, Wengert GJ, Lazar M, Pinker K, et al. Quantitative multiparametric breast ultrasound: Application of contrast-enhanced ultrasound and elastography leads to an improved differentiation of benign and malignant lesions. Invest Radiol (2019) 54(5):257–64. doi: 10.1097/RLI.0000000000000543

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Dorsi CJ, Sickles EA, Mendelson EB, Morris EA. ACR BI-RADS atlas,breast imaging reporting and data system. 5th ed. Reston: American College of Radiology (2013) p. 123–32.

Google Scholar

14. Rao AA, Feneis J, Lalonde C, Ojeda-Fournier H. A pictorial review of changes in the BI-RADS fifth edition. Radiographics (2016) 36(3):623–39. doi: 10.1148/rg.2016150178

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Nakano S, Otsuka M, Mibu A, Oinuma T. Significance of fine needleaspiration cytology and vacuum-assisted core needle biopsy forsmall breast lesions. Clin Breast Cancer (2015) 15(1):e23–6. doi: 10.1016/j.clbc.2014.07.001

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Han S, Kang HK, Jeong JY, Park MH, Kim W, Bang WC, et al. A deep learning framework for supporting the classification of breast lesions in ultrasound images. Phys Med Biol (2017) 62(19):7714–28. doi: 10.1088/1361-6560/aa82ec

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Bartolotta TV, Orlando AAM, Spatafora L, Dimarco M, Gagliardo C, Taibbi A, et al. S-detect characterization of focal breast lesions according to the US BI RADS lexicon: a pictorial essay. J Ultrasound (2020) 23(2):207–15. doi: 10.1007/s40477-020-00447-w

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Li J, Sang T, Yu WH, Jiang M, Hunag SY, Cao CL, et al. The value of s-detect for the differential diagnosis of breast masses on ultrasound: a systematic review and pooled meta-analysis. Med Ultrason (2020) 22(2):211–9. doi: 10.11152/mu-2402

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Zhao C, Xiao M, Liu H, Wang M, Wang H, Zhang J, et al. Reducing the number of unnecessary biopsies of US-BI-RADS 4a lesions through a deep learning method for residents-in-training: a cross-sectional study. BMJ Open (2020) 10(6):e035757. doi: 10.1136/bmjopen-2019-035757

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Kim K, Song MK, Kim EK, Yoon JH. Clinical application of s-detect to breast masses on ultrasonography: a study evaluating the diagnostic performance and agreement with a dedicated breast radiologist. Ultrasonography (2017) 36(1):3–9. doi: 10.14366/usg.16012

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Fujioka T, Kubota K, Mori M, Kikuchi Y, Katsuta L, Kasahara M, et al. Distinction between benign and malignant breast masses at breast ultrasound using deep learning method with convolutional neural network. Japanese J Radiol (2019) 37(6):466–72. doi: 10.1007/s11604-019-00831-5

CrossRef Full Text | Google Scholar

22. Choi JS, Han BK, Ko ES, Bae JM, Ko EY, Song SH, et al. Effect of a deep learning framework-based computer-aided diagnosis system on the diagnostic performance of radiologists in differentiating between malignant and benign masses on breast ultrasonography. Korean J Radiol (2019) 20(5):749–58. doi: 10.3348/kjr.2018.0530

PubMed Abstract | CrossRef Full Text | Google Scholar

23. Choi JH, Kang BJ, Baek JE, Lee HS, Kim SH. Application of computer-aided diagnosis in breast ultrasound interpretation: Improvements in diagnostic performance according to reader experience. Ultrasonography (2018) 37(3):217–25. doi: 10.14366/usg.17046

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: S-detect, breast imaging reporting and data system, BI-RADS, breast cancer, ultrasound

Citation: Xing B, Chen X, Wang Y, Li S, Liang Y-K and Wang D (2022) Evaluating breast ultrasound S-detect image analysis for small focal breast lesions. Front. Oncol. 12:1030624. doi: 10.3389/fonc.2022.1030624

Received: 17 September 2022; Accepted: 21 November 2022;
Published: 13 December 2022.

Edited by:

Pasquale Cianci, Azienda Sanitaria Localedella Provincia di Barletta Andri Trani (ASL BT), Italy

Reviewed by:

Bilgin Kadri Aribas, Bülent Ecevit University, Turkey
Marina Minafra, Ospedale civile “Lorenzo Bonomo”, Italy

Copyright © 2022 Xing, Chen, Wang, Li, Liang and Wang. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Dawei Wang, nit05wangdawei@163.com; Ying-Kui Liang, liangyingkui2012@sina.com

These authors have contributed equally to this work and share first authorship

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.