A functional magnetic resonance imaging (fMRI)-compatible fiber-optic bend sensor was investigated to assess whether the device could be used effectively to monitor opening and closing of the jaw during an fMRI experiment at 3 T. In contrast to surface electromyography, a bend sensor fixed to the chin of the participant is fast and easy to use and is not affected by strong electromagnetic fields. Bend sensor recordings are characterized by high validity (compared with concurrent video recordings of mouth opening) and high reliability (comparing two independent measurements). The results of this study indicate that a bend sensor is able to record the opening and closing of the jaw associated with different overt speech conditions (producing the utterances /a/, /pa/, /pataka/) and the opening of the mouth without speech production. Data post-processing such as filtering was not necessary. There are several potential applications for bend sensor recordings of speech-related jaw movements. First, bend sensor recordings are a valuable tool to assess behavioral performance, such as response latencies, accuracies, and completion times, which is particularly important in children, seniors, or patients with various neurological or psychiatric conditions. Second, the timing information provided by bend sensor data may improve the predicted hemodynamic response that is used for fMRI analysis based on the general linear model (GLM). Third, bend sensor recordings may be included in GLM analyses not for statistical contrast purposes, but as a covariate of no interest, accounting for part of the data variance to model fMRI artifacts due to motion outside the field of view.
In recent years, a growing number of functional magnetic resonance imaging (fMRI) studies on the neural correlates of overt speaking (Abrahams et al., 2003
; Gracco et al., 2005
; Bohland and Guenther, 2006
; Sörös et al., 2006a
; Christoffels et al., 2007
; Riecker et al., 2008
) and singing (Riecker et al., 2000
; Kleber et al., 2007
) have been published. A major advantage of investigating overt speech in contrast to silent (or covert) speech is the possibility to monitor the behavioral performance during the experiment (Abrahams et al., 2003
). Assessing task performance such as response latency or movement amplitude is especially important in studies involving children, seniors, or patients with various neurological or psychiatric conditions who might give delayed responses or even omit responses.
The objective of the present study was to demonstrate the use of a fiber-optic bend sensor, fixed to the chin, to record speech-related jaw movements during fMRI. Fiber-optic based bend sensors have been used in various fields of biomedical research, including the monitoring of finger (Ku et al., 2003
; Gorbet et al., 2004
; Jindrich et al., 2004
; Gorbet and Sergio, 2007
) and ankle movements (Seto, 2000
; Seto et al., 2001
; MacIntosh et al., 2004
).
To study speech-related movements outside MRI systems, surface electromyography (EMG) is widely used (Smith, 1992
). In positron emission tomography (Murphy et al., 1997
) and magnetoencephalography (Sörös et al., 2003
), EMG has also been used frequently and successfully to monitor muscle function associated with overt speech production. The use of EMG in fMRI studies, however, is challenging (for a recent review on the recording of electrophysiological data during fMRI, see Laufs et al., 2008
). Time-varying radiofrequency pulses and magnetic field gradients may induce artifacts in EMG recordings and, conversely, movements of EMG electrodes and leads inside the magnetic field may cause artifacts in fMRI data (Ganesh et al., 2007
; MacIntosh et al., 2007
). In addition, the use of EMG in an fMRI study is limited by the need of an fMRI-compatible EMG system (e.g., BrainAmp, Brain Products, Gilching, Germany
1
), specialized EMG data post-processing (van Duinen et al., 2005
; MacIntosh et al., 2007
), and by time constraints. Setting up a combined fMRI–EMG experiment requires extra time for the preparation of the skin, the mounting of the electrodes and the checking of the electrode impedance. Increasing the length of an fMRI experiment is especially problematic in studies involving participants for whom the monitoring of behavioral performance is most important, such as the populations named above.
To record high-quality acoustic signals associated with overt speech production during an fMRI experiment, noise cancelling optical microphones have been utilized (e.g., FOMRI II microphone, Optoacoustics Ltd., Or-Yehuda, Israel
2
). These microphones reduce the intense acoustic noise, often exceeding 100 dB SPL, during fMRI experiments by real-time adaptive noise cancelling (Chambers et al., 2007
) and can be used even during continuous scanning. Acoustic recordings of overt responses are essential for the assessment of response accuracy, but do not represent oro-facial movements.
To record oro-facial movements associated with speech production, MRI-compatible cameras have been used successfully (e.g., MRC Systems, Heidelberg, Germany
3
) (Graham et al., 2009
). Such cameras, which must be integrated with a dedicated video capture system, can be difficult to mount in the magnet bore without obstructing the patient and head coil as they move to and from isocenter, and require appropriate line-of-sight from the camera to the mouth through apertures in the head coil. To determine the onset and the amplitude of movements based on video recordings, extensive image processing is also required.
Bend Sensor
Bend sensor recordings were performed using a flexible, fMRI-compatible fiber-optic bend sensor (S700 ShapeSensor, Measurand Inc., Fredericton, NB, Canada
4
) that allows one degree of freedom measurements (Figure 1
) (Seto et al., 2001
).
Figure 1. The fiber-optic bend sensor to monitor jaw movements. The active zone of the sensor (right, indicated in blue) is attached to the participant’s chin. The electronics (middle) are located outside the magnet room and consist of an LED and two photodiodes for transmission and detection of the light, respectively. A matchstick is shown for scale.
This sensor contains a single strand plastic optical fiber (transmit and return paths, diameter = 0.25 mm) encased in a plastic cylindrical sheath of sufficient length (i.e., 6 m) to route from the operator console to the center of the magnet bore through an appropriately located wave guide in the radiofrequency shield enclosing the magnet room. The reflectiveness of one side of the optical fiber cladding is purposely degraded in the “U-shaped” active zone within the cylindrical sheath, such that attenuation of light in the fiber is proportional to how much the fiber is bent. Bend sensor signals were recorded with an associated electronics box, consisting primarily of one light emitting diode (LED) for transmission and two photodiodes for signal reception, which was located near the operator console outside the magnet room.
Characteristics of the Bend Sensor Signal
To determine the warm-up characteristics of the bend sensor, the sensor was fixed in position outside the magnet. Measurements of the sensor signal were acquired (USB-6008, National Instruments, Austin, TX, USA) at 400 Hz for 20 min.
To assess the stability of bend sensor measurements after the warm-up phase, the bend sensor was fixed in place and allowed to warm up for over 30 min. The sensor signal was then sampled at 400 Hz for 5 min to assess stability. Measurements were conducted inside the magnet during a routine fMRI scan (EPI, TR/TE/FA = 2000 ms/30 ms/70°, 64 × 64 matrix, 200 mm FoV, 28 axial slices, 5 mm thick), with a phantom, in a 3-T MR imager (Magnetom Tim Trio, Siemens, Erlangen, Germany). The tip of the sensor was about 35 cm from isocenter such that it was possible to position and manipulate the sensor tip appropriately by reaching into the magnet bore and without moving the patient table, which would disrupt scanning.
To assess the dynamic performance and MRI-compatibility of the bend sensor, the sensor tip was manually slid between two points in space that were 2 cm apart on a ruler to simulate a moderate mouth movement, using music for temporal reference. Movements were repeated at two speeds (approximately 0.25 and 0.5 Hz), both inside the magnet and outside the MRI room. Inside-magnet measurements were conducted during a routine fMRI scan (see above).
To assess the reliability of the bend sensor signal, a Pearson correlation coefficient was computed between these two independent bend sensor recordings (inside the magnet and outside the MRI room).
To study the validity of bend sensor recordings, the relationship between the bend sensor signal and mouth movements was assessed. The bend sensor was taped to the chin of a volunteer (Figure 2
) and allowed to warm up for several minutes. The signal was measured at 400 Hz, and an MRI-safe video camera (12M, MRC Systems GmbH, Heidelberg, Germany) in conjunction with a specially developed video capture system (Graham et al., 2009
) was used to record video at 30 Hz within the magnet bore. An electrical trigger was used to synchronize the sensor and video recordings before the volunteer performed several mouth movements. The video analysis consisted of manually selecting the center of a tape mark on the chin in each video frame and noting the pixel location.
Figure 2. Mounting of the bend sensor. The sensor (red) is fixed to the chin and chest of the participant with adhesive tape (3M Durapore, white).
Bend Sensor Recordings During Overt Speech Production
Participants
To study jaw movements in an actual fMRI experiment involving overt speech production, bend sensor signals were recorded in seven healthy younger volunteers (three women, four men; mean age = 25 years) and in eight healthy older adults (four women, four men; mean age = 70 years). All participants were right-handed and fluent speakers of English. The study was approved by the Research Ethics Board at Sunnybrook Health Sciences Centre. Informed consent for participation in the project was obtained from all subjects according to the Declaration of Helsinki. The present study is part of a broader project investigating the neural correlates of overt speech production (Sörös et al., 2006a
,b
, 2009
).
Tasks and fMRI data acquisition
Subjects were asked to repeat acoustically presented sub-lexical speech sounds of different complexity and to perform oral movements without articulation. The required responses were the long vowel /a/, a consonant-vowel syllable (/pa/), a three-syllabic utterance (/pataka/), and oral movements (opening the mouth or protruding the lips). Instructions were delivered through the audio Silent Scan Audio System at a constant onset-to-onset interstimulus interval of 10 s using padded headphones with a noise reduction rating of 30 dB to reduce acoustic fMRI noise. Subjects were asked to perform the given task or articulate the required response immediately after the end of the verbal instruction. Six experimental sessions were performed. Each session comprised six separate blocks of speech (five events per block, 50 s duration each), two blocks of oral movement (five events, 50 s duration each) and three blocks of baseline (30 s, no verbal cues). Blocks were grouped in a pseudo-randomized order with sessions 1, 3, 5 and 2, 4, 6 having the same order. For fMRI, clustered image acquisition (sparse temporal sampling) was performed (Edmister et al., 1999
). All instructions were delivered and all responses were made within the silent interval between the acquisition of the fMR images thus separating the hemodynamic brain activation signals associated with acoustic fMRI noise from the speech-related behaviors of interest. The experiment is illustrated in Figure 3
. For a discussion of the details of the experimental setup, imaging parameters, and the subsequent functional brain maps the reader is referred to previous papers of our group (Sörös et al., 2006a
, 2009
).
Figure 3. Recordings of jaw movements and speech during the acquisition of fMRI data. The figure displays five events in a representative data set. (A) The pre-recorded auditory instruction used for the vowel condition (“say a”). (B) Microphone recording of the subject’s overt responses (/a/) and of the fMRI acoustic noise during image acquisition. Both the auditory cue (A) and the overt response (B) fall within the silent interval between multislice data acquisition. (C) Bend sensor recording. The opening of the jaw precedes the speech signal. The bend sensor recording is not affected by fMRI data acquisition.
Bend sensor recordings
The sensor was attached to the chin and the chest of each participant immediately before the fMRI experiment using adhesive tape (Figure 2
). Paper tape (3M, St. Paul, MN, USA) was used in the initial experiments, which was not adhesive enough and led to a dislocation of the sensor tip at the chin. Medical silk tape (3M Durapore, St. Paul, MN, USA), in contrast, was used in all following measurements and provided sufficient adhesion.
Bend sensor signals were digitized at 40 Hz and recorded using a custom-written Labview program (National Instruments, Austin, TX, USA
5
). Collection of bend sensor and fMRI data were synchronized by the stimulation software E-Prime (Psychology Software Tools, Pittsburgh, PA, USA
6
). The onset latency and the amplitude of jaw opening were calculated using a custom-written program for the statistical package R
7
. Bend sensor recordings were baseline corrected and normalized to the largest amplitude in the entire session. The relative peak amplitude and the onset of jaw opening were determined for each epoch separately. To assess the onset of jaw opening, the mean and standard deviation (SD) of the data points prior to the first instruction in each experimental run were calculated to quantify the bend sensor signal baseline. The length of the baseline was 1600 points (40 s). Task timing details are given below. Movement onset was defined as the time point at which the signal exceeded the mean value of the baseline + 5 SD. For comparison between jaw movements and the acoustic speech signal, the participants’ overt responses were recorded via the microphone channel of the fMRI-compatible Silent Scan Audio System (Avotec, Stuart, FL, USA
8
), digitized at 44.1 kHz, and stored as an audio file on a PC.
Characteristics of the Bend Sensor Signal
During the warm-up phase with the sensor fixed, the signal drifts slowly by about 4% over approximately 15 min (Figure 4
). After the warm-up phase, the SD of the signal is 1.56% (inside the magnet) and 1.86% (outside the magnet room) during a 5-min recording (sampling rate 400 Hz, 120,000 data points; data not shown).
Figure 4. Warm-up characteristics of the bend sensor signal.
A comparison of bend sensor recordings inside the magnet and outside the magnet room with different frequencies (0.25 and 0.5 Hz) revealed no artefacts or distortions of the sensor signal due to fMRI scanning (Figure 5
).
Figure 5. Bend sensor recordings during manual movements. Recordings were performed outside the magnet room (A,C) and inside the magnet (B,D) with a frequency of approximately 0.5 Hz (A,B) and 0.25 Hz (C,D).
A Pearson correlation was performed between these bend sensor recordings (Figure 5
). For an alpha level of 0.05, the correlation between the two recordings at a frequency of 0.25 Hz was found to be statistically significant [r(7998) = 0.98, p < 0.0001, R2 = 0.96]. A statistically significant correlation was also found between two recordings at a higher frequency of approximately 0.5 Hz [r(7998) = 0.92, p < 0.0001, R2 = 0.84].
A Pearson correlation addressed the relationship between mouth opening as determined by the analysis of a video recording of chin movement and the concurrent bend sensor signal (Figure 6
). For an alpha level of 0.05, the correlation between mouth opening and sensor signal was found to be statistically significant [r(164) = 0.94, p < 0.0001, R2 = 0.89]. Figure 6
also shows that the onset and offset of mouth movement are measured very similarly by both recordings.
Figure 6. Comparison of a video recording and a concurrent bend sensor recording of mouth opening. The blue curve represents the opening and closing of the mouth, determined with an fMRI-safe video camera. The red curve represents the bend sensor signal in arbitrary units.
Bend Sensor Recordings During Overt Speech Production
The MRI-compatible bend sensor was able to record the opening and closing of the jaw for all overt speech conditions (/a/, /pa/, /pataka/) used in this experiment. Bend sensor recordings from 6 individual participants are shown in Figure 7
. In two initial measurements, however, the sensor was dislocated during the experiment, probably due to perspiration and an insufficiently adhesive tape. In the movement condition, two different oral movements were required, opening the mouth and protruding the lips without mouth opening. Mouth opening without speech production was also associated with a reliable signal in all participants, often with the highest amplitude across all conditions. As expected, protruding the lips did not reveal consistent signals (Figure 7
). Figure 8
illustrates the consistency of bend sensor recordings between blocks of responses for one condition (here: /a/ condition, data of one participant). Inter-individual variability of the onset of the bend sensor signal during opening of the jaw is shown in Figure 9
. Signal onset times were not significantly different between younger and older participants.
Figure 7. Bend sensor recording of 6 individual participants. Data from one run of three younger (A) and three older individuals (B) are shown in blue. In the runs shown here, the participants performed all verbal tasks correctly as verified by the concurrent audio recording. All bend sensor recordings are baseline-corrected (0–40 s prior to the beginning of the first block) and normalized to the largest deflection in the entire data set which was set to a relative amplitude of 100%. The dotted lines mark the beginning and end of blocks of five responses. Kiss denotes protruding the lips. Mouth denotes opening the mouth without producing an utterance.
Figure 8. Overlay of bend sensor data containing all 12 blocks of the /a/ condition recorded in one participant. The recordings are baseline-corrected (0–5 s prior to the beginning of the block).
Figure 9. Inter-individual variability of signal onset. Individual data is shown for all four experimental conditions. Mouth denotes the opening of the mouth without articulation. /a/, /pa/, /pataka/ denote the overt production of the respective utterance. Data of younger participants are shown in red, of older participants in blue. Average signal onset times in the younger participants were (mean ± SD): 1.27 ± 0.17 s (mouth), 1.77 ± 0.20 s (/a/), 1.97 ± 0.26 s (/pa/) and 2.94 ± 0.26 s (/pataka). Signal onset times in the older participants were: 1.09 ± 0.18 s (mouth), 1.50 ± 0.17 s (/a/), 1.71 ± 0.31 s (/pa/) and 2.80 ± 0.42 s (/pataka/). Signal onset times were not significantly different between age groups (ANOVA).
In many speech production experiments, monitoring the behavioral performance of the participants is desired. Outside the scanner, recording of speech-related muscle activity and of overt speech can be easily performed using surface EMG and a microphone connected to a computer. Inside the scanner, the use of these techniques is limited by the acoustic noise and the strong time-varying electromagnetic fields during fMRI (Ganesh et al., 2007
). In this study, recording the opening and closing of the jaw using a fiber-optic, fMRI-compatible bend sensor proved to be a valuable alternative to the established methods, demonstrating good reliability and validity (Figures 5 and 6
). Bend sensor recordings were found to be valid as assessed by a comparison between video recordings and concurrent bend sensor recordings of mouth opening (Figure 6
). There was a close agreement between the onset of mouth opening in both methods, emphasizing the usefulness of the bend sensor to monitor response latencies. The shape of the bend sensor signal curve, however, showed some slight differences compared to the analysis of the video recording, particularly a more rounded time course during the period when the mouth was most open. These differences could be due to several effects. First, the sensor signal is sensitive to bend angle and does not record linear displacements. Consequently, a deviation would be expected between bend sensor signals and displacement data, when displacements become large. However, in Figure 6
the bend sensor signals continue to increase toward a signal plateau during periods when the video data record a quite consistent displacement indicative of holding the mouth open in a fixed position. This is more likely explained by the fact that the bend angle and thus the sensor signal is influenced not only by the actual mouth movement but also by the fixation of the sensor and its potential movement when being bent. Parameters that influence the fixation of the sensor include the adhesiveness of the tape used, the angle of the sensor tip relative to the chin, the participant’s skin surface environment and the condition of the participant’s clothing on which the sensor is fixed.
The bend sensor assesses the opening and closing of the mouth, which appears to be more closely related to the activation of cortical and subcortical cognitive and motor areas than voice onset. In a variety of sounds, mouth movements can start substantially earlier than sound production, e.g., in the stop consonant /p/. Thus, the bend sensor signal is useful to determine the response latency for mouth movement, but, in general, cannot be used to differentiate between the mouth movements associated with different sounds produced during an experiment. It has to be noted that the onset of the bend sensor signal showed considerable inter-individual variability (Figure 9
), which should be minimized to improve comparisons across conditions or subject groups. Subject variability could be reduced by giving an explicit instruction (“respond as rapidly as possible with equal amplitude”) and adding a short training session before starting with the actual experiment.
The preparation of the volunteer, in particular fixing the sensor with adhesive tape and testing the signal, was considerably faster than the preparation usually needed for EMG studies. The bend sensor, however, has to be carefully fixed at the chin. Probably due to perspiration and insufficiently adhesive paper tape, the sensor was dislocated during two of the initial measurements. In the following experiments, the sensor was fixed with silk tape without dislocation during the entire measurement. To ensure a tight contact between the adhesive tape and the skin, the participants should be asked to wash or shave their chin before the experiment. In addition, participants should wear light clothing to avoid excessive perspiration. In contrast to EMG (Ganesh et al., 2007
), Figure 5
indicates that fMRI data acquisition has no impact on the quality of bend sensor recordings. Previous work has shown that use of fiber-optic bend sensing technology has no statistically significant impact on fMRI data quality (Di Diodato et al., 2007
).
Based on these specifications, a bend sensor is not only advantageous for recording speech-related jaw movements, but also for monitoring oral movements that serve as control conditions in speech production experiments (Sörös et al., 2006a
). The use of a chin-mounted bend sensor, however, is limited to speech sounds and oral movements involving the opening of the mouth such as the vowel /a/. Close vowels without considerable mouth opening such as /i/ or mouth movements such as protruding the lips cannot be monitored with a single bend sensor fixed at the chin. Other mounting arrangements of a one degree of freedom sensor or the use of multiple sensors or even a higher degree of freedom device (e.g., ShapeTape, Measurand Inc., Fredericton, NB, Canada) are possible and should be tested.
Bend sensor recordings can also be used as purely behavioral measurements to assess response latencies and accuracies. Bend sensor recordings have already been employed to compare the latency of verbal responses between younger and older adults with the experimental setup presented here (Sörös et al., 2009
). Bend sensor recordings are also valuable for assessing responses of patients who, when an overt response is required, perform articulatory movements without vocalization (e.g., in aphasia).
For fMRI data analysis, bend sensor recordings could help to improve the predicted hemodynamic response function that is used for statistical analysis based on the general linear model by incorporating the actual timing of movements and by omitting missing responses. In addition, bend sensor recordings may be included in the linear model as covariate of no interest (confound explanatory variable). Covariates of no interest are expected to account for part of the data variance without being used for statistical contrasts. Covariates of no interest were used, e.g., to model the effect of head motion (Johnstone et al., 2006
) and laryngeal motion (Sörös et al., 2008
) on fMRI data. In experiments on speech production, bend sensor recordings could be employed to model artifacts associated with jaw movements (Birn et al., 1999
). Motion outside the field of view, such as movement of the oral cavity, the sinuses or the pharynx, might cause magnetic field inhomogeneities masking brain activation or generating artifactual intensity changes (Yetkin et al., 1996
). For future research, the effect of introducing jaw motion as covariate of no interest in a fMRI analysis should be compared with the effect of other techniques of artifact reduction, such as the use of head motion parameters as covariates of no interest or the removal of artifactual signal components based on independent component analysis (Sörös et al., 2008
).
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
The authors wish to thank Nicole Baker for developing the data acquisition software used for bend sensor recordings and Dr. Gary Glover for providing his spiral-in/out pulse sequence. The authors also wish to thank two reviewers for insightful comments and suggestions that improved substantially the manuscript. This study was supported by the Heart and Stroke Foundation of Ontario through its Centre for Stroke Recovery, the Ontario Research and Development Challenge Fund and GE Healthcare Canada.
Abrahams, S., Goldstein, L. H., Simmons, A., Brammer, M. J., Williams, S. C., Giampietro, V. P., Andrew, C. M., and Leigh, P. N. (2003). Functional magnetic resonance imaging of verbal fluency and confrontation naming using compressed image acquisition to permit overt responses. Hum. Brain Mapp. 20, 29–40.