Skip to main content

ORIGINAL RESEARCH article

Front. Neurol., 18 May 2022
Sec. Endovascular and Interventional Neurology

Application of the Willis Covered Stent in the Treatment of Blood Blister-Like Aneurysms: A Single-Center Experience

\nWei Fang
&#x;Wei Fang1*Jia Yu&#x;Jia Yu1Yufeng Liu&#x;Yufeng Liu1Peng SunPeng Sun1Zijian YangZijian Yang1Zhenwei ZhaoZhenwei Zhao1Yue HeYue He2Jianping Deng
Jianping Deng1*Tao Zhang
Tao Zhang1*
  • 1Department of Neurosurgery, Tangdu Hospital, Air Force Medical University, Xi'an, China
  • 2Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China

Objective: To evaluate the effectiveness of the Willis covered stent (WCS) in the treatment of ruptured blood blister-like aneurysms (BBAs) of the internal carotid artery (ICA).

Method: The clinical data of 16 patients consecutively treated with WCSs from December 2015 to January 2019 were retrospectively analyzed. Clinical data and angiographic findings were analyzed by two experienced neuroradiologists and neurosurgeons, including age, sex, Hunt and Hess (H&H) grade at admission, modified Rankin scale (mRS) score, aneurysm size, and location, the diameter of the patent artery in proximal and distal ends, stent size, rate of aneurysm occlusion, procedure-related complications, and follow-up.

Results: All the 16 patients (five males, 11 females) with ICA BBAs underwent WCS deployment successfully. The median age was 49 years (range, 29–72). All patients had complete aneurysm occlusion on immediate postoperative angiography. Anterior choroidal artery (AChA) was occluded in one patient accidentally while no obvious neurological dysfunction was observed. However, this patient underwent subarachnoid hemorrhage 1 day after the treatment; endoleak and aneurysm recurred, and the patient died 10 days later. Therefore, the effective rate of WCS treatment was 93.8% (15/16), and procedure-related complications rate was 6.3% (1/16). Moreover, one patient was urgently treated because of accidental aneurysm rupture after anesthesia, and external ventricular drainage was then performed postoperatively. Another patient developed coma and hemiplegia 3 days after treatment, with emergency angiography showing in-stent thrombosis and ICA occlusion which was recanalized with arterial rt-PA thrombolysis; the patient recovered completely. The clinical follow-up period was 3–30 months in 14 patients. The mRS scores were 0 in 12 patients (85.7%) and 4 in 1 case (7.1%), while 1 patient (7.1%) died 6 months postoperatively for unknown reasons. Angiographic follow-up was performed in 13 patients, and no recurrence was observed. However, ICA occlusion without neurological deficit was observed in one patient.

Conclusion: Based on careful preoperative evaluation, appropriate WCS size selection, and precise surgical operation, WCSs may provide an alternative and effective solution for blood BBAs via aneurysm isolation and ICA reconstruction immediately; However, further follow-up studies with larger samples are required.

Introduction

Blood blister-like aneurysm (BBA) is often characterized with a fragile wall, a wide neck, and lack of distinguishable normal boundary between the aneurysm neck and parent vessel (1). Due to its special pathological features, neurosurgical treatment for BBA has been a big challenge compounded with the high odds of intraoperative rupture and postoperative recurrence. Different treatment modalities, including direct clipping, suturing, wrapping, coiling, stent-assisted coiling, multiple overlapping stents, and flow–diverter, have been applied for this terrible disease. Endovascular treatments have a lower risk of intraprocedural rupture but lower rates of aneurysm obliteration, while microsurgical treatments have higher rates of aneurysm obliteration but higher risk of intraprocedural rupture and subsequent ICA sacrifice (25). Among all these treatments, the most optimal management option for BBAs remains uncertain (1, 36).

In recent years, Willis covered stents (WCSs) have been specially designed for intracranial application, which brings a new option for the treatment of aneurysms, especially for BBAs (7). The WCS deployment has unique advantages compared with other treatments. It can isolate aneurysm and reconstruct the parent artery instantly without manipulations in the aneurysm (8). Several retrospective studies have evaluated the effectiveness of WCSs for BBAs with promising results (3, 9, 10). However, whether it is a reasonable choice for the BBA remains debatable considering the technical details. Here, we report our experience and evaluate the effectiveness in the treatment of BBAs of the internal carotid artery (ICA) with the WCS in 16 patients.

Materials and Methods

Patients

A retrospective study was performed in 16 patients with ruptured ICA BBAs treated with WCSs from December 2015 to January 2019. The examined data included patient's age, sex, Hunt and Hess (H&H) grade at admission, modified Rankin scale (mRS) score, aneurysm size and location, the diameter of the patent artery in proximal and distal ends, stent size, rate of aneurysm occlusion, procedure-related complications, and follow-up. This study was approved by the Ethics Committee of Tandu Hospital. Written informed consent was obtained from each patient for the purpose of research.

The CT and DSA Diagnostic Criteria of BBA

All patients were examined clinically and radiologically by cerebral computed tomography (CT) to confirm subarachnoid hemorrhage before treatment. Aneurysms arising from the anterior wall of the ICA may be saccular aneurysms or blister-like aneurysms. However, these two subgroups are difficult to be identified angiographically. The most reliable method to diagnose BBA is microsurgical morphology and pathology. As the technical limitation of endovascular therapy, we adopt six criteria to diagnose BBA according to the following literature: (1) Aneurysms located at the non-branching sites of the supraclinoid ICA. (2) Typical aneurysm features, such as a wide neck and a typical blister-like shape. (3) Initially small size (maximum diameter <10 mm). (4) Ruptured, presented with SAH. (5) Rapid growth (<2 weeks) on repeated angiograms. (6) Irregular wall of the aneurysm or the parent artery. An aneurysm was diagnosed as BBA when criteria 1–4 were fulfilled and either of criteria 5 or 6 was matched (2, 1113).

WCS and Stent Deployment Procedure

The WCS consists of three parts, including a bared balloon-expandable stent, an expandable polytetrafluoroethylene (ePTFE) membrane, and a low-pressure flexible balloon catheter. Various diameters (3.5, 4.0, and 4.5 mm) and different lengths (7, 10, 13, and 16 mm) have been designed for different conditions. The general principle for stent selection is that the length of the stent must exceed the aneurysm neck at least by 2 mm for both sides, with the diameter of the stent imperatively exceeding the parent artery by 0–0.5 mm (3, 9).

Endovascular treatment was performed under general anesthesia and systemic venous heparinization. An 8F guide catheter (Envoy, Codman Neurovascular, USA) was placed into the common carotid artery (CCA) or the initial part of the ICA. A 5F Navien (ev3/Covidien, USA) supporting the catheter was generally advanced to the M1 segment of the middle cerebral artery (MCA) with an XT-27 microcatheter (Stryker, USA) and a 0.014'-microguidewire (Sychro, Stryker, USA) under roadmap guidelines. In case of a too tortuous proximal vessel, the 5F Navien was then advanced to the C4 segment of the ICA at least. With a microguidewire placed on the M2 segment of the MCA, the WCS was then carefully delivered upward to the target location. After angiographic confirmation was made that the stent had completely covered the neck of the aneurysm and not covering any important perforators, including the ophthalmic (OA) and anterior choroidal arteries, the stent was then deployed by balloon inflation at 5–6 atm. In case of an endoleak, the balloon was re-inflated with higher pressure until it was eliminated. However, the manipulation for re-inflation would not be performed for more than 3 times. If the endoleak persisted, another WCS could be deployed. The postoperative CT and neurologic examination were performed routinely to rule out intracranial hemorrhage or ischemic events.

Antithrombotic Treatment

Intravenous tirofiban hydrochloride was administered immediately after stent deployment at an initial bolus of 10 μg/kg within 3 min and maintained at a rate of 0.15 μg/kg/min for 36 h. A loading dose of 300-mg or 75-mg clopidogrel and 300-mg or 100-mg aspirin was administered 6 h before tirofiban withdrawal. All patients were maintained on dual antiplatelet therapy (75-mg/day clopidogrel and 100-mg/day aspirin) for a minimum of 6 months. Aspirin (100 mg/day) was then taken for at least 1 year. Before discharge, all patients underwent thrombosis elastography (TEG) to confirm the drug effect.

Follow-Up and Postoperative Outcome Evaluation

Among all 16 patients, 14 were assessed by angiography 7 days post-treatment. The other 2 patients received emergency angiography 1- and 3-days post-treatment, respectively, for deteriorative neurological symptoms. All patients were then followed up at 1, 3, 6, and 12 months, via clinical and angiographic assessments.

Results

General Characteristics

Our cohort consisted of 16 patients (11 women and 5 men) with a median age of 49 years (range, 29–72). As shown in Table 1, patients with primary H&H grade II accounted for 56.25%, and the BBAs were mainly located at the ICA C6 segment. One patient received decompressive craniectomy before admission to our hospital. One patient received an additional WCS after primary stent deployment for continuous endoleak.

TABLE 1
www.frontiersin.org

Table 1. Demographic, clinical, CT data, endovascular treatment, and follow-up outcomes of the 16 patients.

Procedural Results and Complications

A total of 17 stents were successfully deployed and all 16 aneurysms were completely occluded for immediate angiography postoperatively. In our definition, procedure-related complication only means complications that occurred during stent delivery and deployment procedure. Thus, the procedure-related complications occurred in one case (6.3%). In this patient, the AChA was covered by the WCS accidentally and was subsequently occluded (see Patient No. 13 in Table 1) during immediate angiography, while no newly neurological deficit occurred postoperatively. Three patients experienced perioperative complications. The patient with AChA occlusion fell into coma on the next day, and CT scan showed a significant increase in cerebral hemorrhage. Previously occluded BBA and AChA were visualized again in subsequent angiography, the reason for which was considered to be the shrinkage of the WCS membrane. Although we advised performing a craniotomy to save his life, the patient's family refused any further treatment. This patient died 10 days later. Moreover, one patient (Patient No. 3 in Table 1) suffered an unexpected aneurysm rupture during the roadmap procedure after anesthesia. After an urgent WCS deployment, the aneurysm was completely occluded, and lateral ventricle external drainage was performed immediately post-treatment. No neurological deficit occurred postoperatively and the mRS score was 1 at discharge. However, ipsilateral ICA occlusion was found in a follow-up Magnetic Resonance Angiography (MRA) 12 months postoperatively in this patient. In another case, in-stent thrombosis was found by angiography 2 days after the treatment (Patient No. 6 in Table 1). However, revascularization was achieved by intra-arterial administration of 3.5 mg of tirofiban and 9 mg of alteplase, and the patient had no neurological deficit at discharge.

Follow-Up Results

As described in Table 1, 14 patients received clinical and radiographic evaluations at follow-up ranging from 3 to 30 months, and 12 (85.7%) obtained good outcomes (mRS score = 0). One case (7.1%) improved from the mRS score of 5 to 4. Another patient died 6 months after treatment for an unknown reason. Of all 16 patients, 13 (81.25%) achieved good recovery after treatment and one was lost to follow-up.

Angiographic follow-up was performed in 13 patients from 3 to 14 months after the treatment. No aneurysm recurrence was found, but ipsilateral ICA occlusion was found in case 2 (Figure 2).

Special Cases

Patient 1

An adult patient (see Patient No. 4 in Table 1) was admitted to our hospital with intense headache (Figure 1). Emergency angiography demonstrated a BBA (3.0*1.5 mm) located on the anterior wall of the C6 segment of the R-ICA; the diameters of the patent artery in proximal and distal ends were 3.4 and 3.2 mm, respectively. A 3.5*7-mm WCS was deployed successfully, and immediate postoperative angiography showed complete aneurysm occlusion. The 13-month follow-up angiography showed complete occlusion with good parent artery patency. The patient had a satisfactory clinical status with the mRS score of 0 at the last follow-up examination (Figure 1).

FIGURE 1
www.frontiersin.org

Figure 1. Angiographic images of a 50-year-old woman (see Patient No. 4 in Table 1) with a right ICA C6 BBA. (A) A BBA (3.0*1.5 mm) located on the anterior wall of the C6 segment. (B) A 3.5*7-mm WCS was deployed (arrow shows). (C) Immediate postoperative angiography showed complete aneurysm occlusion, and the OA and ACHA were not affected. (D) The 3D-DSA showing complete aneurysm occlusion and the parent artery's morphology. (E) Dual volume technology showing the morphology of the stent and its relationship with the parent artery. (F) Angiography images showing complete aneurysm occlusion at the 13-month follow-up examination.

Patient 2

An adult patient (see Patient No. 3 in Table 1) was admitted to our hospital with intense headache for 3 days (Figure 2). The aneurysm ruptured during preoperative angiography unexpectedly. A BBA (3.8*6.9 mm) located on the anterior wall of the C6 segment of the R-ICA; the diameters of the patent artery in proximal and distal ends were 3.2 and 2.9 mm, respectively. After a 3.5*10 mm WCS was deployed immediately, the aneurysm was completely occluded, and the lateral ventricle external drainage was performed. Although the mRS score was 0 at 12 months follow-up, ipsilateral ICA occlusion was found by an MRA examination (Figure 2).

FIGURE 2
www.frontiersin.org

Figure 2. Angiographic images of a 43-year-old woman (see Patient No. 3 in Table 1) with a L-ICA C6 BBA. (A) A BBA (3.8*6.9 mm) located on the anterior wall of the C6 segment. (B) The aneurysm ruptured during preoperative angiography (black arrow). (C,D) A 3.5*10-mm WCS deployed immediately (white arrow). (E) Postoperative angiography showing complete aneurysm occlusion, and the OA was also affected. (F) The MRA images showing the L-ICA was occluded completely.

Patient 3

An adult patient (see Patient No. 8 in Table 1) was admitted to our hospital with intense headache and a history of vomiting for 6 h (Figure 3). The BBA (2.2*1.9 mm) was located on the anterior wall of the C6 segment of the left internal carotid artery (L-ICA); the diameters of the patent artery in proximal and distal ends were 3.5 and 2.2 mm, respectively. After a 3.5*10-mm WCS deployment and two balloon re-inflation procedures, the endoleak phenomenon remained. Therefore, another 3.5*7 mm WCS was joined proximally, and the aneurysm was completely occluded (Figure 3).

FIGURE 3
www.frontiersin.org

Figure 3. Angiographic images of a 65-year-old man (see Patient No. 8 in Table 1) with a L-ICA C6 BBA. (A) A BBA (2.2*1.9 mm) located on the anterior wall of the C6 segment (white arrow). (B,C) A 3.5*10-mm WCS was deployed (black arrow). (D) After re-inflation, the endoleak phenomenon remained (white arrow). (E) A 3.5*7-mm WCS was joined proximally. (F) Angiographic images showing complete aneurysm occlusion.

Patient 4

An adult patient (see Patient No. 13 in Table 1) was admitted to our hospital with intense headache for 1 day. The BBA (1.8*5.1 mm) was located on the lateral wall of the C6 segment of the L-ICA; the diameters of the patent artery in proximal and distal ends were 3.5 and 3.3 mm, respectively. After a 3.5*10-mm WCS deployment, the aneurysm and the AChA were both occluded. The occlusion of the AChA was an accident that was not part of our surgical plan, as we had already considered avoiding this vessel when selecting the stent length and designing the stent deployment. However, the stent migration during balloon dilation and occlude the AChA. Although the patient had no new neurological deficit postoperatively, he fell into coma on the second day. Emergency angiography showed aneurysm recurrence, and the AChA was visualized again (Figure 4).

FIGURE 4
www.frontiersin.org

Figure 4. Angiographic images of a 49-year-old man (Patient no. 13 in Table 1) with a L-ICA C6 BBA. (A) A BBA (1.8*5.1 mm) located on the lateral wall of the C6 segment (white thin arrow); the AHCA was indicated by the white fat arrow. (B,C) A 3.5*10-mm WCS was deployed (black arrow). (D) Immediate postoperative angiography showing the aneurysm and the ACHA were both occluded. (E,F) DSA and dual volume technology showing aneurysm recurrence, and the AChA appeared again. (G) Preoperative CT scan showing subarachnoid hemorrhage. (H) The CT scan showing significantly increased hematoma as the patient fell in coma.

Discussion

Although neurosurgical techniques and instruments have made great progress in the past decades. Both treatments for BBAs are still greatly challenging for neurosurgeons and neuroradiologists due to the special features of BBAs. Currently, the main treatment options for BBAs include surgical clipping, artery suturing, wrapping and clipping, stent-assisted coiling embolization, multi-stent implantation, and flow-diverting stents (2, 6, 1417). However, a rescue or second treatment was required in 21% patients, and the overall morbidity and mortality rates were 17 and 15%, respectively. As endovascular treatment offers a lower procedural-related complication rate compared with surgical approaches, interventional therapy appears to be a more optimal treatment option for BBAs, with a more favorable patient outcome (16, 18). It should be pointed out that flow–diverter stent application in BBAs has achieved relatively more satisfactory results compared with traditional endovascular treatment in recent years (1922). However, the lower rate of the immediate aneurysm occlusion and the use of antiplatelet agents make it controversial.

Since Li et al. first reported the treatment of intracranial pseudoaneurysms with a Willis coated stent in 2006 (7), more and more studies have reported the efficacy of WCSs in the treatment of BBAs, with the majority of patients having promising results (3, 9, 10, 23). The prominent advantage of WCS deployment is the immediate occlusion of aneurysms from the circulation, especially compared with the flow–diverter stents. Furthermore, the deployment of the WCS involves no manipulations focusing on the aneurysm cavity, thus reducing the risk of procedural-related rupture. As described here, all the 16 patients achieved immediate aneurysm occlusion after stent deployment, and no procedural-related rupture occurred. Compared with surgical procedures and other endovascular treatments, WCS deployment is technically simple and safe.

Although WCSs have many advantages, their limitations should also be considered. As WCS is stiffer than the ordinary stents due to its membrane characteristics and the microguidewire transportation system, its delivery is much more difficult than for another stent, especially during tortuous vascular access (24). Therefore, this stent is not recommended for the tortuous vascular approach. According to our experience, the combined use of an 8F guiding catheter and a 5F Navien support catheter is an optional choice. In most cases, we preferred to advance the 5F Navien beyond the aneurysmal neck to the M1 segment of the MCA; thus, WCSs could be delivered within the Navien to the target position. In some cases, with tortuous proximal vessels, a 5F Navien would be advanced to the C4 segment of the ICA for best support. Moreover, vasospasm, calcification, and stenosis should also be considered cautiously as they may increase the procedure risks. Vasospasm may influence the accurate selection for stent size while inappropriate stent may lead to poor stent apposition and aneurysm recurrence. Calcification and stenosis may influence stent delivery, balloon inflation, and stent apposition. As for the choice of stent length, branching and perforating artery occlusion is a major concern. Therefore, the appropriate size and deployment precisely of the cover stent is the key for the operation. For the clinoidal segment of the ICA, the OA, posterior communicating (PComA), and AChA are the three major arteries close to the BBA site. During WCS application, we tried our best not to sacrifice these arteries by choosing appropriate stents. The PComA was the primary concern in this study; if the PComA is absent or hypoplastic, it could be sacrificed without hesitation. As for the OA, because the deep and superficial anastomotic network will ensure the blood supply of the central retina artery, only a few patients experience blindness when OA is occluded (3, 25). The AChA provides the blood supply to many vital structures, including the posterior two-thirds of the internal capsule, adjacent optic nerve and auditory radiation, the medial portion of the globus pallidus, and the tail of the caudate nucleus (3, 26, 27). Acute occlusion of the AChA would result in catastrophic consequences that could not be predicted. In this study, the AChA was occluded in one patient owing to stent migration during balloon dilation. Although this patient had no neurological deficit after the treatment, he died several days later because of BBA recurrence and rebleeding.

As for the choice of stent diameter, endoleak is another major concern, although it is more likely to happen in an application for giant aneurysms (8, 10, 28). Endoleak issues may include stent shrinkage, inappropriate choice of stent, and irregular appearance of the parent artery. Based on our experience in preventing endoleak, a stent equal to or a little bigger than the artery in diameter is preferred. In this study, immediate endoleak was observed in two cases which were solved by balloon re-inflation with higher pressure in one case and deployment of a second stent proximally in another case. The significance of the early angiographic review should also be emphasized for WCS application to prevent BBA recurrence. In this study, one patient (see Patient No. 13 in Table 1) had aneurysm recurrence and serious re-hemorrhage the next day post-treatment, which was confirmed by angiographic review. Stent shrinkage caused by inappropriate selection of stent diameter might be the reason. Another possible reason may be the vasospasm caused by subarachnoid hemorrhage. The diameter of the parent artery may be not accurate. Therefore, when the vasospasm is relieved, the increase in the artery diameter may cause poor stent apposition. To our knowledge, the stent should be equal to or slightly larger than the target vessel in diameter. If the stent applied is smaller than the target vessel (although the balloon dilated with higher pressure), the stent might retract to its normal size and the covered membrane could not isolate the lesion, which would result in aneurysm recurrence and re-bleeding. Furthermore, the vessel spasm should also be cautiously considered in the measurement of vessels. As ipsilateral vessel spasm might be serious, we suggest referring to a contralateral vessel in diameter during measurement.

In this study, in-stent stenosis was found in one patient (see Patient No. 3 in Table 1) during the angiographic follow-up 6 months later. The main reason is that WCSs are more thrombogenic than the ordinary stents. Stent malposition may also increase the risk of in-stent thrombosis.

Although this study showed a promising result for WCS application in BBA treatment, there were still two major limitations. First, the long-term follow-up should be carried out to confirm the safety and efficacy of WCSs in BBA. Second, the number of patients was relatively small, and larger clinical trials are needed to confirm these results.

Conclusion

The WCSs may provide an alternative and effective solution for BBAs by aneurysm isolation and immediate ICA reconstruction. However, further follow-up studies with larger samples are required for confirmation.

Data Availability Statement

The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author/s.

Ethics Statement

The studies involving human participants were reviewed and approved by Institutional Review Board, Tang DU Hospital. The patients/participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author Contributions

WF acquired and analyzed data, and wrote the first draft of the manuscript. JY, YL, and PS performed imaging analyses and acquired related data. ZY acquired clinical data. ZZ and YH contributed to the database. JD and TZ contributed to the conception and design of the study and reviewed the manuscript. All authors contributed to manuscript revision, read, and approved the submitted version.

Conflict of Interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Zhang X, Shen R, Zhao J, Sun J, Zhang Q, Chen Z. Using overlapping low-profile visualized intraluminal support stent-assisted coil embolization for treating blood blister-like aneurysms of the internal carotid artery. Neurosurg Rev. (2020) 44:1053–60. doi: 10.1007/s10143-020-01284-5

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Meling TR, Sorteberg A, Bakke SJ, Slettebø H, Hernesniemi J, Sorteberg W. Blood blister-like aneurysms of the internal carotid artery trunk causing subarachnoid hemorrhage: treatment and outcome. J Neurosurg. (2008) 108:662–71. doi: 10.3171/JNS/2008/108/4/0662

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Liu LX, Zhang CW, Xie XD, Wang CH. Application of the Willis covered stent in the treatment of blood blister-like aneurysms: a single-center experience and systematic literature review. World Neurosurg. (2019) 123:e652–60. doi: 10.1016/j.wneu.2018.11.245

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Wu YQ, Li LZ, Wang ZY, Zhang T, Xu M, Cheng MX. Endovascular intervention with a low-profile visualized intraluminal support stent versus surgical clipping for blood blister-like aneurysms: a retrospective study. Clin Neuroradiol. (2021) 31:417–24. doi: 10.1007/s00062-020-00886-2

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Capocci R, Shotar E, Di Maria F, Rolla-Bigliani C, Al Raaisi A, Andre A, et al. Delayed treatment (>/=5 days) by flow diversion of ruptured blister-like cerebral aneurysms: case series of 8 consecutive patients. Clin Neuroradiol. (2020) 30:287–96. doi: 10.1007/s00062-019-00758-4

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Meling TR, Patet G. Clip-wrapping of ruptured blood blister-like aneurysms of the internal carotid artery. Neurosurg Rev. (2020) 43:1365–71. doi: 10.1007/s10143-019-01172-7

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Li MH, Gao BL, Wang YL, Fang C, Li YD. Management of pseudoaneurysms in the intracranial segment of the internal carotid artery with covered stents specially designed for use in the intracranial vasculature: technical notes. Neuroradiology. (2006) 48:841–6. doi: 10.1007/s00234-006-0127-7

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Liu LX, Zhang CW, Lin S, Wu C, Wang T, Zhou LX, et al. Application of the willis covered stent in the treatment of ophthalmic artery segment aneurysms: a single-center experience. World Neurosurg. (2019) 122:e546–52. doi: 10.1016/j.wneu.2018.10.098

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Fang C, Tan HQ, Han HJ, Feng H, Xu JC, Yan S, et al. Endovascular isolation of intracranial blood blister-like aneurysms with Willis covered stent. J Neurointerv Surg. (2017) 9:963–8. doi: 10.1136/neurintsurg-2016-012662

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Liu Y, Yang HF, Xiong ZY, Zheng J, Liu CY, Zhao HY, et al. Efficacy and safety of willis covered stent for treatment of complex vascular diseases of the internal carotid artery. Ann Vasc Surg. (2019) 61:203–11. doi: 10.1016/j.avsg.2019.05.027

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Fang YB Li Q, Yang PF, Huang QH, Zhao WY, Xu Y, et al. Treatment of blood blister-like aneurysms of the internal carotid artery with stent-assisted coil embolization. Clin Neurol Neurosurg. (2013) 115:920–5. doi: 10.1016/j.clineuro.2012.09.022

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Ji T, Guo Y, Huang X, Xu B, Xu K, Yu J. Current status of the treatment of blood blister-like aneurysms of the supraclinoid internal carotid artery: a review. Int J Med Sci. (2017) 14:390–402. doi: 10.7150/ijms.17979

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Qi Y, Xu T, Jiang C, Wang Y, Liu H. Application of the Willis covered stent in the treatment of internal carotid artery blood blister-like aneurysms. Neurosurg Rev. (2021) 45:1513–9. doi: 10.1007/s10143-021-01666-3

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Nishi T, Kaji M, Koga K, Yamashiro S, Mizuno T, Hamasaki K, et al. Clipping on crossed wrapping method for ruptured blood blister-like aneurysm of the internal carotid artery: technical note and long-term results. World Neurosurg. (2019) 2:100005. doi: 10.1016/j.wnsx.2018.100005

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Yu J, Xu B, Guo Y, Xu K. Direct clipping of a blister-like aneurysm in the supraclinoid segment of the internal carotid artery: a clinical analysis of nine cases. Int J Clin Exp Med. (2015) 8:21786–95.

PubMed Abstract | Google Scholar

16. Gonzalez AM, Narata AP, Yilmaz H, Bijlenga P, Radovanovic I, Schaller K, et al. Blood blister-like aneurysms: single center experience and systematic literature review. Eur J Radiol. (2014) 83:197–205. doi: 10.1016/j.ejrad.2013.09.017

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Song J, Oh S, Kim MJ, Chung J, Lim YC, Kim BS, et al. Endovascular treatment of ruptured blood blister-like aneurysms with multiple (≥3) overlapping enterprise stents and coiling. Acta Neurochir. (2016) 158:803–9. doi: 10.1007/s00701-016-2721-8

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Liu Z, Zhang S, Wang S, Chen Q. Management of blood blister-like aneurysms of the internal carotid artery: lessons learned from direct clipping in 22 cases. World Neurosurg. (2017) 108:618–26. doi: 10.1016/j.wneu.2017.09.080

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Zhu D, Yan Y, Zhao P, Duan G, Zhao R, Liu J, et al. Safety and efficacy of flow diverter treatment for blood blister-like aneurysm: a systematic review and meta-analysis. World Neurosurg. (2018) 118:e79–86. doi: 10.1016/j.wneu.2018.06.123

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Nerva JD, Morton RP, Levitt MR, Osbun JW, Ferreira MJ, Ghodke BV, et al. Pipeline embolization device as primary treatment for blister aneurysms and iatrogenic pseudoaneurysms of the internal carotid artery. J Neurointerv Surg. (2015) 7:210–6. doi: 10.1136/neurintsurg-2013-011047

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Mokin M, Chinea A, Primiani CT, Ren Z, Kan P, Srinivasan VM, et al. Treatment of blood blister aneurysms of the internal carotid artery with flow diversion. J Neurointerv Surg. (2018) 10:1074–8. doi: 10.1136/neurintsurg-2017-013701

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Cerejo R, Bain M, John S, Hardman J, Moore N, Hussain MS, et al. Flow diverter treatment of cerebral blister aneurysms. Neuroradiology. (2017) 59:1285–90. doi: 10.1007/s00234-017-1936-6

PubMed Abstract | CrossRef Full Text | Google Scholar

23. Zhang Y, Zhang Y, Liang F, Jiang C. Procedure-related complication of willis covered stent in the treatment of blood blister-like aneurysm: stent detachment from dilating balloon. Front Neurol. (2017) 8:639. doi: 10.3389/fneur.2017.00639

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Ma L, Feng H, Yan S, Xu JC, Tan HQ, Fang C. Endovascular treatment of complex vascular diseases of the internal carotid artery using the willis covered stent: preliminary experience and technical considerations. Front Neurol. (2020) 11:554988. doi: 10.3389/fneur.2020.554988

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Perrini P, Cardia A, Fraser K, Lanzino G, A microsurgical study of the anatomy and course of the ophthalmic artery and its possibly dangerous anastomoses. J Neurosurg. (2007) 106:142–50. doi: 10.3171/jns.2007.106.1.142

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Tolla AS, Schichtel LT, Alboudi A, Sanders AK. Anterior choroidal artery infarct presenting as right hemichoreoathetosis. BMJ Case Rep. (2020) 13:e234861. doi: 10.1136/bcr-2020-234861

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Cheng Z, Duan H, Meng F, Du H, Zhang W, Li H, et al. Acute anterior choroidal artery territory infarction: a retrospective study. Clin Neurol Neurosurg. (2020) 195:105826. doi: 10.1016/j.clineuro.2020.105826

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Ma L, Xu JC, Yan S, Feng H, Han HJ, Tan HQ, et al. A single-center experience in the endovascular treatment of carotid siphon aneurysms using the Willis covered stent: a retrospective analysis. J Neurointerv Surg. (2018) 10:1197–202. doi: 10.1136/neurintsurg-2017-013695

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: blood blister-like aneurysm, endovascular, covered stent, internal carotid artery, subarachnoid hemorrhage

Citation: Fang W, Yu J, Liu Y, Sun P, Yang Z, Zhao Z, He Y, Deng J and Zhang T (2022) Application of the Willis Covered Stent in the Treatment of Blood Blister-Like Aneurysms: A Single-Center Experience. Front. Neurol. 13:882880. doi: 10.3389/fneur.2022.882880

Received: 24 February 2022; Accepted: 12 April 2022;
Published: 18 May 2022.

Edited by:

Qinghai Huang, Changhai Hospital, China

Reviewed by:

Yueqi Zhu, Shanghai Sixth People's Hospital, China
Li Pan, Wuhan People's Liberation Army General Hospital, China
Yingkun He, Henan Provincial People's Hospital, China

Copyright © 2022 Fang, Yu, Liu, Sun, Yang, Zhao, He, Deng and Zhang. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Tao Zhang, baltimore@163.com; Jianping Deng, 13991139395@163.com; Wei Fang, mailfangwei@163.com

These authors have contributed equally to this work

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.