SYSTEMATIC REVIEW article

Front. Genet., 14 June 2023
Sec. Genetics of Common and Rare Diseases

Population screening for glucose-6-phosphate dehydrogenase deficiency using quantitative point-of-care tests: a systematic review

  • 1Department of Pathology, Faculty of Medicine, Universiti Kebangsaan Malaysia (UKM), Kuala Lumpur, Malaysia
  • 2UKM Medical Molecular Biology Institute (UMBI), Universiti Kebangsaan Malaysia Medical Centre, Kuala Lumpur, Malaysia
  • 3Department of Biological Sciences Dan Biotechnology, Faculty of Science and Technology, Universiti Kebangsaan Malaysia, Bangi, Malaysia
  • 4Department of Pathology, Faculty of Medicine and Health Sciences, Universiti Sains Islam Malaysia, Nilai, Malaysia

Background: Glucose-6-phosphate dehydrogenase (G6PD) deficiency is an X-linked hereditary disorder and a global public health concern that is most prevalent in malaria-endemic regions including Asia, Africa, and the Mediterranean. G6PD-deficient individuals are at high risk of developing acute hemolytic anemia following treatment with antimalarial drugs including Primaquine and Tafenoquine. However, the currently available tests for G6PD screening are complex and often have been misclassifying cases, particularly for females with intermediate G6PD activity. The latest innovation of quantitative point-of-care (POC) tests for G6PD deficiency provides an opportunity to improve population screening and prevent hemolytic disorders when treating malaria. Aim(s): To assess the evidence on the type and performance of quantitative point-of-care (POC) tests for effective G6PD screening and hence, radical elimination of Plasmodium malaria infections.

Methods: Relevant studies published in English language confined from two databases, Scopus and ScienceDirect were searched from November 2016 onwards. The search was conducted using keywords including “glucosephosphate dehydrogenase” or “G6PD”, “point-of-care”, “screening” or “prevalence”, “biosensor” and “quantitative”. The review was reported following the PRISMA guidelines.

Results: Initial search results yielded 120 publications. After thorough screening and examination, a total of 7 studies met the inclusion criteria, and data were extracted in this review. Two types of quantitative POC tests were evaluated, namely, the CareStartTM Biosensor kit and the STANDARD G6PD kit. Both tests showed promising performance with high sensitivity and specificity ranging mostly from 72% to 100% and 92%–100%, respectively. The positive and negative predictive values (PPV and NPV) ranged from 35% to 72% and 89%–100%, with accuracy ranging from 86% to 98%.

Conclusion: In areas with a high prevalence of G6PD deficiency that overlap with malaria endemicity, availability and validation of the diagnostic performance of quantitative POC tests are of absolute importance. Carestart™ biosensor and STANDARD G6PD kits showed high reliability and performed well in comparison to the spectrophotometric reference standard.

1 Introduction

Glucose-6-phosphate dehydrogenase (G6PD) deficiency is a genetic disorder that results in an inadequate amount of G6PD enzyme, a biological catalyst that is important to produce the reduced form of nicotinamide adenine dinucleotide phosphate that protects red blood cells against oxidative stress (Luzzatto et al., 2020). G6PD-deficient individuals are at high risk of developing acute hemolytic anemia and severe neonatal hyperbilirubinemia leading to brain damage upon exposure to exogenous agents, including certain food intakes such as fava beans, diseases such as bacterial infections, and drugs such as Primaquine and Tafenoquine (Chu et al., 2019; Lee et al., 2022). This inherited condition affects approximately 500 million individuals worldwide and is present most frequently in malaria-endemic regions such as parts of Asia, Africa, the Mediterranean, and the Middle East (Zeng et al., 2022). For instance, in Malaysia, a Southeast Asian country where malaria remains a serious public health concern due to its high forest coverage area, the prevalence of G6PD deficiency was 3.4%; among which 5.3% are males and 1.1% are females (Ainoon et al., 2003).

The disease is an X-linked hereditary disorder, meaning that it is most common in males. Phenotypically males can be either hemizygous G6PD deficient or normal, and females can be either homozygous or heterozygous G6PD deficient (Au et al., 2007). In heterozygote females, the range of G6PD activity is wide, ranging from completely deficient to perfectly normal activity levels. Normal G6PD activity is defined as the median G6PD activity level in subjects with no evidence of G6PD mutations. Based on the World Health Organization’s (WHO) latest recommendation, the current case definitions of G6PD deficiency are as follows: males and females with <30% of normal enzyme activity are G6PD deficient, heterozygous females with enzyme activity of 30%–80% are G6PD intermediate, and males with >30% and females with >80% activity levels are G6PD normal (World Health Organization, 2016).

Historically, a different classification was used back in the year 1985 which was known as the WHO G6PD Classification of Variants (Class I–V) (World Health Organization, 2019; He et al., 2020). However, there had been a gradual shift throughout the years from biochemical to mutation analysis in which it was later found that many G6PD variants which were classified as class II and class III have the same clinical manifestations (World Health Organization, 2016). Based on mutation analyses that were conducted in studies worldwide, most G6PD gene mutations (84%) are point mutations that affect a single nucleotide, while the remaining variants are multiple nucleotide substitutions, deletions, and intronic mutations. Different G6PD mutations produce a wide range of biochemical phenotypes such as decreased stability or decreased catalytic activity of the G6PD enzyme. In field application, there were high requirements to inform the G6PD status and the performance of point-of-care (POC) G6PD assays to guide clinicians on the treatment using antimalarial drugs such as Primaquine and Tafenoquine, which then have redefined a ‘normal’ G6PD activity to be either >70% or >80% of normal. Therefore, based on these findings, the G6PD classification was revised, modified, and updated as per the previously mentioned case definitions (World Health Organization, 2014).

The diagnosis of G6PD deficiency requires an appropriate and effective testing method, particularly for malarial patients who need to be treated with 8-aminoquinoline drugs including Primaquine and Tafenoquine. Malaria infection is caused by Plasmodium species including P. ovale, P. vivax, P. falciparum, P. malariae, and P. knowlesi. According to the WHO, the most common and lethal malaria parasite on the continent of Africa is P. falciparum, while the most common malaria parasite outside of sub-Saharan Africa is P. vivax (World Health Organization, 2023). Hypnozoites, which are liver-dormant stages, is a part of P. ovale and P. vivax’s life cycle. After the initial infection, these stages might reactivate and lead to malaria relapse in weeks, months, or even years. Relapses are substantially less common for P. ovale than for P. vivax. Nonetheless, to cure these patients, both the asexual stages of the parasites in the blood and the hypnozoites in the liver require to be eliminated–a process known as radical cure. Two important hypnozoiticidal agent that can be used for the radical cure of malaria and malaria chemoprophylaxis are Primaquine and Tafenoquine, which can cause severe hemolysis in individuals with G6PD deficiency (Commons et al., 2020).

Primaquine has a short half-life of 6 h and is rapidly metabolized and eliminated from the body. This medication can be prescribed to patients with ≥30% G6PD activity and thus can be prescribed based on a qualitative test result (Baird, 2018). However, a radical cure of malaria using Primaquine requires daily administration for 14 days and it is unclear that the Primaquine treatment is effective against the P. ovale hypnozoites due to limited studies. In contrast, Tafenoquine has a longer half-life of 14 days and can be prescribed as a single-dose for radical cure of P. vivax and P. ovale malaria infection such as using the Kozenis (GlaxoSmithKline, Brentford, United Kingdom) 150 mg tablets or weekly regiments using the Kodatef 100 mg tablets (60 Degrees Pharmaceuticals, Washington, DC) for chemoprophylaxis. Both Primaquine and Tafenoquine should be discontinued early by the prescriber at the first signs of hemolysis. However, Tafenoquine discontinuation cannot reduce oxidative stress exposure in G6PD-deficient patients since the drug is removed slowly from the body with a lengthy half-life duration. As a result, the recommended G6PD activity threshold for Tafenoquine prescription is higher, and all females with moderate G6PD activity of less than 70% should be excluded (Dean et al., 2012). Hence, the safe use of Tafenoquine requires a precise quantitative test that can demonstrate G6PD activity within the normal G6PD reference range.

In addition, the strong link between G6PD deficiency and severe neonatal jaundice also led to the deployment of a national program of mandatory screening for G6PD deficiency in all infants since the early 1980s, using the WHO-recommended qualitative/semiquantitative Fluorescent Spot Test (FST) (Beutler and Mitchell, 1968). Although the method costs far less and was simpler to apply, it requires laboratory infrastructure and expertise in its interpretation. Additionally, studies found that there was a lack of sensitivity of the FST to detect G6PD-deficient individuals, particularly females with moderate enzyme activity ranging between 20% and 60% of the normal mean (Henriques et al., 2018; Thielemans et al., 2018). Henceforth, the condition necessitates the introduction of a quantitative method to measure and screen for G6PD deficiency among the population.

A gold standard diagnostic laboratory method at present is the quantitative ultraviolet (UV) spectrophotometry assay, which is complex, costly, and requires several hours of turn-around time, special equipment, a source of electricity, and experienced personnel. These various limiting factors make the spectrophotometry assay unsuitable for routine G6PD testing in most malaria-endemic countries and resource-limited settings (Pfeffer et al., 2020; Roper et al., 2020). Over the past few decades, all developed field applications for diagnosing G6PD deficiency including rapid diagnostic test (RDT) kit provide a qualitative result, which only reliably identifies whether the individual is normal or severely G6PD deficient with less than 30% residual enzyme activity (Ley et al., 2015). Although the tests were affordable, feasible, easily performed at the patient’s side, and better diagnostic than the conventional FST, its performance for G6PD carrier or G6PD deficient female heterozygotes remained poor and inaccurate in many cases (World Health Organization, 2015). The previous studies showed that most females with an intermediate enzyme activity level were often falsely classified as “normal” by the qualitative test (Yu et al., 2020; Djigo et al., 2021). These misclassifications were alarming as they increased the patients’ risk of hematological disorder upon administration of antimalarial drugs (Chu et al., 2018).

In addressing the diagnostic challenges and gender disparity issues by the currently available qualitative POC method, particularly for radical malaria elimination, the latest technological innovation of POC devices was developed to quantitatively measure G6PD enzyme activity. The assay was designed to be able to discriminate between the full range of deficiencies and therefore could accurately diagnose G6PD-deficient individuals. The novelty of the assay led to the aim of our current study which was to identify and assess available types of quantitative POC G6PD tests, compare the evaluation of performance from diverse laboratory and geographic contexts, and evaluate the strength and limitations of the quantitative assay.

2 Materials and methods

2.1 Study design

This is a systematic review of literature. This study was conducted and reported by following the principles of Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines (Supplementary Material 1) (Page et al., 2021).

2.2 Literature search strategy

The following electronic databases were searched from January 2017 to recent investigations in December 2021: Scopus and ScienceDirect, with restriction to English language publications. The last date of the database search was 26 December 2022. The keywords used were “glucosephosphate dehydrogenase”, “G6PD”, “point-of-care”, “screening”, “prevalence”, “biosensor”, and “quantitative”. The subject terms were combined using the Boolean operator “AND,” and “OR” for a comprehensive search. Additional relevant studies found from the references of selected articles were also retrieved and screened. An example of a search strategy for Scopus is as follows: TITLE-ABS-KEY [“glucose phosphate dehydrogenase deficiency” AND quantitative AND (“point-of-care” OR “biosensor”)].

2.3 Screening of articles for eligibility and quality assessment

Selected articles identified through the databases and references were screened for eligibility. The process started with a screening of the title and abstract and was followed by the selection of eligible studies that met the inclusion criteria developed from the research question using the PICOS (Population, Intervention, Comparator, Outcome, Study) component (Table 1). Exclusion criteria included studies that were performed using a qualitative type of point-of-care test for G6PD deficiency and irrelevant studies using other than cross-sectional methods such as economic evaluation and user perspective. Throughout the abstract screening, full-text articles were retrieved and read in the event of any uncertainty pertaining to the content relevance of the study. Once the list of abstracts was compiled and full articles were retrieved, all studies were evaluated by one researcher, while the second researcher compiled and reviewed the results. A third reviewer would be consulted if there was a disagreement on the research selection in order to come to a decision. The quality assessment and risk of bias in each study were assessed using the Appraisal tool for Cross-Sectional Studies (AXIS) (Downes et al., 2016).

TABLE 1
www.frontiersin.org

TABLE 1. PICOS criteria for eligibility of studies.

2.4 Data extraction

The following data was manually extracted from each study: name of the first author, year of publication, country or study location, study design, sample size, type of POC test, type of reference assay, results of the screening procedure including G6PD prevalence, levels of enzyme activity, and diagnostic performance of the POC tests. The extracted data was organized and collated into tables using Microsoft Excel Spreadsheet software.

2.5 Data synthesis

The tabulated data was then analyzed and interpreted by the first and second researchers. The primary outcomes assessed were the prevalence or the number of identified subjects with G6PD deficiency, the levels of G6PD activity for the studied population, the performance of the POC test including sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV), and results of correlation analysis. This information was synthesized using a narrative (descriptive) method.

3 Results

3.1 Selection of articles

The selection process of articles for this study was summarized using the PRISMA flow diagram for systematic review (Figure 1). Overall, 117 publications were found during our initial search of the electronic databases. Three additional papers were found for this review from other sources, such as references from the searched articles. Following a duplicate removal, a total of 103 publications were screened for content relevance from the title and abstract. In case of doubt, the full-text versions of these articles were examined, and eligibility and full-text assessment were conducted on those that fulfill the inclusion criteria and were English publications. The final result of this study identified 7 articles that were included in this review, consisting of cross-sectional studies conducted from the year 2017 until 2021. An extracted data summarization from all studies was shown in Table 2.

FIGURE 1
www.frontiersin.org

FIGURE 1. Flowchart showing inclusion in systematic review of studies reporting on diagnostic performance of quantitative point-of-care tests for screening of G6PD deficiency.

TABLE 2
www.frontiersin.org

TABLE 2. Summary of studies reviewing the diagnostic performance of the quantitative point-of-care test for screening of G6PD deficiency.

3.2 Risk of bias

In general, all selected studies had a low to moderate risk of bias (Table 2). The studies applied an adequate approach to the research question and the findings were coherent in their sources, data collection, and analysis.

3.3 Main findings

This systematic review reports the evidence of the diagnostic performance of POC testing for G6PD deficiency across the globe, with a focus on a quantitative approach. A total of 3,605 individuals from seven different studies were universally screened for G6PD deficiency using quantitative POC tests. These studies demonstrated significant methodological diversity and statistical heterogeneity. It was notably found that the available type of quantitative POC tests was limited to two assay kits: CareStart™ Biosensor (CB) and STANDARD G6PD (SG). Four studies evaluated the CB, two studies evaluated the SG, and one study evaluated both CB and SG (Ley et al., 2017; Weppelmann et al., 2017; Alam et al., 2018; Bancone et al., 2018; Pal et al., 2019; Pengboon et al., 2019; Zobrist et al., 2021). Each study compared the POC diagnostic performance to a gold-standard spectrophotometric assay. They were conducted independently in six countries: Thailand, Myanmar, Bangladesh, Brazil, Haiti, and the United States (US).

According to the included articles, the prevalence of G6PD deficiency in the studied population ranged from 14.8% to 99.5%. Different lower and upper cut-off values were used in their evaluation including 10%, 30%, 60%, 70%, and 80% of normal enzyme activity. The 10% and 60% thresholds correspond to the WHO G6PD Classification of Variants’ Classes I, II, and III (World Health Organization, 2019). The 30% and 80% thresholds referred to the most recent WHO recommendation of the current case definitions of G6PD deficiency (World Health Organization, 2022), while the 70% threshold corresponded to the cut-off point for exclusion criteria of Tafenoquine treatment (Commons et al., 2020). For evaluation of the CB, the sensitivity, specificity, and positive and negative predictive values of the analyzer ranged from 0% to 100%, 91%–100%, 59%–73%, and 90.6%–100%, respectively. The corresponding values for the SG were 85%–100%, 74.4%–98.6%, 35.2%, and 99.8%. These results’ diversity reflected the variability in the study design and ethnography, as well as the rapid technological progress of the POC device over the years. Finally, the diagnostic accuracy was described only by two studies, which ranged from 86.6% to 98.8% (Weppelmann et al., 2017; Pengboon et al., 2019). All studies, except research conducted by (Alam et al., 2018; Bancone et al., 2018) reported excellent values of area under the receiver operating characteristic curve (AUC) which ranged from 0.88 to 1, denoting an outstanding overall diagnostic performance of the POC tests (Ley et al., 2017; Weppelmann et al., 2017; Pal et al., 2019; Pengboon et al., 2019; Zobrist et al., 2021).

4 Discussion

4.1 Summary of evidence

Many countries have promoted screening for G6PD deficiency in the general population, particularly in malaria-endemic areas and among newborns (Marasini et al., 2020). The most effective management strategy for malaria elimination and neonatal jaundice in G6PD-deficient neonates is early diagnosis of the hereditary condition. The informed G6PD status not only makes patients or the parents of affected newborns aware of the deficiency in order to avoid exposure to oxidative agents, but it also guides clinicians in the safe administration of medications such as antimalarial drugs and antibiotics. This systematic review expresses grave concern about the low reliability of currently available screening assays for G6PD deficiency, which is the FST, and advocates for the development of novel point-of-care diagnostics.

In contrast to the spectrophotometric assay, which requires trained personnel, a UV spectrophotometer, and other laboratory equipment, the POC devices were simple, compact, easily operated with minimal training, and convenient to be used in field application (Pal et al., 2019). One of the quantitative POC tests that are available in the market is the CareStartTM Biosensor (CB). Since 2017, five studies have been conducted to assess the robustness of this assay. This method requires a volume of 7–10 microliters (μL) of capillary or venous blood specimen, which could be directly applied to the analyzer’s strip. When the G6PD enzyme in the specimen combines with a substrate, this biosensor measures G6PD enzyme activity electrochemically by transferring electrons between donor and acceptor molecules. The magnitude of the generated electric current is directly proportional to the level of G6PD activity in the blood sample. Previous CB models, as used in studies by Weppelmann et al. (2017); Ley et al. (2017); Alam et al. (2018), could only quantify G6PD activity without hemoglobin analyses. The G6PD activity was later standardized to hemoglobin concentration using a separate digital hemoglobin meter or the results of a complete blood count analysis. Then, as used in studies by Bancone et al. (2018); Pengboon et al. (2019), a newer and more compact model of the CB test was developed which could perform both G6PD enzyme and Hb analyses in the same device.

In the earlier research, the CB performed poorly with a very low sensitivity of 5.9% at the 30% threshold for diagnosing moderate-to-severe G6PD deficiency (Weppelmann et al., 2017). It was also demonstrated that the CB only had a slight, or fair agreement between the assay and reference spectrophotometry in correctly determining the ordinal G6PD activity classification. The sensitivity of the CB was later proven to be higher in subsequent studies, which ranged between 19% and 72% using a similar biosensor model, indicating device technological improvement (Ley et al., 2017; Pengboon et al., 2019). In Alam et al. (2018) demonstrated that the CB performed well at 70% upper cut-off level with 71% sensitivity and 98% specificity, a threshold that is important for identifying heterozygous females and those contraindicated for tafenoquine treatment. Bancone et al. (2018) were the first to use the newer model of CB. The study revealed that the hemoglobin concentration measured by the new biosensor correlated well with the results from the complete blood count analysis of the study (R2 = 0.88). The ranges of G6PD activities assessed through the study were almost identical between the CB and the gold standard spectrophotometric assay. The G6PD results of this study were found to be very similar in G6PD normal males but much higher in G6PD-deficient males and females. Despite the CB’s significant improvement over previous evaluations, the threshold points for distinguishing samples with activity ranging from 30% to 80% normal were demonstrated to be alarmingly narrow (Bancone et al., 2018). Further studies are warranted to investigate the small difference in “intermediate” G6PD activity between the two critical thresholds.

The STANDARD G6PD (SG) test is another quantitative POC G6PD assay that was designed to measure G6PD enzymatic level and total hemoglobin level simultaneously using the reflectometry assays method. Alam et al. (2018); Pal et al. (2019); Zobrist et al. (2021) used this test to evaluate its diagnostic performance in population screening. A total of 10 μL of capillary or venous blood specimen was required for this assay, which was mixed with a special extraction buffer solution provided by the manufacturer before being transferred to the designated device’s strip for biochemical detection. A study revealed that the correlation between the SG and spectrophotometric assay for both normalized G6PD activity and hemoglobin measurement was good, with squared correlation coefficient (R2) values of 0.92 and 0.75, respectively (Pal et al., 2019). The strong and positive correlations of the test with high sensitivity ranged from 85% to 100% and specificity ranged from 74% to 100%, implying that it is a promising tool for screening for G6PD deficiency.

In addition, the SG performed well under a wide range of operating conditions, including temperatures ranging from 17°C to 43°C and humidity levels reaching 75% (Pal et al., 2019). This is significant because G6PD deficiency variants are widely distributed across malaria-endemic regions, including tropical Asian countries. According to the WHO prequalification technical specifications, an in vitro diagnostic test for G6PD deficiency must be able to distinguish between normal, G6PD-deficient with enzyme level below 30%, and G6PD-intermediate with 30%–80% of normal enzymatic activity (World Health Organization, 2016). Current findings proved that the SG’s performance was excellent at the 30% threshold and acceptable at the 70% and 80% thresholds.

It is important to note that each of the seven studies included in this systematic review had its own set of limitations. Small to moderate sample sizes, potential selection bias, and lack of diversity among participants were among the few limitations. For example, Weppelmann et al. (2017) highlighted the use of convenience sampling from a single Haiti’s department in their study design which resulted in findings that might not be representative of another department or the whole country, and Pengboon et al. (2019) discovered that the diagnostic efficacy of the CB was higher in their study than in previous evaluation studies, which could be attributed to their small sample size (n = 216). The limitation was also emphasized by Zobrist et al. (2021), who described that there were a lower-than-expected number of participants with deficient and intermediate G6PD status in their study, which subsequently reflected in the sensitivity and specificity values at 95% confidence intervals.

Another limitation is that most evaluation studies were performed in laboratory settings with highly regulated environmental temperature and humidity. This limitation was highlighted by Bancone et al. (2018); Pal et al. (2019), who suggested that future clinical studies be conducted in near-patient or intended settings in order to be more reflective of the intended use and essential to further validate the device’s performance and robustness. Nonetheless, all reported studies in this systematic review have considered the distribution and genotypic-phenotypic map of G6PD deficiency in their study designs. This correlation is normally demonstrated in chronic nonspherocytic hemolytic anemia (CNSHA) G6PD-deficient individuals, which present in individuals with Class I G6PD variants. However, most G6PD mutations worldwide are Class II and II, and these mutations show ethnogeographic variability with a specific spectrum of variants in different ethnicities. As an example, a study by Koromina et al. (2021) revealed the large extent of variability in mutations of G6PD deficiency across worldwide population and highlight its population-specific genetic composition.

Similarly, the studies included in this review to evaluate the performance of the quantitative device were performed across multiple study sites with a high prevalence of G6PD deficiency and carefully selected specific ethnic groups such as the Afro-Haitian population in the Republic of Haiti by Weppelmann et al. (2017), Tibeto-Asian and Bengali descent in Chittagong Hill Tracts, Bangladesh by Ley et al. (2017); Alam et al. (2018), Burman and Karen ethnic groups in northwestern Thailand by Bancone et al. (2018); Pengboon et al. (2019); Pal et al. (2019), and Manaus and Porto Velho population in Brazil by Zobrist et al. (2021).

4.2 CareStartTM Biosensor (CB) versus STANDARD G6PD (SG) tests

Only a study conducted by Alam et al. (2018) provides a direct comparison of both POC tests. The study was performed in Bangladesh, and the sample size was 158 participants. Both devices demonstrated a high correlation with the reference spectrophotometric assay and performed well under field and laboratory conditions with comparable accuracy. At a 30% cut-off, the AUC for both tests did not differ significantly, indicating that they have comparable discriminatory power. However, at 70% cut-off activity, only the SG achieved high sensitivity, reaching to 90%, while the sensitivity of the CB was 70%. According to this study, the CB had a smaller difference in the clinically relevant window between 30% and 70% G6PD activity than the SG, with values ranging from 4.6 to 6.8 IU/gHb and 2.5–6.4 IU/gHb, respectively. This slightly wider window suggested that the SG had more granularity than the CB.

4.3 Strengths and limitations

The strength of our study is that it is a systematic review that looks at novel quantitative point-of-care testing for population screening for G6PD deficiency. It is useful in guiding policymaking in relation to screening for hereditary conditions, prescribing important medications, and combating malaria infection. One limitation of this review is the heterogeneity of the studies retrieved. Although all studies included are cross-sectional studies, high methodological diversity in their evaluation of device performance was expected with different populations and laboratory settings. Apart from that, we did not prospectively register the protocol of this study in any international database due to concerns about the expected major delays in checking and publishing registrations (Puljak, 2021).

5 Conclusion

This review looks at the outcome of quantitative point-of-care testing methods for G6PD deficiency in terms of their diagnostic performance, strength, and limitation. Given the high prevalence of G6PD deficiency in certain regions of the world that overlap with malaria endemicity, effective point-of-care diagnostics may provide enhanced safety to all affected individuals in these areas including males, females, and neonates, improve management of P. vivax, P. ovale, and P. falciparum malaria cases in resource-limited settings, and avoid drug resistance evolution. The CB and the SG demonstrated high reliability as a POC test, and further research into their effectiveness and feasibility should be conducted before real-life clinical implementation in health facilities.

Data availability statement

The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author.

Author contributions

MZ led the investigation, was responsible for managing the synthesis, data analysis, and drafted the report. RR and AO provided expert clinical advice (Pathology), while EI and SS provided expert biosciences and biotechnology advice. HA and AI provided constructive feedback that aided in the interpretation of the findings, editing, and writing sections of the report. All authors contributed to the article and approved the submitted version.

Funding

This study was supported by the Fundamental Research Grant Scheme (FRGS) from the Ministry of Higher Education (MOHE), Malaysia (Grant number: FRGS/1/2020/SKK0/UKM/01/4).

Acknowledgments

We would like to express our gratitude to our research colleagues at the Universiti Kebangsaan Malaysia (UKM) Medical Centre in Kuala Lumpur, Malaysia for their invaluable assistance and support throughout the study. Our heartfelt gratitude goes to the Malaysian Ministry of Higher Education for providing the FRGS funding for this study.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

Ainoon, O., Yu, Y. H., Amir Muhriz, A. L., Boo, N. Y., Cheong, S. K., and Hamidah, N. H. (2003). Glucose-6-phosphate dehydrogenase (G6PD) variants in Malaysian Malays. Hum. Mutat. 21 (1), 101. doi:10.1002/humu.9103

PubMed Abstract | CrossRef Full Text | Google Scholar

Alam, M. S., Kibria, M. G., Jahan, N., Thriemer, K., Hossain, M. S., and Douglas, N. M. (2018). Field evaluation of quantitative point of care diagnostics to measure glucose-6-phosphate dehydrogenase activity. PLoS ONE 13 (11), e0206331. doi:10.1371/journal.pone.0206331

PubMed Abstract | CrossRef Full Text | Google Scholar

Au, W. Y., Pang, A., Lam, K., Song, Y-Q., Lee, W-M., So, J., et al. (2007). G6PD deficiency from lyonization after hematopoietic stem cell transplantation from female heterozygous donors. Bone marrow Transplant. 40, 677–681. doi:10.1038/sj.bmt.1705796

PubMed Abstract | CrossRef Full Text | Google Scholar

Baird, J. K. (2018). Tafenoquine for travelers’ malaria: Evidence, rationale and recommendations. J. Travel Med. 25, tay100. doi:10.1093/jtm/tay110

PubMed Abstract | CrossRef Full Text | Google Scholar

Bancone, G., Gornsawun, G., Chu, C. S., Porn, P., Pal, S., and Bansil, P. (2018). Validation of the quantitative point-of-care CareStart biosensor for assessment of G6PD activity in venous blood. PLoS ONE 13 (5), e0196716.

PubMed Abstract | CrossRef Full Text | Google Scholar

Beutler, E., and Mitchell, M. (1968). Special modifications of the fluorescent screening method for glucose-6-phosphate dehydrogenase deficiency. Blood 32 (5), 816–818. PMID: 4386875.

PubMed Abstract | CrossRef Full Text | Google Scholar

Chu, C. S., Bancone, G., Nosten, F., Nicholas, J. W., and Lucio, L. (2018). Primaquine-induced haemolysis in females heterozygous for G6PD deficiency. Malar. J. 17, 101. doi:10.1186/s12936-018-2248-y

PubMed Abstract | CrossRef Full Text | Google Scholar

Chu, C. S., Bancone, G., Soe, N. L., Carrara, V. I., Gornsawun, G., and Nosten, F. (2019). The impact of using primaquine without prior G6PD testing: A case series describing the obstacles to the medical management of haemolysis. Wellcome Open Res. 4, 25. doi:10.12688/wellcomeopenres.15100.2

PubMed Abstract | CrossRef Full Text | Google Scholar

Commons, R. J., McCarthy, J. S., and Price, R. N. (2020). Tafenoquine for the radical cure and prevention of malaria: The importance of testing for G6PD deficiency. Med. J. Aust. 212 (4), 152–153.e1. doi:10.5694/mja2.50474

PubMed Abstract | CrossRef Full Text | Google Scholar

Dean, L., and Kane, M. (2012). “Tafenoquine therapy and G6PD genotype. 2020 oct 13,” in Medical genetics summaries [internet]. V. M. Pratt, S. A. Scott, M. Pirmohamed, B. Esquivel, B. L. Kattman, A. J. Malheiroet al. (Bethesda (MD): National Center for Biotechnology Information US). Available online at: https://www.ncbi.nlm.nih.gov/books/NBK562890/.

Google Scholar

Djigo, O. K. M., Ould, K. Y., Ould, A. S., Mohamed Salem, O. A. S., Gomez, N., Basco, L., et al. (2021). Assessment of CareStart G6PD rapid diagnostic test and CareStart G6PD biosensor in Mauritania. Infect. Dis. Poverty 10, 105. doi:10.1186/s40249-021-00889-2

PubMed Abstract | CrossRef Full Text | Google Scholar

Downes, M. J., Brennan, M. L., Williams, H. C., and Dean, R. S. (2016). Development of a critical appraisal tool to assess the quality of cross-sectional studies (AXIS). BMJ Open 6 (12), e011458.

PubMed Abstract | CrossRef Full Text | Google Scholar

Engel, N., Ghergu, C., Matin, M. A., Kibria, M. G., Thriemer, K., Price, R. N., et al. (2021). Implementing radical cure diagnostics for malaria: User perspectives on G6PD testing in Bangladesh. Malar. J. 20 (1), 217. doi:10.1186/s12936-021-03743-w

PubMed Abstract | CrossRef Full Text | Google Scholar

He, Y., Zhang, Y., Chen, X., Wang, Q., Ling, L., and Xu, Y. (2020). Glucose-6-phosphate dehydrogenase deficiency in the han Chinese population: Molecular characterization and genotype-phenotype association throughout an activity distribution. Sci. Rep. 10 (1), 17106. doi:10.1038/s41598-020-74200-y

PubMed Abstract | CrossRef Full Text | Google Scholar

Henriques, G., Phommasone, K., Tripura, R., Peto, T. J., Raut, S., Snethlage, C., et al. (2018). Comparison of glucose-6 phosphate dehydrogenase status by fluorescent spot test and rapid diagnostic test in Lao PDR and Cambodia. Malar. J. 17 (1), 243. doi:10.1186/s12936-018-2390-6

PubMed Abstract | CrossRef Full Text | Google Scholar

Koromina, M., Pandi, M. T., van der Spek, P. J., Patrinos, G. P., and Lauschke, V. M. (2021). The ethnogeographic variability of genetic factors underlying G6PD deficiency. Pharmacol. Res. 173, 105904. doi:10.1016/j.phrs.2021.105904

PubMed Abstract | CrossRef Full Text | Google Scholar

Lee, H. Y., Ithnin, A., Azma, R. Z., Othman, A., Salvador, A., and Cheah, F. C. (2022). Glucose-6-Phosphate dehydrogenase deficiency and neonatal hyperbilirubinemia: Insights on pathophysiology, diagnosis, and gene variants in disease heterogeneity. Front. Pediatr. 10, 875877. doi:10.3389/fped.2022.875877

PubMed Abstract | CrossRef Full Text | Google Scholar

Ley, B., Luter, N., Espino, F., Devine, A., Kalnoky, M., Lubell, Y., et al. (2015). The challenges of introducing routine G6PD testing into radical cure: A workshop report. Malar. J. 14, 377. doi:10.1186/s12936-015-0896-8

PubMed Abstract | CrossRef Full Text | Google Scholar

Ley, B., Alam, M. S., O'Donnell, J. J., Hossain, M. S., Kibria, M. G., Jahan, N., et al. (2017). A comparison of three quantitative methods to estimate G6PD activity in the Chittagong Hill Tracts, Bangladesh. PLoS ONE 12 (1), e0169930. doi:10.1371/journal.pone.0169930

PubMed Abstract | CrossRef Full Text | Google Scholar

Luzzatto, L., Ally, M., and Notaro, R. (2020). Glucose-6-phosphate dehydrogenase deficiency. Blood 136 (11), 1225–1240. doi:10.1182/blood.2019000944

PubMed Abstract | CrossRef Full Text | Google Scholar

Marasini, B., Lal, B. K., Thapa, S., Awasthi, K. R., Bajracharya, B., Khanal, P., et al. (2020). G6PD deficiency in malaria endemic areas of Nepal. Malar. J. 19 (1), 287. doi:10.1186/s12936-020-03359-6

PubMed Abstract | CrossRef Full Text | Google Scholar

Page, M. J., McKenzie, J. E., Bossuyt, P. M., Boutron, I., Hoffmann, T. C., Mulrow, C. D., et al. (2021). The PRISMA 2020 statement: An updated guideline for reporting systematic reviews. BMJ 372, n7.1. doi:10.1136/bmj.n71

CrossRef Full Text | Google Scholar

Pal, S., Bansil, P., Bancone, G., Hrutkay, S., Kahn, M., Gornsawun, G., et al. (2019). Evaluation of a novel quantitative test for glucose-6-phosphate dehydrogenase deficiency: Bringing quantitative testing for glucose-6-phosphate dehydrogenase deficiency closer to the patient. Am. J. Trop. Med. Hyg. 100 (1), 213–221. doi:10.4269/ajtmh.18-0612

PubMed Abstract | CrossRef Full Text | Google Scholar

Pengboon, P., Thamwarokun, A., Changsri, K., Kaset, C., and Chomean, S. (2019). Evaluation of quantitative biosensor for glucose-6-phosphate dehydrogenase activity detection. PLoS ONE 14, e0226927. doi:10.1371/journal.pone.0226927

PubMed Abstract | CrossRef Full Text | Google Scholar

Pfeffer, D. A., Ley, B., Howes, R. E., Adu, P., Alam, M. S., Bansil, P., et al. (2020). Quantification of glucose-6-phosphate dehydrogenase activity by spectrophotometry: A systematic review and meta-analysis. PLOS Med. 17 (5), e1003084. doi:10.1371/journal.pmed.1003084

PubMed Abstract | CrossRef Full Text | Google Scholar

Puljak, L. (2021). Delays in publishing systematic review registrations in PROSPERO are hindering transparency and may lead to research waste. BMJ Evidence-Based Med. 26, e4. doi:10.1136/bmjebm-2020-111474

PubMed Abstract | CrossRef Full Text | Google Scholar

Roper, D., Layton, M., Rees, D., Lambert, C., Vulliamy, T., De la Salle, B., et al. (2020). Laboratory diagnosis of G6PD deficiency: A British society for haematology guideline. Br. J. Haematol. 189 (1), 24–38. doi:10.1111/bjh.16366

PubMed Abstract | CrossRef Full Text | Google Scholar

Runjic, E., Rombey, T., Pieper, D., and Puljak, L. (2019). Half of systematic reviews about pain registered in PROSPERO were not published and the majority had inaccurate status. J. Clin. Epidemiol. 116, 114–121. doi:10.1016/j.jclinepi.2019.08.010

PubMed Abstract | CrossRef Full Text | Google Scholar

Solla, F., Bertoncelli, C. M., and Rampal, V. (2021). Does the PROSPERO registration prevent double review on the same topic? BMJ Evid. Based Med. 26, 140. doi:10.1136/bmjebm-2020-111361

PubMed Abstract | CrossRef Full Text | Google Scholar

Thielemans, L., Gornsawun, G., Hanboonkunupakarn, B., Paw, M. K., Porn, P., Moo, P. K., et al. (2018). Diagnostic performances of the fluorescent spot test for G6PD deficiency in newborns along the Thailand-Myanmar border: A cohort study. Wellcome Open Res. 3, 1. doi:10.12688/wellcomeopenres.13373.1

PubMed Abstract | CrossRef Full Text | Google Scholar

Weppelmann, T. A., von Fricken, M. E., Wilfong, T. D., Aguenza, E., Philippe, T. T., and Okech, B. A. (2017). Field trial of the carestart biosensor analyzer for the determination of glucose-6-phosphate dehydrogenase activity in Haiti. Am. J. Trop. Med. Hyg. 97 (4), 1262–1270. doi:10.4269/ajtmh.16-0714

PubMed Abstract | CrossRef Full Text | Google Scholar

World Health Organization. Point-of-care G6PD testing to support safe use of primaquine for the treatment of vivax malaria. WHO evidence review group meeting report. Geneva, World Health Organization. (2014) 8–9 October 2014. Available online at: https://www.who.int/malaria/mpac/mpac-march2015-erg-g6pd.pdf (accessed June 24, 2021)

Google Scholar

World Health Organization (2015). Point-of-care G6PD testing to support safe use of primaquine for the treatment of vivax malaria. Geneva, Switzerland: WHO. Available online at: https://www.who.int/malaria/mpac/mpac-march2015-erg-g6pd.pdf?ua=1 (accessed Nov 24, 2021).

Google Scholar

World Health Organization (2016). Technical specifications series for submission to WHO prequalification—diagnostic assessment: In vitro diagnostics medical devices to identify glucose-6-phosphate dehydrogenase (g6pd) activity. Geneva, Switzerland: World Health Organization. Available online at: http://apps.who.int/iris/bitstream/handle/10665/252628/9789241511865-eng.pdf?sequence=1 (accessed May 3, 2021).

Google Scholar

World Health Organization (2019). Revision of WHO classification of G6PD variants and international classification of diseases (ICD)-11. Geneva, Switzerland: World Health Organization. Available online at: https://www.who.int/docs/default-source/malaria/mpac-documentation/mpac-october2019-session7-g6pd-presentation.pdf?sfvrsn=3696032c_2 ((accessed June 21, 2021).

Google Scholar

World Health Organization (2022). in Malaria policy advisory group meeting (Geneva, Switzerland: WHO). Available online at: https://cdn.who.int/media/docs/default-source/malaria/mpac-documentation/mpag-mar2022-session2-technical-consultation-g6pd-classification.pdf?sfvrsn=1f36be5e_7&download=true ((accessed Oct 15, 2021).Technical consultation to review the classification of glucose-6-phosphate dehydrogenase (G6PD)

Google Scholar

World Health Organization (2023). Malaria. Available online at: https://www.who.int/news-room/fact-sheets/detail/malaria ((accessed April 20, 2023).

Google Scholar

Yu, F., Zhang, S., Chen, B., Zhou, Y., Ma, C., Yang, S., et al. (2020). Evaluation of the diagnostic accuracy of the CareStart™ glucose-6-phosphate dehydrogenase deficiency rapid diagnostic test among Chinese newborns. J. Trop. Pediatr. 66 (5), 495–503. doi:10.3389/fped.2022.875877

PubMed Abstract | CrossRef Full Text | Google Scholar

Zeng, W., Liu, N., Li, Y., Gao, A., Yuan, M., Ma, R., et al. (2022). Prevalence of glucose-6-phosphate dehydrogenase deficiency (G6PDd) and clinical implication for safe use of primaquine in malaria-endemic areas of Hainan Province, China. Front. Public Health 10, 1010172. doi:10.3389/fpubh.2022.1010172

PubMed Abstract | CrossRef Full Text | Google Scholar

Zobrist, S., Brito, M., Garbin, E., Monteiro, W. M., Clementino Freitas, S., and Macedo, M. (2021). Evaluation of a point-of-care diagnostic to identify glucose-6-phosphate dehydrogenase deficiency in Brazil. PLoS Negl. Trop. Dis. 15 (8), e0009649. doi:10.1371/journal.pntd.0009649

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: glucose-6-phosphate dehydrogenase, point-of-care, quantita tive, screening, systematic review

Citation: Zailani MAH, Raja Sabudin RZA, Ithnin A, Alauddin H, Sulaiman SA, Ismail E and Othman A (2023) Population screening for glucose-6-phosphate dehydrogenase deficiency using quantitative point-of-care tests: a systematic review. Front. Genet. 14:1098828. doi: 10.3389/fgene.2023.1098828

Received: 19 December 2022; Accepted: 22 May 2023;
Published: 14 June 2023.

Edited by:

Rajesh Parsanathan, Central University of Tamil Nadu, India

Reviewed by:

Giovanni Mario Pes, University of Sassari, Italy
Kristan Alexander Schneider, Hochschule Mittweida, Germany

Copyright © 2023 Zailani, Raja Sabudin, Ithnin, Alauddin, Sulaiman, Ismail and Othman. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Raja Zahratul Azma Raja Sabudin, zahratul@ppukm.ukm.edu.my

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.