Efficacy and safety of advanced hybrid closed loop systems in children with type 1 diabetes younger than 6 years

Novella RapiniNovella Rapini1Mariangela Martino,*Mariangela Martino1,2*Claudia ArnaldiClaudia Arnaldi3Annalisa Deodati,Annalisa Deodati1,4Lilian AnagnostopoulouLilian Anagnostopoulou1Maria Elisa AmodeoMaria Elisa Amodeo1Paolo CiampaliniPaolo Ciampalini1Valentina PampaniniValentina Pampanini1Antonella LorubbioAntonella Lorubbio1Davide TosiniDavide Tosini3Stefano Cianfarani,,Stefano Cianfarani1,4,5Riccardo Schiaffini*Riccardo Schiaffini1*
  • 1Endocrinology and Diabetes Unit, Bambino Gesù Children’s Hospital, IRCCS, Scientific Institute for Research, Hospitalization and Health Care, Rome, Italy
  • 2PhD Program in Immunology, Molecular Medicine and Applied Biotechnologies, University of Rome ‘Tor Vergata’, Rome, Italy
  • 3Pediatric Diabetes Unit, ASL Viterbo, Viterbo, Italy
  • 4Department of Systems Medicine, University of Rome ‘Tor Vergata’, Rome, Italy
  • 5Department of Women’s and Children’s Health, Karolinska Institutet, Stockholm, Sweden

Background: Tight glycemic control is essential for the normal growth and development of preschool children. The aim of our study was to evaluate the impact of advanced hybrid closed loop (AHCL) systems in a real-life setting in children younger than 6 years.

Methods: We conducted a two-center prospective study. We enrolled 19 patients with a median age at disease onset of 2.6 years [interquartile range (IQR) 1.6; 4.4] and a median disease duration of 1.4 years (IQR 0.9; 2.8) who were switched to AHCL from multiple daily injections or open-loop insulin therapy and with a 6-month follow-up. Clinical data, sensor glycemic metrics, and pump settings were collected and analyzed.

Results: After 6 months of follow-up, there was a significant reduction in median HbA1c (p = 0.0007) and glucose management indicator (p = 0.03). A reduction in both mild (>180 mg/dL) (p = 0.04) and severe (>250 mg/dL) (p = 0.01) hyperglycemia was observed after 1 month of auto mode, and in mild hyperglycemia, it persisted up to 6 months (p = 0.02). A small increase in time below range (<70 mg/dL) was observed (p = 0.04) without a significant difference in time <54 mg/dL (p = 0.73). Time in range increased significantly, reaching a 10% increment (p = 0.03) compared with baseline. A significant reduction in the average sensor glucose was observed (p = 0.01) while coefficient of glucose variability (CV%) remained stable (p = 0.12). No episodes of ketoacidosis or severe hypoglycemia have been recorded.

Conclusion: AHCL systems are effective and safe for children younger than 6 years and should be considered as a valid therapeutic option from diabetes onset.

Introduction

Glycemic control in preschool children is challenging and glucose management is burdened by high glycemic variability (1) due to the reduced predictability of daily activities and meals. Tight glycemic control is mandatory, as toddlers diagnosed with type 1 diabetes (T1D) are expected to be exposed to long diabetes duration to reduce complications (2, 3), minimizing at the same time the hypoglycemic risk (4). Furthermore, neuroimaging studies identified alterations particularly affecting white matter, suggesting that during toddler and preschool years, the brain is highly sensitive to metabolic disturbances (both hypo- and hyperglycemia) (5) with implications for cognitive and executive functions, intelligence quotient, delayed memory, and processing speed (6). Advanced hybrid closed loop (AHCL) devices have proven useful in improving disease management and time in range (TIR) (7), but they have been currently approved above 6 or 7 years of age with a minimum total daily dose (TDD). The aim of our study was to evaluate the impact of AHCL on glycemic control over time in children younger than 6 years in a real-life setting, compared to the previous conventional multiple daily injection (MDI) insulin therapy and open-loop continuous subcutaneous insulin infusion (CSII) treatment [with or without a sensor augmented pump (SAP)].

Methods

We enrolled 19 pediatric patients (11 boys; 8 girls) with T1D {with a median age at disease onset of 2.6 [interquartile range (IQR) 1.6; 4.4] years and a median disease duration of 1.4 (IQR 0.9; 2.8) years} from the Bambino Gesù Children’s Hospital Diabetes Unit in Rome and at the Viterbo Pediatric Diabetes Unit, Italy, between January 2021 and June 2023. All involved families gave their consent for the use of algorithm-driven automated insulin delivery, although this therapeutic approach is currently approved in children at the age of 6 years or above with a minimum TDD of 10 units or >25 kg of weight (Tandem t:slim X2 Control-IQ) and above the age of 7 years with a minimum TDD of 8 units (Medtronic MiniMed™ 780G).

We considered the following inclusion criteria:

- age < 6 years

- diagnosis of T1D according to ISPAD guidelines (8)

- being monitored by isCGM (intermittent scanning continuous glucose monitoring) or rtCGM (real time continuous glucose monitoring) at baseline

- being on MDI insulin therapy or open-loop CSII treatment

Exclusion criteria were as follows:

- conditions or use of medications known to affect glycemic levels

- being already on the AHCL system

The study was approved by the Institutional Ethics Committees of participating centers, and all participants’ parents provided informed consent.

Weight (kg), height (cm), body mass index (BMI), glycated hemoglobin (HbA1c), and TDD in international units per body weight (IU/kg) have been evaluated. Participants’ demographic and anthropometric characteristics are shown in Table 1.

Table 1
www.frontiersin.org

Table 1 Participants’ demographic and anthropometric characteristics.

Sensor glucose reading data for 30 days were collected from rtCGM, and glycemic control was evaluated, analyzing data obtained from Carelink and Glooko platforms, considering percentage of TIR (TIR%, between 70 and 180 mg/dL), time above range in mild (TAR% >180 mg/dL) and severe (TAR% >250 mg/dL) hyperglycemia, and time below range in mild (TBR% <70 mg/dL) and very low (TBR% <54 mg/dL) hypoglycemia (9). Glucose management indicator (GMI), glucose average (mg/dL), coefficient of variability (CV%), and recorded dietary carbohydrate (9) were also evaluated.

All participants reported >90% of time with sensor in use.

The last available 30 days of sensor glucose data on MDI treatment or the open-loop system, before switching to AHCL, were considered as baseline, and 30 days of sensor glucose data were considered for each of the other time points (1, 3, and 6 months) during the follow-up period.

During follow-up, nine patients used Tandem t:slim X2 in Control-IQ mode, nine used MiniMed 780G with SmartGuard mode, and one shifted from MiniMed 640G to MiniMed 670G.

The MiniMed 780G pump was initially used in manual mode for 2 weeks before switching to auto mode with a 120 mg/dL glycemic target and a 100 mg/dL glycemic target after 3 months with a mean active insulin time of 2 h.

After switching to AHCL systems, subjects were followed up for 6 months and glycemic metrics were recorded after 1, 3, and 6 months. At each visit, the family was asked whether ketoacidosis or severe hypoglycemia occurred during the reported period. Furthermore, differences in anthropometric parameters were analyzed (weight, height, and BMI) as well as the HbA1c level at the beginning and at the end of the follow-up (Table 2).

Table 2
www.frontiersin.org

Table 2 Participants’ glycemic metrics at baseline and during follow-up.

Statistical analysis

Participants’ characteristics are reported as median and IQR for continuous variables and as absolute frequency and percentage for categorical variables.

We performed Wilcoxon signed-rank test to check whether the differences between paired data were statistically significant. We considered p-value below 0.05 as statistically significant.

Analyses were performed using GraphPad Prism ver. 9.00.

Results

After 6 months of follow-up with the AHCL systems, there was a significant reduction in both median HbA1c from 56.3 (52, 62.5) to 55 (44.8, 58.7) mmol/mol (Figure 1A) (p = 0.0007) and median GMI from 7.2 (6.9, 7.8) to 7 (6.8, 7.2) % (p = 0.03) without changes in BMI (p = 0.27, not shown).

Figure 1
www.frontiersin.org

Figure 1 In (A) difference in HbA1c from baseline and 6 months of follow-up. Difference in TIR, in TAR (>180 mg/dl) and in TBR (<70 mg/dl) from baseline and 1-3-6 months follow-up are shown in (B, D, E). In (C) difference in the average of sensor glucose. *=p<0.05; **=p<0.001; ***=p<0.0001.

Time in range (TIR%) increased significantly during the 6 months of follow-up, reaching a difference from baseline of +10% (p = 0.03) and +6% after the first (p = 0.007) and after the third (p = 0.03) month of use (Figure 1B). Insulin requirement presented a slight increment at 1 and 3 months (p = 0.02 and p = 0.004), without a significant change at the end of follow-up period compared to baseline (0.6 IU/kg/day, p = 0.4).

A significant reduction in the glucose average was observed, during the entire 6 months (p = 0.01) (Figure 1C).

A reduction in both mild (>180 mg/dL) (p = 0.04) and severe (>250 mg/dL) hyperglycemia (p = 0.01) was observed 1 month after AHCL systems, which persisted up to 6 months for mild hyperglycemia (p = 0.02) (Figure 1D).

Furthermore, time below range <70 mg/dL presented a small increment (p = 0.04) (Figure 1E) from baseline to 6 months without a significant difference in time <54 mg/dL (p = 0.73).

No significant differences were found in CV during follow-up (p = 0.12).

Recorded carbohydrates per day remained stable during the study period (p = 0.12, 0.43, and 0.62, respectively) with a significant reduction between 3 and 6 months (p = 0.02).

Both AHCL systems have always worked in auto mode during the observational period despite the low insulin demand (the minimum reported daily dose was 6 IU).

No episodes of ketoacidosis or severe hypoglycemia were reported during follow-up.

Discussion

Our study, involving preschool children with T1D treated with AHCL systems for a follow-up period of 6 months, showed a +10% increment of TIR from baseline, reflecting a reduction of 2.5 h per day spent in a hyperglycemic state. Interestingly, improvement in TIR was already evident after 1 month and was sustained during the 6-month period, suggesting a precocious and stable effect of these systems on glycemic control. Furthermore, they led to a reduction in HbA1c, which was more evident for individuals with worse baseline levels.

We also observed an important reduction of time spent in hyperglycemia using AHCL systems, compared to baseline MDI or SAP therapy.

These results are relevant whereas detectable changes in brain volumes and cognitive scores in children with T1D are associated with hyperglycemic metrics (10).

Literature confirms these results. Recently, a 12-week open-label prospective study with MiniMed 780G™ in children 2–6 years old (TDD ≥8 IU/day) demonstrated that this device is also safe in this age group, improving glycemic control and reducing parental distress (11).

Similar lines of evidence were gathered from a retrospective analysis conducted by Tornese et al. on 12 children <7 years (minimum TDD 4 IU/day) with MiniMed 780G™ with SmartGuard who have been followed up for 12 months (12).

In another multicenter 13-week trial with Tandem t:slim X2 insulin pump with Control-IQ Technology, 102 children with a mean age of 4 years (TDD ≥5 IU/day) were assigned to receive an advanced hybrid closed-loop system of insulin delivery or standard care. TIR increased from 56.7% ± 18.0% to 69.3% ± 11.1% during follow-up in the closed-loop group. There was also a significant reduction of TAR and HbA1c, while TBR remained stable (13).

Consistent with these data, we did not observe a reduction of TBR. This finding could be explained by the fear of hypoglycemia in very young children, which led the caregivers to try to maintain a strict control on low glucose values. These systems, however, allow parents to become more confident, thanks to the automatic insulin delivery suspension, leading to a reduction of excessive sugar correction and, consequently, a reduction in time spent in hyperglycemia as well.

CV, above the threshold already at baseline, showed a slight increasing trend, similar to the results of Tornese et al. (12) at the end of the follow-up. The median values above the threshold could be attributed to the unpredictability of meals and activity in this age group while the increasing trend could be due to more frequent bolus performed by the pump.

Because of this variability, even if 48 h is sufficient, we decided to prolong the manual period before transitioning to auto mode with MiniMed 780G until 14 days to allow the algorithm to calculate a more precise basal rate.

The strengths of our study include the real-world setting and the evaluation of two different AHCL systems. A limitation of this 6-month study was its relatively short duration of follow-up, although it seems clear that the benefits on glycemic control persist.

Another limit is the small sample size. Our data should be confirmed and further assessed in other studies with extended follow-up periods and more participants.

Conclusion

Our results confirm that AHCL systems are effective in improving glycemic control in preschool children as already shown in previous studies.

Furthermore, these pumps have proven to be safe tools, improving TIR and TAR, working in auto mode also with a TDD <8 IU/day.

In conclusion, AHCL systems, such as MiniMed 780G with SmartGuard and Tandem t:slim X2 with Control-IQ technology, are an effective therapeutic option for children younger than 6 years.

Data availability statement

The original contributions presented in the study are included in the article/supplementary material. Further inquiries can be directed to the corresponding authors.

Ethics statement

The studies involving humans were approved by Ospedale Pediatrico Bambino Gesù, Rome. The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation was not required from the participants or the participants’ legal guardians/next of kin in accordance with the national legislation and institutional requirements.

Author contributions

NR: Conceptualization, Supervision, Writing – original draft. MM: Formal analysis, Methodology, Writing – original draft. CA: Data curation, Investigation, Writing – review & editing. AD: Writing – review & editing, Supervision. LA: Data curation, Formal analysis, Writing – original draft. MA: Conceptualization, Investigation, Writing – review & editing. PC: Supervision, Writing – review & editing. VP: Conceptualization, Validation, Writing – review & editing. AL: Data curation, Writing – original draft. DT: Data curation, Writing – original draft. SC: Resources, Supervision, Writing – review & editing. RS: Conceptualization, Data curation, Resources, Supervision, Validation, Writing – review & editing.

Funding

The author(s) declare financial support was received for the research, authorship, and/or publication of this article. This work was supported by the Italian Ministry of Health with Current Research funds.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

The handling editor MD declared a past co-authorship with the authors NR and RS.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Daneman D, Frank M, Perlman K, Wittenberg J. The infant and toddler with diabetes: challenges of diagnosis and management. Paediatr Child Health. (1999) 4:57–63. doi: 10.1093/pch/4.1.57

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Sundberg F, de Beaufort CE, Krogvold L, Patton S, Piloya T, Smart C, et al. ISPAD Clinical Practice Consensus Guidelines 2022: Managing diabetes in preschoolers. Pediatr Diabetes. (2022) 23:1496–511. doi: 10.1111/pedi.13427

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Nathan DM. The diabetes control and complications trial/epidemiology of diabetes interventions and complications study at 30 years: overview. Diabetes Care. (2014) 37:9–16. doi: 10.2337/dc13-2112

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Barnard K, Thomas S, Royle P, Noyes K, Waugh N. Fear of hypoglycaemia in parents of young children with type 1 diabetes: a systematic review. BMC Pediatr. (2010) 10:50. doi: 10.1186/1471-2431-10-50

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Barnea-Goraly N, Raman M, Mazaika P, Marzelli M, Hershey T, Weinzimer Stuart A, et al. Alterations in white matter structure in young children with type 1 diabetes. Diabetes Care. (2014) 37:332–40. doi: 10.2337/dc13-1388

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Gaudieri PA, Chen R, Greer TF, Holmes CS. Cognitive function in children with type 1 diabetes: a meta-analysis. Diabetes Care. (2008) 31:1892–7. doi: 10.2337/dc07-2132

PubMed Abstract | CrossRef Full Text | Google Scholar

7. American Diabetes Association. Diabetes technology: standards of medical care in diabetes—2022. Diabetes Care. (2022) 45:S97–S112. doi: 10.2337/dc22-S007

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Libman I, Haynes A, Lyons S, Rwagasor E, Yuet-ling Tung J, Jefferies Craig A, et al. ISPAD Clinical Practice Consensus Guidelines 2022: Definition, epidemiology, and classification of diabetes in children and adolescents. Pediatr Diabetes. (2022) 23:1160–74. doi: 10.1111/pedi.13454

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Battelino T, Danne T, Bergenstal RM, Amiel SA, Beck R, Biester T, et al. Clinical targets for continuous glucose monitoring data interpretation: recommendations from the international consensus on time in range. Diabetes Care. (2019) 42:1593–603. doi: 10.2337/dci19-0028

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Mauras N, Buckingham B, White NH, Tsalikian E, Weinzimer SA, Joet B, et al. Impact of type 1 diabetes in the developing brain in children: A longitudinal study. Diabetes Care. (2021) 44(4):983–92. doi: 10.2337/dc20-2125

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Pulkkinen MA, Varimo TJ, Hakonen ET, Harsunen MH, Hyvönen ME, Janér JN, et al. MiniMed 780G™ in 2- to 6-year-old children: safety and clinical outcomes after the first 12 weeks. Diabetes Technol Ther. (2023) 25(2):100–7. doi: 10.1089/dia.2022.0313

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Tornese G, Carletti C, Lanzetta MA, Tamaro G, Barbi E, Faleschini E, et al. Safety of real-life usage of advanced hybrid closed-loop system miniMed 780G in children with type 1 diabetes younger than 7 years old. Diabetes Care. (2023) 46:e123–5. doi: 10.2337/dc22-2046

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Wadwa RP, Reed ZW, Buckingham BA, DeBoer DM, Ekhlaspour L, Forlenza GP, et al. Trial of hybrid closed-loop control in young children with type 1 diabetes. N Engl J Med. (2023) 388:991–1001. doi: 10.1056/NEJMoa2210834

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: T1D (type 1 diabetes), CSII (continuous subcutaneous insulin infusion), children, insulin, AHCL

Citation: Rapini N, Martino M, Arnaldi C, Deodati A, Anagnostopoulou L, Amodeo ME, Ciampalini P, Pampanini V, Lorubbio A, Tosini D, Cianfarani S and Schiaffini R (2024) Efficacy and safety of advanced hybrid closed loop systems in children with type 1 diabetes younger than 6 years. Front. Endocrinol. 15:1382920. doi: 10.3389/fendo.2024.1382920

Received: 06 February 2024; Accepted: 19 April 2024;
Published: 21 May 2024.

Edited by:

Maurizio Delvecchio, University of L’Aquila, Italy

Reviewed by:

Joseph Isidore Wolfsdorf, Boston Children’s Hospital and Harvard Medical School, United States
Barbara Piccini, Meyer University Hospital, Italy

Copyright © 2024 Rapini, Martino, Arnaldi, Deodati, Anagnostopoulou, Amodeo, Ciampalini, Pampanini, Lorubbio, Tosini, Cianfarani and Schiaffini. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Mariangela Martino, bWFyaWFuZ2VsYS5tYXJ0aW5vLjAxQHN0dWRlbnRzLnVuaXJvbWEyLmV1; Riccardo Schiaffini, cmljY2FyZG8uc2NoaWFmZmluaUBvcGJnLm5ldA==

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.