Skip to main content

ORIGINAL RESEARCH article

Front. Endocrinol., 24 August 2023
Sec. Thyroid Endocrinology
This article is part of the Research Topic 2023 Symposium on Parathyroid Fluorescence View all 14 articles

TOETVA parathyroid autofluorescence detection: hANDY-i endoscopy attachment

Stefanie SeoStefanie Seo1Khalid Mohamed AliKhalid Mohamed Ali1Samantha A. WolfeSamantha A. Wolfe2Nimesh V. NagururuNimesh V. Nagururu1Andy S. DingAndy S. Ding1Dipan DesaiDipan Desai1R. Alex HarbisonR. Alex Harbison1Yoseph KimYoseph Kim3Bo NingBo Ning4Richard Jaepyeong Cha,,Richard Jaepyeong Cha3,4,5Jonathon O. Russell*Jonathon O. Russell1*
  • 1Department of Otolaryngology – Head and Neck Surgery, Johns Hopkins School of Medicine, Baltimore, MD, United States
  • 2Department of Endocrine Surgery, University Health Network, Toronto, ON, Canada
  • 3Department of Research, Optosurgical, LLC, Columbia, MD, United States
  • 4Sheikh Zayed Institute, Children’s National Hospital, Washington, DC, United States
  • 5Department of Pediatrics, George Washington University School of Medicine and Health Sciences, Washington, DC, United States

Background: Treatment options for thyroid pathologies have expanded to include scarless and remote access methods such as the transoral endoscopic thyroidectomy vestibular approach (TOETVA). Currently, no standardized methods exist for locating parathyroid glands (PGs) in patients undergoing TOETVA, which can lead to parathyroid injury and subsequent hypocalcemia. This early feasibility study describes and evaluates the hANDY-i endoscopic attachment for detecting PGs in transoral thyroidectomy.

Methods: We used a prototype parathyroid autofluorescence imager (hANDY-i) that was mounted to a 10-mm 0-degree endoscope. The device delivers a split screen view of Red-green-blue (RGB) and near-infrared autofluorescence (NIRAF) which allows for simultaneous anatomical localization and fluorescence visualization of PGs during endoscopic thyroid dissection.

Results: One cadaveric case and two patient cases were included in this study. The endoscopic hANDY-i imaging system successfully visualized PGs during all procedures.

Conclusion: The ability to leverage parathyroid autofluorescence during TOETVA may lead to improved PG localization and preservation. Further human studies are needed to assess its effect on postoperative hypocalcemia and hypoparathyroidism.

Introduction

Accurate intraoperative identification of parathyroid glands (PG) is critical for gland preservation and prevention of hypoparathyroidism in transoral endoscopic thyroidectomy vestibular approach (TOETVA), a recently described procedure that has gained global popularity for thyroid and parathyroid procedures (1). TOETVA owes its attractiveness in particular to its avoidance of virtually all cutaneous scars (2). Compared to the standard transcervical thyroidectomy approach (TCA), TOETVA is predicted to decrease psychological morbidity associated with cervical scars and increase overall quality of life (35). While complication rates in this newer technique are comparable to those in TCA, permanent or transient hypoparathyroidism remains the most commonly reported complication (6).

Currently, no standardized methods exist for locating PGs in patients undergoing thyroidectomies. Indocyanine green (ICG) was previously a widely used exogenous dye to help improve PG visualization, but its setbacks were allergic reactions to dye and slow onset of fluorescence signals (7). Near-infrared autofluorescence (NIRAF) is a novel technique that utilizes the intrinsic biological fluorophores of PG cells for noninvasive and automated identification (8). While there are FDA-approved NIRAF devices, Fluobeam LX and PTeye, these systems are not compatible with the laparoscopic surgical approach to the central neck. Our team has recently prototyped and reported feasibility of a handheld optical module called hANDY-i in open thyroidectomy procedures (9). The hANDY-i endoscopic attachment is an adaptation of our device for transoral procedures and allows identification of PGs without the need for any additional incisions. In the landscape of many newly emerging PG detection devices, this paper describes one of the earliest transoral applications of a PG imaging system.

Materials and methods

hANDY-i is one of the earliest reported non-invasive, handheld devices that produces a combined RGB and NIRAF view of PGs during thyroid or parathyroid surgery (9). This prototype employs two imaging modalities to reveal PGs in a superimposable, split screen view within the surgical workflow and can be used for both open and transoral cases. hANDY-i consists of a coaxial 785 nm fiber laser, dual-sensors for RGB and NIRAF imaging, and an embedded processing board (Figure 1A). This camera module was connected to a 10mm 0-degree endoscope that allows for use during transoral surgeries (Figure 1B). The connected display monitor shows two superimposable images: a true-color view of the operating field in mode 1 and label-free NIRAF localization of PGs in mode 2.

FIGURE 1
www.frontiersin.org

Figure 1 The TOETVA autofluorescence imaging system consists of four parts as shown (A). The camera head (ITS Model-L) includes both visible and near-infrared (NIR) sensors (B). The NIR sensor detects autofluorescence emission light (>820nm) from parathyroid glands. A custom-designed bifurcated fiber light-guide was used to transmit both visible and near-infrared laser light into a 0-degree Hopkins telescope.

Our early feasibility study consisted of one cadaveric case and two patient cases. This study was approved by the Johns Hopkins Institutional Review Board (IRB00224302). The operating surgeon followed the standard TOETVA procedure consisting of a surgical incision to the oral vestibule, expansion of the vestibular access, use of blunt dissectors to expand the subplatysmal space with enlargement of the surgical field, and placement of trocars. The hANDY-i endoscopic attachment was inserted through the 10mm central trocar (Figure 2). Mode 1 displayed the RGB color view of the operating field. Mode 2 was used to isolate PGs, seen as superimposable, hyperfluorescent spots on a hypofluorescent background. PG location was confirmed by visual inspection. The visual cues used by the surgeon were (1) yellow-brown color, (2) typical ovoid shape, and (3) distinct vasculature. Parathyroid vascularization was qualitatively evaluated by assessing the color of the gland and visualizing evidence of bleeding from the microcirculation following removal of the thyroid.

FIGURE 2
www.frontiersin.org

Figure 2 Standard set-up of hANDY-i for TOETVA thyroidectomy: the camera endoscope was inserted through the 10mm central trocar.

Results

Case 1: cadaver

The left superior (Figure 3A) and a left subcapsular, intrathyroidal (Figure 3B) glands were detected with hANDY-i. The second gland was difficult to visualize by the surgeon’s naked eye, but hANDY-i showed an area of fluorescence along the inferolateral edge of the thyroid lobe corresponding to a subcapsular, intrathyroidal PG. Next, the entire thyroid lobe specimen was removed and re-examined ex-vivo with hANDY-i. The fluorescent tissues were dissected from the surrounding thyroid parenchyma and inspected with Fluobeam LX, which confirmed positive fluorescent results.

FIGURE 3
www.frontiersin.org

Figure 3 RGB (left) and NIRAF (right) split screen views of six parathyroid glands accessed via TOETVA. Row 1: left superior (A) and left subscapular (B) from cadaver. Row 2: left superior (C) and right inferior (D) from patient 1. Row 3: left superior (E) and left inferior (intracapsular) (F) parathyroid glands from patient 2. STP, Superior thyroid pedicle; SM, strap muscles; ITC, inferior thyroid capsule.

Case 2: patient 1

The patient was a 38-year-old female with Graves’ disease with nodular evolution who underwent a total thyroidectomy for a 7 cm left benign nodule and a 2 cm right benign nodule. The left superior (Figure 3C), right superior, and right inferior (Figure 3D) glands were detected with hANDY-i and preserved in situ. Re-inserting the endoscope, it was noted that the left superior and right inferior glands appeared devascularized upon visual inspection. They were subsequently removed from the patient, minced, and autotransplanted into the respective sternohyoid muscles. Post-operative PTH and calcium were 9 pg/mL and 9.3 mg/dL respectively, and the patient was started on 30 days of calcium supplementation. At one month follow-up, her PTH corrected to 27 pg/mL.

Case 3: patient 2

The patient was a 47-year-old female who underwent a left lobectomy for a 4 cm benign thyroid nodule. The left superior (Figure 3E) and left inferior glands were detected with hANDY-i. Initially, a separate PG-like tissue on the laterally reflected portion was thought to be the left inferior gland, but hANDY-i did not show autofluorescence (AF) in this tissue and instead located AF in the capsule (Figure 3F), indicating that the true PG was still intracapsular. The inferior PG was meticulously dissected from the thyroid gland and anatomically preserved maintaining the tissue encompassing the pedicle of the inferior thyroid artery. The parathyroid gland was therefore excluded from the final thyroid specimen. Maintaining anatomic orientation, both glands were well-perfused upon visual inspection based on the color and presence of blood flow and left in situ. Calcium and PTH values were not drawn per standard practice for thyroid lobectomies at our institution.

Discussion

Recent developments in remote access thyroidectomy techniques have necessitated imaging modalities that meet new operative needs. Currently there are two FDA approved NIRAF parathyroid detection devices: Fluobeam LX and PTeye. While Fluobeam has a sensitivity and specificity of 94.1% and 80% respectively, the device’s high light sensitivity requires a dark operating field, potentially hindering its incorporation into the workflow (10). PTeye is a similar device with the distinction of a probe-based design. PTeye has a comparable overall accuracy of 94.3%, but the lack of a video feedback and need for patient-specific calibration complicate its use in the operating room (11). While Fluobeam LX and PTeye can only be utilized in open transcervical thyroidectomy, there have been recent developments in endoscopic PG detection devices. The da Vinci Firefly and EleVision are the first two reported devices designed to be compatible with endoscopic thyroidectomy. The da Vinci Firefly is an ICG fluorescence-based technology that was first introduced in 2017. While the authors showed its applicability in bilateral axillo-breast approach (BABA) robotic thyroidectomy, Firefly is not PG specific and utilizes ICG fluorescence instead of NIRAF (12). EleVision bears more similarities to hANDY-i in that it displays NIRAF and ICG angiography and can be mounted to an endoscopic device (13). EleVision successfully located PGs in the first three transoral thyroidectomy cases as described in a 2022 communication (14). To date, the endoscopic applicability of Firefly and EleVision, as with hANDY-i, is still in early testing phases with promising results but few reported cases available in literature.

In this feasibility study, the endoscopic attachment of hANDY-i was used to achieve detection of PGs in both cadaveric and clinical scenarios of transoral thyroid surgery. In all seven of the PGs identified, the hANDY-i fluorescence signal correctly corresponded to the true PG confirmed by the surgeon’s visual inspection. Of these, two PGs were initially discovered only by AF, then subsequently confirmed to be PGs with Fluobeam LX AF and surgeon inspection. This outcome highlights a strength of NIR technologies in detecting even PGs that are buried in other tissue or lack standard visual cues. However, as this is an early feasibility report, future studies with larger case series and randomized controlled trials will be necessary to clearly show the advantage of hANDY-i for reducing post-operative hypoparathyroidism.

A limitation of any image-based identification system is that only PGs that are within the visual field can be identified. For example, in the cadaveric case, the left inferior PG was initially not visualized on endoscopy and subsequently excised with the thyroid specimen. In normal surgical scenarios, this PG would have been devitalized. A second limitation is that PG localization with this method does not inform the surgeon of its viability, which limits its clinical utility in prognosticating the risk of postoperative hypocalcemia. Further improvements in technology are needed to overcome both limitations.

Advancements in video-assisted endoscopic surgery hold a wealth of opportunities for patient outcome improvement beyond the capacity of open surgeries. Remote access thyroidectomy methods have significantly improved cosmetic outcomes (15, 16) while achieving comparable complication rates (1722), garnering increasing attention globally since the first completely video-assisted endoscopic thyroid lobectomy was introduced in 1997 (2325). Additional features that allow assessment of PG viability in real time without the need for ICG or other injectables will add value to patients and surgeons. Moreover, the ability to display both NIR and tissue viability within an endoscopic system would allow for rapid transition between technologies that promises improved imaging and clinical decision making. Pairing these technologies in a single system that does not disrupt normal operative workflow may lead to improved patient safety and decreased costs in a manner that improves adoption.

Conclusion

In conclusion, the endoscopic prototype of hANDY-i extends the applicability of our device to transoral thyroid surgery, a minimally invasive technique that continues to rise in demand. This is one of the first early feasibility studies on a TOETVA-specific PG detection system. Notably, hANDY-i achieved identification of two nontraditional PGs which were initially missed during standard visual inspection. Larger scale clinical applications are necessary to corroborate these findings.

Author’s note

The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving human participants were reviewed and approved by Johns Hopkins Institutional Review Board (IRB00224302). The patients/participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

SS: Conception/design, data acquisition, data analysis, manuscript drafting, data presentation, final approval. KA: Conception/design, data acquisition, manuscript drafting, final approval. SW: Conception/design, data acquisition, manuscript drafting, final approval. NN: Conception/design, data acquisition, manuscript drafting, final approval. AD: Conception/design, data acquisition, manuscript drafting, final approval. DD: Conception/design, data acquisition, manuscript drafting, final approval. AH: Conception/design, data acquisition, manuscript drafting, final approval. YK: Conception/design, data acquisition, manuscript drafting, final approval. BN: Conception/design, data acquisition, manuscript drafting, final approval. RC: Conception/design, data acquisition, manuscript drafting, final approval. JR: Conception/design, data acquisition, manuscript drafting, final approval. All authors contributed to the article and approved the submitted version.

Funding

Funding for this work was provided by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health under Award Number R44EB030874.

Acknowledgments

Authors thank InTheSmart Co. Ltd for a device loan for this work.

Conflict of interest

YK is an employee of Optosurgical LLC. RC has ownership stake in Optosurgical LLC. JR is a consultant for Baxter Scientific.

The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fendo.2023.1233956/full#supplementary-material

References

1. Banuchi VE, Ballakur SS, Russell JO, Tufano RP. Benefits and risks of scarless thyroid surgery. Ann Thyroid (2020) 5:24–4. doi: 10.21037/aot-20-52

CrossRef Full Text | Google Scholar

2. Anuwong A, Sasanakietkul T, Jitpratoom P, Ketwong K, Kim HY, Dionigi G, et al. Transoral endoscopic thyroidectomy vestibular approach (TOETVA): indications, techniques and results. Surg Endosc (2018) 32(1):456–65. doi: 10.1007/s00464-017-5705-8

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Russell JO, Razavi CR, Shaear M, Liu RH, Chen LW, Pace-Asciak P, et al. Transoral thyroidectomy: safety and outcomes of 200 consecutive North American cases. World J Surg (2021) 45(3):774–81. doi: 10.1007/s00268-020-05874-8

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Tae K, Ji YB, Song CM, Park JS, Park JH, Kim DS. Safety and efficacy of transoral robotic and endoscopic thyroidectomy: The first 100 cases. Head Neck (2020) 42(2):321–9. doi: 10.1002/hed.25999

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Xuan Nguyen H, Nguyen HX, Thi Hoang H, Van Le Q. Quality of life and surgical outcome of transoral endoscopic thyroidectomy vestibular approach (TOETVA) versus open thyroid surgery: Experience from a single center in Vietnam. J Thyroid Res (2022) 2022:1–8. doi: 10.1155/2022/2381063

CrossRef Full Text | Google Scholar

6. Akritidou E, Douridas G, Spartalis E, Tsourouflis G, Dimitroulis D, Nikiteas NI. Complications of trans-oral endoscopic thyroidectomy vestibular approach: A systematic review. In Vivo (Brooklyn) (2022) 36(1):1–12. doi: 10.21873/INVIVO.12671

CrossRef Full Text | Google Scholar

7. Fanaropoulou NM, Chorti A, Markakis M, Papaioannou M, Michalopoulos A, Papavramidis T. The use of Indocyanine green in endocrine surgery of the neck. Med (United States) (2019) 98(10):1–8. doi: 10.1097/MD.0000000000014765

CrossRef Full Text | Google Scholar

8. Paras C, Keller M, White L, Phay J, Mahadevan-Jansen A. Near-infrared autofluorescence for the detection of parathyroid glands. J BioMed Opt (2011) 16(6):067012. doi: 10.1117/1.3583571

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Kim Y, Lee HC, Kim J, Oh E, Yoo J, Ning B, et al. A coaxial excitation, dual-red-green-blue/near-infrared paired imaging system toward computer-aided detection of parathyroid glands in situ and ex vivo. J Biophotonics (2022) e202200008. doi: 10.1002/jbio.202200008

PubMed Abstract | CrossRef Full Text | Google Scholar

10. De Leeuw F, Breuskin I, Abbaci M, Casiraghi O, Mirghani H, Lakhdar A, et al. Intraoperative near-infrared imaging for parathyroid gland identification by auto-fluorescence: A feasibility study. World J Surg (2016) 40(9):2131–8. doi: 10.1007/s00268-016-3571-5

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Kiernan CM, Thomas G, Baregamian N, Solόrzano CC. Initial clinical experiences using the intraoperative probe-based parathyroid autofluorescence identification system—PTeyeTM during thyroid and parathyroid procedures. J Surg Oncol (2021) 124(3):271–81. doi: 10.1002/jso.26500

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Yu HW, Chung JW, Yi JW, Song RY, Lee JH, Kwon H, et al. Intraoperative localization of the parathyroid glands with indocyanine green and Firefly(R) technology during BABA robotic thyroidectomy. Surg Endosc (2017) 31(7):3020–7. doi: 10.1007/s00464-016-5330-y

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Makovac P, Muradbegovic M, Mathieson T, Demarchi MS, Triponez F. Preliminary experience with the EleVision IR system in detection of parathyroid glands autofluorescence and perfusion assessment with ICG. Front Endocrinol (Lausanne). (2022) 13:1030007(October). doi: 10.3389/fendo.2022.1030007

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Vidal Fortuny J, Mennet A, Ott V, Michel J-M, MorenoLlorente P. Autofluorescence detection and perfusion assessment of parathyroid glands in transoral thyroid excision. VideoEndocrinology (2022). doi: 10.1089/ve.2021.0026. (Case 1).

CrossRef Full Text | Google Scholar

15. Chen LW, Razavi CR, Hong H, Fondong A, Ranganath R, Khatri S, et al. Cosmetic outcomes following transoral versus transcervical thyroidectomy. Head Neck (2020) 42(11):3336–44. doi: 10.1002/hed.26383

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Liao D, Ishii LE, Chen LW, Chen J, Juarez M, Darrach HM, et al. Transoral neck surgery prevents attentional bias towards the neck compared to open neck surgery. Laryngoscope (2020) 130(6):1603–8. doi: 10.1002/lary.28305

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Razavi CR, Russell JO. Indications and contraindications to transoral thyroidectomy. Ann Thyroid (2017) 2(5):12–2. doi: 10.21037/aot.2017.10.01

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Russell JO, Razavi CR, Shaear M, Chen LW, Lee AH, Ranganath R, et al. Transoral vestibular thyroidectomy: current state of affairs and considerations for the future. J Clin Endocrinol Metab (2019) 104(9):3779–84. doi: 10.1210/jc.2019-00116

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Razavi CR, Tufano RP, Russell JO. Completion thyroidectomy via the transoral endoscopic vestibular approach. Gland Surg (2018) 7(Suppl 1):S77–9. doi: 10.21037/gs.2018.02.01

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Dionigi G, Bacuzzi A, Lavazza M, Inversini D, Boni L, Rausei S, et al. Transoral endoscopic thyroidectomy: preliminary experience in Italy. Updates Surg (2017) 69(2):225–34. doi: 10.1007/s13304-017-0436-x

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Guo F, Wang W, Zhu X, Xiang C, Wang P, Wang Y. Comparative Study between Endoscopic Thyroid Surgery via the Oral Vestibular Approach and the Areola Approach. J Laparoendosc Adv Surg Tech (2020) 30(2):170–4. doi: 10.1089/lap.2019.0562

CrossRef Full Text | Google Scholar

22. Arikan M, Riss P, Scheuba C, Schopf S, Karakas E, Klein G, et al. Transoral thyroidectomy: initial results of the European TOETVA study group. World J Surg (2023) 47:1201–8. doi: 10.1007/s00268-023-06932-7

PubMed Abstract | CrossRef Full Text | Google Scholar

23. Chong KH, Wu MH, Lai CW. Comparison of surgical outcome between conventional open thyroidectomy and endoscopic thyroidectomy through axillo-breast approach. Tzu Chi Med J (2020) 32(3):286–90. doi: 10.4103/tcmj.tcmj_109_19

CrossRef Full Text | Google Scholar

24. Hüscher CS, Chiodini S, Napolitano C, Recher A. Endoscopic right thyroid lobectomy. Surg Endosc (1997) 11(8):877. doi: 10.1007/s004649900476

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Bellantone R, Lombardi CP, Raffaelli M, Rubino F, Boscherini M, Perilli W. Minimally invasive, totally gasless video-assisted thyroid lobectomy. Am J Surg (1999) 177(4):342–3. doi: 10.1016/S0002-9610(99)00054-9

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: parathyroid glands, near-infrared autofluorescence, endoscopy, remote-access thyroidectomy, thyroidectomy

Citation: Seo S, Ali KM, Wolfe SA, Nagururu NV, Ding AS, Desai D, Harbison RA, Kim Y, Ning B, Cha RJ and Russell JO (2023) TOETVA parathyroid autofluorescence detection: hANDY-i endoscopy attachment. Front. Endocrinol. 14:1233956. doi: 10.3389/fendo.2023.1233956

Received: 03 June 2023; Accepted: 12 July 2023;
Published: 24 August 2023.

Edited by:

Paolo Miccoli, University of Pisa, Italy

Reviewed by:

Marco Stefano Demarchi, Hôpitaux universitaires de Genève (HUG), Switzerland
Pietro Princi, Ospedale Cristo Re, Italy
Pornpeera Jitpratoom, Police General Hospital, Thailand

Copyright © 2023 Seo, Ali, Wolfe, Nagururu, Ding, Desai, Harbison, Kim, Ning, Cha and Russell. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Jonathon O. Russell, jrusse41@jhmi.edu

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.