Front. Cardiovasc. Med., 08 January 2024
Sec. Aortic Surgery and Endovascular Repair Archive
This article is part of the Research Topic Case Reports in Aortic Surgery and Endovascular Repair: 2023 View all 10 articles

Case Report: The application of amplatzer vascular plug to repair aortic dissection intimal tears and false lumen

  • Department of General Surgery, Vascular Surgery, Qilu Hospital of Shandong University, Jinan, Shandong, China

In recent years, significant advancements have been made in endovascular therapy for aortic dissection, resulting in the development of various treatment methods. Nevertheless, there is a contentious discussion regarding the suitability of different treatment methods for addressing the unique features of individual lesions. Specifically, the matter of whether the presence of small intimal tears or multiple distal tears requires extensive aortic coverage with graft stents, which may block the blood supply to important organs, intercostal arteries, or lumbar arteries, remains unresolved. Further research is necessary to determine the need for complex fenestrated or branched endovascular aortic repair (F/B-EVAR) in the management of chronic post-dissection aneurysms involving visceral arteries. Our report highlights the successful endovascular repair of three patients with aortic dissection, encompassing one Stanford A type and two Stanford B types. Utilizing the Amplatzer Vascular Plug (AVP) alone or in combination with the coil embolization technique, we achieved favorable clinical outcomes. This article aims to provide valuable insights and new perspectives on the tailored management of aortic dissection by conducting a comprehensive examination of the subtle differences in the treatment processes and techniques applied to these three patients.

Introduction

Aortic dissection is a severe cardiovascular emergency (1), occurring at a rate of 2–3 cases per 100,000 individuals in China, leading to a mortality rate of 7.3% (2). Guidelines recommend various aortic replacement surgical techniques for Stanford Type A, while Type B is treated with endovascular aortic repair (EVAR) (3, 4). Advancements in technology prompt exploration of endovascular intervention for Type A. However, both open and endovascular surgeries involve intricate procedures with significant mortality rates (5). Hence, embolization is a technique that medical professionals carefully consider for treating commonly occurring intimal tears or stable chronic aortic dissection.

The Amplatzer Vascular Plug (AVP), a vascular occlusion device, is designed to optimize embolization in various endovascular surgery applications (6, 7). Available in four models, each tailored to different vascular anatomy and clinical scenarios, AVP's application for the direct occlusion of aortic dissection entry tears has limited study, and evidence supporting its safety and efficacy is insufficient. Our center selectively treated three patients with Stanford A and Stanford B type aortic dissection, using AVP II (Abbott, Minnesota, USA) to close intimal tears or combining it with coils for false lumen embolization (Table 1). Our initial idea was to use embolization techniques to reconstruct the aneurysm neck, addressing post-dissection aortic aneurysms involving visceral arteries. This approach aimed to avoid complex surgeries for preserving visceral arteries and to create conditions for standard EVAR procedures (Figure 1). To our delight, the results of the first surgery exceeded our expectations, with both patients showing complete thrombosis of the false lumen, eliminating the need for subsequent procedures in the true lumen. In this paper, we delve into the analysis of the treatment process and technical intricacies, aiming to provide valuable experiences and fresh perspectives for the surgical management of aortic dissection.

Table 1
www.frontiersin.org

Table 1. Demographics of 3 patients with chronic dissection treated with amplatzer vascular plug II.

Figure 1
www.frontiersin.org

Figure 1. Aortic dissection treated using AVPs and coils.

Case report

Case 1: A 51-year-old female underwent thoracic endovascular aortic repair (TEVAR) for aortic dissection 2 years ago. The patient was readmitted because of persistent chest discomfort. Computed tomography angiography (CTA) revealed multiple tears at the origin of the ascending aorta and around the renal artery, with post-dissection aneurysm formation (Supplementary Figures S1A,C).

Under general anesthesia, a 6 F catheter was inserted through the right femoral artery to perform false lumen angiography in the ascending aorta (Figure 2A). A 4 mm × 7 mm AVP II was placed in the tear, with two parts in the false lumen and one in the true lumen. Aortography confirmed proper placement of AVP and absence of false lumen opacification (Figure 2B). Abdominal aortography used to find tear near renal artery (Figure 2C). A 7 × 25 mm Viabahn stent (W. L. Gore & Associates, Flagstaff, AZ, USA) was placed in the right renal artery to block the tear at the opening of the renal artery. Then, using a 7F guide catheter, we inserted an AVP II measuring 10 × 7 mm into the tear in the abdominal aorta. Aortography confirmed proper AVP placement with smooth blood flow in the right renal artery stent, and no false lumen opacification (Figure 2D). After 6 months, the follow-up CTA showed complete thrombosis in the false lumen of both the ascending and abdominal aorta without endoleaks. The AVP position remained stable. The abdominal aorta diameter decreased from 51 to 38 mm, while the true lumen diameter increased from 20 to 27 mm (Supplementary Figures S1B,D).

Figure 2
www.frontiersin.org

Figure 2. (A) Intraoperative imaging of aneurysm cavity. (white arrow) (B) Use AVP to embolize the tear of ascending aortic. (blue arrow) (C) The false lumen of the abdominal aorta. (white arrow) (D) Final intraoperative angiography showed the AVP (blue arrow) and the stent (white arrow).

Case 2: A 62-year-old male underwent fenestrated or branched endovascular aortic repair (F/B-EVAR) for aortic arch and thoracic aorta 9 months ago for type B aortic dissection. Post-surgery CTA revealed a 57 mm Post-dissection abdominal aortic aneurysm extending to both iliac arteries. Multiple tears were located at the level of the celiac trunk and bilateral iliac arteries (Figure 3A).

Figure 3
www.frontiersin.org

Figure 3. (A) Preoperative CTA showed blood flow in false lumen of aorta. (B) Intraoperative angiography showed the false lumen opacification in abdominal aorta. (C) Angiography showed no opacification of the false lumen after AVPs (white arrow) and coils (blue arrow) placement. (D) 3-month post-surgery CTA shows AVP in abdominal aorta.

To avoid complex endovascular techniques, the goal is to reconstruct the aneurysm's neck as a healthy anchoring site before implementing standard EVAR in the second phase. Under local anesthesia, the false lumen is accessed through an intimal tear in the left iliac artery (Figure 3B). Due to limited space, several Interlock mechanically detachable coils (Boston Scientific, Natick, MA, USA) were inserted at the level of the celiac trunk artery. After that, AVP II with diameters of 14 mm, 18 mm, 20 mm, and 22 mm were released sequentially in the false lumen. Finally, multiple coils (COOK, Bloomington, IN, USA) were placed at the distal false lumen to aid with embolization. Angiography confirmed no false lumen opacification (Figure 3C). Postoperative CTA examination after 3 months showed complete thrombosis of the false lumen in the abdominal aorta. The diameter of the abdominal aorta decreased from 57 mm to 47 mm, while the diameter of the true lumen increased from 22 mm to 37 mm (Figure 3D). Therefore, a second-stage EVAR surgery is not necessary.

Case 3: A 38-year-old male underwent F/B-EVAR for type B aortic dissection 2 years ago. Postoperative CTA follow-up: The patient has a post-dissection abdominal aortic aneurysm, with a maximum diameter of approximately 52 mm. The intimal tears are located at the level of the infrarenal abdominal aorta and the right common iliac artery respectively. All visceral arteries originate from the true lumen (Supplementary Figure S2A).

Procedure 1: Under local anesthesia, the false lumen was accessed via a tear in the right iliac artery (Figure 4A). Subsequently, AVP II with diameters of 16 mm, 20 mm, and 22 mm were placed to embolize the false lumen (Figure 4B). Intraoperatively, angiography revealed that several lumbar arteries arose from the false lumen. Since the therapeutic goal had been attained, the procedure was terminated to prevent acute spinal cord ischemia. The second phase of the surgery is scheduled to take place in a few months. Follow-up CTA revealed persistent blood flow in the mid to distal false lumen of the abdominal aorta, with distal narrowing, indicating the possibility of re-embolization therapy (Supplementary Figures S2B,C). Procedure 2: Under local anesthesia, right femoral artery access. Angiography showed post-dissection aortic aneurysm formation, and the tear was located in the infrarenal abdominal aorta and right iliac artery (Figure 4C). The 8F guiding catheter passed through iliac artery tear into false lumen, then entered true lumen via abdominal aorta tear. Placed 8 mm × 6 mm AVP II at tear site. A 20 × 16 mm AVP II was inserted into the stenosed false lumen, and Interlock mechanically detachable coils (Boston Scientific, Natick, MA, USA) was used to embolize the right internal iliac artery. Subsequently, a 13 mm × 50 mm Viabahn stent (W. L. Gore & Associates, Flagstaff, AZ, USA) and a 10 mm ×80 mm Fluency stent (Bard, Temple, AZ, USA) were placed in the right iliac artery to block the tear. Intraoperative angiography reveals patent visceral arteries and no false lumen opacification (Figure 4D). The 6-month postoperative CTA examination showed stable positions of the AVP and stent graft, with complete thrombosis in the false lumen. The abdominal aorta diameter decreased from 52 mm to 47 mm, while the true lumen increased from 24 mm to 28 mm (Supplementary Figure S2D).

Figure 4
www.frontiersin.org

Figure 4. (A) Intraoperative angiography showed abdominal aortic false lumen opacification. (B) AVPs were inserted into the false lumen. (C) Opacification of false lumen in infrarenal aorta (white arrow); Aneurysmal dilatation of right iliac artery (blue arrow). (D) Final intraoperative angiography showed the AVP (white arrow) and the stent (blue arrow).

Discussion

After TEVAR, the persistence of distal tears impedes the complete closure of the false lumen, leading to inadequate thrombus formation and facilitating the advancement of distal dissection. This becomes a primary reason for secondary surgical treatment and sparks debates on whether to simultaneously seal the distal tear during TEVAR (8, 9). Currently, various techniques exist in the field, such as F/B-EVAR, chimney technique, embedded branch stent, provisional extension to induce complete attachment (PETTICOAT) technique, stent-assisted balloon-induced intimal disruption and relamination in aortic dissection repair (STABILISE) technique, Knickerbocker, etc (10, 11).

Although significant progress has been made in endovascular treatment of aortic dissection, there remain numerous unresolved issues including etiology, pathophysiology, anatomical changes, prognosis, as well as treatment strategies, techniques, and various materials and tools. Currently, clinical research is primarily focused on the improvement of treatment modalities and materials. Due to factors such as aortic compliance and compatibility, the use of covered stents in treatment leads to a relatively high rate of long-term reintervention. For anatomically complex aortic dissections, there has been an emergence of numerous surgical approaches in recent years. The selection of appropriate individualized treatment should be based on a deep understanding of the disease, extensive clinical experience, and advanced technical skills. Currently, commonly chosen techniques for selective patients include chimney technique and F/B-EVAR technique (12, 13). While these methods can achieve the desired therapeutic effects, they also come with a high rate of re-intervention (14, 15) and significant economic costs. Therefore, it is crucial to simplify treatment strategies and procedures based on patient benefit.

At present, there have been only a few case reports that have described the application of AVP for selectively sealing tears and filling false lumens. Compared to other surgical methods, the use of AVP closure for tears has advantages such as less trauma and lower economic costs. Moreover, the most important aspect is that this technique is relatively simple in terms of operation. In 2012, Yeom et al. (16) published the first documented case of employing AVP Ⅱ for the occlusion of a distal tear in chronic type B aortic dissection. Subsequent studies have demonstrated a high success rate of AVP occlusion in sealing the tear, as well as favorable long-term aortic remodeling (1720). AVP is a nickel-titanium mesh occlusion device with multiple layers. It is designed to be precise in positioning, controllable in release, and retrievable. AVP selectively blocks the tears, thereby preventing extensive ineffective coverage of the graft within the true lumen. This significantly lowers the risk of spinal cord ischemia and eliminates the possibility of complications related to the stent.

Using AVP to occlude the tears in combination with false lumen embolization effectively solves the problem of incomplete tears occlusion and small hidden endoleaks. These two methods can also be applied alone. It's worth exploring the use of false lumen embolization to treat post-dissection aortic aneurysm that involve the visceral arteries. This technique can effectively rebuild a healthy anchor zone, which means we can avoid complicated surgeries and even obviating the need for further EVAR.

Based on the concept of simplifying surgery, we employed embolization techniques for selective treatment in three patients with aortic dissection, including one case of Stanford type A, achieving better-than-expected outcomes. The technique procedure is relatively simple, and there's no need for excessive placement of stents within the true lumen.

The AVP occlusion of the intimal tears is applicable for the treatment of select patients with type A and type B aortic dissection. However, strict patient selection is required due to limitations in materials and intimal stability. Firstly, the diameter of the AVP should correspond to the size of the intimal tear, so it's important to choose a tear that is both more regular in shape and smaller than the diameter of the AVP. Furthermore, in the acute phase, the endarterium is fragile and morphologically unstable, with the size of the tears prone to change. There is a potential risk of device displacement and endarterium tearing when AVP is used, making thorough preoperative assessment extremely important. The chronic phase of aortic dissection exhibits a stable intimal flap, which makes it suitable for embolization technique treatment. However, there are also issues such as a relatively large diameter of the entry tears. Therefore, precise measurements should be taken before surgery to select an appropriate AVP. Thirdly, in the area of visceral arteries with true and false lumens blood supply, embolization of the false lumen should be combined with the treatment of graft stents for the visceral arteries to avoid organ ischemia. In the occlusion of AVP with intimal tears located at the opening of cervical artery and visceral artery, it is crucial to exercise caution in order to prevent stenosis or unintentional closure of the arterial opening. Additionally, the use of embolization techniques alone is not suitable for ruptured or impending rupture of aortic dissection.

Conclusion

Simplification of endovascular treatment is one of the important goals in the design of procedures for complicated aortic dissections. The selective application of embolization techniques can yield significant benefits in cases of aortic dissection or post-dissection aortic aneurysms that involve visceral arteries. Although embolization technology is not a recent technique, with continuous improvements and innovations in materials, its application will become more and more widespread. In the future, the development of specialized embolic devices for complex morphologies of aortic dissection will have significant practical value.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving humans were approved by Scientific Research Ethics Committee of Qilu Hospital of Shandong University. The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation was not required from the participants or the participants’ legal guardians/next of kin because this study was a retrospective study, and informed consent of patients was not required. Written informed consent was obtained from the participants for the publication of our study.

Author contributions

RL: Writing – original draft, Writing – review & editing. YL: Conceptualization, Funding acquisition, Supervision, Writing – review & editing. JJ: Conceptualization, Data curation, Funding acquisition, Methodology, Project administration, Resources, Supervision, Validation, Writing – review & editing.

Funding

The author(s) declare financial support was received for the research, authorship, and/or publication of this article.

This report is supported by the National Natural Science Foundation of China (Grant No. 82000451).

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcvm.2023.1337430/full#supplementary-material

References

1. Mussa FF, Horton JD, Moridzadeh R, Nicholson J, Trimarchi S, Eagle KA. Acute aortic dissection and intramural hematoma: a systematic review. JAMA. (2016) 316(7):754–63. doi: 10.1001/jama.2016.10026

PubMed Abstract | Crossref Full Text | Google Scholar

2. Tang X, Lu K, Liu X, Jin D, Jiang W, Wang J, et al. Incidence and survival of aortic dissection in urban China: results from the national insurance claims for epidemiological research (NICER) study. Lancet Reg Health West Pac. (2021) 17:100280. doi: 10.1016/j.lanwpc.2021.100280

PubMed Abstract | Crossref Full Text | Google Scholar

3. Riambau V, Böckler D, Brunkwall J, Cao P, Chiesa R, Coppi G, et al. Editor’s choice - management of descending thoracic aorta diseases: clinical practice guidelines of the European society for vascular surgery (ESVS). Eur J Vasc Endovasc Surg. (2017) 53(1):4–52. doi: 10.1016/j.ejvs.2016.06.005

PubMed Abstract | Crossref Full Text | Google Scholar

4. Erbel R, Aboyans V, Boileau C, Bossone E, Bartolomeo RD, Eggebrecht H, et al. 2014 ESC guidelines on the diagnosis and treatment of aortic diseases: document covering acute and chronic aortic diseases of the thoracic and abdominal aorta of the adult. The task force for the diagnosis and treatment of aortic diseases of the European society of cardiology (ESC). Eur Heart J. (2014) 35(41):2873–926. doi: 10.1093/eurheartj/ehu281

PubMed Abstract | Crossref Full Text | Google Scholar

5. Gaudino M, Khan FM, Rahouma M, Naik A, Hameed I, Spadaccio C, et al. Spinal cord injury after open and endovascular repair of descending thoracic and thoracoabdominal aortic aneurysms: a meta-analysis. J Thorac Cardiovasc Surg. (2022) 163(2):552–64. doi: 10.1016/j.jtcvs.2020.04.126

PubMed Abstract | Crossref Full Text | Google Scholar

6. Rimon U, Heldenberg E, Golan G, Shinfeld A, Garniek A. Amplatzer vascular plug: expanding the applications spectrum. Cardiovasc Intervent Radiol. (2008) 31(Suppl 2):S84–7. doi: 10.1007/s00270-007-9042-5

PubMed Abstract | Crossref Full Text | Google Scholar

7. Lopera JE. The amplatzer vascular plug: review of evolution and current applications. Semin Intervent Radiol. (2015) 32(4):356–69. doi: 10.1055/s-0035-1564810

PubMed Abstract | Crossref Full Text | Google Scholar

8. Zhang S, Chen Y, Zhang Y, Shi D, Shen Y, Bao J, et al. Should the distal tears of aortic dissection be treated? The risk of distal tears after proximal repair of aortic dissection. Int J Cardiol. (2018) 261:162–6. doi: 10.1016/j.ijcard.2018.01.028

PubMed Abstract | Crossref Full Text | Google Scholar

9. Hu Z, Zhang M, Li B, Hong J, Wang Z. “Wait and watch” is a rational strategy for treating distal tears of acute aortic dissection. Int J Cardiol. (2018) 271:322. doi: 10.1016/j.ijcard.2018.05.104

PubMed Abstract | Crossref Full Text | Google Scholar

10. Zeng Z, Zhao Y, Wu M, Bao X, Li T, Feng J, et al. Endovascular strategies for post-dissection aortic aneurysm (PDAA). J Cardiothorac Surg. (2020) 15(1):287. doi: 10.1186/s13019-020-01331-8

PubMed Abstract | Crossref Full Text | Google Scholar

11. Spanos K, Kölbel T. Role of endoluminal techniques in the management of chronic type B aortic dissection. Cardiovasc Intervent Radiol. (2020) 43(12):1808–20. doi: 10.1007/s00270-020-02566-7

PubMed Abstract | Crossref Full Text | Google Scholar

12. He Y, Jia S, Sun G, Cao L, Wang X, Zhang H, et al. Fenestrated/branched endovascular repair for postdissection thoracoabdominal aneurysms: a systematic review with pooled data analysis. Vasc Endovascular Surg. (2020) 54(6):510–8. doi: 10.1177/1538574420927131

PubMed Abstract | Crossref Full Text | Google Scholar

13. Oikonomou K, Kopp R, Katsargyris A, Pfister K, Verhoeven EL, Kasprzak P. Outcomes of fenestrated/branched endografting in post-dissection thoracoabdominal aortic aneurysms. Eur J Vasc Endovasc Surg. (2014) 48(6):641–8. doi: 10.1016/j.ejvs.2014.07.005

PubMed Abstract | Crossref Full Text | Google Scholar

14. Spear R, Sobocinski J, Settembre N, Tyrrell MR, Malikov S, Maurel B, et al. Early experience of endovascular repair of post-dissection aneurysms involving the thoraco-abdominal aorta and the arch. Eur J Vasc Endovasc Surg. (2016) 51(4):488–97. doi: 10.1016/j.ejvs.2015.10.012

PubMed Abstract | Crossref Full Text | Google Scholar

15. Spear R, Hertault A, Van Calster K, Settembre N, Delloye M, Azzaoui R, et al. Complex endovascular repair of postdissection arch and thoracoabdominal aneurysms. J Vasc Surg. (2018) 67(3):685–93. doi: 10.1016/j.jvs.2017.09.010

PubMed Abstract | Crossref Full Text | Google Scholar

16. Yeom SK, Lee SH, Chung HH, Shin JS. Endovascular repair of growing chronic type B aortic dissection with a vascular plug. Acta Radiol. (2012) 53(6):648–51. doi: 10.1258/ar.2012.110605

PubMed Abstract | Crossref Full Text | Google Scholar

17. Huh U, Song S, Kim E, Kim CW. Thoracic endovascular aortic repair using vascular plug for chronic type IIIb aortic dissection. Ann Thorac Surg. (2017) 104(6):e413–5. doi: 10.1016/j.athoracsur.2017.08.003

PubMed Abstract | Crossref Full Text | Google Scholar

18. Falkenberg M, Roos H, Lepore V, Svensson G, Zachrisson K, Henrikson O. Endovascular closure of chronic dissection entries in the aortic arch using the amplatzer vascular plug II as a sealing button. J Endovasc Ther. (2016) 23(2):378–83. doi: 10.1177/1526602816632351

PubMed Abstract | Crossref Full Text | Google Scholar

19. De Boo DW, Mott N, Kavnoudias H, Walton A, Lyon SM. Endovascular closure of ascending aortic pseudoaneurysm with a type II amplatzer vascular plug. Vasc Endovascular Surg. (2014) 48(4):329–32. doi: 10.1177/1538574414522329

PubMed Abstract | Crossref Full Text | Google Scholar

20. Coscas R, Capdevila C, Colacchio G, Goëau-Brissonnière O, Coggia M. Hybrid repair of type II dissecting thoracoabdominal aneurysm using amplatzer vascular plugs for entry tear closure. Ann Vasc Surg. (2013) 27(7):972.e1–5. doi: 10.1016/j.avsg.2013.03.004

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: endovascular treatment, embolization technique, amplatzer vascular plug, coils, aortic dissection

Citation: Li R, Liu Y and Jiang J (2024) Case Report: The application of amplatzer vascular plug to repair aortic dissection intimal tears and false lumen. Front. Cardiovasc. Med. 10:1337430. doi: 10.3389/fcvm.2023.1337430

Received: 13 November 2023; Accepted: 21 December 2023;
Published: 8 January 2024.

Edited by:

Gianni Angelini, University of Bristol, United Kingdom

Reviewed by:

Antonio Maria Calafiore, Henry Dunant Hospital, Greece
Walter Gomes, Federal University of São Paulo, Brazil

© 2024 Li, Liu and Jiang. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Jianjun Jiang amlhbmdqaWFuanVuMTk5NUAxNjMuY29t

These authors have contributed equally to this work and share first authorship

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.