ORIGINAL RESEARCH article

Front. Cardiovasc. Med., 14 December 2023
Sec. Cardiovascular Surgery

Standardized in vitro bleeding tests in a non-coated novel hybrid prosthesis for frozen elephant trunk demonstrates minimal oozing during full heparinization, supported by clinical data

\r\nHeinz Jakob,
Heinz Jakob1,2*Timo Leins,Timo Leins3,4Meltem Avci-AdaliMeltem Avci-Adali5Christian SchlensakChristian Schlensak5Daniel Wendt,Daniel Wendt6,2Yatin MehtaYatin Mehta7Matthias ThielmannMatthias Thielmann2Klaus GrlingerKlaus Görlinger8Suk-Won SongSuk-Won Song9Konstantinos Tsagakis\r\nKonstantinos Tsagakis2
  • 1Diagnosticum Mülheim, Mülheim, Germany
  • 2Department of Thoracic- and Cardiovascular Surgery, West German Heart and Vascular Center, Essen, Germany
  • 3Medira GmbH, Balingen, Germany
  • 4JOTEC-Artivion, Hechingen, Germany
  • 5Department of Thoracic and Cardiovascular Surgery, University Hospital Tuebingen, Tuebingen, Germany
  • 6CytoSorbents Europe, Berlin, Germany
  • 7Medanta Institute of Critical Care and Anesthesiology, Medanta - The Medicity, Gurgaon, India
  • 8Tem Innovations GmbH, Munich, Germany
  • 9Department of Thoracic and Cardiovascular Surgery, College of Medicine, Ewha Womans University, Seoul, Republic of Korea

Introduction: Recent reports have questioned the blood impermeability of the novel frozen elephant trunk (FET) device E-vita Open NEO© (EO-NEO). Therefore, standardized in vitro bleeding tests using porcine heparinized blood were performed, as well as stress testing on the blood tightness of the collar suture line, to investigate this observation.

Material and methods: EO-NEO prostheses were examined in vitro for blood permeability in three test series. Initially, antegrade perfusion with heparinized porcine blood [activated clotting time (ACT) of 500 s, with a 60 min duration] was performed, followed by ante/retrograde testing via the EO-NEO side port. Testing of the collar suture line under a tension of 10 Newton (N) within a suspension device (blood pressure 120 mmHg, ACT of 560 s, 1 min duration) was carried out with the suture material force fiber white (FFWs) yarn, using standard fixation (5 stitches/cm), FFWh yarn in hemostatic fixation (15 stitches/cm), and flow weave yarn (FWYh).

Results: Blood permeability testing of EO-NEO through the prosthetic lumen or via the side port demonstrated minor leakage without statistical difference between the standard and hemostatic suture lines or suture materials used, or positioning on the crimped or tapered portion (p > 0.05). The specific collar anastomosis testing demonstrated leakage volumes of 140 ml/min for FFWs vs. 16 ml/min for FFWh (p = 0.02), vs. 9 ml/min with the FWYh (p = 0.01).

Conclusion: Different blood leakage tests showed minimal oozing and no difference in blood loss through the fabric and different collar suture lines, but unphysiological pressurized retrograde perfusion of the collar region showed significantly less leakage using FWYh and FFWh, prompting production modification of EO-NEO. Clinical results confirmed low blood loss using this novel FET device.

Introduction

In 2005, the first commercially available hybrid aortic arch, the descending aorta stent graft, was introduced by us, the authors, in collaboration with JOTEC GmbH in Hechingen, Germany (1). The woven prosthesis was non-coated to be intussusceptive into the stented portion, which itself was compressed within a polyester sleeve and mounted on the application system. This offered the possibility of introducing the hybrid prosthesis into the descending aorta through the opened arch and releasing the stented portion. Thereafter, the anastomosis with the distal arch in zone 3 prior to unfolding the intussuscepted arch portion, by retracting it into the arch position for anastomosing with the head vessels and the ascending aortic graft, was completed. Another important aspect of this new concept was the avoidance of bovine-origin collagen coating. It is known that bovine coating might theoretically be associated with the potential transfer of bovine spongiform encephalopathy (BSE, mad cow disease). Therefore, the graft fabric required meticulous pre-sealing with human-derived fibrin glue (Tisseel©, Baxter, Germany) for blood impermeability during full heparinization (2). To simplify its application, the later named E-vita Open Plus© (EO PLUS) graft was manufactured by a tighter weaving process and introduced in 2008 with a significant reduction in oozing through the graft’s pores during full heparinization, which could be proven in experimental and clinical studies (3). With the introduction of the CE-marked EO-NEO in 2020, the tightly woven material was maintained with the principle of non-precoating (4). This applies to all three new variations of frozen elephant trunk (FET) prostheses—straight, branched, and trifurcated versions—to offer easier surgical adaptation to any kind of pathoanatomic situation.

A casuistic report by Czerny et al. on three massive oozing incidents, and a consecutive in vitro study by Tan and Bashir, as well as a systematic coagulopathy review by Bashir and colleagues, raised concerns (57). Those reports, as well as individual case reports, prompted us, in collaboration with the JOTEC CryoLife/ARTIVION company and the Clinical Research Laboratory of Department of Cardiac, Thoracic and Vascular Surgery, University Hospital Tübingen, to scrutinize those observations by performing in vitro experiments with the novel hybrid prosthesis. The following questions had to be addressed: Where does the reported casuistic bleeding come from? Is there a mechanical defect? Finally, the results were related to the clinical experiences reported thus far.

Material and methods

The prerequisite basic facts are: there is no difference in the fabric between EO PLUS and EO-NEO. The fabric tubes for EO Plus and EO-NEO are produced on the same loom, the collar material is identical (laser cut out of the same fabric tube), and the suture material has the same specification (5.0 braided surgical suture) in both. For basic orientation, a TM-1000 (Hitachi, Tokyo, Japan) scanning electron microscope (SEM) was used to study suture hole width resulting from the manufacturing after a tension of 10 N was placed on the collar anastomosis of the native EO-NEO and was compared with EO PLUS at a 60× magnification and preceded blood stress testing (Figure 1). The test circuit consisted of a roller pump connected to silicone tubing, porcine blood bag, manometer, and flow probe. The silicone tubing was partitioned into four branches, all connected to four containment chambers, in which the prostheses were mounted to sample blood loss over defined time slots. The EO-NEO devices tested are depicted in Figure 2. Owing to the COVID-19 pandemic lockdown in 2021/2022, the originally planned use of human blood had to be abandoned. Thus, easily available fresh heparinized porcine blood had to be used.

Figure 1
www.frontiersin.org

Figure 1. SEM demonstration of suture hole stress testing (10 N) on a Hitachi TM-1000 scanning electron microscope. C, collar; S, standard suture line for collar fixation; A, arch portion fabric.

Figure 2
www.frontiersin.org

Figure 2. E-vita open - NEO© (CE-mark 3/2020) in its three variations. Straight, branched, and trifurcated including a perfusion port to accomodate implantation at all pathoanatomic situations. Anastomotic options according to Ishimaru zones 0–III.

First test set up: whole prosthesis blood permeability testing by antegrade perfusion

The test set up consisted of four in-series placed containment chambers in which four straight full-length EO-NEO prostheses without a perfusion port were clamped in using PTFE bands and silk ties for fixation into inflow and outflow hollow silicone plugs with integrated 3/8' tubing. These were connected with adapters to the tubing circuit with integrated flow probes and manometers (Figures 3A,B).

Figure 3
www.frontiersin.org

Figure 3. (A) Schematic drawing of antegrade perfusion with fresh heparinized porcine blood through EO-NEO. 1. Porcine blood bag; 2. 3/8" Slicone tubing – flow direction; 3. Roller pump; 4. Manometer; 5. Flow probe; 6. Branched inflow into EO NEO straight, no side arm; 7. Containment chamber; 8. EO NEO Stented portion; 9. Prosthetic outflow; 10. Blood pressure regulation with clamp. (B) Experimental set up. (C) Perfusion with heparinized porcine blood. Sample 1: FFWh, hemostatic fixation; Sample 2: FWYh, hemostatic fixation; Sample 3: EEs, standard fixation; Sample 4: FFWBs, standard fixation.

The collar was expanded with two stay sutures horizontally fixed at the chamber’s upper edges under mild tension. Blood loss was documented in separate measuring cylinders. Two bags of heparinized porcine blood (ACT of 500 s) were used to fill the test circuit and pump flow was started with a conventional roller pump after deairing at a full flow of 2 L/min at 40 mmHg for equilibration within 5 min, followed by the augmentation of blood pressure to 120 mmHg by partial clamping of the tubing for the ensuing 55 min. Oozing through the prostheses’ pores was collected and measured at 5-min intervals for 60 min (Figure 3C).

The prosthetic material consisted of tapered EO-NEO full-length prostheses, and different 5–0 sewing material, with the collar anastomosed either in standard fashion (5 stitches per cm) or hemostatic (15 stitches per cm):

Sample 1: Suture material A (force fiber white, FFWh), hemostatic fixation

Sample 2: Suture material B (flow weave yarn, FWYh) hemostatic fixation

Sample 3: Suture material C (Ethibond Excel, EEs), standard green filament from EO PLUS, standard fixation

Sample 4: Suture material D (force fiber white-blue, FFWBs), standard fixation

Second test set up: whole prosthesis blood permeability perfused ante-retrogradely via the side port

To control the suture integrity of the fabric and branches, including the collar and perfusion port, dynamic blood permeability testing was performed as in series 1. The test set up consisted of four in-series placed containment chambers with four EO-NEO prostheses clamped in. The flow rate was set at 2 L/min at a pressure primarily at 40 mmHg for 5 min for equilibration, and thereafter at 120 mmHg. ACT was equilibrated at 500 s. Inflow was directed into the straight EO-NEO prostheses via the side arm perfusion port with the proximal prosthetic end blocked (Figure 4A). Oozing through the prostheses’ pores was collected and measured at 5-min intervals for 60 min after the primary 5-min equilibration phase. To scrutinize whether the position of the collar suture line was more effective either at the crimped arch portion of the EO-NEO or on the tapered portion of the stented part of the prosthesis, the test set up was constructed as follows (Figure 4B):

Sample 1: Collar position moved to the first crimp, suture material A1 (FFWs), standard fixation suture technique

Sample 2: Standard collar positioning at the tapered portion, suture material B1 (FWYh), hemostatic suture technique

Sample 3: Standard collar position at the tapered portion, suture material A2 (FFWs), standard fixation suture technique

Sample 4: Collar position moved to the first crimp, suture material B2 (FWYh), hemostatic suture technique

Figure 4
www.frontiersin.org

Figure 4. (A) Schematic drawing of ante-retrograde perfusion through the side port at the prosthesis’ arch portion. 1. Porcine blood bag; 2. 3/8" Slicone tubing – flow direction; 3. Roller pump; 4. Manometer; 5. Flow probe; 6. Inflow into EO NEO via side arm; 7. Containment chamber; 8. EO NEO Stented portion; 9. Prosthetic outflow; 10. Blood pressure regulation with clamp. (B) Perfusion with heparinized porcine blood (ACT of 500). 1. FFWs – Collar at crimp, standard fixation; 2. FWYh – Collar at tapering, hemostatic fixation; 3. FFWs – Collar at tapering, standard fixation; 4. FWYh – Colar at crimp, hemostatic fixation.

Third test set up: simulation of retrograde flow onto the collar only

Five hundred milliliters of heparinized (ACT of 560 s) porcine blood was brought through 3/8' silicone tubing vertically mounted and manometrically controlled at a pressure of 120 mmHg for 1 min onto the collar region of three test sets with five samples each of EO-NEO prostheses reduced to a length of 10 cm (crimped status), with the prosthetic lumen obliterated. The collar was put under a tension of 10 N to overemphasize a clinical scenario (Figure 5). The sewing material and sewing technique tested was:

Sample 1: Suture material A (FFWs), standard suture technique

Sample 2: Suture material A (FFWh), hemostatic suture technique

Sample 3: Suture material B FWYh), hemostatic suture technique

Figure 5
www.frontiersin.org

Figure 5. Sole collar blood permeability testing under tension (10 N) with heparinized porcine blood at 120 mmHg for 1 min (ACT of 559 s). (A) Schematic drawing of collection vessel with plugged prosthetic lumen (B) Leakage measuring unit for collar anastomosis. (C) Demonstration of sole collar leaking into collection vessel.

The leakage was measured within a collection vessel in milliliters per 1 min.

Statistical analysis

Data were analyzed using SPSS version 27 software (IBM, Armonk, NY, USA). Leakage volumes (continuous variables) were expressed as mean ± standard deviation (SD) and compared using Student's t-test. Clinical data (categorical data) were expressed as the number of patients and frequencies or mean ± standard deviation (SD). P-values <0.05 were considered significant.

Results

SEM analyses and the comparison of suture hole width under a stress tension of 10 N stress did not show any microscopic difference between the classic EO PLUS and EO-NEO fabrics (Figure 1). In the first test series with antegrade perfusion of full-length EO-NEO prostheses without perfusion ports or branches, there was no significant difference (p > 0.05) in blood loss by oozing through the prostheses’ pores between the four different suture techniques of collar anastomoses, as demonstrated in Table 1 and Figure 6. Within a few minutes, primary oozing ceased through the prosthetic pores and no difference in collar suture bleeding was observed. After the observation time span, longitudinal transection of the prostheses allowed measurement of the total prosthetic area. Blood loss was marginal (between 0.0021 and 0.0037 ml/cm2 per minute) with the four different suture materials regardless of technique (standard or hemostatic) (p > 0.05) (Table 1A).

Table 1
www.frontiersin.org

Table 1. Blood loss measurements (ml) through full-length EO NEO at 120 mmHg over 60 min after 5 min of equilibration at a pressure of 40 mmHg (ACT of 500 s).

Figure 6
www.frontiersin.org

Figure 6. Full-length EO NEO blood permeability testing at 120 mmHg H2O over 60 min (ml/min) with different collar suture variants.

Table 1A
www.frontiersin.org

Table 1A. Average leakage/oozing results (ml per cm2 per minute; ACT of 500 s, full length prostheses).

In test series 2, ante-retrogradely perfusion via the perfusion port again did not show any significant difference between the location on either the crimped or tapered portion of the EO-NEO, no matter whether the suturing technique was standard or hemostatic, as demonstrated in Table 2 and in the graphic representation in Figure 7 (p > 0.05).

Table 2
www.frontiersin.org

Table 2. Whole prosthesis blood permeability perfused ante-retrogradely via the side port.

Figure 7
www.frontiersin.org

Figure 7. Antegrade perfusion through the side port for 60 min (first 5 min at 40 mmHG, thereafter at 120 mmHg; ACT of 614 s).

In test series 3, the unphysiological stress test directing pressurized heparinized porcine blood at 120 mmHg precisely on the expanded collar under a tension of 10 N with five samples each demonstrated significantly less permeability with the hemostatic suture lines for collar fixation than with the standard suture fixation FFWs (mean 140 ml/min compared with 16 ml/min for FFWh, p =0.02) or with FWYh (mean blood loss 9 ml/min, p = 0.01, Figure 8 and Table 3)

Figure 8
www.frontiersin.org

Figure 8. Stress test for collar anastomosis by retrograde pressurized exposure to 120 mmHg for 1 min with heparinized porcine blood (ACT of 559 s).

Table 3
www.frontiersin.org

Table 3. Stress test of collar anastomosis by retrograde pressurized exposure to 120 mmHg for 1 min with heparinized porcine blood (ACT of 559 s).

Clinical data from Essen University Hospital (Table 4) and Ewha Womans University Aorta and Vascular Hospital, Seoul, South Korea (Table 5) confirm low perioperative blood product use due to minor blood loss within 24 hours. The low postoperative re-operation rate for bleeding was based on collar surrounding big bites over a teflon felt supporting close attachment of the aortic wall to the collar site of the EO NEO (Essen), or a double row suture technique (Seoul). All prostheses used had standard collar anastomosis fabrication.

Table 4
www.frontiersin.org

Table 4. Clinical results of the university of Essen.

Table 5
www.frontiersin.org

Table 5. Clinical results university of Ewha womans university.

Discussion

The FET concept certainly represents an important step forward, aimed at reducing the previous two or even three step approaches needed to cope with multilevel thoracic aortic disease, but it does require advanced surgical skills, especially when all-comer pathologies, including emergency situations, are to be handled (8). Currently, four devices are available globally: the Cronus open (MicroPort Medical, Shanghai, China) from 2004 used mainly in China (9); the Frozenix-J graft (LifeLine, Tokyo, Japan) used mainly in Japan from 2014 (10); and the E-vita open (plus) graft from 2005 (1) and the Thoraflex hybrid graft (Vascutek, Inchinnan, United Kingdom) from 2010 (11), both used mainly in the western world. A first report of a significant improvement regarding long term freedom from reoperation after acute type A aortic dissection surgery, in FET vs. proximal repair, was reported by us in 2020 (12). Based on the same manufacturing process as for the EO PLUS, the device becomes blood impervious after a brief period of equilibration even in the heparinized state. However, the aforementioned casuistic reports on disturbing pinhole bleeding at the collar site or severe diffuse oozing through the fabric prompted us to restudy the new device in a standardized in vitro set up.

There are three main observations from the current study: first, scanning electron microscopic orientation did not show any difference in suture hole size at the collar anastomotic sites compared with the decade-long proven blood impermeability of the EO Plus fabric (3) and the EO-NEO structure (Figure 1). This was also proven for the transluminal orthograde perfusion of the straight EO NEO, regardless of the suture material applied or the standard vs. hemostatic suture fixation of the collar during 60-min perfusion with heparinized porcine blood at 120 mmHg. The oozing volumes through the fabrics, ranging between 0.0026 and 0.0033 ml/cm2/min, were negligible and certainly do not play a role in the clinical field (Figure 6 and Table 1A). In the ante-retrograde perfusion test series via the side port, again there was no significant difference in blood loss over the 60-min test period, regardless of whether the collar was attached to the crimped (arch portion of the graft) or to the non-crimped proximally tapered stented portion, irrespective of the suture fixation technique. The resulting blood loss between 37 and 80 ml over 60 min in the fully heparinized state is not clinically relevant (>0.05, Figure 7 and Table 2). However, in series 3, unphysiological stress testing only of the expanded collar region with 10 N for targeted identification of a potential weak spot revealed a significant difference between standard suture line fixation and hemostatic anastomoses with 15 stitches per 1 cm (Figure 8 and Table 3). This finding explains the few personal casuistic reports regarding continuous pinhole bleeding at the collar anastomotic sites, probably associated with a large periaortic space and enormous back bleeding during reperfusion. For prevention of this bleeding problem, two surgical approaches are recommended: first, the avoidance of a hollow space between the proximal stented portion of the EO-NEO and the surrounding aortic tissue. Trimming of the collar to a seam height of 5 mm for a tight anastomosis with the surrounding aortic tissue, reinforced by a close-fitting fixed Teflon strip outside, ensures the blood tightness at the anastomotic site. In case of the inevitable hollow space around the prosthesis, with large backflow and stress onto the suture line, as expected in aneurysms or dissections with a large distal false lumen, this problem can be overcome through a v-shaped resection of the pathological aortic tissue and a provisional double row 3–0 polypropylene suture to achieve a close proximity of the aortic tissue to the EO-NEO. Then, placement of wide surrounding 3–0 polypropylene continuous suture bites on the Teflon-bolstered approximated aortic tissue through the stented graft portion to the arch part of the prosthesis, ignoring the collar, will result in absolute blood tightness, as experienced with the classic EOP for many years (2). Second, another efficient surgical but more time consuming approach is to position additional Teflon pledged 5–0 U-stitches, secured by the application of BioGlue, as propagated by the Hongkong and Seoul groups (14, 15).

In addition to the surgical aspects of this complex surgery, we have often had to deal with intraoperative hyperinflammation, metabolic changes, and hypothermia that lead to bleeding complications. In particular, the hemostasis system and coagulation cascade might be impaired by (deep) hypothermia and the cardiopulmonary bypass itself (13, 1619). In addition to these mandatory but potentially harmful intraoperative prerequisites, the coagulation system could also be significantly influenced or impaired in patients treated with P2Y12 inhibitors or direct oral anticoagulants (DOACs) (2022). In some patients, an aortic syndrome or dissection might be misdiagnosed as an acute coronary syndrome being loaded with potent P2Y12 inhibitors such as ticagrelor. Moreover, the current increased prescription of DOACs, e.g., in chronic atrial fibrillation, has increased bleeding complications in urgent cardio-surgical patients without the recommended washout period (2326). To overcome and attenuate these potentially harmful effects of hyperinflammation, coagulation, and antithrombotic effects, intraoperative hemoadsorption, e.g., using the CytoSorb® hemoadsorption device, has been introduced into the clinical routine for complex cardiosurgical cases, both intraoperatively and postoperatively (2734). Mehta et al. (35) showed significantly improved postoperative hemodynamics in aortic surgery patients by applying intraoperative hemoadsorption. Another large propensity score-matched pairs analysis (168 vs. 168 patients) from Munich, Germany, demonstrated that the use of intraoperative hemoadsorption during aortic surgery was associated with a significant reduction in the intraoperative transfusion of blood products, a more stable acid-base balance, and a lower requirement for vasopressors (36). Moreover, it has been shown that in the worst-case scenario of operating on emergent aortic surgery patients with aortic dissections under antithrombotic medication, intraoperative antithrombotic removal by hemoadsorption is effective and results in significantly less bleeding events, blood product use, and surgical re-explorations. Accordingly, a recently published review recommended antithrombotic removal in the cardiovascular surgery setting (37). The removal of direct factor Xa inhibitors, such as rivaroxaban and apixaban, during cardiopulmonary bypass can also be monitored by thromboelastometry (38). This new opportunity for intraoperative antithrombotic drug removal has now also been added to and recommended by the updated intraoperative guidelines of the European Society of Anaesthesiology and Intensive Care (39).

Regarding the serious questions raised by Czerny et al. (5) of continuous oozing through the graft pores in three cases operated on in Freiburg, Germany, and Bologna, Italy, even after protamine application, the hemostatic therapy required huge amounts of red blood cell concentrate, fresh frozen plasma, and platelet transfusion, but not cryoprecipitate or fibrinogen concentrate based on the thromboelastometry results. Accordingly, coagulopathic bleeding cannot be excluded as a reason for the excessive oozing, as reported in this case series (4042). Bleeding and transfusion of red blood cells and “yellow blood products” such as fresh frozen plasma and platelets, are well known risk factors for increased morbidity [transfusion-associated circulatory overload (TACO), transfusion-associated lung injury (TRALI), and transfusion-related immunomodulation (TRIM) with nosocomial infection and multiple organ failure (MOF)] and mortality in major cardiovascular surgery and other clinical set ups (18, 43).

Limitations

We believe that the use of porcine heparinized blood for in vitro blood permeability testing of the new EO NEO hybrid graft was a successful surrogate for human blood, as shown by an ACT beyond 500 s throughout the experiments. This was required because the COVID-19 lockdown resulted in a scarcity of human blood. Owing to the stability of the results achieved with sample sizes of only four to five, it seemed appropriate not to perform more experiments, especially in light of the production controls applied by the manufacturer for the quality control of the fabric in regard to homogeneous weaving and the placement of branch attachments using hemostatic suture techniques.

In conclusion

Even in worst case scenarios such as Penn Class BC dissection and/or devastating preoperative conditions of the patient, the combination of rapid surgery and the described optimized surgical techniques, as well as optimized coagulation treatment with cryoprecipitate, fibrinogen concentrate, and prothrombin complex concentrate guided by thromboelastometry, is promising improved results. The anecdotally reported pinhole bleeding instances have been corrected meanwhile by the manufacturer by modifying the standard anastomotic suture line at the collar to a hemostatic one using FWY (Figure 9). The results of the in vitro tests clearly demonstrated the blood tightness of the EO-NEO grafts in different set ups, thus the principle of the tightly woven non-covered polyester graft design can be maintained. Excellent clinical results that rely on the aforementioned principles with few instances of perioperative bleeding can be achieved, as demonstrated at Essen University Hospital, Germany, and Ewha Womans University Aorta and Vascular Hospital, South Korea, as well as by other groups.

Figure 9
www.frontiersin.org

Figure 9. Consequences of the in vitro test series (A–D). Pinhole leakage (red ellipse) during pre-experiment pressurized saline testing of the standard suture line (A) Optimization of standard suture lines (A,C) to the hemostatic suture line (B,D, blue arrows) using flow weave yarn. (B) the red circle demonstrates the absolute water tightness of the collar suture line post correction.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Author contributions

HJ: Conceptualization, Investigation, Methodology, Resources, Supervision, Visualization, Writing – original draft, Writing – review & editing. TL: Conceptualization, Data curation, Formal Analysis, Investigation, Methodology, Writing – review & editing. MA-A: Validation, Writing – review & editing, Investigation. CS: Visualization, Writing – review & editing. DW: Conceptualization, Formal Analysis, Investigation, Methodology, Project administration, Validation, Visualization, Writing – original draft, Writing – review & editing. YM: Resources, Validation, Writing – review & editing. MT: Conceptualization, Validation, Visualization, Writing – review & editing. KG: Methodology, Validation, Visualization, Writing – review & editing. S-WS: Investigation, Validation, Writing – review & editing. KT: Investigation, Data curation, Resources, Writing – review & editing.

Funding

The author(s) declare that no financial support was received for the research, authorship, and/or publication of this article.

Conflict of interest

HJ was a consultant for Artivion and received royalties until 12/2019. KG is a full-time employee of TEM Innovations GmbH, and DW is a full-time employee of CytoSorbents Europe. TL was employed by Medira GmbH.

The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

The author(s) declared that they were an editorial board member of Frontiers, at the time of submission. This had no impact on the peer review process and the final decision.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Jakob H, Tsagakis K, Leyh R, Buck T, Herold U. Development of an integrated stent graft-dacron prosthesis for intended one-stage repair in complex thoracic aortic disease. Herz. (2005) 30(8):766–8. doi: 10.1007/s00059-005-2765-7

PubMed Abstract | Crossref Full Text | Google Scholar

2. Jakob H, Tsagakis K, Tossios P, Massoudy P, Thielmann M, Buck T, et al. Combining classic surgery with descending stent grafting for acute DeBakey type I dissection. Ann Thorac Surg. (2008) 86(1):95–101. doi: 10.1016/j.athoracsur.2008.03.037

PubMed Abstract | Crossref Full Text | Google Scholar

3. Tsagakis K, Pizanis N, Baba HA, Wendt D, Wisbrun N, Altenbernd J, et al. Impermeability to blood of the E-vita open plus hybrid stent-graft: experimental and clinical evaluation. J Endovasc Ther. (2010) 17(3):340–8. doi: 10.1583/09-2758MR.1

PubMed Abstract | Crossref Full Text | Google Scholar

4. Jakob H, Ho JYK, Wong RHL, Idhrees M, Velayudhan B, Matalanis G, et al. Paving the way for E-vita open NEO hybrid prosthesis implantation for complex aortic arch disease in Asia-Pacific. J Card Surg. (2021) 36(10):3963–7. doi: 10.1111/jocs.15882

PubMed Abstract | Crossref Full Text | Google Scholar

5. Czerny M, Beyersdorf F, Murana G, Pacini D. Excessive oozing through the fabric of the branched Cryolife-Jotec Evita open NEO hybrid prosthesis. Eur J Cardiothorac Surg. (2021) 60(2):423–4. doi: 10.1093/ejcts/ezab058

PubMed Abstract | Crossref Full Text | Google Scholar

6. Tan S, Bashir M. Prevention versus cure: is BioGlue priming the optimal strategy against E-vita NEO graft oozing? J Card Surg. (2022) 37(3):555–60. doi: 10.1111/jocs.16179

PubMed Abstract | Crossref Full Text | Google Scholar

7. Bashir M, Abo Aljadayel H, Mousavizadeh M, Daliri M, Rezaei Y, Tan SZ, et al. Correlation of coagulopathy and frozen elephant trunk use in aortic arch surgery: a systematic review and meta-analysis. J Card Surg. (2021) 36(12):4699–714. doi: 10.1111/jocs.16001

PubMed Abstract | Crossref Full Text | Google Scholar

8. Tsagakis K, Osswald A, Weymann A, Demircioglu A, Schmack B, Wendt D, et al. The frozen elephant trunk technique: impact of proximalization and the four-sites perfusion technique. Eur J Cardiothorac Surg. (2021) 61(1):195–203. doi: 10.1093/ejcts/ezab295

PubMed Abstract | Crossref Full Text | Google Scholar

9. Ma WG, Zhang W, Wang LF, Zheng J, Ziganshin BA, Charilaou P, et al. Type a aortic dissection with arch entry tear: surgical experience in 104 patients over a 12-year period. J Thorac Cardiovasc Surg. (2016) 151(6):1581–92. doi: 10.1016/j.jtcvs.2015.11.056

PubMed Abstract | Crossref Full Text | Google Scholar

10. Okita Y. Frozen elephant trunk with frozenix prosthesis. Ann Cardiothorac Surg. (2020) 9(3):152–63. doi: 10.21037/acs.2020.03.13

PubMed Abstract | Crossref Full Text | Google Scholar

11. Shrestha M, Pichlmaier M, Martens A, Hagl C, Khaladj N, Haverich A. Total aortic arch replacement with a novel four-branched frozen elephant trunk graft: first-in-man results. Eur J Cardiothorac Surg. (2013) 43(2):406–10. doi: 10.1093/ejcts/ezs296

PubMed Abstract | Crossref Full Text | Google Scholar

12. Jakob H, Moughal S, Bashir M. Frozen elephant trunk with straight vascular prosthesis: single-center experience with a review of current trends. J Cardiovasc Surg. (2020) 61(3):301–7. doi: 10.23736/S0021-9509.20.11401-0

Crossref Full Text | Google Scholar

13. Ortmann E, Klein AA, Sharples LD, Walsh R, Jenkins DP, Luddington RJ, et al. Point-of-care assessment of hypothermia and protamine-induced platelet dysfunction with multiple electrode aggregometry (multiplate®) in patients undergoing cardiopulmonary bypass. Anesth Analg. (2013) 116(3):533–40. doi: 10.1213/ANE.0b013e31827cee88

PubMed Abstract | Crossref Full Text | Google Scholar

14. Ho JYK, Kim CH, Chow SCY, Kwok MWT, Lee H, Kim TH, et al. Initial Asian experience of the branched E-vita open NEO in complex aortic pathologies. J Thorac Dis. (2023) 15(2):484–93. doi: 10.21037/jtd-22-1055

PubMed Abstract | Crossref Full Text | Google Scholar

15. Ho JYK, Lim K, Fujikawa T, Wong RHL. Pre-emptive BioGlue priming of the branched evita open NEO hybrid prosthesis: tackling excessive oozing. Eur J Cardiothorac Surg. (2022) 61(3):728–9. doi: 10.1093/ejcts/ezab398

PubMed Abstract | Crossref Full Text | Google Scholar

16. Karkouti K, Callum J, Crowther MA, McCluskey SA, Pendergrast J, Tait G, et al. The relationship between fibrinogen levels after cardiopulmonary bypass and large volume red cell transfusion in cardiac surgery: an observational study. Anesth Analg. (2013) 117(1):14–22. doi: 10.1213/ANE.0b013e318292efa4

PubMed Abstract | Crossref Full Text | Google Scholar

17. Mace H, Lightfoot N, McCluskey S, Selby R, Roy D, Timoumi T, et al. Validity of thromboelastometry for rapid assessment of fibrinogen levels in heparinized samples during cardiac surgery: a retrospective, single-center, observational study. J Cardiothorac Vasc Anesth. (2016) 30(1):90–5. doi: 10.1053/j.jvca.2015.04.030

PubMed Abstract | Crossref Full Text | Google Scholar

18. Bartoszko J, Callum J, Karkouti K, investigators Fs. The association of prothrombin complex concentrates with postoperative outcomes in cardiac surgery: an observational substudy of the FIBRES randomized controlled trial. Can J Anaesth. (2021) 68(12):1789–801. doi: 10.1007/s12630-021-02100-4

PubMed Abstract | Crossref Full Text | Google Scholar

19. Baryshnikova E, Di Dedda U, Ranucci M. Are viscoelastic tests clinically useful to identify platelet-dependent bleeding in high-risk cardiac surgery patients? Anesth Analg. (2022) 135(6):1198–206. doi: 10.1213/ANE.0000000000006231

PubMed Abstract | Crossref Full Text | Google Scholar

20. Hansson EC, Geirsson A, Hjortdal V, Mennander A, Olsson C, Gunn J, et al. Preoperative dual antiplatelet therapy increases bleeding and transfusions but not mortality in acute aortic dissection type A repair. Eur J Cardiothorac Surg. (2019) 56(1):182–8. doi: 10.1093/ejcts/ezy469

PubMed Abstract | Crossref Full Text | Google Scholar

21. Xiao FC, Ma WG, Ge YP, Zhu JM, Sun LZ. Does preoperative dual antiplatelet therapy affect bleeding and mortality after total arch repair for acute type A dissection? Interact Cardiovasc Thorac Surg. (2022) 34(1):120–7. doi: 10.1093/icvts/ivab226

PubMed Abstract | Crossref Full Text | Google Scholar

22. Sromicki J, Van Hemelrijck M, Schmiady MO, Kruger B, Morjan M, Bettex D, et al. Prior intake of new oral anticoagulants adversely affects outcome following surgery for acute type A aortic dissection. Interact Cardiovasc Thorac Surg. (2022) 35(1):ivac037. doi: 10.1093/icvts/ivac037

PubMed Abstract | Crossref Full Text | Google Scholar

23. Hassan K, Bayer N, Schlingloff F, Oberhoffer M, Wohlmuth P, Schmoeckel M, et al. Bleeding complications after use of novel oral anticoagulants in patients undergoing cardiac surgery. Ann Thorac Surg. (2018) 105(3):702–8. doi: 10.1016/j.athoracsur.2017.11.066

PubMed Abstract | Crossref Full Text | Google Scholar

24. Erdoes G. Martinez Lopez De Arroyabe B, Bolliger D, Ahmed AB, Koster A, Agarwal S, et al. International consensus statement on the peri-operative management of direct oral anticoagulants in cardiac surgery. Anaesthesia. (2018) 73(12):1535–45. doi: 10.1111/anae.14425

PubMed Abstract | Crossref Full Text | Google Scholar

25. Lee MY, Han S, Bang OY, On YK, Jang SW, Han S, et al. Drug utilization pattern of oral anticoagulants in patients with atrial fibrillation: a nationwide population-based study in Korea. Adv Ther. (2022) 39(7):3112–30. doi: 10.1007/s12325-022-02151-z

PubMed Abstract | Crossref Full Text | Google Scholar

26. Zeitler EP, Ronk CJ, Cockerham A, Huse S, McKindley DS, Kim MH. Healthcare resource utilization in patients with newly diagnosed atrial fibrillation in the United States. Expert Rev Pharmacoecon Outcomes Res. (2022) 22(5):763–71. doi: 10.1080/14737167.2022.2045955

PubMed Abstract | Crossref Full Text | Google Scholar

27. Javanbakht M, Trevor M, Rezaei Hemami M, Rahimi K, Branagan-Harris M, Degener F, et al. Ticagrelor removal by CytoSorb® in patients requiring emergent or urgent cardiac surgery: a UK-based cost-utility analysis. Pharmacoecon Open. (2020) 4(2):307–19. doi: 10.1007/s41669-019-00183-w

PubMed Abstract | Crossref Full Text | Google Scholar

28. Hassan K, Bruning T, Caspary M, Wohlmuth P, Pioch H, Schmoeckel M, et al. Hemoadsorption of rivaroxaban and ticagrelor during acute type A aortic dissection operations. Ann Thorac Cardiovasc Surg. (2022) 28(3):186–92. doi: 10.5761/atcs.oa.21-00154

PubMed Abstract | Crossref Full Text | Google Scholar

29. Hassan K, Thielmann M, Easo J, Kamler M, Wendt D, Haidari Z, et al. Removal of apixaban during emergency cardiac surgery using hemoadsorption with a porous polymer bead sorbent. J Clin Med. (2022) 11(19):5889. doi: 10.3390/jcm11195889

PubMed Abstract | Crossref Full Text | Google Scholar

30. Traeger K, Skrabal C, Fischer G, Datzmann T, Schroeder J, Fritzler D, et al. Hemoadsorption treatment of patients with acute infective endocarditis during surgery with cardiopulmonary bypass - a case series. Int J Artif Organs. (2017) 40(5):240–9. doi: 10.5301/ijao.5000583

PubMed Abstract | Crossref Full Text | Google Scholar

31. Haidari Z, Demircioglu E, Boss K, Tyczynski B, Thielmann M, Schmack B, et al. Intraoperative hemoadsorption in high-risk patients with infective endocarditis. PLoS One. (2022) 17(7):e0266820. doi: 10.1371/journal.pone.0266820

PubMed Abstract | Crossref Full Text | Google Scholar

32. Haidari Z, Leiler S, Mamdooh H, Fittkau M, Boss K, Tyczynski B, et al. Effect of intraoperative haemoadsorption therapy on cardiac surgery for active infective endocarditis with confirmed Staphylococcus aureus bacteraemia. Interdiscip Cardiovasc Thorac Surg. (2023) 36(1):ivad.010. doi: 10.1093/icvts/ivad010

Crossref Full Text | Google Scholar

33. Traeger K, Fritzler D, Fischer G, Schroder J, Skrabal C, Liebold A, et al. Treatment of post-cardiopulmonary bypass SIRS by hemoadsorption: a case series. Int J Artif Organs. (2016) 39(3):141–6. doi: 10.5301/ijao.5000492

PubMed Abstract | Crossref Full Text | Google Scholar

34. Boss K, Jahn M, Wendt D, Haidari Z, Demircioglu E, Thielmann M, et al. Extracorporeal cytokine adsorption: significant reduction of catecholamine requirement in patients with AKI and septic shock after cardiac surgery. PLoS One. (2021) 16(2):e0246299. doi: 10.1371/journal.pone.0246299

PubMed Abstract | Crossref Full Text | Google Scholar

35. Mehta Y, Singh A, Singh A, Gupta A, Bhan A. Modulating the inflammatory response with hemadsorption (CytoSorb) in patients undergoing Major aortic surgery. J Cardiothorac Vasc Anesth. (2021) 35(2):673–5. doi: 10.1053/j.jvca.2020.06.028

PubMed Abstract | Crossref Full Text | Google Scholar

36. Saller T, Hagl C, Woitsch S, Li Y, Niedermayer S, Born F, et al. Haemadsorption improves intraoperative haemodynamics and metabolic changes during aortic surgery with hypothermic circulatory arrest. Eur J Cardiothorac Surg. (2019) 56(4):731–7. doi: 10.1093/ejcts/ezz074

PubMed Abstract | Crossref Full Text | Google Scholar

37. Matejic-Spasic M, Hassan K, Thielmann M, Geidel S, Storey RF, Schmoeckel M, et al. Management of perioperative bleeding risk in patients on antithrombotic medications undergoing cardiac surgery-a systematic review. J Thorac Dis. (2022) 14(8):3030–44. doi: 10.21037/jtd-22-428

PubMed Abstract | Crossref Full Text | Google Scholar

38. Roed-Undlien H, Schultz NH, Lunnan A, Husebraten IM, Wollmann BM, Molden E, et al. In vitro apixaban removal by CytoSorb whole blood adsorber: an experimental study. J Cardiothorac Vasc Anesth. (2022) 36(6):1636–44. doi: 10.1053/j.jvca.2022.01.038

PubMed Abstract | Crossref Full Text | Google Scholar

39. Kietaibl S, Ahmed A, Afshari A, Albaladejo P, Aldecoa C, Barauskas G, et al. Management of severe peri-operative bleeding: guidelines from the European society of anaesthesiology and intensive care: second update 2022. Eur J Anaesthesiol. (2023) 40(4):226–304. doi: 10.1097/EJA.0000000000001803

PubMed Abstract | Crossref Full Text | Google Scholar

40. Hanke AA, Herold U, Dirkmann D, Tsagakis K, Jakob H, Gorlinger K. Thromboelastometry based early goal-directed coagulation management reduces blood transfusion requirements, adverse events, and costs in acute type A aortic dissection: a pilot study. Transfus Med Hemother. (2012) 39(2):121–8. doi: 10.1159/000337723

PubMed Abstract | Crossref Full Text | Google Scholar

41. Desborough M, Sandu R, Brunskill SJ, Doree C, Trivella M, Montedori A, et al. Fresh frozen plasma for cardiovascular surgery. Cochrane Database Syst Rev. (2015) 2015(7):CD007614. doi: 10.1002/14651858.CD007614.pub2

PubMed Abstract | Crossref Full Text | Google Scholar

42. Monaco F, Barucco G, Nardelli P, Licheri M, Notte C, De Luca M, et al. Editor’s choice - a rotational thromboelastometry driven transfusion strategy reduces allogenic blood transfusion during open thoraco-abdominal aortic aneurysm repair: a propensity score matched study. Eur J Vasc Endovasc Surg. (2019) 58(1):13–22. doi: 10.1016/j.ejvs.2019.02.009

PubMed Abstract | Crossref Full Text | Google Scholar

43. Bartoszko J, Martinez-Perez S, Callum J, Karkouti K, Investigators FS. Impact of cardiopulmonary bypass duration on efficacy of fibrinogen replacement with cryoprecipitate compared with fibrinogen concentrate: a post hoc analysis of the fibrinogen replenishment in surgery (FIBRES) randomised controlled trial. Br J Anaesth. (2022) 129(3):294–307. doi: 10.1016/j.bja.2022.05.012

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: dissection, frozen elephant trunk (FET), in vitro, bleeding, aortic surgery, hemoadsorption

Citation: Jakob H, Leins T, Avci-Adali M, Schlensak C, Wendt D, Mehta Y, Thielmann M, Görlinger K, Song S-W and Tsagakis K (2023) Standardized in vitro bleeding tests in a non-coated novel hybrid prosthesis for frozen elephant trunk demonstrates minimal oozing during full heparinization, supported by clinical data. Front. Cardiovasc. Med. 10:1303816. doi: 10.3389/fcvm.2023.1303816

Received: 29 September 2023; Accepted: 23 November 2023;
Published: 14 December 2023.

Edited by:

Wei-Guo Ma, Yale University, United States

Reviewed by:

Guangmin Yang, Nanjing Drum Tower Hospital, China
Tomas Holubec, University Hospital Frankfurt, Germany
Sven Peterss, Ludwig Maximilian University of Munich, Germany

© 2023 Jakob, Leins, Avci-Adali, Schlensak, Wendt, Mehta, Thielmann, Görlinger, Song and Tsagakis. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Heinz Jakob aC5qYWtvYkBkaWFnbm9zdGljdW0uZGU=

Abbreviations FET, frozen elephant trunk; EO-NEO, e-vita open NEO; ACT, activated clotting time; N, Newton; FFWs, force fiber white standard fixation; FFWh, force fiber white hemostatic fixation; FWYh, flow weave yarn hemostatic fixation; BSE, bovine spongiform encephalopathy; SEM, scanning electron microscopy; EO-PLUS, e-vita open PLUS; PTFE, polytetrafluorethylene; EEs, ethibond excel standard fixation; FFWBs, force fiber white-blue standard fixation; “, inch; DOACs, direct oral anticoagulants; TACO, transfusion-associated circulatory overload; TRALI, transfusion-associated lung injury; TRIM, transfusion-related immunomodulation; MOF, multi-organ failure.

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.