Skip to main content

ORIGINAL RESEARCH article

Front. Cardiovasc. Med., 08 September 2023
Sec. Pediatric Cardiology

\r\nRoland Heck,Roland Heck1,2Bjrn Peters,Björn Peters2,3Pia Lanmüller,Pia Lanmüller1,2Joachim Photiadis,Joachim Photiadis2,4Felix Berger,,Felix Berger2,3,5Volkmar Falk,,,Volkmar Falk1,2,5,6Christoph Starck,,,Christoph Starck1,2,5,7Peter Kramer,
\r\nPeter Kramer2,3*
  • 1Department of Cardiothoracic and Vascular Surgery, Deutsches Herzzentrum der Charité (DHZC), Berlin, Germany
  • 2Charité—Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Berlin, Germany
  • 3Department of Congenital Heart Disease—Pediatric Cardiology, Deutsches Herzzentrum der Charité (DHZC), Berlin, Germany
  • 4Department of Congenital and Pediatric Heart Surgery, Deutsches Herzzentrum der Charité (DHZC), Berlin, Germany
  • 5Partner Site Berlin, DZHK (German Center for Cardiovascular Research), Berlin, Germany
  • 6Translational Cardiovascular Technologies, Institute of Translational Medicine, Department of Health Sciences and Technology, Swiss Federal Institute of Technology (ETH) Zurich, Switzerland
  • 7Steinbeis Hochschule, Steinbeis-Transfer-Institut Kardiotechnik, Berlin, Germany

Objectives: Due to the limited longevity of endovascular leads, children require thoughtful lifetime lead management strategies including conservation of access vessel patency. Consequently, there is an increasing interest in transvenous lead extraction (TLE) in children, however, data on TLE and the use of powered mechanical dissection sheaths is limited.

Methods: We performed a retrospective cohort study analyzing all children <18 years that underwent TLE in our institution from 2015 to 2022. Procedural complexity, results and complications were defined as recommended by recent consensus statements.

Results: Twenty-eight children [median age 12.8 (interquartile range 11.3–14.6) years] were included. Forty-one leads were extracted [median dwell time 85 (interquartile range 52–102) months]. Extractions of 31 leads (76%) in 22 patients (79%) were complex, requiring advanced extraction tools including powered bidirectional rotational dissection sheaths in 14 children. There were no major complications. Complete procedural success was achieved in 18 (64%) and clinical success in 27 patients (96%), respectively. Procedural success and complexity varied between lead types. The Medtronic SelectSecure™ lead was associated with increased odds of extraction by simple traction (p = 0.006) and complete procedural success (p < 0.001) while the Boston Scientific Fineline™ II lead family had increased odds of partial procedural failure (p = 0.017).

Conclusions: TLE with the use of mechanical powered rotational dissection sheaths is feasible and safe in pediatric patients. In light of rare complications and excellent overall clinical success, TLE should be considered an important cornerstone in lifetime lead management in children. Particular lead types might be more challenging and less successful to extract.

1. Introduction

Children with congenital dysrhythmias, channelopathies or congenital heart disease (CHD) frequently require implantation of a pacemaker (PM) or implantable cardioverter defibrillator (ICD) in an early stage of life. In smaller children, cardiac implantable electronic devices (CIED) with epicardial leads are often used, due to the limited possibilities of transvenous lead access and the risk for thrombotic vascular complications given the comparably large diameter of lead bodies (1, 2). While short-term performance is comparable to transvenous leads, epicardial leads may have inferior long-time performance, especially in patients with CHD, and lead replacement will require thoracotomy (36). Therefore, primary implantation of transvenous leads or conversion to transvenous leads in case of epicardial lead malfunction is generally preferred in older children (2, 3). However, due to somatic growth and higher levels of physical activity, among other causes, resulting in increased mechanical strain, failure of transvenous leads is frequent in children (7). CIED and lead revisions may also be indicated due to system infection, system upgrade and lead recall or advisory status. There is a large experience with lead extractions in adult patients with CIED and various techniques and extractions tools were developed over recent decades (811). Indications for lead extractions for adult patients have been recommended by current consensus statements (12, 13). In children, however, data on transvenous lead extractions is limited to a small number of observational studies (1420). Accordingly, due to the limited evidence, current consensus statements for CIED therapy in children have largely adopted recommendations on lead extractions from the experience in adult patients but emphasize important knowledge gaps (21). Comparable to adults, the use of various advanced extraction tools has been reported including radiofrequency, laser and mechanical powered sheaths (14, 16, 17, 18, 22). However, in particular data on the use of new generation mechanical rotational sheaths is limited to small numbers of pediatric patients (17, 23, 24).

Clinical practice concerning the decision of extracting or abandoning malfunctioning transvenous leads in children is variable and long-term prospective studies on lead abandonment vs. extraction are missing (14, 15, 25). However, as children dependent on CIED therapy will require functional leads for decades to come, lifetime lead management strategies and conservation of access vessel patency are mandatory. Procedural complexity of lead extraction and the need for advanced extraction techniques are associated with increasing lead age among other factors (16, 19). The incidence of major complications of lead extractions in children is fortunately very low with reported rates of 0%–4% and exceptionally rare procedure related mortality (1419). With these considerations, it has become our institutional policy to attempt transvenous lead extraction in children whenever lead revision is required.

In light of the limited available data, the aim of our present study was to investigate our institutional results with lead extraction procedures in children.

2. Material and methods

We performed a retrospective cohort study reviewing patients that underwent transvenous lead extraction (TLE) in our institution during the study period 2015–2022. Patients were included in the study if they were <18 years of age at the time of TLE. Exclusion criteria were lead dwell time <12 months and surgical lead extraction. If patients underwent more than one extraction procedure during the study period, only the most recent was analyzed. The development of the final cohort is displayed in Figure 1. Demographic, clinical, CIED related and TLE procedural data were retrieved from patients' electronic charts and analyzed. The study was approved by the institutional review board and ethics committee (EA2/278/20), individual informed consent was not required due to the retrospective nature of this study.

FIGURE 1
www.frontiersin.org

Figure 1. Study cohort. LE, Lead extraction.

2.1. Definitions

Procedural results were defined according to the definitions of the 2017 Heart Rhythm Society and the 2018 European Heart Rhythm Association expert consensus (12, 13). Complete procedural success for the extraction of an individual lead was defined as the removal of all components of that lead from the vascular space without the occurrence of a fatal or permanently disabling complication. A procedure was considered a clinical success if it attained the intended clinical outcome and did not involve the retention of any lead portion >4 cm in length. Complications were defined as undesired occurrences related to the procedure. They were classified as minor complications if they required medical intervention or minor procedural intervention or as major complications if they were life-threatening, resulted in death or caused significant or persistent disability, prolongation of the hospital admission or required additional significant therapeutic intervention such as cardiac surgery, pericardiocentesis, or vascular surgery (12, 13). Extraction procedures were classified as simple if only manual traction with non-locking or locking stylets was used or as complex, if advanced extraction tools were required, respectively (14, 16, 19).

2.2. Transvenous lead extraction

All procedures were performed by pediatric cardiologists experienced in CIED therapy and TLE according to current consensus statements recommendations (12). TLE was performed in a hybrid operating suite with monoplane fluoroscopic guidance. Patients were intubated and managed under general anesthesia. A radial arterial line as well as a jugular central venous line were placed and the patient's chest was prepped for emergency sternotomy. Additionally, arterial and venous femoral vascular access was established to allow emergency femoral cardiopulmonary bypass cannulation as well as to permit temporary transvenous pacing and access for a possible femoral extraction approach or possible interventions such as emergency vessel occlusion or lead access vessel revascularization. Transesophageal echocardiography was in place throughout the procedure to allow intraprocedural monitoring and exclusion of cardiovascular complications due to TLE. During TLE procedure, a cardiac surgical team as well as a perfusionist were on standby until leads were extracted and cardiovascular complications ruled out. For lead extraction, a multistep approach was applied. In general, TLE was performed primarily via the lead access vessel. Initially, simple manual traction with a non-locking stylet inserted and, if possible, retraction of the active fixation helix was attempted. If unsuccessful, a locking stylet and compression coil (Liberator®, One-tie®, Cook Medical, USA) were used to support manual traction. In case manual traction was insufficient for lead extraction, advanced extraction tools were employed such as polypropylene telescope dissection sheaths (Byrd Dilator Sheath, Cook Medical, USA) or mechanical powered rotational dissection sheaths (Evolution® RL, Cook Medical, USA). Lumenless leads were extended (Bulldog™ lead extender, Cook Medical, USA). If primary complete extraction of leads failed, extraction of remaining lead components via femoral access was attempted using snare catheters (Needle's Eye Snare®, Cook Medical, USA; EN Snare®, Merit Medical Systems, USA; Multi-Snare®, PFM Medical, Germany).

2.3. Outcomes

Primary outcome was procedural success; secondary outcomes were complexity of extraction procedures and complications.

2.4. Statistics

Statistical analyses were performed using SPSS statistics software (version 25, IBM Corp., NY, USA). Data distribution was tested using D'Agostino-Pearson test. Variables are expressed as figures (percentages) and median [interquartile range, IQR]. Continuous variables were compared using non-parametric Mann–Whitney test. Chi–Square-Test was used for comparison of categorical data. Potential factors associated with primary or secondary outcome were evaluated with univariable logistic regression analysis. P-values <0.05 were considered statistically significant.

3. Results

3.1. Patient cohort

Patient data are provided in Table 1. Twenty-eight pediatric patients who underwent TLE were included in the study (Figure 1). The median age at the first transvenous lead implant was 4.3 [IQR 2.3–7.9] and 6.2 [IQR 3.1–8.5] years at the implantation of indwelling leads, respectively. The median age at TLE was 12.8 [IQR 11.3–14.6] years. Ten patients (35%) had CHD, 8 patients (29%) had a channelopathy and 7 patients (25%) had a congenital dysrhythmia. The remaining patients had acquired dysrhythmias of unknown or presumably inherited origin (n = 4, 14%) and hypertrophic cardiomyopathy with recurrent ventricular tachycardia (n = 1, 4%). Two patients had more than one diagnosis. The most frequent indications for CIED treatment included complete AV block (n = 17, 61%), long QT syndrome (n = 6, 21%), sinus node dysfunction (n = 6, 21%) and ventricular tachycardia (n = 3, 11%), several patients had more than one indication.

TABLE 1
www.frontiersin.org

Table 1. Patient characteristics.

3.2. Procedural details

In the 28 included patients, a total of 41 leads were extracted with a median dwell time of 85 [IQR 52–102] months (Table 2). The most frequent primary indications for lead revision were malfunction (n = 17, 61%) and somatic growth related lead distortion (n = 5, 18%) while CIED infection was rare (n = 1, 4%). Growth related lead distortion as presumable cause was also present in the majority of patients with lead malfunction (13/17, 76%) (Figure 2). In those patients with distorted leads without malfunction, advancing present leads within access vessels [n = 8 leads, median dwell time of 88 (IQR 51–120) months] was primarily intended but unsuccessful in all cases due to vascular adhesions.

TABLE 2
www.frontiersin.org

Table 2. Characteristics of extracted leads.

FIGURE 2
www.frontiersin.org

Figure 2. Chest radiographs of a 12-year-old boy before (A) and after (B) TLE. (A): depleted length reserve (arrow) of the right ventricular lead due to somatic growth. (B): lead length reserve for anticipated further somatic growth of the new lead implanted after TLE. The lead “reserve loop” is positioned in the right atrium (arrow). A fragment of the lead tip from the previously implanted 4472 model (Boston Scientific Fineline™ II) lead (dotted arrow) remained in the right ventricle. TLE, transvenous lead extraction.

Median duration of all TLE procedures (skin-to-skin time) was 162 [IQR 118–200] minutes, median fluoroscopy time was 12.8 [IQR 6.7–22.3] minutes. New leads and, if indicated, generators were implanted in all but two patients during TLE procedures. One patient with hypertrophic cardiomyopathy underwent bi-ventricular assist device implantation due to terminal heart failure. Explantation of her single chamber ICD and lead extraction prior to pending cardiac transplantation was favored and TLE was performed concomitantly with assist device implantation. The second patient had CIED infection. His single chamber pacemaker had been implanted due to intermittent sinus node dysfunction, however, he had low rates of atrial pacing and was therefore judged to be stable enough without temporary pacing. A new pacemaker was implanted subsequently 8 days later.

Extractions of 31 leads (76%) in 22 patients (79%) were considered complex, requiring advanced extraction tools. In 14/22 patients (64%), powered bidirectional rotational dissection sheaths (Cook Evolution® RL and/or Evolution® Shortie RL) were used. Snare catheters to retrieve lead remnants via femoral approach were used in 8 of these 22 patients (36%) in 11 leads. Extraction characteristics are provided in Table 3.

TABLE 3
www.frontiersin.org

Table 3. Procedural details.

3.3. Procedural success

There were no major complications. Tricuspid valve/subpulmonary atrioventricular valve function was unchanged in all patients as documented by transesophageal echocardiography. All patients survived to hospital discharge, median length of hospital stay was 4 [IQR 3–6] days. Minor complications occurred in three patients (11%). One experienced an episode of bradycardia during TLE, which was managed medically without further consequences. The second patient developed episodes of non-sustained ventricular tachycardia during the attempt of extracting remaining fragments of the extracted ventricular lead with a snare catheter. The attempts were abandoned and the dysrhythmias resolved without further interventions required. In a third patient, small lead fragments dislodged to a hepatic vein and could not be retrieved. This, however, was without clinical consequences.

Outcome characteristics are summarized in Table 4. Complete procedural success was achieved in 18 patients (64%) while clinical success was achieved in almost all patients (n = 27, 96%). Partial procedural success with retention of small lead fragments <4 cm were in particular observed with the Boston Scientific Fineline lead models 447x (4471, 4472, 4473, 4474) in our cohort. Frequently, these lead types fragmented just at the tip during extraction, and the lead tips could not be retrieved (Figure 2). The other lead type with only partial procedural success was a bipolar Osypka K5Y lead, which disintegrated at the tip during extraction and small fragments of the lead tip embolized to a hepatic vein. The additional use of femoral access for the extraction of lead fragments with snare catheters was significantly more often required with the 447x lead types compared to other lead types (10/21 vs. 1/20, Chi² p = 0.002).

TABLE 4
www.frontiersin.org

Table 4. Outcome characteristics.

Angiography prior to TLE revealed thrombotic obstructions of access vessels, mostly of the innominate vein, in a considerable number of patients (partial n = 6, 21%, complete n = 2, 7%). In all but one patient, access vessel patency was maintained or achieved by TLE. Two patients (7%), however, required additional balloon angioplasty due to stenosis of the innominate vein after recanalization of complete (n = 1) or partial (n = 1) thrombotic obstruction. In one patient during extraction with a powered rotational dissection sheath, the leads fragmented at the level of the completely obstructed innominate vein. Leads could be successfully extracted via femoral approach by snare catheters; however, recanalization of the innominate vein was unsuccessful.

3.4. Factors associated with procedural success and procedural complexity

Analysis by type of the two most frequently used leads in our cohort shows that the 3830 model was significantly associated with complete procedural success (Chi² p = 0.039, univariable regression p < 0.001). In contrast, the 447x models were significantly associated with decreased odds of complete success (p = 0.017). None of the additional parameters analyzed was associated with procedural success (Table 5).

TABLE 5
www.frontiersin.org

Table 5. Univariable analysis of predictors for procedural success and complex extraction.

While median dwell times did not differ significantly between the 3830 model leads and other leads (72 [IQR 49–95] vs. 85 [IQR 54–111] months, p = 0.501), TLE by simple manual traction was significantly more frequent in the 3830 lead model (Chi² p = 0.003, univariable regression p = 0.006). In addition, older age at TLE was significantly associated with increased odds of complex extraction while additionally analyzed parameters did not show a significant association (Table 5).

4. Discussion

In our study summarizing our institutional experience with TLE in children, we demonstrate that TLE can be successfully and safely performed in the pediatric population. There were no major complications or mortalities despite the high rate of complex extractions and the frequent use of powered rotational dissection sheaths in our cohort. Clinical success was achieved in 97% of patients. This compares well with the results from previous studies with clinical success rates of 80%–98% in children or mixed cohorts and approximately 98% in large registries of adult patients reported in recent years (8, 11, 1417).

The numbers of children and young adults with CIED therapy is increasing (26, 27). Children in particular have a limited durability of transvenous PM and ICD leads (3, 4, 7, 28). One reason for that is somatic growth related lead distortion and subsequent malfunction, which is frequently observed in children even though lead reserve loops, anticipating growth, are generally placed during CIED implantations (Figure 2) (7, 29). However, children with CIED therapy are dependent on functional leads for decades of life expectancy. Consequently, life time lead management strategies are mandatory and conservation of access vessel patency is paramount in particular. In fact, 8 of our patients (29%) had partial or complete thrombotic obstruction of the subclavian and/or innominate vein prior to TLE. While some authors have described uncomplicated short-term outcomes of abandonment of malfunctioning leads in young patients, long-term prospective studies are missing (25). Abandoning superfluent leads, in particular in children with smaller vessel dimensions, might impede later vascular access due to thrombotic venous occlusion (15, 30). Moreover, lead abandonment may result in interactions with other leads, cause regurgitation of the tricuspid valve/subpulmonary atrioventricular valve and increase the risk of infection (31). Also, abandoned leads may pose a contraindication for magnetic resonance imaging, although these concerns might require reconsideration in the light of more recent clinical data (32, 33).

In apprehension of life time lead management, there has been an increasing interest in TLE in children. However, in contrast to adult patients, the overall data on TLE in children is restricted to a small number of single institutional studies with limited cohort sizes or mixed pediatric/adult populations (1420, 22). TLE procedures in children are often complex and require advanced extraction tools in approximately 30%–70% (14, 16, 17). Accordingly, the use of powered sheaths, including radiofrequency, laser and mechanical powered sheaths, has previously been reported with good success in pediatric patients (14, 16, 17, 18, 22). However, in particular data on the use of new generation mechanical rotational sheaths, such as the Evolution® RL (Cook Medical) is limited to very small numbers of pediatric patients (17, 23, 24). One study with a larger, mixed cohort of children and young adults reports the use of mechanical powered sheaths in 48/118 complex TLE, however the study did not specify in which age group mechanical sheaths were used (16). In our cohort, the Evolution® RL (sizes from 9–13F) mechanical rotational dissection sheath was used in 14 pediatric patients without complications. Our youngest patient in whom a mechanical powered sheath was used was 10 years of age (weight 31.8 kg); however, a previous case report describes the use in a child as young as 7 years (weight 17 kg) (24). In 7/8 patients with thrombotic obstruction of access vessels prior to TLE, patency was achieved successfully by TLE; two of these patients required balloon angioplasty while no patient required stent implantation. Our limited data suggests that the Evolution® RL can effectively and safely be used for TLE in children. Thrombotic occlusion of access vessel may be successfully recanalized by the use of powered mechanical sheaths.

Interestingly, in our study we demonstrate differences in extraction complexity and procedural success according to lead types. The 3830 model (Medtronic SelectSecure™) showed a decreased odd for complex TLE. After a median dwell time of 72 [IQR 49–95] months, 6/10 leads could be extracted by manual traction and all leads could be extracted with complete procedural success. The 3830 model is a polyurethane insulated lead that differs from other lead models by its low diameter (4.1F) and lumenless isodiametric design with a central cable, probably offering more tensile strength. Due to its small lead body dimensions it is frequently used in the pediatric population. A good extractability of the 3830 lead with a lower number of complex extractions and a higher procedural success rate has also previously been observed in cohorts of children and young adults as well as pediatric and adult CHD patients (16, 34). The 4471–4474 lead models (Boston Scientific Fineline™ II) are also small diameter leads (5–6F) often used in children with a coaxial design. In our cohort, we observed an increased incidence of partial procedural failure with the 447x lead family. In 9/21 leads, small fragments <4 cm length remained that could not be extracted. Typically, the lead fractured at the tip (Figure 2) Our findings are in line with a recent study that also reported a high rate of lead fracture during TLE in the Fineline™ family (35). Moreover, extraction of these lead model types might more frequently require complex extraction techniques (12). This was also observed in our cohort. A femoral approach with additional snaring of lead fragments was significantly more frequently required with Fineline™ leads.

4.1. Limitations

Limitations of this study are inherent to its retrospective design and restriction to a single institution. Importantly, the number of patients in our study is small, limiting the options of statistical analyses such as multivariable analyses and therefore the possibilities of confident conclusions. Most extractions in children were performed by two extractors (BP, PK) and preferences for approaches may have varied. Other frequently used advanced extractions techniques such as laser powered sheaths were not employed in our study. Due to our good institutional experience with mechanical powered sheaths for TLE and with more recent data suggesting higher complications rates and lower success rates with laser powered sheaths, this technique was not adopted in our institution (36). A considerable number of patients were referred to our center by other institutions, thus previous implantation techniques and choice of lead types varied, which might have influenced the procedural success of lead extractions.

5. Conclusions

TLE can be safely and effectively performed in children. A large number of procedures require complex extraction techniques which emphasizes the requirement of adequate experience and resources for TLE in children. The use of advanced extraction tools such as the Evolution® RL mechanical powered rotational dissection sheath is feasible and safe in the pediatric population. While overall clinical success of TLE is excellent, extraction of particular lead types such as the Boston Scientific Fineline™ lead family might be more difficult and frequently result in only partial procedural success. A considerable number of thrombotic venous obstructions observed in children with transvenous CIED leads suggests that lead abandonment might not be an adequate strategy in light of the requirement of life-long CIED therapy. Larger studies and preferably multi-institutional prospective registries are required to reveal patient-, lead- and technique-related aspects that might have impact on the success and complexity of TLE in children.

Data availability statement

The data that support the findings of this study are available on request from the corresponding author in accordance with applying data protection legislation. The data are not publicly available due to privacy or ethical restrictions.

Ethics statement

The studies involving humans were approved by Ethics Committee Charité—Universitätsmedizin Berlin. The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation was not required from the participants or the participants’ legal guardians/next of kin in accordance with the national legislation and institutional requirements.

Author contributions

RH: Conceptualization, Investigation, Writing – review & editing. BP: Conceptualization, Writing – review & editing. PL: Investigation, Writing – review & editing. JP: Conceptualization, Supervision, Writing – review & editing. FB: Conceptualization, Supervision, Writing – review & editing. VF: Conceptualization, Supervision, Writing – review & editing. CS: Conceptualization, Supervision, Writing – review & editing. PK: Conceptualization, Investigation, Writing – original draft, Writing – review & editing.

Funding

The author(s) declare that no financial support was received for the research, authorship, and/or publication of this article.

Conflict of interest

VF, CS and FB declare relevant financial activities unrelated to the submitted work with several commercial entities. VF has relevant financial activities with Medtronic, Biotronik, Abbott, Boston Scientific, Edwards Lifesciences, Berlin Heart, Novartis Pharma, JOTEC and Zurich Heart in relation to educational grants (including travel support), fees for lectures and speeches, fees for professional consultation and research and study funds. CS has received consulting fees and travel expenses from Medtronic; consulting fees and research support from Biotronik; research support from Abbott; workshop fees, consulting fees, educational grants, and research support from Cook Medical; consulting fees from Spectranetics/Philips; consulting fees from Angiodynamics. FB has received consulting fees and travel expenses from Medtronic; consulting fees, travel expenses and research support from Abbott; consulting fees from Venus Medtech Europe.

The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Vos LM, Kammeraad JAE, Freund MW, Blank AC, Breur J. Long-term outcome of transvenous pacemaker implantation in infants: a retrospective cohort study. Europace. (2017) 19:581–7. doi: 10.1093/europace/euw031

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Konta L, Chubb MH, Bostock J, Rogers J, Rosenthal E. Twenty-seven years experience with transvenous pacemaker implantation in children weighing <10 kg. Circ Arrhythm Electrophysiol. (2016) 9:e003422. doi: 10.1161/CIRCEP.115.003422

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Silvetti MS, Drago F, Di Carlo D, Placidi S, Brancaccio G, Carotti A. Cardiac pacing in paediatric patients with congenital heart defects: transvenous or epicardial? Europace. (2013) 15:1280–6. doi: 10.1093/europace/eut029

PubMed Abstract | CrossRef Full Text | Google Scholar

4. McLeod CJ, Attenhofer Jost CH, Warnes CA, Hodge D 2nd, Hyberger L, Connolly HM, et al. Epicardial versus endocardial permanent pacing in adults with congenital heart disease. J Interv Card Electrophysiol. (2010) 28:235–43. doi: 10.1007/s10840-010-9494-4

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Udink ten Cate F, Breur J, Boramanand N, Crosson J, Friedman A, Brenner J, et al. Endocardial and epicardial steroid lead pacing in the neonatal and paediatric age group. Heart. (2002) 88:392–6. doi: 10.1136/heart.88.4.392

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Eliasson H, Sonesson SE, Salomonsson S, Skog Andreasson A, Wahren-Herlenius M, Gadler F. Pacing therapy in children with isolated complete atrioventricular block: a retrospective study of pacing system survival and pacing-related complications in a national cohort. Europace. (2019) 21:1717–24. doi: 10.1093/europace/euz268

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Fortescue EB, Berul CI, Cecchin F, Walsh EP, Triedman JK, Alexander ME. Patient, procedural, and hardware factors associated with pacemaker lead failures in pediatrics and congenital heart disease. Heart Rhythm. (2004) 1:150–9. doi: 10.1016/j.hrthm.2004.02.020

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Bongiorni MG, Kennergren C, Butter C, Deharo JC, Kutarski A, Rinaldi CA, et al. The European lead extraction ConTRolled (ELECTRa) study: a European heart rhythm association (EHRA) registry of transvenous lead extraction outcomes. Eur Heart J. (2017) 38:2995–3005. doi: 10.1093/eurheartj/ehx080

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Brunner MP, Cronin EM, Duarte VE, Yu C, Tarakji KG, Martin DO, et al. Clinical predictors of adverse patient outcomes in an experience of more than 5000 chronic endovascular pacemaker and defibrillator lead extractions. Heart Rhythm. (2014) 11:799–805. doi: 10.1016/j.hrthm.2014.01.016

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Deshmukh A, Patel N, Noseworthy PA, Patel AA, Patel N, Arora S, et al. Trends in use and adverse outcomes associated with transvenous lead removal in the United States. Circulation. (2015) 132:2363–71. doi: 10.1161/CIRCULATIONAHA.114.013801

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Starck CT, Gonzalez E, Al-Razzo O, Mazzone P, Delnoy PP, Breitenstein A, et al. Results of the patient-related outcomes of mechanical lead extraction techniques (PROMET) study: a multicentre retrospective study on advanced mechanical lead extraction techniques. Europace. (2020) 22:1103–10. doi: 10.1093/europace/euaa103

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Kusumoto FM, Schoenfeld MH, Wilkoff BL, Berul CI, Birgersdotter-Green UM, Carrillo R, et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. (2017) 14:e503–51. doi: 10.1016/j.hrthm.2017.09.001

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Bongiorni MG, Burri H, Deharo JC, Starck C, Kennergren C, Saghy L, et al. 2018 EHRA expert consensus statement on lead extraction: recommendations on definitions, endpoints, research trial design, and data collection requirements for clinical scientific studies and registries: endorsed by APHRS/HRS/LAHRS. Europace. (2018) 20:1217. doi: 10.1093/europace/euy050

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Cecchin F, Atallah J, Walsh EP, Triedman JK, Alexander ME, Berul CI. Lead extraction in pediatric and congenital heart disease patients. Circ Arrhythm Electrophysiol. (2010) 3:437–44. doi: 10.1161/CIRCEP.110.957324

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Zartner PA, Wiebe W, Toussaint-Goetz N, Schneider MB. Lead removal in young patients in view of lifelong pacing. Europace. (2010) 12:714–8. doi: 10.1093/europace/euq059

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Pham TDN, Cecchin F, O'Leary E, Fynn-Thompson F, Triedman JK, Gauvreau K, et al. Lead extraction at a pediatric/congenital heart disease center: the importance of patient age at implant. JACC Clin Electrophysiol. (2022) 8:343–53. doi: 10.1016/j.jacep.2021.11.008

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Kutarski A, Miszczak-Knecht M, Brzezinska M, Birbach M, Lipinski W, Polewczyk AM, et al. Transvenous lead extraction in pediatric patients—is it the same procedure in children as in adults? Circ J. (2023) 87(7):990–9. doi: 10.1253/circj.CJ-22-0542

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Dilber E, Karagoz T, Celiker A. Lead extraction in children and young adults using different techniques. Med Princ Pract. (2009) 18:356–9. doi: 10.1159/000226287

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Atallah J, Erickson CC, Cecchin F, Dubin AM, Law IH, Cohen MI, et al. Multi-institutional study of implantable defibrillator lead performance in children and young adults: results of the pediatric lead extractability and survival evaluation (PLEASE) study. Circulation. (2013) 127:2393–402. doi: 10.1161/CIRCULATIONAHA.112.001120

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Friedman RA, Van Zandt H, Collins E, LeGras M, Perry J. Lead extraction in young patients with and without congenital heart disease using the subclavian approach. Pacing Clin Electrophysiol. (1996) 19:778–83. doi: 10.1111/j.1540-8159.1996.tb03359.x

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Writing Committee M, Shah MJ, Silka MJ, Silva JNA, Balaji S, Beach CM, et al. 2021 PACES expert consensus statement on the indications and management of cardiovascular implantable electronic devices in pediatric patients: developed in collaboration with and endorsed by the heart rhythm society (HRS), the American college of cardiology (ACC), the American heart association (AHA), and the association for European paediatric and congenital cardiology (AEPC) endorsed by the Asia pacific heart rhythm society (APHRS), the Indian heart rhythm society (IHRS), and the latin American heart rhythm society (LAHRS). JACC Clin Electrophysiol. (2021) 7:1437–72. doi: 10.1016/j.jacep.2021.07.009

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Moak JP, Freedenberg V, Ramwell C, Skeete A. Effectiveness of excimer laser-assisted pacing and ICD lead extraction in children and young adults. Pacing Clin Electrophysiol. (2006) 29:461–6. doi: 10.1111/j.1540-8159.2006.00376.x

PubMed Abstract | CrossRef Full Text | Google Scholar

23. Migliore F, Cipriani A, Ferretto S, Hadas D, Iliceto S, Leoni L. Effective and safe lead extraction using the bidirectional rotational evolution[(R)] sheath in a child with congenital heart disease. J Arrhythm. (2018) 34:93–5. doi: 10.1002/joa3.12008

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Ergul Y, Ozyilmaz I, Guzeltas A, Tola HT. Successful transvenous lead extraction using the evolution system in a 17-kg child. Turk Kardiyol Dern Ars. (2015) 43:82–5. doi: 10.5543/tkda.2015.65188

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Silvetti MS, Drago F. Outcome of young patients with abandoned, nonfunctional endocardial leads. Pacing Clin Electrophysiol. (2008) 31:473–9. doi: 10.1111/j.1540-8159.2008.01017.x

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Czosek RJ, Meganathan K, Anderson JB, Knilans TK, Marino BS, Heaton PC. Cardiac rhythm devices in the pediatric population: utilization and complications. Heart Rhythm. (2012) 9:199–208. doi: 10.1016/j.hrthm.2011.09.004

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Berul CI, Van Hare GF, Kertesz NJ, Dubin AM, Cecchin F, Collins KK, et al. Results of a multicenter retrospective implantable cardioverter-defibrillator registry of pediatric and congenital heart disease patients. J Am Coll Cardiol. (2008) 51:1685–91. doi: 10.1016/j.jacc.2008.01.033

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Wilhelm BJ, Thone M, El-Scheich T, Livert D, Angelico R, Osswald B. Complications and risk assessment of 25 years in pediatric pacing. Ann Thorac Surg. (2015) 100:147–53. doi: 10.1016/j.athoracsur.2014.12.098

PubMed Abstract | CrossRef Full Text | Google Scholar

29. McLeod KA. Cardiac pacing in infants and children. Heart. (2010) 96:1502–8. doi: 10.1136/hrt.2009.173328

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Figa FH, McCrindle BW, Bigras JL, Hamilton RM, Gow RM. Risk factors for venous obstruction in children with transvenous pacing leads. Pacing Clin Electrophysiol. (1997) 20:1902–9. doi: 10.1111/j.1540-8159.1997.tb03594.x

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Wazni O, Wilkoff BL. Considerations for cardiac device lead extraction. Nat Rev Cardiol. (2016) 13:221–9. doi: 10.1038/nrcardio.2015.207

PubMed Abstract | CrossRef Full Text | Google Scholar

32. Indik JH, Gimbel JR, Abe H, Alkmim-Teixeira R, Birgersdotter-Green U, Clarke GD, et al. 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm. (2017) 14:e97–53. doi: 10.1016/j.hrthm.2017.04.025

PubMed Abstract | CrossRef Full Text | Google Scholar

33. Schaller RD, Brunker T, Riley MP, Marchlinski FE, Nazarian S, Litt H. Magnetic resonance imaging in patients with cardiac implantable electronic devices with abandoned leads. JAMA Cardiol. (2021) 6:549–56. doi: 10.1001/jamacardio.2020.7572

PubMed Abstract | CrossRef Full Text | Google Scholar

34. Shepherd E, Stuart G, Martin R, Walsh MA. Extraction of selectsecure leads compared to conventional pacing leads in patients with congenital heart disease and congenital atrioventricular block. Heart Rhythm. (2015) 12:1227–32. doi: 10.1016/j.hrthm.2015.03.004

PubMed Abstract | CrossRef Full Text | Google Scholar

35. Morita J, Yamaji K, Nagashima M, Kondo Y, Sadohara Y, Hirokami J, et al. Predictors of lead break during transvenous lead extraction. J Arrhythm. (2021) 37:645–52. doi: 10.1002/joa3.12524

PubMed Abstract | CrossRef Full Text | Google Scholar

36. Lee SY, Allen IE, Diaz C, Guo X, Pellegrini C, Beygui R, et al. Efficacy and mortality of rotating sheaths versus laser sheaths for transvenous lead extraction: a meta-analysis. J Interv Card Electrophysiol. (2023) 66:1067–75. doi: 10.1007/s10840-021-01076-x

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: transvenous lead extraction, rotational extraction sheaths, congenital heart disease, cardiac implantable electronic device, children

Citation: Heck R, Peters B, Lanmüller P, Photiadis J, Berger F, Falk V, Starck C and Kramer P (2023) Transvenous lead extraction in children with bidirectional rotational dissection sheaths. Front. Cardiovasc. Med. 10:1256752. doi: 10.3389/fcvm.2023.1256752

Received: 12 July 2023; Accepted: 21 August 2023;
Published: 8 September 2023.

Edited by:

Matthias Sigler, University of Göttingen, Germany

Reviewed by:

Wilson W. Lam, Baylor College of Medicine, United States
Mark Gallagher, St George's University Hospitals NHS Foundation Trust, United Kingdom

© 2023 Heck, Peters, Lanmüller, Photiadis, Berger, Falk, Starck and Kramer. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Peter Kramer cGV0ZXIua3JhbWVyQGRoemMtY2hhcml0ZS5kZQ==

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.