Skip to main content

SYSTEMATIC REVIEW article

Front. Cardiovasc. Med., 13 September 2023
Sec. Structural Interventional Cardiology

Comparison of middle-term valve durability between transcatheter aortic valve implantation and surgical aortic valve replacement: an updated systematic review and meta-analysis of RCTs

\r\nTsahi T. Lerman,,
Tsahi T. Lerman1,2,3*Amos Levi,Amos Levi2,3Troels Hjsgaard JrgensenTroels Højsgaard Jørgensen4Lars SndergaardLars Søndergaard4Yeela Talmor-Barkan,Yeela Talmor-Barkan2,3Ran Kornowski,\r\nRan Kornowski2,3
  • 1Department of Internal Medicine F-Recanati, Rabin Medical Center, Beilinson Hospital, Petah Tikva, Israel
  • 2Department of Cardiology, Rabin Medical Center, Petah Tikva, Israel
  • 3The Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel
  • 4Department of Cardiology, The Heart Center, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark

Background: This study aims to compare valve durability between transcatheter aortic valve implantation (TAVI) and surgical aortic valve replacement (SAVR).

Methods: We conducted a systematic review and meta-analysis using data from randomized controlled trials (RCTs). The primary outcome was structural valve deterioration (SVD). Secondary outcomes were bioprosthetic valve failure, reintervention, effective orifice area (EOA), mean pressure gradient, and moderate–severe aortic regurgitation (AR, transvalvular and/or paravalvular).

Results: Twenty-five publications from seven RCTs consisting of 7,970 patients were included in the analysis with follow-up ranges of 2–8 years. No significant difference was found between the two groups with regard to SVD [odds ratio (OR) 0.72; 95% CI: 0.25–2.12]. The TAVI group was reported to exhibit a statistically significant higher risk of reintervention (OR 2.03; 95% CI: 1.34–3.05) and a moderate–severe AR (OR 6.54; 95% CI: 3.92–10.91) compared with the SAVR group. A trend toward lower mean pressure gradient in the TAVI group [(mean difference (MD) −1.61; 95% CI: −3.5 to 0.28)] and significant higher EOA (MD 0.20; 95% CI: 0.08–0.31) was noted.

Conclusion: The present data indicate that TAVI provides a comparable risk of SVD with favorable hemodynamic profile compared with SAVR. However, the higher risk of significant AR and reintervention was demonstrated.

Systematic Review Registration: PROSPERO (CRD42022363060).

Introduction

Since its clinical introduction in 2002 by Dr. Alain Cribier, transcatheter aortic valve implantation (TAVI) has become a procedure of choice for treating severe symptomatic aortic stenosis (AS) in high-risk and elderly patients (13). Continuous research and development of the valve systems improved the outcomes of the patients over the years and reduced complication rates (48). As a result, using TAVI to treat intermediate- and lower-risk patients with longer life expectancy is continuously expanding (4, 7, 914). The increasing lifespan of patients who underwent TAVI, combined with the desire to minimize recurrence of symptoms and the need for reintervention, had placed valve durability at the focus of attention (1517). The data from randomized controlled trials (RCTs) examining the comparison of valve durability between TAVI and surgical aortic valve replacement (SAVR) were limited, and only one network meta-analysis had examined this issue (18). This meta-analysis focused on the comparison between the two most common TAVI valve systems, such as balloon-expandable valve and self-expanding valve, and compared each of these with SAVR without comparing TAVI with SAVR (18). In addition, since its publication, important new data have been published that were not included in the previous analysis (11, 1922). Therefore, we aimed to conduct an updated meta-analysis comparing the durability between TAVI and SAVR using all data currently available from RCTs.

Methods

The study protocol was written by TL and AL. We conducted a comprehensive search to identify studies in PubMed, Embase, and Cochrane CENTRAL, until December 2022, using a combination of keywords and MeSH terms for transcatheter aortic valve replacement, transcatheter aortic valve implantation, and surgical aortic valve implantation. References of all included trials and reviews identified were scanned for additional studies. All titles and abstracts were screened, and those thought to possibly meet the inclusion criteria were screened for eligibility using full text. The primary outcome was structural valve deterioration (SVD) that is defined as an intrinsic and permanent structural valvular change, causing AS or transvalvular aortic regurgitation (AR) (23, 24). Variability in definitions and criteria were found, and the data used in the pooled analysis were based on the criteria used in the individual studies included. Secondary outcomes were bioprosthetic valve failure (BVF; valve-related death, reintervention, or severe hemodynamic SVD), reintervention, aortic valve effective orifice area (EOA), mean pressure gradient (MG), and moderate–severe AR (transvalvular and/or paravalvular) (23, 24). We included studies with a follow-up period longer than 1 year. The longest follow-up data available were used in the pooled analysis.

Two reviewers (TL and AL) independently extracted the data and resolved conflicts by discussion. In studies where the data of interest were not available, we contacted the authors of the article with a request to share the information. In cases of data presented as figures without numerical data, extraction was done using Plot Digitizer software (PlotDigitizer, 3.1.5, 2023, https://plotdigitizer.com) as recommended by the Cochrane Collaboration.

Two authors (TL and AL) assessed the risk of bias. Cochrane's handbook tool was used to assess the studies (25). A funnel plot was used to assess publication bias. The systematic review and meta-analysis were conducted in compliance with the Cochrane Collaboration and Preferred Reporting Items for Systematic Reviews and Meta-Analyses statement (25). The meta-analysis was conducted using ReviewManager (RevMan) (Version 5.4. Copenhagen: The Nordic Cochrane Centre, The Cochrane Collaboration, 2020).

Heterogeneity between the included trials was assessed using the chi-squared test, and the I2 test was used to assess inconsistency. We used a fixed-effect model with Mantel–Haenszel method for pooling trial results throughout the review unless statistically significant heterogeneity was found (P < 0.10 or I2 > 50%), in which case we chose a random-effects model and used the inverse variance method. The values reported were two-tailed, and the hypothesis-testing results were considered significant at P < 0.05.

A sensitivity analysis was carried out by examining the effect of the exclusion of each study on the pooled results (“leave-one-out” analysis). The study was pre-registered in PROSPERO with ID number CRD42022363060.

Results

A flow chart representing the study selection process is illustrated in Figure 1. Our initial search yielded 921 citations, 30 of which were judged to be potentially eligible and underwent a full-text review. Twenty-five publications from seven RCTs were found to be eligible for inclusion after the full-text review (47, 9, 11, 1922, 2640). Overall, our meta-analysis included data on 7,970 patients: 4,007 treated by TAVI and 3,963 treated by SAVR. Their mean age was 79%, and 59% were males. The longest follow-up available per study ranged from 2 to 8 years. The characteristics of the studies and patients included in this meta-analysis are presented in Table 1.

FIGURE 1
www.frontiersin.org

Figure 1. Study selection process for inclusion in the meta-analysis (PRISMA flow diagram).

TABLE 1
www.frontiersin.org

Table 1. Characteristics of studies and patients included in the meta-analysis.

Structural valve deterioration

Five studies reported data with regard to SVD. Their follow-up ranged from 1 to 8 years. A total of 84 patients out of 2,810 (3.0%) were enrolled in the TAVI group, and 91 out of 2,443 (3.7%) in the SAVR group. Overall, no significant difference were found between the two groups (OR 0.72; 95% CI: 0.25–2.12; P = 0.55, I2 = 87%; Figure 2). However, in a subgroup analysis, a statistically significant lower risk was observed among patients who underwent self-expanding TAVI compared with SAVR (OR 0.44; 95% CI: 0.30–0.67; P < 0.01, I2 = 0%).

FIGURE 2
www.frontiersin.org

Figure 2. Forest plots for structural valve deterioration.

Secondary outcomes

Only two studies reported data with regard to BVF. No statically significant difference was noted between the two groups (OR 1.50; 95% CI: 0.46–4.90); P = 0.5, I2 = 77%).

Seven studies reported data with regard to reintervention. A statistically significant higher risk was noted for reintervention in the TAVI group compared with SAVR (OR 2.03; 95% CI: 1.34–3.05); P < 0.01, I2 = 44%). BVF and reintervention results are presented in Figure 3.

FIGURE 3
www.frontiersin.org

Figure 3. Forest plots for bioprosthetic valve failure (A) and reintervention (B).

All RCTs provided data with regard to valve performances with follow-up ranges of 1–8 years. No significant difference was seen in mean pressure gradients between the two groups [(mean difference (MD) −1.61; 95% CI: −3.5 to 0.28); P = 0.1, I2 = 95%)]. Importantly, a significant heterogeneity was observed between the self-expanding subgroup and the balloon-expandable subgroup (I2 = 97%, P < 0.01). In the self-expanding group, the mean gradient was significantly lower than in the SAVR group (MD −3.25; 95% CI: −4.23 to −2.28; P < 0.01, I2 = 38%). In the balloon-expandable group, a statistically non-significant trend toward a higher mean gradient was reported in the TAVI group compared with the SAVR group (MD 1; 95% CI: −0.06 to 2.05; P = 0.06, I2 = 71%).

EOA was higher in the TAVI group than in the SAVR group (MD 0.20; 95% CI: 0.08–0.31); P < 0.01, I2 = 89%). This difference was mainly driven by the self-expanding subgroup (MD 0.3; 95% CI: 0.21–0.4; P < 0.01, I2 = 38%).

The rate of moderate–severe AR was higher in the TAVI group than that in the SAVR group, 112 patients out of 1,796 (6.2%) vs. 16 patients out of 1,607 (1%), respectively (OR 6.54; 95% CI: 3.92–10.91; P < 0.01, I2 = 0%), which was mainly driven by paravalvular leak (PVL). Valve performances are presented in Figure 4. MG and EOA during a 5-year follow-up period are presented in Figure 5.

FIGURE 4
www.frontiersin.org

Figure 4. Forest plots for mean gradients (A), effective orifice area (B), moderate–severe aortic regurgitation (C), and moderate–severe paravalvular leak (D).

FIGURE 5
www.frontiersin.org

Figure 5. Mean pressure gradient and effective orifice area during a 5-year follow-up.

Sensitivity analysis

With the exclusion of the PARTNER 2 study, the SVD rate was reported to be significantly lower in the TAVI group (OR 0.44; 95% CI: 0.29–0.66; P < 0.01, I2 = 0%). The exclusion of PARTNER 1 study that used different SVD criteria did not change the statistical significance of the analysis.

After the exclusion of the PARTNER 3 study, the mean gradient was found to be significantly lower in the TAVI group (MD −2.24; 95% CI: −4.18 to −0.31; P = 0.02, I2 = 92%).

Exclusion of all other studies did not affect the statistical significance of the analysis.

Discussion

We conducted the systematic review and meta-analysis including all RCTs that compared the valve durability and hemodynamic performances between TAVI and SAVR for a follow-up period longer than 1 year. We decided to use SVD as the primary outcome in our meta-analysis. SVD is one of the proposed categories of standardized definition of valve dysfunction, i.e., valvular changes causing AS or transvalvular AR. This contrasts with non-structural valve dysfunction (PVL, patient prosthesis mismatch), endocarditis, and valve thrombosis which are not caused by intrinsic valvular dysfunction (23, 24). Since the publication of the previous network meta-analysis examining valve durability, two important studies had been published comparing the SVD rate between TAVI and SAVR. The first study was an 8-year follow-up study of the Nordic Aortic Valve Intervention Trial (NOTION) which showed a lower risk of SVD in patients treated with self-expanding TAVI (21). The second study was a post hoc analysis of 5-year pooled data from two RCTs: US High-Risk Pivotal and SURTAVI trials (20). This study enrolled 4,762 patients and found a lower SVD rate among the self-expanding TAVI group compared with SAVR group. Importantly, this large-scale study found an association between SVD and all-cause mortality as well as cardiovascular mortality and therefore emphasized the clinical importance of this outcome as a surrogate marker for a worse prognosis of the patients. Interestingly, the advantage of TAVI was most pronounced in smaller annuli, and therefore further study of patient-level meta-analysis examining this aspect was suggested. In addition, it was well established that SVD rate was affected by the age of the patient, patient–prosthesis mismatch, and the use of lipid-lowering agent further emphasizing the need for analyses using patient-level data (4143). It was worth mentioning that the studies included in the analysis used different criteria for SVD. However, these criteria generally overlapped, and therefore a pooled analysis was feasible. One exception was the PARTNER 1 study. However, excluding this study in a “leave-one-out” scrutiny did not significantly change the results. Another durability outcome we examined was BVF which was not reported in the previously published meta-analysis (18). This outcome was only reported in three studies: the NOTION study, PARTNER 2 study, and PARTNER 3 study (11, 21, 39). However, the latter did not report raw data with regard to the event rate and was therefore not included in our analysis. The two studies included used different, yet, overlapping BVF criteria, the EAPCI/ESC/EACT and the VARC-3 criteria, respectively. According to the pooled analysis, the rate of BVF was also comparable between TAVI and SAVR. However, due to a limited number of studies examining this outcome, it should be cautiously interpreted over a longer period of follow-up. In addition, it should be noted that in the PARTNER 2 study a higher rate of BVF was demonstrated in the TAVI group using the Sapien XT percutaneous valve that was not in clinical use anymore. Therefore, additional studies examining this outcome using contemporary devices are necessary (39).

In line with the previous network meta-analysis published by Ueyama et al., our meta-analysis found a favorable forward-flow hemodynamic profile of TAVI compared with SAVR with stable valve performances during the long-term follow-up (18). The hemodynamic advantage of TAVI was most prominently demonstrated in the self-expanding subgroup perhaps owing to their supra-annular functional design which facilitates larger effective valve diameter. This finding was consistent with a recently published meta-analysis from our group, comparing the latest-generation self-expanding valve with the balloon-expandable valve (44). Despite the favorable forward hemodynamic profile of TAVI, a significant inferiority was demonstrated with regard to the rate of moderate–severe AR and the rate of reintervention. Our pooled analysis was related to a total number of moderate–severe AR (i.e., transvalvular and PVL), and the higher rate of significant AR was mainly driven by PVL without a significant difference between the valve-type subgroups. This explains the fact that despite the higher rate of significant AR the rate of SVD was comparable between the two study arms since PVL is considered as non-SVD valve dysfunction. Since SVD and PVL are the two most common indications for reintervention, one can assume that the higher rate of PVL encountered in the TAVI group led to a higher reintervention rate (45). It has to be acknowledged that the characteristics of the surgical population were not always advantageous for optimized SAVR durability. For instance, in NOTION, 34% of patients received a valve prosthesis with lower durability performance (24% Saint Jude Trifecta, 10% Sorin Mitroflow). Despite the well-known negative effect of patient–prosthesis mismatch on valve durability, SAVR patients in PARTNER 3 received a labeled prosthesis of ≤21 mm in 20% of patients (9, 40, 46, 47). The advantage of SAVR with regard to reintervention should be emphasized in light of these considerations. The higher reintervention rate in the TAVI group was in line with a recently published meta-analysis (48).

It has been previously established that balloon-expandable and self-expanding valves have a different efficacy and safety profile. This is reflected by differences in pacemaker implantation rate, periprocedural bleeding rate, and hemodynamic profile (44) Combined with the results of the current study examining the durability of TAVI valves, we believe that a set of considerations are obtained that should be incorporated into the heart team decision-making (44).

Our meta-analysis has several strengths. First, our analysis has the longest follow-up period available today of up to 8 years. In addition, this is the most comprehensive analysis conducted including five follow-up studies that were not included in the previous meta-analysis published (11, 1822). In addition, the pooled analysis included only data from RCTs without data from observational studies, therefore reflecting high-quality data and reliable analysis. Yet, several limitations should be mentioned. First, the patients included in the RCTs were at a wide range of surgical risk, and it is possible that the final study population in the meta-analysis does not reflect the real-world population. However, since we mainly tested valvular outcomes and not “hard outcomes,” the effect is expected to be minimal; yet, long-term real-life registries data are needed to examine valve durability and confirm our results. Second, follow-up echocardiography was not available for a significant number of patients. The impact of selective echocardiography on results is unpredictable. Sicker patients with poorer valve performance could have been underrepresented if they were unable to complete the follow-up. On the other hand, patients feeling well, with a lower rate of valve degeneration, may also feel disinclined to continue the follow-up, leading to their underrepresentation. Third, older-generation valves were used in some studies and may not reflect modern-era TAVI. Fourth, we did not have access to individual patient data, and the analysis is based on aggregate data and therefore should be interpreted with caution. Fifth, in some of the analyses, a significant heterogeneity was demonstrated between the studies.

Lastly, although only RCTs were included in the analysis, few biases should be considered. In RCTs comparing TAVI with SAVR, systematic imbalances in the proportion of deviation from randomly assigned treatment (DAT), high loss to follow-up (8.9% at 5 years), and receipt of additional procedures and additional myocardial revascularization that can pose a serious threat to internal validity due to a high risk of performance and attrition biases were noted (49).

Conclusions

In conclusion, according to our analysis, TAVI provided a comparable risk of SVD with favorable hemodynamic profile compared with SAVR. However, a higher rate of significant AR and reintervention was demonstrated. These results provide an important insight regarding valve durability that should be considered when tailoring treatment to the individual patient. Additional real-life and expanded data over a longer period of time are needed to confirm our results.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding author.

Author contributions

TL, AL, YT, and RK contributed to the conception and design of the study. TL performed the statistical analysis. TJ, LS, TL, AL, YT, and RK contributed to the preparation of the manuscript. All authors contributed to the article and approved the submitted version.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcvm.2023.1242608/full#supplementary-material

References

1. Leon MB, Smith CR, Mack M, Miller DC, Moses JW, Svensson LG, et al. Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. N Engl J Med. (2010) 363:1597–607. doi: 10.1056/NEJMoa1008232

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Cribier A, Eltchaninoff H, Bash A, Borenstein N, Tron C, Bauer F, et al. Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis. Circ. (2002) 106:3006–8. doi: 10.1161/01.CIR.0000047200.36165.B8

CrossRef Full Text | Google Scholar

3. Kalogeropoulos AS, Redwood SR, Allen CJ, Hurrell H, Chehab O, Rajani R, et al. A 20-year journey in transcatheter aortic valve implantation: evolution to current eminence. Front Cardiovasc Med. (2022) 9. doi: 10.3389/fcvm.2022.971762

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Makkar RR, Thourani VH, Mack MJ, Kodali SK, Kapadia S, Webb JG, et al. Five-year outcomes of transcatheter or surgical aortic-valve replacement. N Engl J Med. (2020) 382:799–809. doi: 10.1056/NEJMoa1910555

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Smith CR, Leon MB, Mack MJ, Miller DC, Moses JW, Svensson LG, et al. Transcatheter versus surgical aortic-valve replacement in high-risk patients. N Engl J Med. (2011) 364:2187–98. doi: 10.1056/NEJMoa1103510

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Adams DH, Popma JJ, Reardon MJ, Yakubov SJ, Coselli JS, Deeb GM, et al. Transcatheter aortic-valve replacement with a self-expanding prosthesis. N Engl J Med. (2014) 370:1790–8. doi: 10.1056/NEJMoa1400590

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Reardon MJ, van Mieghem NM, Popma JJ, Kleiman NS, Søndergaard L, Mumtaz M, et al. Surgical or transcatheter aortic-valve replacement in intermediate-risk patients. N Engl J Med. (2017) 376:1321–31. doi: 10.1056/NEJMoa1700456

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Lerman TT, Levi A, Talmor-Barkan Y, Kornowski R. Early and mid-term outcomes of transcatheter aortic valve implantation versus surgical aortic valve replacement: updated systematic review and meta-analysis. J Cardiovasc Dev Dis. (2023) 10:157. doi: 10.3390/jcdd10040157

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Mack MJ, Leon MB, Thourani VH, Makkar R, Kodali SK, Russo M, et al. Transcatheter aortic-valve replacement with a balloon-expandable valve in low-risk patients. N Engl J Med. (2019) 380:1695–705. doi: 10.1056/NEJMoa1814052

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Siontis GCM, Overtchouk P, Cahill TJ, Modine T, Prendergast B, Praz F, et al. Transcatheter aortic valve implantation vs. surgical aortic valve replacement for treatment of symptomatic severe aortic stenosis: an updated meta-analysis. Eur Heart J. (2019) 40:3143–53. doi: 10.1093/eurheartj/ehz275

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Leon MB, Mack MJ, Hahn RT, Thourani VH, Makkar R, Kodali SK, et al. Outcomes 2 years after transcatheter aortic valve replacement in patients at low surgical risk. J Am Coll Cardiol. (2021) 77:1149–61. doi: 10.1016/j.jacc.2020.12.052

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Popma JJ, Deeb GM, Yakubov SJ, Mumtaz M, Gada H, O’Hair D, et al. Transcatheter aortic-valve replacement with a self-expanding valve in low-risk patients. N Engl J Med. (2019) 380:1706–15. doi: 10.1056/NEJMoa1816885

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Witberg G, Lador A, Yahav D, Kornowski R. Transcatheter versus surgical aortic valve replacement in patients at low surgical risk: a meta-analysis of randomized trials and propensity score matched observational studies. Catheter Cardiovasc Interv. (2018) 92:408–16. doi: 10.1002/ccd.27518

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Voigtländer L, Seiffert M. Expanding TAVI to low and intermediate risk patients. Front Cardiovasc Med. (2018) 5. doi: 10.3389/fcvm.2018.00092

CrossRef Full Text | Google Scholar

15. Spitzer E, Mylotte D, Lauten A, O’Sullivan CJ. Editorial: TAVI and the challenges ahead. Front Cardiovasc Med. (2020) 7. doi: 10.3389/fcvm.2020.00149

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Montarello NJ, Willemen Y, Tirado-Conte G, Travieso A, Bieliauskas G, Sondergaard L, et al. Transcatheter aortic valve durability: a contemporary clinical review. Front Cardiovasc Med. (2023) 10. doi: 10.3389/fcvm.2023.1195397

PubMed Abstract | CrossRef Full Text | Google Scholar

17. De Backer O, Søndergaard L. Challenges when expanding transcatheter aortic valve implantation to younger patients. Front Cardiovasc Med. (2018) 5. doi: 10.3389/fcvm.2018.00045

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Ueyama H, Kuno T, Takagi H, Kobayashi A, Misumida N, Pinto DS, et al. Meta-analysis comparing valve durability among different transcatheter and surgical aortic valve bioprosthesis. Am J Cardiol. (2021) 158:104–11. doi: 10.1016/j.amjcard.2021.07.046

PubMed Abstract | CrossRef Full Text | Google Scholar

19. van Mieghem NM, Deeb GM, Søndergaard L, Grube E, Windecker S, Gada H, et al. Self-expanding transcatheter vs surgical aortic valve replacement in intermediate-risk patients: 5-year outcomes of the SURTAVI randomized clinical trial. JAMA Cardiol. (2022) 7:1000–8. doi: 10.1001/jamacardio.2022.2695

PubMed Abstract | CrossRef Full Text | Google Scholar

20. O’Hair D, Yakubov SJ, Grubb KJ, Oh JK, Ito S, Deeb GM, et al. Structural valve deterioration after self-expanding transcatheter or surgical aortic valve implantation in patients at intermediate or high risk. JAMA Cardiol. (2023) 8(2):111–9. doi: 10.1001/jamacardio.2022.4627

CrossRef Full Text | Google Scholar

21. Jørgensen TH, Thyregod HGH, Ihlemann N, Nissen H, Petursson P, Kjeldsen BJ, et al. Eight-year outcomes for patients with aortic valve stenosis at low surgical risk randomized to transcatheter vs. surgical aortic valve replacement. Eur Heart J. (2021) 42:2912–9. doi: 10.1093/eurheartj/ehab375

CrossRef Full Text | Google Scholar

22. Forrest JK, Deeb GM, Yakubov SJ, Rovin JD, Mumtaz M, Gada H, et al. 2-year outcomes after transcatheter versus surgical aortic valve replacement in low-risk patients. J Am Coll Cardiol. (2022) 79:882–96. doi: 10.1016/j.jacc.2021.11.062

PubMed Abstract | CrossRef Full Text | Google Scholar

23. Généreux P, Piazza N, Alu MC, Nazif T, Hahn RT, Pibarot P, et al. Valve academic research consortium 3: updated endpoint definitions for aortic valve clinical research. J Am Coll Cardiol. (2021) 77:2717–46. doi: 10.1016/j.jacc.2021.02.038

CrossRef Full Text | Google Scholar

24. Capodanno D, Petronio AS, Prendergast B, Eltchaninoff H, Vahanian A, Modine T, et al. Standardized definitions of structural deterioration and valve failure in assessing long-term durability of transcatheter and surgical aortic bioprosthetic valves: consensus statement from the European Association of Percutaneous Cardiovascular Interventions (EAPCI) endorsed by the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS). Eur J Cardiothorac Surg. (2017) 52:408–17. doi: 10.1093/ejcts/ezx244

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Higgins JPT, Thomas J, Chandler J, Cumpston M, Li T, Page MJ, et al., editors. Cochrane handbook for systematic reviews of interventions version 6.0 (updated July 2019). Cochrane. Hoboken, NJ, US: Wiley-Blackwell (2019).

26. Pibarot P, Salaun E, Dahou A, Avenatti E, Guzzetti E, Annabi M-S, et al. Echocardiographic results of transcatheter versus surgical aortic valve replacement in low-risk patients: the PARTNER 3 trial. Circulation. (2020) 141:1527–37. doi: 10.1161/CIRCULATIONAHA.119.044574

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Daubert MA, Weissman NJ, Hahn RT, Pibarot P, Parvataneni R, Mack MJ, et al. Long-term valve performance of TAVR and SAVR: a report from the PARTNER I trial. JACC Cardiovasc Imaging. (2017) 10:15–25. doi: 10.1016/j.jcmg.2016.11.004

CrossRef Full Text | Google Scholar

28. Leon MB, Smith CR, Mack MJ, Makkar RR, Svensson LG, Kodali SK, et al. Transcatheter or surgical aortic-valve replacement in intermediate-risk patients. N Engl J Med. (2016) 374:1609–20. doi: 10.1056/NEJMoa1514616

PubMed Abstract | CrossRef Full Text | Google Scholar

29. Gleason TG, Reardon MJ, Popma JJ, Deeb GM, Yakubov SJ, Lee JS, et al. 5-year outcomes of self-expanding transcatheter versus surgical aortic valve replacement in high-risk patients. J Am Coll Cardiol. (2018) 72:2687–96. doi: 10.1016/j.jacc.2018.08.2146

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Reardon MJ, Adams DH, Kleiman NS, Yakubov SJ, Coselli JS, Deeb GM, et al. 2-year outcomes in patients undergoing surgical or self-expanding transcatheter aortic valve replacement. J Am Coll Cardiol. (2015) 66:113–21. doi: 10.1016/j.jacc.2015.05.017

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Thyregod HG, Steinbrüchel DA, Ihlemann N, Nissen H, Kjeldsen BJ, Petursson P, et al. Transcatheter versus surgical aortic valve replacement in patients with severe aortic valve stenosis: 1-year results from the all-comers NOTION randomized clinical trial. J Am Coll Cardiol. (2015) 65:2184–94. doi: 10.1016/j.jacc.2015.03.014

PubMed Abstract | CrossRef Full Text | Google Scholar

32. Søndergaard L, Steinbrüchel DA, Ihlemann N, Nissen H, Kjeldsen BJ, Petursson P, et al. Two-year outcomes in patients with severe aortic valve stenosis randomized to transcatheter versus surgical aortic valve replacement: the all-comers Nordic aortic valve intervention randomized clinical trial. Circ Cardiovasc Interv. (2016) 9. doi: 10.1161/CIRCINTERVENTIONS.115.003665

CrossRef Full Text | Google Scholar

33. Hahn RT, Pibarot P, Stewart WJ, Weissman NJ, Gopalakrishnan D, Keane MG, et al. Comparison of transcatheter and surgical aortic valve replacement in severe aortic stenosis: a longitudinal study of echocardiography parameters in cohort A of the PARTNER trial (placement of aortic transcatheter valves). J Am Coll Cardiol. (2013) 61:2514–21. doi: 10.1016/j.jacc.2013.02.087

PubMed Abstract | CrossRef Full Text | Google Scholar

34. Mack MJ, Leon MB, Smith CR, Miller DC, Moses JW, Tuzcu EM, et al. 5-year outcomes of transcatheter aortic valve replacement or surgical aortic valve replacement for high surgical risk patients with aortic stenosis (PARTNER 1): a randomised controlled trial. Lancet. (2015) 385:2477–84. doi: 10.1016/S0140-6736(15)60308-7

PubMed Abstract | CrossRef Full Text | Google Scholar

35. Deeb GM, Reardon MJ, Chetcuti S, Patel HJ, Grossman PM, Yakubov SJ, et al. 3-year outcomes in high-risk patients who underwent surgical or transcatheter aortic valve replacement. J Am Coll Cardiol. (2016) 67:2565–74. doi: 10.1016/j.jacc.2016.03.506

PubMed Abstract | CrossRef Full Text | Google Scholar

36. Kodali SK, Williams MR, Smith CR, Svensson LG, Webb JG, Makkar RR, et al. Two-year outcomes after transcatheter or surgical aortic-valve replacement. N Engl J Med. (2012) 366:1686–95. doi: 10.1056/NEJMoa1200384

PubMed Abstract | CrossRef Full Text | Google Scholar

37. Douglas PS, Leon MB, Mack MJ, Svensson LG, Webb JG, Hahn RT, et al. Longitudinal hemodynamics of transcatheter and surgical aortic valves in the PARTNER trial. JAMA Cardiol. (2017) 2:1197–206. doi: 10.1001/jamacardio.2017.3306

PubMed Abstract | CrossRef Full Text | Google Scholar

38. Thyregod HGH, Ihlemann N, Jorgensen TH, Nissen H, Kjeldsen BJ, Petursson P, et al. Five-year clinical and echocardiographic outcomes from the NOTION randomized clinical trial in patients at lower surgical risk. Circulation. (2019) 139:2714–23. doi: 10.1161/CIRCULATIONAHA.118.036606

PubMed Abstract | CrossRef Full Text | Google Scholar

39. Pibarot P, Ternacle J, Jaber WA, Salaun E, Dahou A, Asch FM, et al. Structural deterioration of transcatheter versus surgical aortic valve bioprostheses in the PARTNER-2 trial. J Am Coll Cardiol. (2020) 76:1830–43. doi: 10.1016/j.jacc.2020.08.049

PubMed Abstract | CrossRef Full Text | Google Scholar

40. Søndergaard L, Ihlemann N, Capodanno D, Jørgensen TH, Nissen H, Kjeldsen BJ, et al. Durability of transcatheter and surgical bioprosthetic aortic valves in patients at lower surgical risk. J Am Coll Cardiol. (2019) 73:546–53. doi: 10.1016/j.jacc.2018.10.083

CrossRef Full Text | Google Scholar

41. Forcillo J, el Hamamsy I, Stevens L-M, Badrudin D, Pellerin M, Perrault LP, et al. The perimount valve in the aortic position: twenty-year experience with patients under 60 years old. Ann Thorac Surg. (2014) 97:1526–32. doi: 10.1016/j.athoracsur.2014.02.019

PubMed Abstract | CrossRef Full Text | Google Scholar

42. David TE, Armstrong S, Maganti M. Hancock II bioprosthesis for aortic valve replacement: the gold standard of bioprosthetic valves durability? Ann Thorac Surg. (2010) 90:775–81. doi: 10.1016/j.athoracsur.2010.05.034

PubMed Abstract | CrossRef Full Text | Google Scholar

43. Bourguignon T, Bouquiaux-Stablo A-L, Candolfi P, Mirza A, Loardi C, May M-A, et al. Very long-term outcomes of the Carpentier-Edwards perimount valve in aortic position. Ann Thorac Surg. (2015) 99:831–7. doi: 10.1016/j.athoracsur.2014.09.030

PubMed Abstract | CrossRef Full Text | Google Scholar

44. Lerman TT, Levi A, Kornowski R. Meta-analysis of short- and long-term clinical outcomes of the self-expanding Evolut R/pro valve versus the balloon-expandable Sapien 3 valve for transcatheter aortic valve implantation. Int J Cardiol. (2023) 371:100–8. doi: 10.1016/j.ijcard.2022.09.035

PubMed Abstract | CrossRef Full Text | Google Scholar

45. Fukuhara S, Nguyen CTN, Kim KM, Yang B, Ailawadi G, Patel HJ, et al. Aortic valve reintervention after transcatheter aortic valve replacement. J Thorac Cardiovasc Surg. (2023) 165:1321–32.e4. doi: 10.1016/j.jtcvs.2021.03.130

PubMed Abstract | CrossRef Full Text | Google Scholar

46. Sénage T, Le Tourneau T, Foucher Y, Pattier S, Cueff C, Michel M, et al. Early structural valve deterioration of mitroflow aortic bioprosthesis. Circulation. (2014) 130:2012–20. doi: 10.1161/CIRCULATIONAHA.114.010400

CrossRef Full Text | Google Scholar

47. Important information regarding SVD in trifecta family of valves. Available at: https://www.structuralheart.abbott/fileadmin/pdf/FINAL_Abbott_Letter_US_Trifecta_Abbott_Website_signed.pdf (accessed on August 1st, 2023).

48. Swift SL, Puehler T, Misso K, Lang SH, Forbes C, Kleijnen J, et al. Transcatheter aortic valve implantation versus surgical aortic valve replacement in patients with severe aortic stenosis: a systematic review and meta-analysis. BMJ Open. (2021) 11:e054222. doi: 10.1136/bmjopen-2021-054222

PubMed Abstract | CrossRef Full Text | Google Scholar

49. Barili F, Brophy JM, Ronco D, Myers PO, Uva MS, Almeida RMS, et al. Risk of bias in randomized clinical trials comparing transcatheter and surgical aortic valve replacement. JAMA Netw Open. (2023) 6:e2249321. doi: 10.1001/jamanetworkopen.2022.49321

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: TAVI, SAVR, durability, structural valve deterioration, bioprosthetic valve failure

Citation: Lerman TT, Levi A, Jørgensen TH, Søndergaard L, Talmor-Barkan Y and Kornowski R (2023) Comparison of middle-term valve durability between transcatheter aortic valve implantation and surgical aortic valve replacement: an updated systematic review and meta-analysis of RCTs. Front. Cardiovasc. Med. 10:1242608. doi: 10.3389/fcvm.2023.1242608

Received: 19 June 2023; Accepted: 21 August 2023;
Published: 13 September 2023.

Edited by:

Martin Swaans, St. Antonius Hospital, Netherlands

Reviewed by:

Dirk-Jan Van Ginkel, St. Antonius Hospital, Netherlands
Thierry Caus, University of Picardie Jules Verne, France
Andrea Messori, Regione Toscana, Italy

© 2023 Lerman, Levi, Jørgensen, Søndergaard, Talmor-Barkan and Kornowski. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: ‪Tsahi T. Lerman tsahilerman@gmail.com

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.