3D Printed Bioreactor Enabling the Pulsatile Culture of Native and Angioplastied Large Arteries

  • 1Cardiovascular Section, Department of Biochemical Sciences, Guildford, United Kingdom
  • 2Clinical Medicine Section, Department of Clinical and Experimental Medicine, University of Surrey, Faculty of Health and Medical Sciences, University of Surrey, Guildford, United Kingdom

Routine interventions such as balloon angioplasty, result in vascular activation and remodeling, often requiring re-intervention. 2D in vitro models and small animal experiments have enabled the discovery of important mechanisms involved in this process, however the clinical translation is often underwhelming. There is a critical need for an ex vivo model representative of the human vascular physiology and encompassing the complexity of the vascular wall and the physical forces regulating its function. Vascular bioreactors for ex vivo culture of large vessels are viable alternatives, but their custom-made design and insufficient characterization often hinders the reproducibility of the experiments. The objective of the study was to design and validate a novel 3D printed cost-efficient and versatile perfusion system, capable of sustaining the viability and functionality of large porcine arteries for 7 days and enabling early post-injury evaluations. MultiJet Fusion 3D printing was used to engineer the EasyFlow insert, converting a conventional 50 ml centrifuge tube into a mini bioreactor. Porcine carotid arteries either left untreated or injured with an angioplasty balloon, were cultured under pulsatile flow for up to 7 days. Pressure, heart rate, medium viscosity and shear conditions were adjusted to resemble arterial in vivo hemodynamics. Tissue viability, cell activation and matrix remodeling were analyzed by immunohistochemistry, and vascular function was monitored by duplex ultrasound. Culture conditions in the EasyFlow bioreactor preserved endothelial coverage and smooth muscle organization and extracellular matrix structure in the vessel wall, as compared to static culture. Injured arteries presented hallmarks of early remodeling, such as intimal denudation, smooth muscle cell disarray and media/adventitia activation in flow culture. Duplex ultrasound confirmed continuous pulsatile blood flow conditions, dose-dependent vasodilator response to nitroglycerin in untreated vessels and impaired dilator response in angioplastied vessels. The scope of this work is to validate a low-cost, robust and reproducible system to explore the culture of native and injured large arteries under pulsatile flow. While the study of vascular pathology is beyond the scope of the present paper, our system enables future investigations and provides a platform to test novel therapies and devices ex vivo, in a patient relevant system.

Introduction

Vascular disease is often caused by narrowing or occlusion of blood vessels leading to decreased blood supply to important organs like the heart or brain, or to the extremities. Revascularization procedures are common interventions aimed at re-establishing blood supply to compromised tissues. Typically, a guidewire is passed through an accessible artery to reach the narrowed area and a balloon is deployed to reopen the lumen. While the immediate effect on blood flow is readily achieved, the intervention damages the blood vessel wall which triggers a cascade of acute inflammatory and regenerative responses that result in intimal hyperplasia and vascular remodeling (1). The vascular remodeling commonly leads to re-stenosis and ultimately occlusion of the intervened vessels (2). In order to maintain patency in the long term, many times additional devices like stents or balloons coated with anti-proliferative drugs are used on the intervened vessels. Clearly, a better understanding of the processes that underly the vascular response to injury and remodeling is key to develop effective treatments to improve long-term patency of revascularization procedures (1).

Largely, experiments aimed at investigating these mechanisms are conducted in small animals (mice, rats, rabbits), upon mechanical injury of a large artery or interposition of a stent (3). Besides the obvious ethical concerns, small rodent experiments have limited predictive capacity, resulting in the development of sub-optimal therapeutic strategies, leaving this critical medical need unmet (4). In vitro strategies for the study of vascular remodeling focus on 2D culture of vascular cells (typically smooth muscle cells) or the static culture of rings obtained from blood vessels, the latter importantly incorporates an element of intact extracellular matrix component, which is critical in the remodeling process [for an example (5)]. Interestingly, some groups have demonstrated the successful use of flow bioreactors simulating the in vivo hemodynamic conditions, to culture blood vessels ex vivo and study vascular pathophysiology (69). This pioneering work has demonstrated the importance and feasibility of a functional and physiological ex vivo approach to vascular remodeling; however, these studies present some limitations that reduce their translational power and reproducibility, which we aim to address with the present work.

In this study, we developed an open-source 3D printed and economical bioreactor (EasyFlow) which enables the multiplex culture of blood vessels from large animals in small volumes of medium. Furthermore, we have optimized conditions of flow, shear stress, pressure, pulsatility and viscosity to closely mimic the hemodynamic forces applying to an artery in vivo. We comprehensively characterized the effect of pulsatile flow culture and endovascular injury on the survival, activation and function of vascular resident cells by histology, immunofluorescence and by Doppler ultrasound vasoreactivity measurements.

This work will demonstrate the reproducible application of ex vivo bioreactor culture for large animal blood vessels, enabling the long-term culture of healthy vessels and the study of pathological mechanisms. Importantly, the open-source nature of our design enables effortless transfer of the system to other labs and reproducibility, promoting the reduction of the use of animals in vascular studies.

Materials and Methods

Ethics and Sample Preparation

Carotid arteries were obtained from 4–6 weeks old piglet at The Pirbright Institute, Pirbright, UK. Animal procedures were carried out under the Home Office Animals (Scientific Procedures) Act (1986) and approved by the Animal Welfare and Ethical Review Board (AWERB) of The Pirbright Institute. The animals were housed in accordance with the Code of Practice for the Housing and Care of Animals Bred.

Pigs were euthanized by an overdose of 10 ml pentobarbital (Dolethal 200 mg/ml solution for injection, Vetoquinol UK Ltd). All procedures were conducted by Personal License holders who were trained and competent and under the Project License PPL70/8852.

Upon exsanguination the neck was opened from the chest cavity to the base of the skull to expose the common carotid artery separating into the right and left carotid artery. Arteries were excised using a no-touch technique to minimize stress to the vascular tissue. Fresh tissue was immediately placed in pre cooled transport media Dulbecco's Modified Eagle's Medium (DMEM) + 20% fetal bovine serum (FBS) + 2% penicillin and streptomycin (P/S) + 1% Amphotericin B (Amp). Tissue was washed two times in transport media and placed on ice for transportation.

Following arrival tissue was transferred into laminar hood where further preparation took place. The excess connective tissue was removed in a Petri dish using precision surgical equipment, avoiding any strain to the vessel. Once cleaned, the vessel was placed in new container and washed in transport media (2 × 20 min at 4°C), followed by short term storage in DMEM + 10% FBS + 1% P/S + 1% Amp at 4°C.

For each preparation, a sample of 2–4 mm was collected at the time of preparation, as a control.

Easy Flow Design and Production

The EasyFlow insert was produced by additive manufacturing which enables for complex design choices to be manufactured with high reproducibility. The CAD model of the system was prepared using Autodesk Fusion 360 (Autodesk® Fusion 360™ 2.0.5818). Printable file is available in Supplementary File 1.

In planning the design, the major criteria considered were user experience, ease of handling, adaptability and reusability. MultiJet Fusion 3D printing allowed us to create fine inner structures within a space-efficient design and therefore maintain a compact outer structure. The manufacturing was outsourced to an external provider (ProtoLabs), which produced EasyFlow by Polyamide 12 (HP 3D High Reusability PA 12) Black MultiJet Fusion built in 80 μm layers, requiring no additional finish. Considering the constantly expanding list of materials available for manufacturing, the proposed design can be manufactured with alternative materials as long as the additive method resembles a Selective Laser Sintering or Stereolithography type manufacturing.

EasyFlow System Set Up

Details of the EasyFlow bioreactor system are presented in Figure 1 and in the Result section, in details.

FIGURE 1
www.frontiersin.org

Figure 1. EasyFlow system characteristics and setup. Cross-sectional view of the EasyFlow insert visualizing the inner structure and interconnectivity of the reaction space contained within a 50 ml centrifuge tube (A). Schematic of the perfusion loop which includes the EasyFlow bioreactor accommodating the porcine carotid artery, a reservoir, the peristaltic pump, and acquisition system (B). A simplified connection map and true image of system identifying the key elements of the perfusion loop (C).

The EasyFlow insert consists of a 3D printed part that fits as a lid of a 50 ml standard centrifuge tube. The hexagonal EasyFlow insert presents 6 inlet and outlet ports, creating two separate circulations inside and outside the blood vessel, which is lodged in its adaptor (Figure 1A).

The bioreactor system is composed by three main parts: a peristaltic pump (Heidolph, PD 5201, 523-52010-00-2) mounting a 4-cassette head (Heidolph, Multichannel Pump Head C8, 524-40810-00), an EasyFlow insert lodging the blood vessel sample and an equivalent EasyFlow functioning as media reservoir (Figure 1B). The two inserts are connected through a series of tubes. Most of the components are connected with transparent silicon tubing (RS Components, RS PRO Silicone Transparent Silicone Tubing, 3.2 mm Bore Size, 667-8444). Larger bore peristaltic pump-compatible tubing (RS Component, Verderflex Verderprene Yellow Process Tubing, 6.4 mm Bore Size, 125-4042) were used in the pump head. Unidirectional flow is ensured by check valves (Cole-Parmer, Masterflex PVDF and Viton® Barbed Check Valve, 3/16-inch, WZ-98766-50) added downstream the peristaltic pump, and pressure is stabilized by a pressure dampener system created with a syringe loaded with liquid and air. To increase baseline pressure low bore tubing (RS Components, RS PRO Silicone Transparent Silicone Tubing, 0.8 mm Bore Size, 667-8432) was used downstream of the artery. To regulate resistance, an additional clamp (Cole-Parmer, Flow Control Clamps, 08028-00) was applied downstream of the resistance tubing. Sterile gas exchange is enabled by the presence of outlets mounted with sterility filters (Sarstedt, Syringe filter, Filtropur S, PES, pore size: 0.2 μm, 83.1826.001) (Figure 1C).

The system was monitored by pressure sensors downstream of the sample-containing EasyFlow (Parker, SciLog SciPres Pressure Sensor Luer Connection, SCIPRE080699PSX), and through repeated Duplex ultrasound imaging performed with Logiq V2 (GE Healthcare).

To assure the aseptic environment necessary for the long-term incubation of tissues, all components were sterilized by autoclaving. Elements of the circulation system were assembled under a laminar flow hood and primed with perfusion media (DMEM + 10% FBS + 1% P/S + 1 % Amp + 3% Dextran). The circulation system was inspected for leaks, and any bubble trapped in the system was removed by flowing media across the system.

Previously harvested and cleaned tissue was then placed in the EasyFlow unit. Depending on the internal diameter of the vessel segment, an adequate barb connector was selected to facilitate the connection to the perfusion system, typically for porcine carotid arteries 2.4 or 1.6 mm ID connectors were used (Cole-Parmer, Female luer x 5/32" hose barb adapter, 45501-06). The length adaptor within EasyFlow is ultimately adjusted to create a minimal stretch on the longitudinal axis (Figure 1A).

Once the tissue was secured using surgical vessel bond (Vessel Loop Maxi, SKU: AR-722), the residual air was removed from the newly assembled segment before sealing the system. A sterile syringe containing 3 ml of air and 1 ml of media is connected to the port situated above the artery. This addition serves as a pressure dampener assuring that there is no extreme pressure effecting the artery. Following complete assembly, the system is sealed and transferred it to the incubator (5% CO2 at 37°C). Circulation was connected to the peristaltic pump (Figure 1C) and primed overnight with low flow conditions (5 rpm). This allows system equilibrium and enables troubleshooting while minimizing unnecessary strain to the tissue. Flow was gradually increased (+1 rpm every 60 min) until the setpoint is achieved (14 rpm). The pressure of the system is constantly monitored safeguarding the vessel from any extreme condition. Once the final pump speed is reached, resistance upstream of the blood vessel is adjusted by incrementally tightening the downstream screw clamp until satisfactory pressure was reached within the system.

Balloon Injury Procedure

A standard peripheral artery angioplasty balloon catheter (Cook Medical, Advance® 35 LP Low 4 mm) was inserted into the artery. With a standard in deflator pump the balloon was inflated with phosphate buffer saline (PBS) to 6 atm and the pressure was maintained for 3 min, akin to the procedure performed during peripheral arterial angioplasties. After deflation, the balloon was extracted, the tissue rinsed and placed in the perfusion system for culture, or static culture as a control.

Artery Culture in Flow or Static Conditions

Healthy and balloon injured arteries were cultured for 7 days in static or flow conditions. In static culture, 2–4 mm long tissue segments are submerged in medium (DMEM + 10% FBS + 1% P/S + 1% Amp) in Petri dishes, and incubated at 5% CO2, 37°C. Every 3 days 50% of the media was replaced with fresh media.

Perfusion cultures were performed using the EasyFlow bioreactor system, as described above, on segments of 2–4 cm of length. The system is filled with ~100 ml of perfusion media (DMEM + 10% FBS + 1% P/S + 1 % Amp + 3% Dextran). Of note, the addition of dextran exclusively to the flow culture media ensures the appropriate viscosity of the medium which in turn affects the shear stress, while avoiding reducing the diffusion of oxygen in the static cultures, that would represent a limiting factor for the culture. The peristaltic pump is set at the final speed of 14 rpm which, due to the 4 rolls mounted on the peristaltic pump, is equivalent to 56 pulse/min. This pump setting corresponds to 35 ml/min volumetric flow rate (Supplementary Figure 1A) and an average stroke volume of 0.6 ml (calculated as volumetric flow rate/pulse rate). To assure that the tissue is not subject to extreme pressure conditions generated by the peristaltic movement, a pressure dampener was introduced to maintain fluctuations within the physiological range (60–120 mmHg), these values were monitored across the entire experiment in one second intervals, using a pressure acquisition system (Supplementary Figure 1B). Lastly, conceptually similar to a heart valve and to maintain “diastolic pressure,” a one-way valve was introduced after the peristaltic pump (Figure 1C).

Samples in flow were monitored at day 3 and 7 by Duplex ultrasound imaging. At the end of the culture period, all samples were processed for histology and en face staining.

Doppler Ultrasound Imaging and Vasoreactivity Assay

Ultrasound images of the incubated arteries were obtained with a standard clinical vascular ultrasound machine with a 10 MHz linear array transducer (LOGIQ V2, GE Healthcare). Briefly, ultrasound gel was applied to the outside of the bioreactor to minimize acoustic impedance, compensate for surface incongruencies between probe and bioreactor and allow angulation of probe to obtain an insonation angle of <60°. After clear longitudinal images were obtained, the position of the ultrasound probe and B-Mode settings were adjusted until clear vessel borders with typical M and I lines were visible, and several cine video loops were acquired and saved as video files for off-line analysis. Flow velocity measurements were performed with pulsed wave Doppler. The sample volume was placed at the center of the perfused artery, insonation angle adjusted to <60° and angle correction applied. Image analysis was performed offline with an automated edge detection software (Brachial analyzer, Medical Imaging Applications). The luminal diameter was measured on video loops over several pulsation cycles. Each diameter measurement represents the average diameter over 0.5–1.0 cm vessel area. The maximal diameter was designated systolic diameter and minimal diameter as diastolic diameter. Volumetric flow rate (Q) was calculated as follows:

Q = A × V¯

Where V¯ is the mean Flow Velocity and A is the Cross-sectional Area, calculated as π(mean diameter/2)2 and mean diameter = (maximal diameter + minimal diameter)/2.

In our setup, the mean diameter of the artery was 1.2 mm and a mean velocity of ~50 cm/s (Figures 4A,B, and Supplementary Figure 1C) and this resulted in a calculated volumetric flow rate of ~35 ml/min within the artery, which confirms the calibration values obtained empirically Supplementary Figure 1A).

Wall shear stress (WSS) was estimated as:

WSS = 8 × μ × V¯mean diameter

The viscosity of the perfusion medium (μ) was estimated at 0.035 dyn*s/cm2 (10). Intimal media thickness was also measured with this software over the same vessel segment on the far side as the distance between lumen-intima (I line) and media-adventitia interface (M line).

For the functional vasoreactivity assay, syringes containing 1 ml of nitroglycerin solutions were prepared, with concentration ranging from 10−9 to 1 M. These concentrations were calculated so that added to 100 ml of medium contained in the bioreactor, they yielded final concentrations ranging from 10−11 and 10−2 M. Nitroglycerin boluses were injected in order of increasing concentration and the reaction of the blood vessel was recorded with a delay of 1 min, once the vessel normalized the next bolus was injected.

Histological Preparations

Portions of freshly isolated (control) and cultured arteries were fixed in 4% PFA (Paraformaldehyde, Santa Cruz Biotechnology) overnight (O/N) at 4°C. Fixed samples were either used for whole tissue en-face staining or were further processed for histological analysis.

To establish the maintenance of a continuous endothelial layer facing the lumen and quantify the effect of flow on the alignment of the cells, en-face staining was performed on samples washed with PBS and permeabilized with 1v/v% TritonX-100 in PBS for 15 min. Tissues were blocked overnight at 4°C in 20v/v% goat serum in PBS and then incubated with primary antibody solution (CD31, Abcam-ab28364, 1:50) overnight followed by goat anti-Rabbit Alexa Fluor 488® (Thermo Fisher Scientific) secondary antibody diluted 1:200 and Phalloidin-iFluor 594 (Abcam- ab176757) for 1 h at 37°C. Nuclei were stained with DAPI (4′,6-diamidino-2-phenylindole, Merck) for 10 min at room temperature. Tissues were laid flat and mounted with Fluoromount G (Invitrogen eBioscience Fluoromount G, Thermo Fisher Scientific). Imaging was performed with Nikon Eclipse Ti A1-A confocal laser scanning microscope, obtaining stack images at 40X along the Z-axis.

For histology, tissues were washed with PBS and incubated overnight in 30w/v% sucrose (Sigma-Aldrich) solution in PBS. Following sucrose incubation, samples were embedded in OCT Compound (Agar scientific) in an Iso-Pentane bath cooled using liquid nitrogen. Frozen samples were stored at−80°C until later use. Samples were cryo-sectioned along the transverse plane, obtaining 5 μm thick sections. Sections were collected on microscope slides in quadruplicate, where sections were separated by at least 250 μm.

Immunofluorescent staining was performed on prepared sections to quantify luminal coverage, medial integrity and the extent of proliferation. Tissue sections underwent antigen retrieval in a water bath at 80°C for 30 min in tris-EDTA buffer (10 mM Tris Base, 1 mM EDTA Solution, pH 9.0) followed by blocking for 1 h at room temperature with 20v/v% goat serum (Sigma-Aldrich) in PBS. Primary antibodies against CD31 (Abcam-ab28364) diluted 1:200, and Proliferating cell nuclear antigen (PCNA, Sigma Aldrich-MABE288) diluted 1:100 were incubated overnight at 4°C.

Appropriate Goat anti-Mouse and Goat anti-Rabbit Alexa Fluor® (Thermo Fisher Scientific) secondary antibodies 488 and 567 diluted 1:200 were incubated for 1 h at 37°C. Following secondary antibodies, Human α-Smooth Muscle Actin (SMA) Alexa Fluor® 647-conjugated antibody (R&D Systems- IC1420R) diluted 1:200 was introduced for 1 h at 37°C. Nuclei were stained with DAPI (Merck) for 10 min at room temperature. Incubation with 0.1 w/v % Sudan Black (Sudan Black B, Santa Crus Biotechnology) in 70% ethanol for 10 min at room temperature was performed to reduce tissue autofluorescence. Slides were then mounted in Fluoromount G (Invitrogen eBioscience Fluoromount G, Thermo Fisher Scientific) and imaged with Nikon Eclipse Ti A1-A confocal laser scanning microscope. Tile images were obtained at 20X to capture the whole tissue section where possible.

Hematoxylin and Eosin (H&E), and Masson's trichrome staining (MT) were used to estimate the overall maintenance of the tissue structure, at a cellular and extracellular matrix level. These stainings were outsourced to the Veterinary School Pathology Centre at the University of Surrey, and performed by an automated staining system (VIP® 6 Vacuum Infiltration Processor). Immunohistochemical assays for Caspase 3 (1:1,000, R&D systems, AF835), PDGFR-β (Santa Cruz Biotechnology, sc-374573, 1:100) and Vimentin (1:5,000 Dako, Ref M0725, Clone V9, 1:5,000) were similarly outsourced to Veterinary Diagnostic Services, School of Veterinary Medicine, University of Glasgow. Primary antibodies were followed by HRP-conjugated secondary antibody and DAB (3,3′-Diaminobenzidine) staining and counter-staining with hematoxylin. Staining was performed in the Dako Autostainer system. Slices were dehydrated and mounted with Cellpath Mounting Media (SEA-1604-00A).

Whole slice imaging was performed on the NanoZoomer 2.0-HT slide scanner by Hamamatsu.

Image Analysis

All acquired images were processed in ImageJ v1.53c (FiJi). Custom macros were composed to facilitate image analysis. Where necessary, vessel wall areas (lumen, intima, media, adventitia) were defined manually and stored as Regions of Interest (ROIs). Resulting ROIs were used to define tissue dimensions and to measure signals specific to the predefined areas.

Quantification of Cell Activation

The quantification of PCNA was performed on the fluorescent images collected by confocal microscopy. Tile images covering the cross section of the tissue were obtained to measure distribution of PCNA across the tissue for each section (three sections/sample). PCNA+ nuclei counts were normalized against the total number of DAPI+ nuclei within each area.

To define the spatial distribution of PCNA+ cells, images were converted into distance maps where orthogonal distance of each nucleus from the lumen was calculated. Each distance was further normalized against the thickness of the tunica media, to account for differences in tissue morphology between specimens. Individual distance values were organized into a histogram representing the relative frequency of PCNA+ cells against the relative distance from the lumen.

SMA Quantification in Immunofluorescence Samples

SMA intensity in the media was quantified on the immunofluorescence samples by selecting three random, uniformly sized, representative areas within the media of each section. The average signal intensity was quantified in each of the selected areas, and normalized by the total area selected.

Signal Quantification in Histological Samples

For histological staining, whole tissue sections were loaded on ImageJ to measure specific signal expression. First an ROI was defined by high-pass gaussian filter to identify the total area of the section. Using a series of user-defined color thresholds, the specific antigen signal was selected and extracted from the image. The area of specific signal was measured and normalized toward the total tissue area.

Tissue Coherence Analysis in Using Masson's Staining

Histological samples prepared with Masson's trichomic staining were used to analyze the coherence of the tissue. Three random, uniformly sized, representative regions were selected to perform the analysis. In each region, the area occupied by the tissue vs. the empty spaces were measured and normalized toward the total area of measurement.

Endothelial Coverage Quantification

The quantification of endothelial coverage was performed on the fluorescent images obtained by confocal microscopy. To quantify the coverage, the fraction of CD31-expressing lumen over the total length of the portion of lumen perimeter captured in each picture was calculated in vessel cross-sections.

Fiber Alignment

Fiber orientation was quantified in the en-face confocal images compressed into a Z-Max projection. In each picture, three random areas were selected and a vector map of the Phalloidin signal was generated using OrientationJ (ImageJ/FiJi plugin) to obtain the local orientation of the signal within the selected area. Resulting vector values were organized into a histogram to represent the relative frequency of angular orientations, where zero is the representative value of fibers perpendicular to the flow direction.

Statistics

Experiments were repeated in 3 to 5 biological replicates. In some analyses, individual samples have been excluded due to poor quality of the preparation or staining.

Difference among groups were evaluated using one-way ANOVA or Kruskal–Wallis test, based on results from normality tests, followed by Fisher's LSD post-hoc test (GraphPad Prism 8.1.2). A value of p < 0.05 was considered as statistically significant. Data are presented as mean ± SD.

Results

Design and Specifications of the EasyFlow Bioreactor Insert

The resolution of MultiJet fusion 3D printing enabled the efficient and cost-effective production of the EasyFlow insert, incorporating detailed features, and producing a non-porous surface. The resulting product is biocompatible and autoclavable multiple times. The insert is designed to fit as the lid of a 50 ml centrifuge tube, creating a versatile and self-contained bioreactor for vascular culture (Figure 1A). The EasyFlow insert is hexagonal in shape and incorporates the following functional features: de-bubbler to remove any air bubbles trapped in the system, 6 inlet/outlet ports creating the potential for 2 separate circulations (inner circulation-going through the vessel sample, and outer circulation-to exchange medium in the tube), and extra access ports for sensors, gas/pressure exchange, and endovascular probes or catheters (Figure 1A).

The EasyFlow system was created as a self-contained, off-the-shelf solution for a wide range of experimental setups, and therefore was designed to maximize the included features, while maintaining a compact design. The hexagonal shape provides ideal internal space to position the necessary features, and enables to spatially distribute the connectors externally, allowing easy handling and setup, and offering enough space flexibility to connect external components.

The blood vessel is accommodated in an adjustable adaptor, through barb connectors of interchangeable size (Figure 1A and Supplementary Video 1). Once the bioreactor containing the sample is sealed, it can be connected to an equivalent bioreactor functioning as a medium reservoir and to a peristaltic pump, though medical grade tubing (Figures 1B,C). EasyFlow is amenable to Doppler imaging, the probe can be applied directly to the to the exterior wall of the centrifuge tube, enabling longitudinal monitoring of the changes in hemodynamic forces and the structural/functional response of the blood vessel (Figure 1B). Additional features are added to the system to improve the performance, such us one-way valves to ensure unidirectional flow, pressure dampener to regulate the system pressure, and gas exchange filters (Figure 1C). Thanks to these features, we ensure that the blood vessel cultured within EasyFlow would experience physiological “heart rate” (56 pulse/min) and “blood pressure” (60–120 mmHg). To maintain the simplicity of use, pulsatility was imparted by the rotation of the peristaltic pump. Pressure measurements were recorded from a pressure sensor downstream from the artery every second (Figure 1 and Supplementary Figure 1B), and adjusted manually by regulating the clamp tightness according to the continuous pressure readings obtained. The media used in our culture was enriched with dextran, to obtain a media viscosity similar to that of human blood (0.035 dyn/sec*cm2 or 3.5 cP).

EasyFlow Culture Preserves Extracellular Matrix Organization and Tissue Homeostasis

Next, we tested how the EasyFlow environment impacted on cellular physiology and ultrastructural components, by performing en face staining, immunofluorescence imaging and histology in porcine carotid artery segments cultured in EasyFlow or in static culture for 7 days, and compared them to their freshly isolated segments that were used as “untreated” controls.

Hematoxylin and eosin staining revealed that the overall structure of the tissue was preserved during culture under flow conditions. In particular, the vessel wall layers appear unaltered, and no remodeling was observed (Figure 2A). Some of the samples cultured in static presented a distinct displacement of the intima, and loss of structure in the media and adventitia (Figure 2A).

FIGURE 2
www.frontiersin.org

Figure 2. Histological evaluation of porcine carotid arteries cultured in EasyFlow or static conditions. Hematoxylin and Eosin staining H&E (A) and Masson's Trichomic staining MASSON'S, (B) of freshly isolated carotid artery tissue (CONTROL) and samples cultured for 7 days in EasyFlow (PERFUSION) or static culture (STATIC). Scale bars= 500 μm. Quantification of tissue coherence is displayed in the bar graph (C), n = 5. All data are mean ± SD. One-Way ANOVA analysis was performed, *p ≤ 0.05.

Extracellular matrix provides the mechanical support and elasticity necessary for blood vessels to perform their function, cell apoptosis and abnormal activation may affect this structure, weakening the vessel wall. We visualized the collagen content by Masson's trichrome staining and confirmed the maintenance of the extracellular matrix organization in tissues cultured in EasyFlow, while static culture produced a visible loss of tissue integrity (Figure 2B). These observations were reflected in the quantification of tissue coherence which indicated an increase in tissue disarrangement in static cultured samples (Figure 2C).

Immunofluorescence staining revealed the finer changes in cell distribution. Both endothelial coverage of the lumen and expression of smooth muscle actin in the media were slightly reduced in the static condition, as compared to the flow culture, although not to a significant level (Figures 3A–C). These modest changes were reflected by the histological analysis of apoptosis and proliferation. Staining for the apoptosis marker Caspase-3, indicated no significant increase in cell death following culture, although a trend is clearly visible in 3/5 static culture samples (Figure 3D). Quantification of the overall percentage of PCNA+ cells revealed a slight increase in proliferation over culture, but no significant differences compared to the control (Figure 3E). We also detected a decrease in Vimentin+ cells in the static culture, potentially reflecting a loss of adventitia (Figure 3F), and an overall maintenance of PDGFRβ+ expression (Figure 3G).

FIGURE 3
www.frontiersin.org

Figure 3. Immunofluorescence analysis of cultured carotid arteries indicates better tissue preservation in flow. Immunofluorescence staining of freshly isolated arteries (d0, Control/CRT), EasyFlow (Perfusion/BR) and static cultured arteries (Static/S) at day 7, showing expression of endothelial marker CD31, smooth muscle marker smooth muscle actin (SMA), and proliferation marker proliferating cell nuclear antigen (PCNA). Nuclei are stained with DAPI. Representative confocal images are displayed in composites and by individual marker in montage (A). Lumen is identified (L), Scalebar = 500 μm. Quantifications of lumen coverage (B), SMA intensity (C), apoptosis caspase 3, casp-3, (D), proliferation PCNA (E), vimentin VIM (F) and platelet-derived growth factor receptor beta PDGFRβ (G) are displayed in the bar graphs, n = 3–5. All data are mean ± SD. One-Way ANOVA analysis was performed, *p ≤ 0.05, **p ≤ 0.01. En-Face representative images showing endothelial coverage (CD31), stress fiber (Phalloidin), and nuclei (DAPI) in samples cultured for 7 days in perfusion or static (H). Fiber alignment is quantified as vector distribution and plotted in histogram representing the relative frequency (y) against the angle (x), where zero is perpendicular to the flow direction (Orientation). An ideal normal distribution of values is represented in green over the histogram (h). Scalebar = 100 μm.

Next, en face staining was performed to better visualize endothelial coverage and media organization. Overall, en face staining revealed a good luminal coverage in both culture conditions (Figure 3H), however the intensity of the expression of CD31 and the cell organization of cultured samples appeared to differ from the freshly isolated control. This could also represent an artifact of the en face tissue processing which is affected by the tissue stiffness and integrity. On the other hand, culture in flow produced more organized structures in the media, which were comparable to the native tissues, while static culture conditions led to fiber disarray (Figure 3H).

Taken together these results indicate that culture of porcine arteries in pulsatile flow conditions helps maintaining appropriate cellular organization and structural integrity.

EasyFlow Can Be Monitored Using Ultrasound, Revealing Culture Conditions Resembling in vivo Hemodynamics and Physico-Mechanics

We applied an ultrasound probe to the wall of the centrifuge tube, and investigated the mechanical parameters to which the cultured artery was subjected in the bioreactor. By Doppler/Duplex analysis, we established that the carotid arteries within EasyFlow were exposed to physiological arterial hemodynamic conditions. The average diameter of arteries after 7 days of culture under pulsatile flow was 1.24 ± 0.32 mm (“systolic”) and 1.19 ± 0.30 mm (“diastolic”). The ‘peak systolic velocity' was 52 ± 13 cm/s and ‘mean velocity' was 44 ± 8 cm/s, resulting in an estimated volumetric luminal flow of 33 ± 16 ml/min (11) and wall shear stress of 10.1 ± 2.8 dyn/cm2 (12) (Figures 4A,B). The average IMT was 0.32 ± 0.08 mm. Overall repeated measurements of vessel diameter showed that there was an average deviation (AD) of 0.00 ± 0.03 mm between two independent automated measurements of four vessel (Figure 4C). These measurements support that EasyFlow provides reproducible ex vivo hydrodynamic conditions that resemble in vivo hemodynamics and physico-mechanics.

FIGURE 4
www.frontiersin.org

Figure 4. Ultrasound evaluation of native and angioplastied carotid arteries after culture reveal in vivo relevant culture conditions and impairment after injury. Representative ultrasound image of an artery in culture (A), highlighting the morphological parameters captured (B). Mean value (MVal) and average deviation (AD) of vessel parameters in flow culture at day 7. IMT, intima media thickness; PSV, peak systolic velocity; MV, mean velocity; WSS, wall shear stress; Systolic, maximal pulsatile diameter; Diastolic, minimal pulsatile diameter (n = 4). Bland Altman plot shows the consistency of repeated measures within the samples (C). Healthy Perfusion (D) and angioplastied Balloon (E) arteries were cultured for 3 and 7 days in EasyFlow and their systolic and diastolic diameter, the intima-media thickness (IMT), and their dose-response to endothelial-independent vasodilator nitroglycerin (F) were quantified by Doppler ultrasound, n = 2–4.

EasyFlow Cultures Are Amenable to Endovascular Balloon Procedures

The study of vascular pathology ex vivo has the potential to reduce and replace the number of animals used as models, providing results that are more physiologically relevant and may have scope to improve the patient care.

We performed balloon injury in porcine carotid arteries before culturing them for 7 days in flow conditions. An important advantage of the EasyFlow system is the ability to monitor the structural changes and functionality of the vessel longitudinally, by Doppler ultrasound. Here, we used it to determine the luminal diameter and intima media thickness and luminal flow within the artery, providing a closer estimate of the hemodynamic environment within the vessel. We also investigated functional response of the vessel wall to increasing doses of the endothelium-independent vasodilator nitroglycerin. Ultrasound showed that at day 3 the diameter of the vessels was significantly greater after angioplasty as compared to control (systolic: 1.17 ± 0.52 mm vs. 2.79 ± 0.12 mm, p = 0.026, diastolic: 1.09 ± 0.52 mm vs. 2.47 ± 0.04 mm, p = 0.044). While control retained the diameter at day 7, injured arteries significantly decreased diastolic diameter (control:1.19 ± 0.30 mm, p = 0.779; angioplasty: 1.44 ± 0.60 mm, p = 0.041). There were no significant changes in luminal flow (Figures 4D,E, Supplementary Videos 2, 3). We leveraged on our ability to monitor the blood vessel in culture by ultrasound to quantify tissue remodeling using clinically relevant readouts. Intima-Media thickness (IMT) is considered a gold standard for the prediction of subclinical vascular remodeling and propensity to atherosclerosis in patients. We performed a preliminary analysis by measuring IMT in B-mode images, and showed that after injury there was an increase in IMT. While IMT remained stable between day 3 and 7 in control (0.27 ± 0.04 mm and 0.32 ± 0.08 mm, p = 0.355), it was decreased as compared to control in angioplastied arteries at day 3 (0.15 ± 0.01) and increased at day 7 (0.38 ± 0.06 mm) (Figures 4D,E). We also compared the vascular response of the injured and control arteries cultured in flow. This identified a loss of 60–70% endothelium-independent vasodilator response upon stimulation with increasing doses of nitroglycerin (Figure 4F).

Next, we compared the ultrastructural effect of culturing injured arteries in static and flow conditions by histology, immunofluorescence and en face staining. As expected, balloon injury determined a significant loss of cells and disorganization in the extracellular matrix in the intima and the medial layers, as shown in the H&E and Masson's trichrome staining (Figure 5A). The effect of the injury was dramatic in tissues cultured in static, likely due to the poor preservation in these conditions. In arteries cultured in flow conditions, the damaged media after injury corresponded to 21.7 ± 11.6% of the total vessel wall. In order to visualize the spatial effect of the injury, we quantified the distribution of nuclei within the vascular wall in culture, with or without injury. Results show that in non-injured samples, the static culture determined a baseline reduction of cell number in the media, and that the balloon injury almost completely abolished cells in these samples (Figure 5B). In flow conditions, we observed a striking loss of cells close to the lumen upon injury, and an accumulation of cells in the distal part of the media, at the junction with the adventitia (Figure 5B). En face staining confirmed that balloon injury provoked extensive endothelial denudation in both culture system and that the alignment of the F-actin fibers was lost in the flow culture, as a result of extension damage (Figures 5C,D). Immunofluorescence analysis confirmed the significant loss of endothelial coverage, and a mild reduction in medial SMA expression in vessels cultured in flow after injury, which is likely due to large areas of damage produced in the media (Figures 6A–C).

FIGURE 5
www.frontiersin.org

Figure 5. Immuno-histological evaluation of angioplastied carotid arteries after culture reveal functional and structural damage. Injured vessels were cultured in EasyFlow (Perfusion Injury) or in static conditions (Static Injury) for 7 days and analyzed by Hematoxylin and Eosin staining (H&E) and Masson's Trichomic staining (MASSON'S) (A). Scale bars = 500 μm. Distribution of DAPI+ nuclei across the vessel wall is quantified in freshly isolated arteries (CRT), healthy and injured arteries cultured for 7 days in EasyFlow (BR and BRB, respectively) or static conditions (S and SB, respectively) (B). En-Face representative images showing endothelial coverage (CD31), stress fiber (Phalloidin), and nuclei (DAPI) in samples cultured for 7 days in perfusion or static after injury (C). Fiber alignment is quantified as vector distribution and plotted in histogram representing the relative frequency (y) against the angle (x), where zero is perpendicular to the flow direction (Orientation). An ideal normal distribution of values is represented in green over the histogram (D). Scalebar = 100 μm.

FIGURE 6
www.frontiersin.org

Figure 6. Immunofluorescence evaluation reveals tissue injury and activation following balloon angioplasty and culture in flow. Following injury, tissues were cultured for 7 days in either perfusion (Perfusion Injury) of static conditions (Static Injury) and then stained for the endothelial marker (CD31), smooth muscle marker (smooth muscle actin, SMA), and proliferation marker (proliferating cell nuclear antigen, PCNA). Nuclei are stained with DAPI. Representative confocal images are displayed in composites and by individual marker in montage (A). Lumen is identified (L), Scalebar= 500 μm. Quantifications of lumen coverage (B), SMA intensity (C), and proliferation PCNA (D) are displayed in the bar graphs, n = 3–5. All data are mean ± SD. One-Way ANOVA analysis was performed, *p ≤ 0.05, **p ≤ 0.01, ***p ≤ 0.001, ****p ≤ 0.0001. Distribution of proliferating cells (PCNA+) in relation to their relative distance to the lumen is plotted for freshly isolated arteries (CRT), healthy and injured arteries cultured in EasyFlow (BR and BRB, respectively) or static conditions (S and SB, respectively) (E).

Taken together these results show that EasyFlow cultured arteries subjected to endovascular damage present areas of cell loss and functional impairment, but maintain overall structural integrity.

Injured Arteries Cultured in EasyFlow Present Signs of Remodeling

In vivo, vascular injury leads to extensive apoptosis and cell replacement, which triggers the activation of the vascular wall progenitor niche and the synthetic switch of vascular smooth muscle cells. Pathological remodeling generally follows, with the formation of fibrosis, neointimal growth or atherosclerotic plaques.

We analyzed the arteries cultured for 7 days in static or EasyFlow post-injury. Overall proliferation was uniquely increased in the injured samples cultured in flow (Figure 6D), and the PCNA+ cells were mainly localized at the interface between the media and the intima, and in the luminal side of the media (Figures 6A,E). Proliferating cells mainly co-expressed SMA, and were found largely in undamaged area of the media (Figure 7ai), however we also detected a unique population of proliferative SMA+ cells, separated from the injured media, and located beneath the lamina elastica (Figure 7aii). It was also possible to identify the occasional proliferative CD31+ cell in the intima (Figure 7ai) and the adventitia (Figure 7aiii). Organized microvascular structures can also be observed penetrating the media after injury (Figure 7a, red arrowhead), as well as infiltrating mono-nucleated cells which could potentially represent resident immune cells (Figure 7a, orange arrowhead).

FIGURE 7
www.frontiersin.org

Figure 7. Balloon angioplasty stimulates early remodeling in EasyFlow cultured arteries. Representative confocal images of the arterial wall following injury and flow culture (a) reveals proliferation (proliferating cell nuclear antigen, PCNA, magenta) of smooth muscle cells (smooth muscle actin, SMA, white) in the proximity of the injured media (ai). In the lumen area, individual proliferating (PCNA+) smooth muscle cells (SMA+) and endothelial cells (CD31+, green) are located either side of the lamina elastica (aii). In the adventitia, proliferative endothelial cells (CD31+) line the microvasculature of the vasa vasorum (aiii). White arrowheads indicate proliferative cells. Across the vessel wall, groups of infiltrating cells display PCNA+ nuclei (orange arrowheads), and small blood vessels sprout from the adventitia toward the media (red arrowhead). Nuclei are stained with DAPI, Scale bar= 100 μm. Smooth muscle cells near the injury area display high expression levels of SMA white (b), Vimentin (c), and PDGFRβ (d). Scale bars = 100 μm.

Immunostaining also revealed that in the border zone near the medial damage, smooth muscle cells strongly expressed SMA, Vimentin and PDGFRβ (Figures 7b–d). This indicates that smooth muscle cells in these areas were provoked into the synthetic phase by the injury.

Our results showed that culture under flow better recapitulate the early characteristics of post-injury vascular remodeling, as compared to static culture, and that early hallmarks of activation and remodeling can be identified in flow cultures by histology and ultrasound imaging.

Discussion

In this study we presented a 3D printed bioreactor that is economical to produce, easily transferable to other labs and enables small volume cultures and multiplexing. EasyFlow cultures can be imaged longitudinally by Doppler ultrasound, enabling the monitoring of flow conditions and tissue integrity in a non-invasive manner. Using EasyFlow bioreactor, porcine carotid arteries can be cultured in pulsatile flow conditions closely resembling the mammalian arterial circulation. When compared to static cultures, tissue integrity and organization was superior in flow cultures. Endovascular balloon injury provoked intimal denudation and media damage, also detected by Doppler ultrasound imaging. After injury, samples cultured in EasyFlow showed increased proliferation in the media and adventitia, and overexpression of smooth muscle markers near the injury. A small population of highly proliferative SMA+ cells was also observed at the intima/media interface. Samples cultured in static were instead significantly compromised and did not show signs of activation.

The study of vascular physiology and pathology is often pursued with the use of moderate severity animal models such as carotid ligation or wire injury, in some case presenting genetic mutations to emulate specific co-morbidities such as diabetes [db/db mice: homozygous for the diabetes-inducing spontaneous mutation (Leprdb)] and hypercholesterolemia (apolipoprotein E knockout mice; ApoE-/-) (3). These in vivo experiments are routinely supported by in vitro work to help elucidate mechanistic insights, using static monocultures of vascular cells, which struggle to recapitulate the physiologically relevant mechanical cues and the interaction with other cell types.

Despite these limitations, these models delivered great insights in the understanding of the mechanisms of vascular homeostasis and the patterns associated with common diseases, such as atherosclerosis and restenosis (13). The striking physiological differences have however determined disappointing outcomes for some of the therapeutic approaches in clinical settings, sparking new interest in large animal models for these diseases (14). However, ethical and economic implications associated with this type of animal work demands new solutions.

Vascular bioreactors have been developed both at commercial level and in-house to provide a suitable physiological environment to maintain blood vessels in culture or enhance the maturation of tissue engineered constructs.

In this study, we present a 3D printed bioreactor that enables the culture of blood vessels within a 50 ml centrifuge tube, enabling small volume cultures and multiplexing. Previous work has demonstrated the feasibility and practicality of a bioreactor designed to fit as a lid of a 50 ml tube, however our system is distinctive as the whole insert is created entirely by 3D printing (15, 16).

EasyFlow's open-access design allows simple uptake by other labs, and the adjustable vessel compartment enables customization to different vessel sizes and tissue sources by simply exchanging the barbed connectors and adjusting the length adapters.

Dynamic bioreactors supply culture conditions that are as close as possible to the in vivo environment experienced by the blood vessel in its natural location. Pump revolution rate was adjusted to impose a “heart rate” of 56 beats/min to simulate resting heart rate and continue pressure monitoring confirmed a physiological range of 60–120 mmHg, which is compatible to the average healthy blood pressure in the human population. Common carotid arteries experience a mean flow velocity of 15–43 cm/s (end diastolic-peak systolic) and a wall shear stress in the range of 6–21 dyne/cm2 (17, 18). EasyFlow culture conditions provide a in vivo relevant environment in the culture, with a mean flow velocity of 44 ± 8 cm/s, resulting in a calculated volumetric flow rate of 33 ±16 ml/min and a mean wall shear stress of 10.1 ± 2.8 dyne/cm2, which are comparable to the arterial flow forces in vivo. It is important to highlight that more sophisticated systems for the regulation of the pressure loop, and the shape of the pressure curve exist and have been developed for specific bioreactor applications. While EasyFlow is a more simplified version, this is a compromise that fulfills the critical characteristic of an end-user friendly system (15, 16).

In addition, we modified the medium composition to simulate the viscosity of blood (3.5–5.5 cP), this parameter closely affects the shear stress experienced by vascular cells in flow, according to the Hagen-Poiseuille equation (18).

Our in-depth consideration and characterization of the conditions and forces acting on the vessel in culture is uncommon in previously published studies (9, 1921). Importantly, these forces were initially estimated by calculation, but also verified within the cultured blood vessel by Doppler ultrasound imaging. While few examples exist in literature, we present a uniquely exhaustive application of the Doppler ultrasound method to a bioreactor culture. This analysis enables direct comparison with large animal and patient studies, and allows to visualize vessel structure and functionality over time (22).

When compared to static culture, EasyFlow cultured arteries demonstrated an overall better maintenance of tissue integrity and structure. In particular, the organization of the smooth muscle cells and the extracellular matrix were closer to the native tissue. Interestingly, we did not detect a critical difference in lumen coverage, apoptosis or activation of the vessel wall in the two culture systems. This is not completely unexpected as the static culture was designed to replicate the vessel ring culture, with tissue specimens of 2–5 mm in length (5, 2325). Previous studies have demonstrated that static culture of rings can be extended up to 56 days, and can be induced to form neointima post-injury, although we did not detect such a growth in our samples (24). Samples cultured in flow were instead much longer (2–4 cm), and would not likely have survived in static culture due to limited oxygen and nutrient diffusion. We purposely elected to compare our flow culture system with the vessel ring model as this represents the most standardized and user-friendly whole vessel culture system currently available. By demonstrating that EasyFlow can not only achieve similar (or better) survival of much larger portions of the vessel, but also incorporate mechanical factors that provide crucial stimuli absent in the static culture, we effectively propose a superior and equally ingenuous solution for blood vessel culture.

After verifying the feasibility of a pulsatile culture using EasyFlow, we assessed the effect of injury in the flow cultures. We delivered the injury before setting up the static or flow culture, but it is important to note that, thanks to the multiple access ports available on EasyFlow, the catheter-mediated balloon injury could potentially be implemented on the mounted sample at any point of the culture, without disturbing the system.

In vivo, mechanical damage and flow disruption, accompanied by damage to the medial layer (for example by large balloons), results in smooth muscle cells activation, proliferation and migration into the intima layer, contributing to the thickening of the vessel wall and effectively narrowing the lumen (restenosis) (26, 27). Adventitial cells, such as resident progenitor cells and pericytes, macrophages and fibroblasts also become activated after injury, releasing cytokines and growth factors and migrating toward the lumen; their contribution to neointimal lesions and the mechanisms associated with it are still highly debated (28, 29).

Following balloon application and flow culture, we observed significant damage to the intima and media layers of the tissue, which is comparable to what reported in similar animal models where intima denudation and early media apoptosis drive the regeneration process (30). In addition, the surviving smooth muscle cells in the media responded to the injury by initiating a robust proliferative response. This is akin to the results obtained after 1 week in rabbit iliac arteries, even in terms of percentage of proliferative cells (roughly 4%) (31). Other hallmarks of early remodeling included the upregulation of contractile proteins in the smooth muscle cells adjacent the injury, and the proliferation of cells in the adventitia.

Interestingly, we did not observe the formation of neointima. This is likely due to the limited timeline of the experiment, as usually neointima forms in vivo after at least 2 weeks in rabbits, a typical model for vascular remodeling (31). However, we cannot exclude that large animals may react differently to angioplasty injury. For example, in vivo studies in pigs failed to detect significant reduction in lumen area even 2–4 weeks after balloon injury (32, 33).

On the other hand, previously published ex vivo work demonstrated that flow culture of porcine arteries is amenable for detection of restenosis 7 days after placement of a stent, this could be due to the smaller damage provoked by the stent as compared to the balloon, and the permanence of the ‘stimulus' as the stent remains in place after intervention (6). Similarly, low impact stimulation such as non-physiological low flow and intima denudation reportedly led to formation of neointima in vessels cultured under flow (6, 9).

After balloon injury, the loss of contractility in the vessel wall was quantifiable in the ultrasound profile at baseline levels and following challenge with vasoconstrictors. In addition, we were able to quantify an increase in IMT, a commonly used clinical indicator of remodeling (34). In samples cultured for 7 days after vascular injury IMT was increased, a measure that is associated with increased risk of stenosis-atherosclerosis in patients.

One critical characteristic of EasyFlow is that its open-source 3D printed nature enables easy adoption by other labs, and encourages reproducibility (see Supplementary File 1 for printing file). This is fundamentally different from previously proposed systems, which were either based on commercially available bioreactors or in house-built systems (69, 15, 16). Furthermore, bioreactors often leverage on the use of specialized glassware fitted with homemade connectors (6, 9) or more complex arrangements (7, 8), the details of which are regularly insufficiently described and designs undisclosed. The EasyFlow insert fits as the lid of a common laboratory 50 ml centrifuge tube, removing the need for specialized glassware. This approach was previously successfully proposed by Piola et al. (16) as the receptacle for their pulsatile pressure culture system for arterial conditioning of bypass vein material. EasyFlow entirely 3D printed design and simplified yet in vivo relevant culture conditions offer an improved off-the-shelf simplicity, uncommon in previously developed systems. The compactness of the EasyFlow insert enables multiplex cultures within a normal incubator where different flow/pressure conditions or interventions can be run independently by connecting each bioreactor/reservoir pair to a separate pump head. Many other designs are bulkier (6, 9) or driven by a single pump, limiting the options for parallel comparisons. In addition, EasyFlow uses small volumes of medium (~100 ml), similar to previously developed small volume bioreactors (15, 16), but much lower as compared to other published systems which required up to 500 ml (8), although in many cases the total volume of medium is not explicit and therefore cannot be compared (6, 9). Lower volume experiments (~50 ml) are also possible with EasyFlow by establishing a recirculating system without the addition of the reservoir, enabling low-cost pharmacological studies. Lastly, commercial bioreactors exist (i.e., VABIO by LifeTec and 3DCulturePro by TA Instruments) and represent a practical solution to many users. The limitation is the cost, which is upward of £10 K, and the rigidity of the design. Even with the outsourcing of the 3D printing, the reusable EasyFlow inserts can be produced cheaply (< £100 per piece), a price that can be further reduced if in-house 3D printing is available. The multiple ports on the EasyFlow insert also permits customization by offering inlets for additional sensors and other specific applications.

This proof of principle study demonstrated with an in-depth characterization of the quality of the tissue after culture, which is uncommon in other engineering-focused papers (7, 8), the suitability of EasyFlow for pulsatile culture of large animal arteries and the study of pathological processes. Overall, our study highlighted the scientific relevance of our ex vivo model and the versatility for both basic and interventional studies.

Data Availability Statement

All raw data, images, and analyses supporting the findings of this study are available through the online repository Zenodo (https://doi.org/10.5281/zenodo.6584173).

Ethics Statement

The animal study was reviewed and approved by Animal Welfare and Ethical Review Board (AWERB) of the Pirbright Institute. All procedures were conducted by Personal License holders who were trained and competent and under the Project License PPL70/8852.

Author Contributions

RM conceived, planned the experiments, and analyzed the data with contribution by CH. DM and CH provided key experimental expertise. RM and PC wrote the manuscript and prepared the figures with support of CH. PC, CH, and JM conceived the original idea and supervised the project. All authors contributed to the article and approved the submitted version.

Funding

RM was supported by the Doctoral College studentship award (University of Surrey). DM and PC were supported by the National Center for the Replacement, Refinement & Reduction of Animals in Research (grant numbers: NC/R001006/1 and NC/T001216/1).

Conflict of Interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Acknowledgments

The authors wish to thank the Veterinary Pathology Centre at the University of Surrey's School of Veterinary Medicine for histology services and the Histology Research Service, Veterinary Diagnostic Services, University of Glasgow. We also thank Drs L. Dixton, A Reis and M. Henstock from the Pirbright Institute (Pirbright, UK) for their support in procuring the animal tissues, and the Department of Biochemical Sciences at the University of Surrey, especially the technical team, for their continuing support.

Supplementary Material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcvm.2022.864580/full#supplementary-material

References

1. Beckman JA, Schneider PA, Conte MS. Advances in revascularization for peripheral artery disease: revascularization in PAD. Circ Res. (2021):1885–912. doi: 10.1161/CIRCRESAHA.121.318261 (accessed January 25, 2022).

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Frank U, Nikol S, Belch J, Boc V, Brodmann M, Carpentier PH, et al. ESVM guideline on peripheral arterial disease. VASA. (2019) 48:1–80. doi: 10.1024/0301-1526/a000834

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Xu Q. Mouse models of arteriosclerosis : from arterial injuries to vascular grafts. Am J Pathol. (2004) 165:1. Available online at: https://pubmed.ncbi.nlm.nih.gov/15215157/ (accessed January 25, 2022).

PubMed Abstract | Google Scholar

4. Nakazawa G, Finn A V, Joner M, Ladich E, Kutys R, Mont EK, et al. Delayed arterial healing and increased late stent thrombosis at culprit sites after drug-eluting stent placement for acute myocardial infarction patients: an autopsy study. Circulation. (2008) 118:1138–45. doi: 10.1161/CIRCULATIONAHA.107.762047 (accessed January 25, 2022).

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Voisard R, Zellmann S, Müller F, Fahlisch F, von Müller L, Baur R, et al. Sirolimus inhibits key events of restenosis in vitro/ex vivo: Evaluation of the clinical relevance of the data by SI/MPL- and SI/DES-ratio's. BMC Cardiovasc Disord. (2007) 7:1–10. doi: 10.1186/1471-2261-7-15 (accessed January 25, 2022).

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Wang J, Kural MH, Wu J, Leiby KL, Mishra V, Lysyy T, et al. An ex vivo physiologic and hyperplastic vessel culture model to study intra-arterial stent therapies. Biomaterials. (2021) 275:120911. doi: 10.1016/j.biomaterials.2021.120911

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Maurel B, Sarraf C, Bakir F, Chai F, Maton M, Sobocinski J, et al. A new hemodynamic ex vivo model for medical devices assessment. Ann Vasc Surg. (2015) 29:1648–55. doi: 10.1016/j.avsg.2015.06.066

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Conklin BS, Surowiec SM, Lin PH, Chen C. A simple physiologic pulsatile perfusion system for the study of intact vascular tissue. Med Eng Phys. (2000) 22:441–9. doi: 10.1016/s1350-4533(00)00052-7

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Kural MH, Dai G, Niklason LE, Gui L. An ex vivo vessel injury model to study remodeling. Cell Transplant. (2018) 27:1375. Available online at: pmc/articles/PMC6168986/ (accessed January 25, 2022).

PubMed Abstract | Google Scholar

10. Kural MH, Dai G, Niklason LE, Gui L. An ex vivo vessel injury model to study remodeling. Cell Transplant. (2018) 27:1375–89. doi: 10.1177/0963689718792201 (accessed November 12, 2022).

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Meadow W, Rudinsky B, Raju T, John E, Fornell L, Shankararao R. Correlation of flow probe determinations of common carotid artery blood flow and internal carotid artery blood flow with microsphere determinations of cerebral blood flow in piglets. Pediatr Res. (1999) 45:324–30. Available online at: https://www.nature.com/articles/pr199933 (accessed January 25, 2022).

PubMed Abstract | Google Scholar

12. Irace C, Carallo C, De Franceschi MS, Scicchitano F, Milano M, Tripolino C, et al. Human common carotid wall shear stress as a function of age and gender: a 12-year follow-up study. Age. (2012) 34:1553. Available online at: https://pmc/articles/PMC3528365/ (accessed January 25, 2022).

PubMed Abstract | Google Scholar

13. Qiu J, Zheng Y, Hu J, Liao D, Gregersen H, Deng X, et al. Biomechanical regulation of vascular smooth muscle cell functions: from in vitro to in vivo understanding. J R Soc Interface. (2014) 11:6323. Available online at: https://pmc/articles/PMC3836323/ (accessed January 25, 2022).

PubMed Abstract | Google Scholar

14. Shim J, Al-Mashhadi RH, Sørensen CB, Bentzon JF. Large animal models of atherosclerosis–new tools for persistent problems in cardiovascular medicine. J Pathol. (2016) 238:257–66. Available online at: https://pubmed.ncbi.nlm.nih.gov/26414760/ (accessed January 25, 2022).

PubMed Abstract | Google Scholar

15. Piola M, Ruiter M, Vismara R, Mastrullo V, Agrifoglio M, Zanobini M, et al. Full mimicking of coronary hemodynamics for ex-vivo stimulation of human saphenous veins. Ann Biomed Eng. (2016) 45:884–97. doi: 10.1007/s10439-016-1747-7 (accessed April 20, 2022).

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Piola M, Prandi F, Bono N, Soncini M, Penza E, Agrifoglio M, et al. A compact and automated ex vivo vessel culture system for the pulsatile pressure conditioning of human saphenous veins. J Tissue Eng Regen Med. (2016) 10:E204–15. doi: 10.1002/term.1798 (accessed April 10, 2022).

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Sui B, Gao P, Lin Y, Gao B, Liu L, An J. Assessment of wall shear stress in the common carotid artery of healthy subjects using 3.0-tesla magnetic resonance. Acta radiol. (2008) 49:442–9. doi: 10.1080/02841850701877349 (accessed January 25, 2022).

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Jeong SK, Rosenson RS. Shear rate specific blood viscosity and shear stress of carotid artery duplex ultrasonography in patients with lacunar infarction. BMC Neurol. (2013) 13:1–7. doi: 10.1186/1471-2377-13-36 (accessed January 25, 2022).

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Ma X, He Z, Li L, Liu G, Li Q, Yang D, et al. Development and in vivo validation of tissue-engineered, small-diameter vascular grafts from decellularized aortae of fetal pigs and canine vascular endothelial cells. J Cardiothorac Surg. (2017) 12:1–11. doi: 10.1186/s13019-017-0661-x (accessed January 25, 2022).

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Liao D, Lin PH, Yao Q, Chen C. Vascular smooth cell proliferation in perfusion culture of porcine carotid arteries. Biochem Biophys Res Commun. (2008) 372:668. Available online at: https://pmc/articles/PMC2925440/ (accessed January 25, 2022).

PubMed Abstract | Google Scholar

21. Halkos ME, Surowiec SM, Conklin BS, Lin PH, Lumsden AB. Effects of homocysteine on smooth muscle cell proliferation in both cell culture and artery perfusion culture models. J Surg Res. (2000) 88:26–33. doi: 10.1006/jsre.1999.5756

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Sansone R, Baaken M, Horn P, Schuler D, Westenfeld R, Amabile N, et al. Release of endothelial microparticles in patients with arterial hypertension, hypertensive emergencies and catheter-related injury. Atherosclerosis. (2018) 273:67–74. Available online at: https://pubmed.ncbi.nlm.nih.gov/29684662/ (accessed January 25, 2022).

PubMed Abstract | Google Scholar

23. Voisard R, Jensch V, Baur R, Hoher M, Hombach V. A coronary porcine organ culture system for studies of postangioplasty cell proliferation. Coron Artery Dis. (2022) 6:657–65. Available online at: https://pubmed.ncbi.nlm.nih.gov/8574462/ (accessed January 25, 2022).

PubMed Abstract | Google Scholar

24. Voisard R, Eicken J V, Baur R, Gschwend JE, Wenderoth U, Kleinschmidt K, et al. A human arterial organ culture model of post-angioplasty restenosis: results up to 56 days after ballooning. Atherosclerosis. (1999) 144:123–34. doi: 10.1016/s0021-9150(99)00046-5

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Haase D, Otto S, Romeike BFM, Figulla HR, Poerner TC. Development and characterization of an ex vivo arterial long-term proliferation model for restenosis research. ALTEX - Altern to Anim Exp. (2015) 32:307–17. Available online at: https://www.altex.org/index.php/altex/article/view/212 (accessed January 25, 2022).

Google Scholar

26. Mayr U, Zou Y, Zhang Z, Dietrich H, Hu Y, Xu Q. Accelerated arteriosclerosis of vein grafts in inducible NO synthase(-/-) mice is related to decreased endothelial progenitor cell repair. Circ Res. (2006) 3:412–20. Available online at: https://pubmed.ncbi.nlm.nih.gov/16385078/ (accessed January 25, 2022).

PubMed Abstract | Google Scholar

27. Bennett MR, O'Sullivan M. Mechanisms of angioplasty and stent restenosis: implications for design of rational therapy. Pharmacol Ther. (2001) 91:149–66. doi: 10.1016/s0163-7258(01)00153-x

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Tang Z, Wang A, Yuan F, Yan Z, Liu B, Chu JS, et al. Differentiation of multipotent vascular stem cells contributes to vascular diseases. Nat Commun. (2012) 3. Available online at: https://pubmed.ncbi.nlm.nih.gov/22673902/ (accessed January 25, 2022).

PubMed Abstract | Google Scholar

29. Campagnolo P, Katare R, Madeddu P. Realities and misconceptions on the pericytes role in tissue repair. Regen Med. (2018) 13:119–22. doi: 10.2217/rme-2017-0091 (accessed January 25, 2022).

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Perlman H, Maillard L, Krasinski K, Walsh K. Evidence for the rapid onset of apoptosis in medial smooth muscle cells after balloon injury. Circulation. (1997) 95:981–7. doi: 10.1161/01.cir.95.4.981 (accessed January 25, 2022).

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Strauss BH, Chisholm RJ, Keeley FW, Gotlieb AI, Logan RA, Armstrong PW. Extracellular matrix remodeling after balloon angioplasty injury in a rabbit model of restenosis. Circ Res. (1994) 75:650–8. doi: 10.1161/01.RES.75.4.650 (accessed January 25, 2022).

PubMed Abstract | CrossRef Full Text | Google Scholar

32. Houbballah R, Robaldo A, Albadawi H, Titus J, Lamuraglia GM. A novel model of accelerated intimal hyperplasia in the pig iliac artery. Int J Exp Pathol. (2011) 92:422–7. Available online at: https://pubmed.ncbi.nlm.nih.gov/22050434/ (accessed April 20, 2022).

PubMed Abstract | Google Scholar

33. Kong J, Hou J, Ma L, Xing L, Jia H, Liu H, et al. Cutting balloon combined with paclitaxel-eluting balloon for treatment of in-stent restenosis. Arch Cardiovasc Dis. (2013) 106:79. doi: 10.1016/j.acvd.2012.10.004

PubMed Abstract | CrossRef Full Text | Google Scholar

34. Keymel S, Heinen Y, Balzer J, Rassaf T, Kelm M, Lauer T, et al. Characterization of macro-and microvascular function and structure in patients with type 2 diabetes mellitus. Am J Cardiovasc Dis. (2011) 1:68. Available online at: https://pmc/articles/PMC3253507/ (accessed January 25, 2022).

PubMed Abstract | Google Scholar

Keywords: bioreactor and 3D culture, ex vivo culture, organotypic culture model, vascular remodeling, balloon angioplasty, blood vessel

Citation: Matos RS, Maselli D, McVey JH, Heiss C and Campagnolo P (2022) 3D Printed Bioreactor Enabling the Pulsatile Culture of Native and Angioplastied Large Arteries. Front. Cardiovasc. Med. 9:864580. doi: 10.3389/fcvm.2022.864580

Received: 28 January 2022; Accepted: 13 May 2022;
Published: 21 June 2022.

Edited by:

Anthal Smits, Eindhoven University of Technology, Netherlands

Reviewed by:

Monica Soncini, Politecnico di Milano, Italy
Maurizio Pesce, Monzino Cardiology Center (IRCCS), Italy

Copyright © 2022 Matos, Maselli, McVey, Heiss and Campagnolo. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Paola Campagnolo, p.campagnolo@surrey.ac.uk

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.