Subvenous epicardial fat tissue (SEAT), which acts as an electrical insulation, and the venous diameter (VD) both constitute histomorphological challenges for optimal application and lead design in cardiac synchronization therapy (CRT). In this study, we characterized the morphology of human coronary veins to improve the technical design of future CRT systems and to optimize the application of CRT leads. We retrospectively analyzed data from cardiac computed tomography (CT) of 53 patients and did studies of 14 human hearts using the postmortem freeze section technique and micro CT. Morphometric parameters (tributary distances, offspring angles, luminal VD, and SEAT thickness) were assessed. The left posterior ventricular vein (VVSP) had a mean proximal VD of 4.0 ± 1.4 mm, the left marginal vein (VMS) of 3.2 ± 1.5 mm and the anterior interventricular vein (VIA) of 3.9 ± 1.3 mm. More distally (5 cm), VDs decreased to 2.4 ± 0.6 mm, 2.3 ± 0.7 mm, and 2.4 ± 0.6 mm, respectively. In their proximal portions (15 mm), veins possessed mean SEAT thicknesses of 3.2 ± 2.4 (VVSP), 3.4 ± 2.4 mm (VMS), and 4.2 ± 2.8 mm (VIA), respectively. More distally (20–70 mm), mean SEAT thicknesses decreased to alternating low levels of 1.3 ± 1.1 mm (VVSP), 1.7 ± 1.1 mm (VMS), and 4.3 ± 2.6 mm (VIA), respectively. In contrast to the VD, SEAT thicknesses alternated along the further distal vein course and did not display a continuous decrease. Besides the CRT responsiveness of different areas of the LV myocardium, SEAT is a relevant electrophysiological factor in CRT, potentially interfering with sensing and pacing. A sufficient VD is crucial for successful CRT lead placement. Measurements revealed a trend toward greater SEAT thickness for the VIA compared to VVSP and VMS, suggesting a superior signal-to-noise-ratio in VVSP and VMS.
Introduction
In contrast to coronary arteries, which are the main focus of public attention, investigation of and literature on the coronary venous system (CVS) are scarce. Despite this, the clinical importance of the CVS is undisputed (1, 2); this is reflected by several electrophysiological methods employed using the CVS, such as radiofrequency catheter ablation, percutaneous mitral annuloplasty or cardiac resynchronization therapy (CRT) (3, 4). Such interventional procedures rely on detailed knowledge of the coronary venous anatomy to result in satisfactory success rates (5).
As a current standard in CRT, a transvenous quadripolar lead is implanted epicardially in the CVS of the left ventricle (LV) to synchronize the function of both ventricles with a pacemaker (6). CRT is indicated in patients with heart failures who were suffering from a low ejection fraction (LVEF ≤ 35%) or prolonged QRS duration (≥150 ms) (7) and therefore have a higher risk of sudden cardiac death (8).
Successful CRT requires appropriate lead placement, but the individual anatomy of the cardiac veins is often a limiting factor (9). In some cases, CRT nonresponse occurs because of suboptimal lead positioning, e.g., in a myocardial scar with a persistent irregular heart rhythm (10). Subvenous epicardial fat tissue (SEAT), which acts as an electrical insulation, and the venous diameter (VD) both constitute histomorphological challenges for optimal left ventricular lead application and lead design; this has been shown to hold true in studies of the porcine heart (11, 12).
There are a large number of studies dealing with epicardial fat tissue and the associated effects on coronary arteries such as increased risk of ischemic heart disease, cardiac hypertrophy and atrial fibrillation (13). However, morphometric data regarding coronary veins in the human heart is limited, data regarding SEAT is extremely scarce (14).
In clinical routines, various examination modalities are used for the morphological evaluation of the CVS. Computed tomography (CT) is an established technique for preoperative mapping of the cardiac veins to assess the vascular structure (15).
However, even the most modern CT systems are limited in their spatial resolution (16). Therefore, a detailed assessment of small structures such as the SEAT requires an in vitro assessment using, for example, micro computed tomography (mCT) (or X-ray microscopy) or histological approaches (17).
By examining the CVS and the adjacent tissue structures with different modalities, a more detailed anatomical comprehension should be possible. Therefore, the aim of our study was to assess morphometric parameters of the CVS and its adjacent tissues to improve the technical design of future CRT systems and to optimize the application of related pacemaker leads. The different methods in our multimodal approach are discussed and evaluated.
Materials and Methods
This study was approved by the local ethics committee (A 2019-0014) with waiver of informed consent. This study was conducted in accordance with the Declaration of Helsinki on the ethical principles for medical research involving human subjects from October 2013.
Computed Tomography (CT)
We used consecutive cardiac CT scans performed in preparation of an upcoming transcatheter aortic valve implantation (TAVI) at our center between July 2018 and April 2019, employing a 64-detector CT system (Aquilion 64, Canon Medical Systems, Germany). Patients were examined in the supine position and with deep inspiration breath-hold. No premedication was given. The examinations were performed with retrospective ECG gating. Tube voltage was 120 kV. A bolus of 90 mL of iodinated contrast (Imeron® 400MCT, Bracco Imaging Deutschland, Germany) was injected at a flow rate of 4 mL/s, followed by a saline chaser. The scan was started by bolus triggering in the ascending aorta. Transverse images in the end-systolic phase were reconstructed with a slice thickness of 0.5 mm.
Of 63 datasets of consecutive patients, 10 were discarded due to poor image quality or poor opacification of the coronary veins. The remaining 53 datasets were used for the segmentation-based reconstruction of the coronary vein tree and relative tributary positions. Patients had a mean age of 74.4 ± 10.3 years and weighed 80.7 ± 13.9 kg on average. Proportion of women was 39.7%. All subjects were Caucasian.
Detailed measurements of the VD and SEAT thickness were done for 10 consecutive exemplary datasets with 30% females and a mean age of 71.7 ± 10.1 years (see section Reconstruction and measurement of CT- and mCT-datasets).
Dissection and Preparation
For the histological examination by frozen section technique (FST), four hearts were taken from suitable body donors (time of death <36 h before removal). Mean age was 87.8 ± 8.4 years, one was male (details see Supplementary Table 1). After a short rinse (3–5 min) in cold neutral phosphate buffered saline (PBS), the hearts were fixed in 3.7% neutral buffered paraformaldehyde (PFA) and stored at 4°C. Shortly before preparation of the coronary veins, the hearts were transferred to 0.1 M PBS (pH 7.4) and continuously stored at 4°C.
For the examination by means of mCT, 10 previously fixed hearts from body donors (50% male) were prepared. During their lifetimes the body donors (mean age 82.6 ± 10.5 years; details see Supplementary Table 1) had been treated with a pacemaker or an implanted defibrillator or had suffered from coronary heart disease (CHD). The hearts were fixed immediately after removal by immersion in 3.7% PFA or aqueous formaldehyde solution (FA). For further examinations, the hearts were transferred to and stored in 70% ethanol.
Body donors were different individuals than the patients who were CT-scanned.
To identify the individual coronary veins in the epicardial adipose tissue, the coronary vessels were superficially dissected (Figure 1) followed by photographic documentation. The morphology and dimensions of the Thebesian valve, shielding the ostium of the coronary valve, was assessed.
Figure 1. (A,B) PFA-fixed heart from a body donor with exposed coronary veins after preparation and removed epicardial adipose tissue. (A) Left anterolateral view. (B) Right posterolateral view. Specific segments (1–9) of the left marginal vein (VMS) are indicated by dashed lines. LAA, left atrial appendage; LAD, left anterior descending artery; PA, pulmonary artery; VCM, great cardiac vein; VIA, anterior interventricular vein; VVSP, left posterior ventricular vein.
The parameters height, width, and depth of the ventricles were recorded using calipers to assess the respective heart size (mean of LV height and widest lateral and AP-axis). The common clinical parameter of the distance from the apex to the base of the heart was not examined because of the mostly collapsed atria.
Subsequently, the respective main stem of the coronary veins was dissected with preservation of the upper myocardial layer tissue, subvenous adipose tissue and coronary vein wall; these were documented photographically.
Frozen-Section Technique (FST)
For histomorphological imaging based on the frozen-section technique (FST), the fixed tissue segments (n = 19) stored in chilled PBS were crosscut into sub-segments (n = 154) of 1 cm length each along the vascular axis and cryoprotected in 20% saccharose. Subsequently, sub-segments were quick-frozen in isopentane at −80°C for 5 min and then stored at −20°C. The frozen segments were then cut at −17°C and a slice thickness of 50 μm in a cryostat (CM3050 S, Leica Biosystems, Germany). For each segment, several stages (2–6) with > 1 mm distance between sequential stages were cut, mounted and dried. Slices (n = 1,423) were digitized with a Leica DM6 bright field microscope at 10× magnification and using the tiles-merge-application within the device software to assess the whole slice.
Micro Computed Tomography (mCT)
The tissue structure of fixed coronary veins stored in ethanol was examined using mCT. The images were taken with a Phoenix Nanotom-180 (GE sensing and inspection technologies, USA) with 50 kV and 300 μA at a resulting voxel resolution of 23 μm. The *.tiff image stacks (exported with VGStudio Max, Volume Graphics, Germany) were then further analyzed (see section Measurement of cryosections).
Measurements
Reconstruction and Measurement of CT and mCT Datasets
Anonymized CT datasets were exported as Tiff images using MicroDICOM viewer (MicroDicom, Bulgaria). Anisotropic voxel size was at a minimum of 300 μm and a maximum of 782 μm edge length. Subsequently, Tiff images were imported into the 3D reconstruction software Imaris 8.4 (Bitplane, Switzerland). At first, manual segmentation of the coronary vein course, starting at the proximal ostium, was performed, followed by threshold-based generation of a vascular surface model (Figure 2). For 10 datasets, measurements of VD and SEAT were done along the course of the veins by setting orthogonal slices (“oblique slicer”) every 5 mm. For coronary sinus (CS) and great cardiac vein (V. cardiac magna; VCM), VD was measured proximal to the tributary of the posterior interventricular vein (V. cardiac media = V. interventricularis posterior; VIP). For each orthogonal slice, the vein diameter was averaged from the LR and the AP diameter. The thickness of the underlying subvenous fat layers was averaged from three measurements for each slice.
Figure 2. (A–C) Exemplary 3D reconstruction and morphometric analysis of the coronary vein system (CVS) based on CT datasets. (A) Volume rendered CT image of the heart with overlaid surface model of the CVS. (B) Reconstructed surface model of the CVS with overlaid orthogonal measurements every 5 mm. (C) Measurement points and lines along the CVS. CS, coronary sinus; LA, left atrium; LV, left ventricle; RA, right atrium; R1/R2, 1st/2nd ramus; RV, right ventricle; VCM, great cardiac vein; VIA, anterior interventricular vein; VIP, posterior interventricular vein (=middle cardiac vein); VMS, left marginal vein; VVSP, left posterior ventricular vein.
The distances of the main left ventricular tributaries of CS and VCM relative to the ostium of the CS were measured for 53 datasets. Corresponding to the body donor hearts, the parameters height, width and depth of the ventricles were measured in the CT data to assess the respective heart size.
For the examination of the mCT datasets (Figure 3), exported Tiff stacks were loaded into the Imaris software, followed by measurements of VD and SEAT along the course of the veins by setting orthogonal slices (“oblique slicer”) every 5 mm. Since vein walls were collapsed in post-mortem tissue, VD was calculated from the inner vascular perimeter for each orthogonal slice, assuming an ideal circular shape in cross-section.
Figure 3. (A–E) mCT based 3D reconstruction of the anterior interventricular vein (VIA). (A) Volume rendering of the VIA (arrow) with adjacent branches of the left anterior descending artery (LAD; asterisks). (B) Volume from panel (A) with decreased opacity and distance measurement. (C) Orthogonal slices of the VIA for distance measurements every 5 mm. (D) Volume rendering of the VIA (arrow) with branches of the LAD (asterisks), underlying adipose tissue (arrow heads) and ventricular myocardium (MC). (E) Virtual slice with subvenous epicardial adipose tissue (SEAT) and distance measurements between VIA and MC.
Measurement of Cryosections
The digitized frozen sections (Figure 4) were analyzed with the measuring tool “Analyze” of the ImageJ/Fiji software. The inner circumference as well as the distance between the inner vein wall and the interface between left ventricular myocardium and SEAT were measured. Only orthogonal sections were used to measure the inner circumference. Longitudinal or oblique sections were excluded from the calculation of the inner diameter. The SEAT thickness was calculated by the mean value of three measurements in an area with the shortest distance between venous lumen and myocardium. The VD was calculated according to the inner circumference assuming an ideal circular profile. The measurements of each section stage were averaged and then evaluated using the digital images. The values for each vein segment with a respective length of ~1 cm were averaged from the prior averaged cutting-stage values to smooth possible outliers due to inflows in the section plane or oblique sections.
Figure 4. Cryosection (50 μm thick) of left posterior ventricular vein (V. ventriculi sinistri posterior; asterisk) embedded in epicardial adipose tissue and underlying left ventricular myocardium (MC).
Statistics
Significance tests were calculated using GraphPad Prism 8.0.2. Normality distribution was checked using the Shapiro–Wilk test. Mean value comparison was done using the Wilcoxon test for non-parametric data or the ratio t-test for parametric data. Significant differences were accepted for p < 0.05.
Results
Tributary Topography
The distances of the main left ventricular tributaries of CS and VCM relative to the ostium of the CS were measured for 53 CT datasets (Figure 5). The posterior interventricular vein (V. interventricularis posterior, VIP) entered the CS between 1 and 21 mm distal to the ostium of the CS (CSO). The posterior vein of the right ventricle (V. ventriculi sinistri posterior, VVSP) reached the CS/VCM between 14 and 100 mm distal to the CSO. The left marginal vein (V. marginalis sinistra, VMS) entered the CS/VCM between 41 and 113 mm, followed by the branch of the anterior interventricular vein (V. interventricularis anterior, VIA) between 69 and 164 mm. For none of the tributaries did branch distance to the CS ostium correlate with heart size (R2 ≤ 0.18). Double tributaries were detected in the six cases for the VVSP and in one case for the VMS. In three cases no VVSP, in five cases no VMS, and in one case no VIA could be detected in the CT-based volumes—mainly due to insufficient structural resolution in the area concerned.
Figure 5. Mean position ± standard deviation of the venous tributaries as relative distances to the ostium of the coronary sinus (CSO) inspected by CT (n = 53). Values include double tributaries for VVSP (6 × 2) and VMS (1 × 2). Minimum and maximum values in brackets. Values are given in mm. VIA, V. ventricularis anterior; VIP, V. interventricularis posterior; VMS, V. marginalis sinistra; VVSP, V. ventriculi sinistri posterior.
Values for the oblique vein of the left ventricle (V. obliqua atrii sinistri) and the right ventricular small coronary vein (V. cardiaca parva) were not assessed.
The tributary angles θ (between parent and branching daughter vessel) averaged 114.3° for all distal main tributaries (VVSP, VMS, and VIA) of 10 exemplary hearts, with a mean 103 ± 29° for VVSP, 110 ± 23° for VMS and 130 ± 29° for VIA. One VCM was consistently sloped, making angle measurement not feasible, another was strongly meandering into the VIA.
VD and SEAT Thickness
For the morphometric assessment of VD and SEAT thickness, 10 reconstructed coronary venous systems based on clinical CT volumes were analyzed, and tissue samples from 14 hearts derived postmortem from body donors were examined by means of FST and mCT. The VD and SEAT values of the different individuals were compared centimeter for centimeter by means of their relative position within the coronary vein tree (Figure 2).
Venous Diameter (VD) (Figure 6)
At its proximal portion, the CS/VCM had the greatest average VD (13.8 ± 5.2 mm) ranging from 5.1 to 25.0 mm. The VD decreased to a mean of 7.5 ± 1.4 mm in the middle portion (5 cm position) and further decreased to 5.3 ± 1.4 mm more distally (10 cm position). The VVSP possessed a mean VD of 4.0 ± 1.4 mm in the proximal portion, ranging from 1.6 to 6.6 mm. More distally (5 cm position), the VD decreased to 2.4 ± 0.6 mm on average. The VD of the VMS started with 3.2 ± 1.5 mm in the proximal portion, decreasing to 2.3 ± 0.7 mm more distally (5 cm position). The distal-most tributary, the VIA, had a mean VD of 3.9 ± 1.3 mm on average, ranging from 1.4 to 6.5 mm. More distally (5 cm position), the VD decreased to 2.4 ± 0.6 mm. For none of the tributaries did VD correlate with heart size (R2 ≤ 0.18). For a detailed list of the assessed VDs see Supplementary Table 2. Scatter plots with individual data points can be consulted in Supplementary Figure 1.
Figure 6. Averaged venous diameters along the distance of the different coronary veins. Error bars indicate standard deviation. Distal values without error bars are single measurements. CS, coronary sinus; VCM, great cardiac vein; VIA, anterior interventricular vein; VMS, left marginal vein; VVSP, left posterior ventricular vein.
Subvenous Epicardial Adipose Tissue (SEAT) Thickness (Figure 7)
The mean SEAT thickness of the main CS/VCM tributaries was greatest in the proximal portion of the veins. In their proximal 15 mm, the veins had a SEAT layer with a mean thickness of 3.2 ± 2.4 mm for the VVSP (n = 16; min: 0.5 mm, max: 11.1 mm), 3.4 ± 2.4 mm for the VMS (n = 15; min: 0.4 mm, max: 7.9 mm), and 4.2 ± 2.8 mm for the VIA (n = 17; min: 0.3 mm, max: 12.4 mm). More distally (20–70 mm), the SEAT thickness decreased to a mean level of 1.3 ± 1.1 mm for the VVSP (n = 15) and 1.7 ± 1.1 mm for the VMS (n = 9). For the VIA (n = 16), mean seat thickness remained high with a mean level of 4.3 ± 2.6 mm; this decreased only slightly more distally. In a direct comparison of the distal portions, the VIA possessed SEAT at least 1.25 times thicker than the VVSP in 85.7% (n = 12/14; p = 0.0004 for paired differences) and the VMS in 88.9% (n = 8/9; p = 0.001 for paired differences) of the individuals.
Figure 7. Averaged SEAT thickness along the distance of the different coronary veins. Error bars indicate standard deviation. Distal values without error bars are single measurements. CS, coronary sinus; VCM, great cardiac vein; VIA, anterior interventricular vein; VMS, left marginal vein; VVSP, left posterior ventricular vein.
In contrast to the VD, SEAT thicknesses alternated along the further distal vein course and did not decrease continuously. For a detailed list of the assessed SEAT thicknesses see Supplementary Table 3. Scatter plots with individual data points can be consulted in Supplementary Figure 2.
VD and SEAT values for the middle cardiac vein (V. interventricularis posterior), oblique vein of the left ventricle (V. obliqua atrii sinistri), and the right ventricular small coronary vein (V. cardiaca parva) were not assessed.
Thebesian Valve Morphology
The CS guarding Thebesian valves of the body donor hearts examined were largely crescent-shaped. The valve height varied between individuals and averaged 4.3 ± 2.1 mm (min: 1.5 mm, max: 8.0 mm). Valve size did not correlate with ostium size (R2 = 0.0014).
Discussion
The CVS is a rather neglected anatomical structure, at least in the public perception, given the major role of the coronary arteries in coronary heart disease. Nonetheless, the clinical significance of the CVS is not to be underestimated. In contrast to the coronary arteries, the CVS is rather part of the problem solving than of the problem in most clinical cases. Besides catheter ablation of faulty conductive tissue (18) and mitral annuloplasty (19, 20), the implantation of leads for cardiac resynchronization therapy (CRT) is one of the most relevant percutaneous interventions using the CVS. In German hospitals alone, more than 12,000 CRT systems for left ventricular arrhythmia therapy are implanted each year; this represents 12% of all annual cardiac implantable electronic devices and the number is rising (21, 22).
Significance of the Coronary Vein Morphology in CRT
Several aspects in coronary vein morphology are critical for successful lead placement in CRT. The coronary VD limits the maximal diameter of any CRT lead, although dilatation is feasible to some degree (23, 24). On the downside, bigger lead diameters affect venous hemodynamics and hinder blood drainage, thereby increasing the risk of thromboembolic events (25). Moreover, intense contact between the lead and the endothelium promotes thrombo-fibrotic lead encapsulations (26, 27) which can cause complications in the case of the indicated lead explantation (21, 28–30).
Present CRT leads have distal diameters ranging from 4.0 to 5.4 Fr (1.2–1.8 mm) which is rather thin in comparison to right ventricular pacing or defibrillation leads with up to 8.6 Fr (2.8 mm). It has been shown that the VD of the CS in systole is usually greater than in diastole, which can be ascribed to the phasic pattern of the coronary venous flow during the cardiac cycle (3).
As previously shown, close contact between lead electrodes and vein wall lowers the pacing threshold (12). This indicates that, although resulting in venous occlusion, a broader lead fully filling the vein lumen would potentially increase the probability of improved, i.e., lower, thresholds in the preferred pacing region. Via various anastomotic connections, accessory Thebesian veins could compensate the occluded venous drain into the right ventricle (RV) (31).
In a previous study, the CS ostium diameter correlated with heart size (32). We did, however, not observe a correlation between heart size and coronary VD.
Another critical aspect is the topography and prevalence of valves. Although physiologically of rather minor importance (33) and variably present (34–37), valves in the coronary veins are known to interfere with successful lead placement in the retrograde direction (36, 38). They are present either as ostial or parietal valves, most prominently the Thebesian valve of the CS ostium and the Vieussens valve between CS and VCM (39). Although we have occasionally recognized cuts of valve leaflets in our cryo-sections, we did not further quantify their presence due to potential error-proneness related to section clippings.
Besides valves, the venous route can pose another anatomical barrier for the application of the lead. Serpentinous or meandering courses of the vessel, which we have occasionally observed, may hinder the placement of the lead (40). Several data on branching angles of the emanating tributaries exist (41–44); these are in the scale of our measurements. Regarding the angle at which the tributaries emanate from the VCM/CS, an obtuse angle is most advantageous for the application of CRT leads.
Consistent with our findings, previous works have shown that the presence of the coronary veins is variable while the great cardiac and posterior interventricular (=middle cardiac) vein are constantly present (3, 41, 42, 45, 46). This limits the accessibility of the anterolateral, lateral and posterolateral LV regions. The averaged branch positions in our examination are largely in line with those in previous studies (41, 47, 48).
SEAT and CRT Lead Position
Our findings on human hearts are corroborated by previous studies on human and porcine hearts (11, 14): coronary veins are underlaid by epicardial adipose tissue of varying thickness (Table 1). Generally, fat is an effective electrical insulator (51, 52). Due to the insulation properties of SEAT, which affects the sensing and pacing capture threshold (PCT) in CRT, SEAT thickness is suggested to be a relevant anatomical factor for CRT efficiency (11). In an histology-based simulation model, it was shown that SEAT lowers electrical field energy of transvenous electrodes, which in turn increases thresholds for myocardial activation (12). Basically, epicardial fat mass is correlated with body adiposity (53). Due to the sparse data situation, however, a link between SEAT thickness and obesity is unknown. Significantly higher pacing thresholds or sensing problems in obese patients have, to our knowledge, hitherto not been reported.
Consistent with our findings, Anderson et al. (14) found the VIA to have a significantly thicker SEAT than VMS and VVSP. Given this, a superior signal-to-noise ratio in sensing and a lower pacing threshold in CRT can be assumed for VMS and VVSP. Interestingly, a majority of clinical studies suggest better outcomes, e.g., in terms of mortality or reverse LV remodeling, etc., when the CRT lead is placed in the VMS and VVSP [e.g., (54–56)]. Various prospective clinical trials have shown that with quadripolar leads, which represent the current standard of care in CRT, the PCTs for non-apical LV electrodes are found to have “the best myocardial contact” [e.g., (6, 57)]. This is barely surprising regarding the distal anchoring tines in those models, which probably distance the electrode from the tissue. The most crucial aspect for an optimal CRT, however, appears to be pacing of the myocardial area with maximum delay (latest area of activation) (58, 59); in most cases this is the lateral and posterolateral LV region (56), corresponding largely with the coverage of VMS and VVSP. The topographic correlation indicated between veins with lower SEAT thickness and the area of latest activation is presumably coincidental and non-causal. Given this, lead placement and electrode positioning according to certain SEAT conditions are presumably subordinate. Nonetheless, fine adjustment in electrode positioning within the delayed area could optimize threshold values and battery longevity to some extent.
Leadless Pacing in CRT
Current limitations in CRT related to vein morphology and SEAT could be overcome with recent developments of leadless LV pacemaker systems (60–62). Leadless systems generally have the advantage of nullifying the risk of lead failure and thrombo-fibrotic lead encapsulations (TFLE) or lead infections by minimizing implant size and implant-tissue interface equally. In the right ventricle (RV), however, leadless systems bear the risk of an embolic event due to detachment of the implant site (63). Implant detachment in the LV would not result in less severe complications by migrating into the arterial pathway. While early battery depletion is still a challenging risk in leadless RV systems, recent leadless LV systems use external pulse generators with ultrasonic and subcutaneous batteries as a power source (60). Similar to conventional pacemakers, pocket problems, including device-related infections, are post-operative complications with leadless CRT systems (60, 64). The usual retrograde transaortic implantation of the leadless devices into the LV often necessitates general anesthesia (64). A transvenous, transseptal implantation is an alternative and would be less invasive (61).
Recently some authors questioned the necessity of leadless CRT systems (65). However, leadless CRT systems are beneficial for up to half of non-responders to conventional CRT (60, 64).
Although leadless biventricular pacing has been shown to be feasible and effective (61, 62), it is limited to patients without an implantable cardioverter defibrillator (ICD) since endocardial defibrillation has (currently) no leadless alternative. CRT-D patients, however, constitute ~70% of all CRT patients and still rely on a subcutaneous pulse generator and transvenous leads. Thus, a combination of a transvenous ICD with a leadless LV system still retains the risks of pocket infections and TFLE in the afferent veins, in addition to the above-mentioned risks for leadless LV devices. The feasibility of subcutaneous defibrillation (66) and leadless biventricular pacing has yet to be proven.
Strengths and Limitations of the Study
One advantage of our study is the multimodal approach which allowed a comprehensive morphometric characterization of the coronary veins, although not statistically powered due to the limited sample size. Examination of the CT-based volumes provided profound topographic information, whereas resolution limited the small-scale reconstruction and measurement.
Although the CT scans were performed to assess the aortic and arterial morphology, retrograde venography (RVG) would have provided better visualization of the coronary veins. However, image quality in the datasets studied herein was sufficient, and the accessible sample size was bigger than for RVG. FST and mCT in turn allowed high resolution imaging but had their limitations in topographic integrity after dissection. Occasional non-orthogonal slicing in FST could have provided biased absolute measurement values in some cases.
The lengths of the measured vein portion varied between the different individuals, as the distal portion of the veins was not always identifiable during macroscopic preparation or in the reconstructed CT volume. In those cases the measurement was limited to the proximal section along the basal and mid-ventricular region of the heart which are, following previous studies, preferred pacing regions with better outcomes after CRT compared with pacing in the apical region (67).
The selection of CT data from TAVI candidates provides for a certain bias in our study. Although concrete data on TAVI-related new-onset CRT device implantations are still lacking, a number of recent studies have revealed that a significant number of patients with aortic valve stenosis developed persistent left bundle branch block (LBBB) after TAVI due to implantation-related damages of the cardiac conduction system (68, 69), making those patients potential candidates for a new-onset CRT device implantation. Moreover, the indications for TAVI (aortic valve stenosis) and CRT share similar risk factor such as, e.g., age and LV ejection fraction (70, 71). No data exist on whether TAVI or CRT candidates might have coronary vein (diameter, tributary topography) or SEAT morphologies different from other patients. There are no indications that this could be the case.
Conclusions
Knowledge on SEAT thickness and allocation is hitherto scarce, although SEAT is a relevant electrophysiological factor in CRT due to its insulating properties, possibly interfering with both effective sensing and pacing. Besides CRT responsiveness of different areas of the LV myocardium and localization of the latest activated region, consideration of the individual SEAT conditions during electrode positioning might help to decrease undersensing and improve battery longevity. Our measurements and those of previous studies revealed a generally greater average SEAT thickness for the VIA compared to VVSP and VMS. This finding suggests that signal/noise ratios might be superior in VVSP and VMS, thus strengthening the preference for the lateral and posterolateral LV regions as optimal pacing regions for most CRT; the majority of recent clinical studies on CRT lead position indicate this. Leadless CRT systems could avoid problems with vein morphology and SEAT insulation but have their own limitations.
Data Availability Statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.
Ethics Statement
The studies involving human participants were reviewed and approved by Rostock University Medical Center ethics committee (A 2019-0014). Written informed consent for participation was not required for this study in accordance with the national legislation and the institutional requirements.
Author Contributions
JK and FM designed and planned the project. JK analyzed the measurement data and designed concepts for figures. JK, FM, AW, M-AW, JO, and FS interpreted and discussed the data. JK and FS wrote the manuscript. All authors have read and approved the article. All authors contributed to the article and approved the submitted version.
Funding
This work was supported by BIOTRONIK SE & Co KG (Berlin, Germany). JK was funded by the German Federal Ministry of Education and Research (FKZ 03ZZ0908A and FKZ 03ZZ0924A).
Conflict of Interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Acknowledgments
We heartily thank Marko Schulze and Laura Hiepe for assistance with heart dissection and Katharina Nellius for help with vessel preparation. We are grateful to Frauke Winzer and Katharina Nellius for support with FST. Ada Braun is thanked for her help with slice digitalization. Nikita Gorokhov and Jakob Kienzler are thanked for their support with 3D reconstruction. We warmly thank Katharina Nellius and Jakob Kienzler for their great support with measurements. Markus Kipp is appreciated for his support with infrastructure. Thanks also go to Christian Wirkner for the opportunity to use the micro-CT. Two anonymous reviewers are thanked for their constructive comments and suggestions which improved the final version of the manuscript. We thank Ron Beier for proofreading. Lastly, we gratefully appreciate the body donors for contributing to this study by bequeathing their corpses for scientific purposes during their lifetimes.
Supplementary Material
The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcvm.2020.611160/full#supplementary-material
Supplementary Figure 1. Direct comparison of venous diameters along the distance of the individual vein segments of the different studied single coronary vein samples. CS, coronary sinus; VCM, great cardiac vein; VIA, anterior interventricular vein; VMS, left marginal vein; VVSP, left posterior ventricular vein.
Supplementary Figure 2. Direct comparison of the subvenous epicardial adipose tissue (SEAT) thickness along the distance of the individual vein segments of the different studied single coronary vein samples. CS, coronary sinus; VCM, great cardiac vein; VIA, anterior interventricular vein; VMS, left marginal vein; VVSP, left posterior ventricular vein.
References
1. Habib A, Lachman N, Christensen KN, Asirvatham SJ. The anatomy of the coronary sinus venous system for the cardiac electrophysiologist. Europace. (2009) 11(Suppl 5):v15–21. doi: 10.1093/europace/eup270
2. Sun C, Pan Y, Wang H, Li J, Nie P, Wang X, et al. Assessment of the coronary venous system using 256-slice computed tomography. PLoS One. (2014) 9:e104246. doi: 10.1371/journal.pone.0104246
3. Tada H, Kurosaki K, Naito S, Koyama K, Itoi K, Ito S, et al. Three-dimensional visualization of the coronary venous system using multidetector row computed tomography. Circ J. (2005) 69:165–70. doi: 10.1253/circj.69.165
4. Echeverri LF, Herrero MA, Lopez JM, Oleaga G. Early stages of bone fracture healing: formation of a fibrin-collagen scaffold in the fracture hematoma. Bull Math Biol. (2014) 77:156–83. doi: 10.1007/s11538-014-0055-3
5. Mühlenbruch G, Koos R, Wildberger JE, Günther RW, Mahnken AH. Imaging of the cardiac venous system: comparison of MDCT and conventional arigiography. Am J Roentgenol. (2005) 185:1252–7. doi: 10.2214/AJR.04.1231
6. D'Onofrio A, Bertini M, Infusino T, D'Arienzo G, Cipolletta L, Bianchi V, et al. Single- and multi-site pacing strategies for optimal cardiac resynchronization therapy: impact on device longevity and therapy cost. J Interv Card Electrophysiol. (2020). doi: 10.1007/s10840-020-00711-3. [Epub ahead of print].
7. Daubert JC, Saxon L, Adamson PB, Auricchio A, Berger RD, Beshai JF, et al. (2012). EHRA/HRS expert consensus statement on cardiac resynchronization therapy in heart failure: implant and follow-up recommendations and management. Europace. (2012) 14:1236–86. doi: 10.1016/j.hrthm.2012.07.025
8. Dokuni K, Matsumoto K, Tatsumi K, Suto M, Tanaka H, Fukuzawa K, et al. Cardiac resynchronization therapy improves left atrial reservoir function through resynchronization of the left atrium in patients with heart failure with reduced ejection fraction. Int J Cardiovasc Imaging. (2020) 36:1203–12. doi: 10.1007/s10554-020-01813-9
9. Markstad H, Bakos Z, Ostenfeld E, Geijer M, Carlsson M, Borgquist R. Preoperative CT of cardiac veins for planning left ventricular lead placement in cardiac resynchronization therapy. Acta radiol. (2019) 60:859–65. doi: 10.1177/0284185118803796
10. Behar JM, Rajani R, Pourmorteza A, Preston R, Razeghi O, Niederer S, et al. Comprehensive use of cardiac computed tomography to guide left ventricular lead placement in cardiac resynchronization therapy. Hear Rhythm. (2017) 14:1364–72. doi: 10.1016/j.hrthm.2017.04.041
11. Anderson SE, Iaizzo PA. Effects of left ventricular lead positions and coronary venous microanatomy on cardiac pacing parameters. J Electrocardiol. (2010) 43:136–41. doi: 10.1016/j.jelectrocard.2009.08.002
12. Anderson SE, Eggum JH, Iaizzo PA. Modeling of induced electric fields as a function of cardiac anatomy and venous pacing lead location. Cardiovasc Eng Technol. (2011) 2:399–407. doi: 10.1007/s13239-011-0057-3
13. Abel ED, Litwin SE, Sweeney G. Cardiac remodeling in obesity. Physiol Rev. (2008) 88:389–419. doi: 10.1152/physrev.00017.2007
14. Anderson SE, Hill AJ, Iaizzo PA. Microanatomy of human left ventricular coronary veins. Anat Rec. (2009) 292:23–8. doi: 10.1002/ar.20766
15. Mlynarski R, Sosnowski M, Wlodyka A, Kargul W, Tendera M. A user-friendly method of cardiac venous system visualization in 64-slice computed tomography. Pacing Clin Electrophysiol. (2009) 32:323–9. doi: 10.1111/j.1540-8159.2008.02239.x
16. Lin E, Alessio A. What are the basic concepts of temporal, contrast, and spatial resolution in cardiac CT? J Cardiovasc Comput Tomogr. (2009) 3:403–8. doi: 10.1016/j.jcct.2009.07.003
17. Faight E, Verdelis K, Ahearn JM, Shields KJ. 3D MicroCT spatial and temporal characterization of thoracic aorta perivascular adipose tissue and plaque volumes in the ApoE-/- mouse model. Adipocyte. (2018) 7:156–65. doi: 10.1080/21623945.2018.1493900
18. Chen YH, Lin JF. Catheter ablation of idiopathic epicardial ventricular arrhythmias originating from the vicinity of the coronary sinus system. J Cardiovasc Electrophysiol. (2015) 26:1160–7. doi: 10.1111/jce.12756
19. Goldberg SL, Meredith I, Marwick T, Haluska BA, Lipiecki J, Siminiak T, et al. A randomized double-blind trial of an interventional device treatment of functional mitral regurgitation in patients with symptomatic congestive heart failure-Trial design of the REDUCE FMR study. Am Heart J. (2017) 188:167–74. doi: 10.1016/j.ahj.2017.02.032
20. Witte KK Lipiecki J Siminiak T Meredith IT Malkin CJ Goldberg SL . The REDUCE FMR trial: a randomized sham-controlled study of percutaneous mitral annuloplasty in functional mitral regurgitation. JACC Hear Fail. (2019) 7:945–55. doi: 10.1016/j.jchf.2019.06.011
21. Markewitz A, Behrens S, Kolb C, Lemke B, Fröhlig G, Noack F, et al. Annual Report 2017 of the German Pacemaker and Defibrillator-Register. Part 1: Cardiac pacemaker: working group on pacemaker and defibrillators at the IQTIG-Institute for Quality Assurance and Transparency in Healthcare. Herzschrittmacherther Elektrophysiol. (2019) 30:377–88. doi: 10.1007/s00399-019-00647-w
22. Markewitz A, Behrens S, Kolb C, Lemke B, Fröhlig G, Noack F, et al. Annual report 2017 of the German pacemaker- and defibrillator register - Part 2: implantable cardioverter defibrillators (ICD).: Working group on Cardiac pacemaker and implantable cardioverter-defibrillators at the IQTIG - Institute of Quality Assurance and Transparency in Healthcare. Herzschrittmacherther Elektrophysiol. (2019) 30:389–403. doi: 10.1007/s00399-019-00648-9
23. Boey E, Tan ESJ, Yeo WT, Singh D, Lim TW, Kojodjojo P, et al. Coronary venoplasty during cardiac resynchronization therapy device implantations: acute results and clinical outcomes. Hear Rhythm. (2020) 17:736–42. doi: 10.1016/j.hrthm.2019.12.012
24. Yi F, Wu F, Shen M, Wang H, Guo W, Li W, et al. Coronary vein angioplasty to facilitate implantation of left ventricular lead. Europace. (2010) 12:1600–3. doi: 10.1093/europace/euq318
25. Lonyai A, Dubin AM, Feinstein JA, Taylor CA, Shadden SC. New insights into pacemaker lead-induced venous occlusion: simulation-based investigation of alterations in venous biomechanics. Cardiovasc Eng. (2010) 10:84–90. doi: 10.1007/s10558-010-9096-x
26. Keiler J, Schulze M, Sombetzki M, Heller T, Tischer T, Grabow N, et al. Neointimal fibrotic lead encapsulation - clinical challenges and demands for implantable cardiac electronic devices. J Cardiol. (2017) 70:7–17. doi: 10.1016/j.jjcc.2017.01.011
27. Keiler J, Schulze M, Dreger R, Springer A, Öner A, Wree A. Quantitative and qualitative assessment of adhesive thrombo-fibrotic lead encapsulations (TFLE) of pacemaker and ICD leads in arrhythmia patients - a post mortem study. Front Cardiovasc Med. (2020) 7:1–14. doi: 10.3389/fcvm.2020.602179
28. Kasravi B, Tobias S, Barnes MJ, Messenger JC. Coronary sinus lead extraction in the era of cardiac resynchronization therapy: single center experience. Pacing Clin Electrophysiol. (2005) 28:51–3. doi: 10.1111/j.1540-8159.2005.09353.x
29. Hauser RG, Katsiyiannis WT, Gornick CC, Almquist AK, Kallinen LM. Deaths and cardiovascular injuries due to device-assisted implantable cardioverter-defibrillator and pacemaker lead extraction. Europace. (2010) 12:395–401. doi: 10.1093/europace/eup375
30. Zucchelli G, Bongiorni MG, Di Cori A, Soldati E, Solarino G, Fabiani I, et al. Cardiac resynchronization therapy after coronary sinus lead extraction: feasibility and mid-term outcome of transvenous reimplantation in a tertiary referral centre. Europace. (2012) 14:515–21. doi: 10.1093/europace/eur339
31. Ziacchi M, Diemberger I, Martignani C, Boriani G, Biffi M. New left ventricular active fixation lead: the experience of lead extraction. Indian Heart J. (2015) 67:S97–9. doi: 10.1016/j.ihj.2015.10.379
32. Karaca M, Bilge O, Hakan Dinckal M, Ucerler H. The anatomic barriers in the coronary sinus: implications for clinical procedures. J Interv Card Electrophysiol. (2005) 14:89–94. doi: 10.1007/s10840-005-4596-0
33. Maros T, Rácz L, Plugor S, Maros T. Contributions to the morphology of the human coronary sinus. Anat Anz. (1983) 154:133–44.
34. Zawadzki M, Pietrasik A, Pietrasik K, Marchel M, Ciszek B. Endoscopic study of the morphology of Vieussens valve. Clin Anat. (2004) 17:318–21. doi: 10.1002/ca.10229
35. Anh DJ, Eversull CS, Chen HA, Mofrad P, Mourlas NJ, Hardwin Mead R, et al. Characterization of human coronary sinus valves by direct visualization during biventricular pacemaker implantation. Pacing Clin Electrophysiol. (2008) 31:78–82. doi: 10.1111/j.1540-8159.2007.00928.x
36. Anderson SE, Quill JL, Iaizzo PA. Venous valves within left ventricular coronary veins. J Interv Card Electrophysiol. (2008) 23:95–9. doi: 10.1007/s10840-008-9282-6
37. Mak GS, Hill AJ, Moisiuc F, Krishnan SC. Variations in Thebesian valve anatomy and coronary sinus ostium: implications for invasive electrophysiology procedures. Europace. (2009) 11:1188–92. doi: 10.1093/europace/eup179
38. Anderson SE, Hill AJ, Iaizzo PA. Venous valves: unseen obstructions to coronary access. J Interv Card Electrophysiol. (2007) 19:165–6. doi: 10.1007/s10840-007-9161-6
39. Silver MA, Rowley NE. The functional anatomy of the human coronary sinus. Am Heart J. (1988) 115:1080–4. doi: 10.1016/0002-8703(88)90080-4
40. Manlucu J, Yee R. Advanced left-ventricular lead placement techniques for cardiac resynchronization therapy. Curr Opin Cardiol. (2014) 29:53–8. doi: 10.1097/HCO.0000000000000028
41. Spencer JH, Anderson SE, Iaizzo PA. Human coronary venous anatomy: Implications for interventions. J Cardiovasc Transl Res. (2013) 6:208–17. doi: 10.1007/s12265-012-9443-y
42. Spencer J, Fitch E, Iaizzo PA. Anatomical reconstructions of the human cardiac venous system using contrast-computed tomography of perfusion-fixed specimens. J Vis Exp. (2013) 74:1–5. doi: 10.3791/50258
43. Christiaens L, Ardilouze P, Ragot S, Mergy J, Allal J. Prospective evaluation of the anatomy of the coronary venous system using multidetector row computed tomography. Int J Cardiol. (2008) 126:204–8. doi: 10.1016/j.ijcard.2007.03.128
44. Uhm JS, Park JW, Lee H, Kim TH, Youn JC, Joung B, et al. Cardiac vein accessibility according to heart diseases and sex: Implications for cardiac resynchronization therapy. Pacing Clin Electrophysiol. (2016) 39:513–21. doi: 10.1111/pace.12843
45. Marić I, Bobinac D, Ostojić LJ, Petković M, Dujmović M. Tributaries of the human and canine coronary sinus. Cells Tissues Organs. (1996) 156:61–9. doi: 10.1159/000147829
46. Ortale JR, Gabriel EA, Iost C, Márquez CQ. The anatomy of the coronary sinus and its tributaries. Surg Radiol Anat. (2001) 23:15–21. doi: 10.1007/s00276-001-0015-0
47. Manzke R, Binner L, Bornstedt A, Merkle N, Lutz A, Gradinger R, et al. Assessment of the coronary venous system in heart failure patients by blood pool agent enhanced whole-heart MRI. Eur Radiol. (2011) 21:799–806. doi: 10.1007/s00330-010-1961-x
48. Mazur M, Zabówka A, Bolechała F, Kopacz P, Klimek-Piotrowska W, Hołda MK. Variations and angulation of the coronary sinus tributaries: Implications for left ventricular pacing. Pacing Clin Electrophysiol. (2019) 42:423–30. doi: 10.1111/pace.13618
49. Hellerstein HK, Orbison JL. Anatomic variations of the orifice of the human coronary sinus. Circulation. (1951) 3:514–23. doi: 10.1161/01.CIR.3.4.514
50. Hood WB. Regional venous drainage of the human heart. Br Heart J. (1968) 30:105–9. doi: 10.1136/hrt.30.1.105
51. Kaufman W, Johnston FD. The electrical conductivity of the tissues near the heart and its bearing on the distribution of the cardiac action currents. Am Heart J. (1943) 26:42–54. doi: 10.1016/S0002-8703(43)90050-X
52. Faes TJC, Van Der Meij HA, De Munck JC, Heethaar RM. The electric resistivity of human tissues (100 HZ-10 MHZ): a meta- analysis of review studies. Physiol Meas. (1999) 20:R1–10. doi: 10.1088/0967-3334/20/4/201
53. Rabkin SW. Epicardial fat: Properties, function and relationship to obesity. Obes Rev. (2007) 8:253–61. doi: 10.1111/j.1467-789X.2006.00293.x
54. Rossillo A, Verma A, Saad EB, Corrado A, Gasparini G, Marrouche NF, et al. Impact of coronary sinus lead position on biventricular pacing: mortality and echocardiographic evaluation during long-term follow-up. J Cardiovasc Electrophysiol. (2004) 15:1120–5. doi: 10.1046/j.1540-8167.2004.04089.x
55. Dong YX, Powell BD, Asirvatham SJ, Friedman PA, Rea RF, Webster TL, et al. Left ventricular lead position for cardiac resynchronization: a comprehensive cinegraphic, echocardiographic, clinical, and survival analysis. Europace. (2012) 14:1139–47. doi: 10.1093/europace/eus045
56. Oddone D, Solari D, Arena G, Mureddu R, Nangah R, Giorgi D, et al. Optimization of coronary sinus lead placement targeted to right-to-left delay in patients undergoing cardiac resynchronization therapy. Europace. (2019) 21:502–10. doi: 10.1093/europace/euy275
57. Mittal S, Nair D, Padanilam BJ, Ciuffo A, Gupta N, Gallagher P, et al. Performance of anatomically designed quadripolar left ventricular leads: results from the NAVIGATE X4 clinical trial. J Cardiovasc Electrophysiol. (2016) 27:1199–205. doi: 10.1111/jce.13044
58. Murphy RT, Sigurdsson G, Mulamalla S, Agler D, Popovic ZB, Starling RC, et al. Tissue synchronization imaging and optimal left ventricular pacing site in cardiac resynchronization therapy. Am J Cardiol. (2006) 97:1615–21. doi: 10.1016/j.amjcard.2005.12.054
59. Khan FZ, Virdee MS, Gopalan D, Rudd J, Watson T, Fynn SP, et al. Characterization of the suitability of coronary venous anatomy for targeting left ventricular lead placement in patients undergoing cardiac resynchronization therapy. Europace. (2009) 11:1491–5. doi: 10.1093/europace/eup292
60. Reddy VY, Miller MA, Neuzil P, Søgaard P, Butter C, Seifert M, et al. Cardiac resynchronization therapy with wireless left ventricular endocardial pacing: The SELECT-LV study. J Am Coll Cardiol. (2017) 69:2119–29. doi: 10.1016/j.jacc.2017.02.059
61. Montemerlo E, Pozzi M, De Ceglia S, Santini F, Piazzi E, Rovaris G. First-in-man fully leadless transvenous CRT-P with a transseptal implant of WISE-CRT® system and Micra® PM. PACE - Pacing Clin Electrophysiol. (2019) 42:1489–92. doi: 10.1111/pace.13750
62. Funasako M, Neuzil P, Dujka L, Petru J, Sediva L, Simon J, et al. Successful implementation of a totally leadless biventricular pacing approach. Hear Case Rep. (2020) 6:153–7. doi: 10.1016/j.hrcr.2019.12.002
63. Sundaram S, Choe W. The one that got away: A leadless pacemaker embolizes to the lungs. Hear Rhythm. (2016) 13:2316. doi: 10.1016/j.hrthm.2016.09.006
64. Sidhu BS, Porter B, Gould J, Sieniewicz B, Elliott M, Mehta V, et al. Leadless left ventricular endocardial pacing in nonresponders to conventional cardiac resynchronization therapy. Pacing Clin Electrophysiol. (2020) 43:966–73. doi: 10.1111/pace.13926
65. De Maria E, Ziacchi M, Diemberger I, Biffi M. Leadless left ventricular endocardial pacing: A real alternative or a luxury for a few? Cardiovasc Diagn Ther. (2018) 8:530–3. doi: 10.21037/cdt.2018.03.08
66. Kaya E, Rassaf T, Wakili R. Subcutaneous ICD: Current standards and future perspective. IJC Hear Vasc. (2019) 24:100409. doi: 10.1016/j.ijcha.2019.100409
67. Singh JP, Klein HU, Huang DT, Reek S, Kuniss M, Quesada A, et al. Left ventricular lead position and clinical outcome in the multicenter automatic defibrillator implantation trial-cardiac resynchronization therapy (MADIT-CRT) trial. Circulation. (2011) 123:1159–66. doi: 10.1161/CIRCULATIONAHA.110.000646
68. Mangieri A, Montalto C, Pagnesi M, Lanzillo G, Demir O, Testa L, et al. TAVI and post procedural cardiac conduction abnormalities. Front Cardiovasc Med. (2018) 5:85. doi: 10.3389/fcvm.2018.00085
69. Tretter JT, Mori S, Anderson RH, Taylor MD, Ollberding N, Truong V, et al. Anatomical predictors of conduction damage after transcatheter implantation of the aortic valve. Open Hear. (2019) 6:e000972. doi: 10.1136/openhrt-2018-000972
70. Kanwar A, Thaden JJ, Nkomo VT. Management of patients with aortic valve stenosis. Mayo Clin Proc. (2018) 93:488–508. doi: 10.1016/j.mayocp.2018.01.020
Keywords: cardiac anatomy, cardiac veins, epicardial fat, left ventricular pacing, human cardiac histology
Citation: Keiler J, Meinel FG, Ortak J, Weber M-A, Wree A and Streckenbach F (2020) Morphometric Characterization of Human Coronary Veins and Subvenous Epicardial Adipose Tissue—Implications for Cardiac Resynchronization Therapy Leads. Front. Cardiovasc. Med. 7:611160. doi: 10.3389/fcvm.2020.611160
Received: 28 September 2020; Accepted: 16 November 2020;
Published: 08 December 2020.
Edited by:
Matteo Anselmino, University of Turin, ItalyReviewed by:
Alessandro Andreis, University of Turin, ItalyOsmar Antonio Centurion, National University of Asunción, Paraguay
Copyright © 2020 Keiler, Meinel, Ortak, Weber, Wree and Streckenbach. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Jonas Keiler, jonas.keiler@med.uni-rostock.de