Precise Purity Analysis of High-Purity Lanthanum Oxide by Gravimetric Analysis Assisted With Trace Elemental Analysis by Inductively Coupled Plasma Mass Spectrometry

  • National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology, Tsukuba, Japan

Gravimetric analysis was used to determine the purity of high-purity La2O3 by stepwise conversions of the weighing forms. In this study, lanthanum in the sample was converted to La oxalate, La2O3, and La2(SO4)3 to evaluate the stoichiometry of the weighing forms for accurate gravimetric determination. The losses of La in the filtrate, the washing solution of the precipitate, and the mechanical loss of La during filtration were measured using inductively coupled plasma optical emission spectrometry. The weighing forms were evaluated by comparing the observed mass ratio with the theoretical value at each conversion step. The final converted La2(SO4)3 was consistent with the theoretical composition based on the observed mass ratio of La2(SO4)3/La2O3. Additionally, impurities in the high-purity La2O3 were determined by inductively coupled plasma tandem mass spectrometry. The purity of the original La2O3 sample was precisely determined to be 99.977 % ± 0.057% (mass fraction as La2O3; the value following “±” indicates the expanded uncertainty with a coverage factor k = 2.18) by a combination of the gravimetric analysis using the precipitation from the homogeneous solution method and verification of the weighing forms for La compound.

Introduction

The traceability of inorganic analysis is established using high-purity inorganic compounds, the purity of which is accurately determined as primary standards (Vogl et al., 2018). The primary methods of measurement were used for accurate purity determination. The gravimetric analysis is one of the primary methods of measurement recognized by the Consultative Committee for Amount of Substance (CCQM) of the International Committee of Weights and Measures (CIPM) under the Metre Convention (Quin, 1997; Bièvre and Taylor 1997; Milton and Quinn 2001). The gravimetric analysis is a particularly important primary method of measurement in the purity assay analysis of chemical elements to which the coulometric analysis cannot be applied (Vetter et al., 1995; Hioki et al., 1990; Suzuki and Hioki 2006). In the gravimetric analysis, the analyte is selectively and quantitatively precipitated, and the mass of the precipitate or compound converted from the precipitate is weighed. The quantity of the analyte is directly determined by weighed mass. Establishing the exact mass balance of an analyte is essential for accurate and precise gravimetric analysis. Therefore, the weighing forms of the analyte should be evaluated based on stoichiometry. Additionally, the determination of the analyte is not in the weighed precipitate; that is, the mechanical losses (the residue on the beaker), and the quantity of the analyte in the filtrate and the washing solution, are required. However, few investigations have been carried out on the exact mass balance of the gravimetric analysis. For example, Vetter et al. (1995) demonstrated an improvement in its accuracy by combining a BaSO4 gravimetric analysis and an instrumental analytical method for the gravimetric determination of sulfate Vetter et al. (1995). Sulfate dissolved in the filtrate was determined as sulfur using inductively coupled plasma mass spectrometry (ICP-MS). Moreover, sodium and chloride ions occluded in the precipitate were measured using atomic absorption spectrometry and coulometric titration, respectively. Subsequently, the data resulting from the gravimetric analysis were corrected to enhance the accuracy of the gravimetric determination.

Rare earth elements are commonly measured in the fields of industry, geochemistry, and cosmochemistry (Yamamoto et al., 2011; Zhu et al., 2021; Ebihara et al., 2020); so there is a strong demand for the development of primary standard solution (Feng et al., 2019). In this study, gravimetric analysis was used to evaluate the purity of high-purity La2O3 as a raw material for a primary La standard solution based on the stoichiometric evaluation of the stepwise conversions of the weighed La compound. The weighing forms for La determination were evaluated by comparing the observed conversion mass ratios with the theoretical mass ratios for each combination of the weighing forms. First, lanthanum in the sample solution was precipitated as La oxalate by precipitation from a homogeneous solution method using diethyl oxalate (Gordon et al., 1959; Hagiwara 1956). The La-oxalate precipitate was converted by a stepwise method to La2O3 and La2(SO4)3 to evaluate the stoichiometry of the weighing forms. The quantity of La remaining in the filtrate and that dissolved in the solution used to wash the precipitate were determined using inductively coupled plasma optical emission spectrometry (ICP-OES). Finally, the purity of the high-purity La2O3 was determined from the analytical results of the evaluated weighing forms. Furthermore, the impurities in high-purity La2O3 were measured using inductively coupled plasma tandem mass spectrometry (ICP-MS/MS) to correct the gravimetric analytical results.

Materials and Methods

Instruments and Apparatus

A semi-micro balance (Mettler Toledo XP-205) was used to weigh the sample. The balance was calibrated using the Japan Calibration Service System (JCSS). The balances were installed in the balance room controlled at 25°C ± 2°C and 50% ± 10% relative humidity. All weighing was carried out in that balance room. A platinum crucible (nominal volume; 30 ml) was used as a container for ignition of the La-oxalate precipitate, followed by the gravimetric analysis using La2O3 and La2(SO4)3. Water was purified using a Merck Millipore Milli-Q Advantage A10 with a Q-Pod Element purification system. A pH meter (HM-30G, DKK-TOA Co. Tokyo, Japan) was used for pH measurements. An ICP-OES (PerkinElmer Optima 4300DV) was used to determine the unprecipitated La content in the filtrate and the washing solution of the La-oxalate precipitation. An ICP-MS/MS (iCAP TQ, Thermo Fisher Scientific, Bremen, Germany) was used to analyze the metal impurities in the original La2O3 sample.

Reagents

Two hundred fifty grams of high-purity La2O3 (Alfa Aesar, REacton, powder, Lot No. U12A037, informative purity value: 99.999%) was purchased from FUJIFILM Wako Pure Chemicals Corporation (Osaka, Japan). Hydrochloric acid, HNO3, HF, and H2SO4 (Ultrapur-100 grade) were purchased from Kanto Chemicals Co. Inc. (Tokyo, Japan). A 50% diethyl oxalate solution was prepared by dissolving diethyl oxalate (JIS analytical grade, FUJIFILM Wako Pure Chemicals Corporation, Osaka, Japan) in ethanol (JIS analytical grade, FUJIFILM Wako Pure Chemicals Corporation, Osaka, Japan). A 2% oxalic acid solution was prepared by dissolving oxalic acid (JIS analytical grade, FUJIFILM Wako Pure Chemicals Corporation, Osaka, Japan) in water. A NIST SRM 3127a La standard solution was used to determine La in the filtrate, washing solution, and residues resulting from mechanical loss. A Co solution intended as an internal standard for La determination with ICP-OES was prepared by dilution of JCSS Co standard solution. The Co standard solution (Co: 1000 mg dm−3) was purchased from Kanto Chemicals Co. Inc. The NMIJ CRM standard solutions (3635-a Y), NIST SRM standard solutions (3148a Sc, 3110 Ce, 3142a Pr, 3135a Nd, 3147a Sm, 3117a Eu, 3118a Gd, 3157a Tb, 3115a Dy, 3123a Ho, 3116a Er, 3160a Tm, 3166a Yb, and 3130a Lu), and NIST traceable standard solutions of SPEX XSTC-331 (Th) were used for the impurity analysis using ICP-MS/MS.

Measurement Procedure of Metallic Impurities by ICP-MS/MS

Approximately 1.5 g of La2O3 was weighed in a 30-ml quartz crucible for impurity analysis. The sample in the quartz crucible was ignited at 900°C for 2 h using an electric furnace and then cooled in a silica gel desiccator for 1 h in the balance room. Subsequently, 1.2 g of La2O3 was precisely weighed in a 200-ml borosilicate glass beaker. Twenty milliliters of HNO3 (1 + 2) was added to the sample in the beaker. The sample was heated at 120°C on a hot plate until La2O3 was completely dissolved. After complete dissolution, the solution was cooled down to ambient temperature and then diluted as a stock solution to 1 kg with 0.2 mol L−1 HNO3 in a high-density polyethylene bottle (mass fraction of La; ca. 1 g kg−1). Three sub-sample solutions (mass fraction of La of 10 mg kg−1) were prepared for the impurity analysis by diluting the stock solution. Impurities were measured by ICP-MS/MS using the calibration curve method. Calibration standard solutions were prepared with mass fractions of 0 ng kg−1, 50 ng kg−1, 100 ng kg−1, and 250 ng kg−1. Table 1 lists the operating conditions for ICP-MS/MS. The ICP-MS/MS instrument was tuned before measurement in accordance with the manufacturer’s recommendations.

TABLE 1
www.frontiersin.org

TABLE 1. Instrumental conditions of Thermo Fisher iCAP TQ ICP-MS/MS for impurity analysis.

Analytical Procedures of Gravimetric Analysis

The gravimetric analysis procedure comprises three steps: 1) sampling and separation of La using La-oxalate precipitation from homogeneous solution, 2) the gravimetric analysis using La2O3, and 3) conversion of the oxide to sulfate and weighing as the product, La2(SO4)3. A description of the steps is provided in the following text. The buoyancy effect during weighing was corrected. The schematic diagram of the gravimetric procedure is shown in Figure 1.

FIGURE 1
www.frontiersin.org

FIGURE 1. Schematic diagram of the gravimetric analysis procedure for La.

Sampling and Separation of La Using La-Oxalate Precipitation From Homogenous Solution

Before the gravimetric analysis, La2O3 samples in the quartz crucible were ignited at 900°C for 2 h and then allowed to cool in a silica gel desiccator for 1 h in the balance room. The mass of La2O3 samples was approximately 96.5% of the original sample mass in this pre-ignition procedure. After the preignition process, 10 samples of La2O3 of approximately 1 g each were weighed precisely in 200 ml borosilicate glass beakers. Twenty-five milliliters of HCl (1 + 3) was added to the beaker covered with a watch glass. The beaker was heated at 120°C on a hot plate until the La2O3 sample was completely dissolved, after which the solution was cooled to ambient temperature. The watch glass was removed and the solution was gently heated until it evaporated to near dryness. After evaporation, the beaker was cooled to ambient temperature. Then, 60 ml of diluted HCl solution adjusted to pH 1.2 by NH4OH was added to the sample. The sample solution was heated at 120°C on a hot plate until it was completely dissolved. After cooling to ambient temperature, 20 ml of the 50% diethyl oxalate/ethanol solution was added to the sample solution. The sample solution was heated overnight at 120°C on a hot plate for La oxalate precipitation from a homogeneous solution. Diethyl oxalate was slowly producing oxalate anions by heating (Gordon et al., 1959; Hagiwara 1956).

(C2H5)2C2O4+2H2O2C2H5OH+2H++C2O42

After cooling to ambient temperature, the La oxalate precipitate was filtered with a filter paper (No. 5C, diameter; 9 cm, ADVANTEC TOYO, Tokyo, Japan) and washed with water. The filtrate was collected in a 200-ml borosilicate glass beaker. A small amount of the precipitate remaining in the beaker was transferred onto the filter paper with a 2% oxalic acid solution. The La oxalate precipitate on the filter paper was washed with 5 ml of the 2% oxalic acid solution. The filtrate volume was approximately 100 ml at the completion of filtration. The filtered precipitate was then transferred to a 30 ml Pt crucible. After charring the filter paper, the precipitate was ignited at 900°C for 1 h in an electric furnace, allowed to cool for 1 h in a silica gel desiccator in the balance room, and weighed until the mass remained constant. The La oxalate precipitate was converted to the oxide form, La2O3, by this ignition process. The sequence of drying, cooling, and weighing was repeated until the mass remained constant. In this study, the criterion used for a constant mass was a variation of less than 0.1 mg.

Conversion From La2O3 to La2(SO4)3

After a constant mass was reached, the oxide in the Pt crucible was dissolved in 2 ml of H2O and 7 ml of HCl (2 + 1). After the dissolution of the oxide, the solution in the Pt crucible was placed on a hot plate and heated until it became dry. Three milliliters of H2SO4 (1 + 1) were added to the Pt crucible to convert the sulfate from the oxide. The Pt crucible was then heated at 350°C on a hot plate until the excess H2SO4 completely evaporated. After evaporation, the Pt crucible was transferred to an electric furnace. The temperature of the electric furnace was increased stepwise to 450°C for 2 h and 550°C for 4 h. Finally, the sample was heated at 600°C for 1 h to complete the conversion to La2(SO4)3. La2(SO4)3 was cooled in a P2O5 desiccator in the balance room and then weighed. The sequence of heating at 600°C, cooling, and weighing was repeated until constant mass was attained. Thus, in this study, La2(SO4)3, which was converted from La2O3 by igniting the La oxalate precipitate, was heated to 600°C to ensure that volatile compounds were eliminated.

The Measurement of Unprecipitated La by ICP-OES

Unprecipitated La in the filtrate and washing solution was measured using PerkinElmer Optima 4300DV ICP-OES to evaluate the loss of La. The measurement conditions for ICP-OES are listed in Table 2. The analytical results were calculated from the mass of La3O3, La2(SO4)3, and the amount of unprecipitated La in the filtrate, remaining in the washing solution.

TABLE 2
www.frontiersin.org

TABLE 2. Instrumental conditions of PerkinElmer Optima 4300DV ICP-OES.

Results and Discussion

Analytical Results of Impurities

Sixteen elements that possibly coprecipitated on La-oxalate at pH 1.2 solution were measured in the original La2O3 sample by ICP-MS/MS. The precipitation reaction of oxalate in acidic solution at pH 1.2 is selective to rare earth elements and Th (IV). The other elements, such as Fe, Al, Ca, Sr, and Na, form soluble oxalate and are not precipitated in acidic solution at pH 1.2 [Woyski et al., 1964, Hagiwara 1956; Miura and Isogai 2017]. The results of the impurity analysis using ICP-MS/MS are shown in Table 3. The elements for which no quantitative value could be obtained showed an upper limit. In this case, the standard uncertainty was calculated by dividing the upper limit by 2√3. The total mass fraction of impurities in the original La2O3 sample was calculated by summing the measured quantitative values and half values of the upper limit for the elements for which the upper limit was obtained. The obtained total mass fraction of impurities was 1.0 mg kg−1±0.19 mg kg−1 (u; combined standard uncertainty, k = 1).

TABLE 3
www.frontiersin.org

TABLE 3. Analytical results of impurities in the original La2O3 sample.

Selection of Precipitation Reaction

The selection of the precipitation reaction and weighing form is of fundamental importance for a precise gravimetric analysis. Lanthanum fluoride, La-hydroxide, La oxalate, and La-8-quinolinolate complex can also be used for the precipitation of La (Woyski and Harris, 1964). Lanthanum oxalate is a relatively selective precipitate in the aforementioned precipitation reactions because it can be precipitated in a lower pH solution (Hagiwara 1956; Miura and Isogai 2017). Furthermore, the oxalate precipitate obtained using the precipitation from the homogeneous solution method (PFHS method) can produce a crystalline precipitate with a small amount of coprecipitate. In this study, La oxalate was precipitated at pH 1.2 with diethyl oxalate using the PFHS method. After filtration of the La oxalate precipitate, the precipitate was ignited to convert oxide. Then oxide was converted to sulfate for stoichiometric evaluation of the weighing form.

Determination of the Purity of the Original La2O3 Sample Based on the Stoichiometric Evaluation

The results of the gravimetric analysis are presented in Tables 4, 5. The results were corrected by considering the amounts of La remaining in the filtrate and washing solution of the precipitate. In addition, the purity value obtained was corrected by subtracting the total mass fraction of impurities to remove their influence. The combined standard uncertainty of purity was estimated from the sensitivity and linearity of the balance, the repeatability of weighing the oxide form as La2O3, and the ICP-OES determination of unprecipitated La and impurities. However, the uncertainty related to the evaluation of stoichiometry of the weighing form was not taken into consideration.

TABLE 4
www.frontiersin.org

TABLE 4. Analytical results for purity of the original La2O3 sample by the oxide weighing methoda,b.

TABLE 5
www.frontiersin.org

TABLE 5. Analytical results for purity of original La2O3 by the La2(SO4)3 conversion methoda,b.

Table 4 shows the results of the gravimetric analysis using the oxide form of La2O3. The quantities of unprecipitated La are also listed in the table. The quantity of unprecipitated La ranged from 0.037 % to 0.064% under these conditions. This indicates that the PFHS method using diethyl oxalate is quantitatively precipitated La-oxalate in a pH 1.2 acidic solution. It was revealed that 99.9% or more of La could be precipitated by the PFHS method using diethyl oxalate. As is clear from Table 4, the purity of the original La2O3 sample was measured to 99.982 % ± 0.012% (u; combined standard uncertainty, k = 1) by the gravimetric analysis using the oxide form as La2O3.

Table 5 presents the results of the gravimetric analysis using the sulfate form, La2(SO4)3. The combined standard uncertainty of the purity was estimated from the sensitivity and linearity of the balance, the repeatability of weighing the sulfate form as La2(SO4)3, and the ICP-OES determination of unprecipitated La and impurities. However, the uncertainty related to the evaluation of stoichiometry of the weighing form was not taken into consideration.

According to the results, the purity of the original La2O3 sample was measured as 99.973 % ± 0.013% (u; combined standard uncertainty, k = 1). This purity value is consistent with the results obtained by the gravimetric analysis of the oxide form as La2O3 (Table 4). Furthermore, the mass ratios of La2(SO4)3/La2O3 in Table 5 are close to the theoretical value; the mean of the mass ratio of La2(SO4)3/La2O3 to the theoretical value is 0.99991. These facts validated that the compositions of converted La2(SO4)3 and La2O3 were close to the theoretical values. Therefore, the purity values of the original La2O3 sample could be determined by the arithmetic mean of the value obtained from the gravimetric analysis using the oxide form as La2O3 and the gravimetric analysis using the sulfate form as La2(SO4)3.

Calculation of Measurement Uncertainty

The measurement uncertainties derived from this method are listed in Table 6. In this study, the reproducibility of the gravimetric analysis using the oxide form of La2O3 (Table 4) and the gravimetric analysis using the sulfate form of La2(SO4)3 (Table 5), as well as the uncertainty of the stoichiometry of the weighing forms, the uncertainty of weighing the original La2O3 sample, the uncertainty in the impurity analysis, and the uncertainty in molecular weights of La2O3, and La2(SO4)3 were considered as uncertainty components. The uncertainties based on the reproducibility of the gravimetric analysis using the oxide form as La2O3 (Table 4) and the gravimetric analysis using the sulfate form as La2(SO4)3 (Table 5) were determined by calculating the relative standard deviation of the mean values of the gravimetric analysis results and number of samples. The uncertainty of the stoichiometry of the weighing form was calculated from the mean of the absolute value of the difference between the analytical values by the gravimetric analysis using the oxide form as La2O3 and the gravimetric analysis using the sulfate form as La2(SO4)3. The uncertainty regarding the stoichiometry of the weighing forms was calculated by the root mean square of the deviation for each sample obtained by considering the difference between the purity values by the two weighing forms of each sample as a rectangular distribution and dividing it by 3. The uncertainties regarding the reproducibility of the gravimetric analysis and the stoichiometry of weighing forms are shown in Table 6.

TABLE 6
www.frontiersin.org

TABLE 6. Uncertainties about the reproducibility of the gravimetric analysis and the stoichiometry of weighing forms.

The calculation results of measurement uncertainties are listed in Table 7, where the main component of the combined standard uncertainty is the uncertainty of stoichiometry of the weighing form in the gravimetric analysis. The effective degree of freedom calculated by the Welch-Satterthwaite equation was 12. The relative expanded uncertainty of the purity value of lanthanum oxide was evaluated to be 0.057% using coverage factor k = 2.18, which represents a 95% confidence interval with the t-distribution from the calculated effective degrees of freedom.

TABLE 7
www.frontiersin.org

TABLE 7. Contribution of measurement uncertainty in the purity analysis of the original La2O3 sample.

Conclusion

A gravimetric analysis method using a stepwise conversion of weighing forms was used to determine the purity of the high-purity La2O3 sample. Lanthanum in the sample was converted stepwise to La oxalate, La2O3, and La2(SO4)3 to evaluate the accuracy of the weighing forms. In addition, the impurities in the original La2O3 sample, that is, possible coprecipitate on La oxalate, were determined by ICP-MS/MS. The observed mass ratios of the weighed forms were compared to the corresponding theoretical values. The composition of converted La2(SO4)3 was close to the theoretical value because the mass ratio of observed La2(SO4)3/La2O3 to the theoretical value was close to unity (0.99991). The purity of the original La2O3 sample was precisely determined to be 99.977 % ± 0.057% (expanded uncertainty, k = 2.18) using the gravimetric analysis including stoichiometric evaluation.

Data Availability Statement

The original contributions presented in the study are included in the article/Supplementary Material; further inquiries can be directed to the corresponding author.

Author Contributions

All authors listed have made a substantial, direct, and intellectual contribution to this work, and approved it for publication.

Conflict of Interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors, and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

Bièvre, P. D., and Taylor, P. D. P. (1997). Traceability to the SI of Amount-Of-Substance Measurements: From Ignoring to Realizing, a Chemist's View. Metrologia 34, 67–75. doi:10.1088/0026-1394/34/1/10

CrossRef Full Text | Google Scholar

Ebihara, M., Hayano, K., and Shirai, N. (2020). Determination of Trace Rare Earth Elements in Rock Samples Including Meteorites by ICP-MS Coupled with Isotope Dilution and Comparison Methods. Anal. Chim. Acta 1101, 81–89. doi:10.1016/j.aca.2019.12.031

PubMed Abstract | CrossRef Full Text | Google Scholar

Feng, L., Wu, B., Li, H., and Shi, N. (2019). Development of Dy, Ho, Er, Tm, and Sc Mono-Element Standard Solution Certified Reference Materials (GBW08680-08684). Anal. Bioanal. Chem. 411, 1365–1374. doi:10.1007/s00216-018-01567-3

PubMed Abstract | CrossRef Full Text | Google Scholar

Gordon, L., Salutsky, M. L., and Willard, H. (1959). Precipitation from Homogeneous Solution. New York: John Wiley & Sons.

Google Scholar

Hagiwara, Z. (1956). New Quantitative Precipitation of Thorium and Rare Earths from a Homogeneous Solution, and its Application to the Extraction of Some Elements from Ores. J. Soc. Chem. Industry, Jpn. 59, 1378–1383. doi:10.1246/nikkashi1898.59.1378

CrossRef Full Text | Google Scholar

Hioki, A., Watanabe, T., Terajima, K., Fudagawa, N., Kubota, M., and Kawase, A. (1990). Accuracy in Gravimetric Determination of Nitrate and Nitrite as Nitron Nitrate. Anal. Sci. 6, 757–762. doi:10.2116/analsci.6.757

CrossRef Full Text | Google Scholar

Milton, M. J. T., and Quinn, T. J. (2001). Primary Methods for the Measurement of Amount of Substance. Metrologia 38, 289–296. doi:10.1088/0026-1394/38/4/1

CrossRef Full Text | Google Scholar

Miura, T., and Isogai, K. (2017). Determination of Th Isotopes (232Th, 230Th and 228Th) in Titanium Ore Certified Reference Material by Alpha-Ray Spectrometry. Bunseki Kagaku 66, 857–861. doi:10.2116/bunsekikagaku.66.857

CrossRef Full Text | Google Scholar

Quinn, T. J. (1997). Primary Methods of Measurement and Primary Standards. Metrologia 34, 61–65. doi:10.1088/0026-1394/34/1/9

CrossRef Full Text | Google Scholar

Suzuki, T., and Hioki, A. (2006). Gravimetric Analysis of High Purity Tellurium for Purity Evaluation. Anal. Chim. Acta 555, 391–394. doi:10.1016/j.aca.2005.09.028

CrossRef Full Text | Google Scholar

Vetter, T. W., Pratt, K. W., Turk, G. C., Beck, C. M., and Butler, T. A. (1995). Using Instrumental Techniques to Increase the Accuracy of the Gravimetric Determination of Sulfate. Analyst 120, 2025–2032. doi:10.1039/AN9952002025

CrossRef Full Text | Google Scholar

Vogl, J., Kipphardt, H., Richter, S., Bremser, W., del Rocío Arvizu Torres, M., Lara Manzano, J. V., et al. (2018). Establishing Comparability and Compatibility in the Purity Assessment of High Purity Zinc as Demonstrated by the CCQM-P149 Intercomparison. Metrologia 55, 211–221. doi:10.1088/1681-7575/aaa677

CrossRef Full Text | Google Scholar

Woyski, M. M., and Harris, R. E. (1964). “The Rare Earths” in Treatise on Analytical Chemistry, Part II, Analytical Chemistry of the Elements, Section A, Systematic Analytical Chemistry of the Elements. Editors I. M. Kolthoff, and P. J. Elving (John Wiley & Sons), Vol. 8, 1–68.

Google Scholar

Yamamoto, K., Miura, K., Yonetani, A., Sakamoto, H., and Shirasaki, T. (2011). Determination of Rare-Earth Elements in Rare-Earth Magnets by Polarized Zeeman Flame Atomic Absorption Spectrometry. Bunseki Kagaku 60, 873–876. doi:10.2116/bunsekikagaku.60.873

CrossRef Full Text | Google Scholar

Zhu, Y., Nakano, K., Shikamori, Y., and Itoh, A. (2021). Direct Determination of Rare Earth Elements in Natural Water Samples by Inductively Coupled Plasma Tandem Quadrupole Mass Spectrometry with Oxygen as the Reaction Gas for Separating Spectral Interferences. Spectrochim. Acta Part B At. Spectrosc. 179, 106100. doi:10.1016/j.sab.2021.106100

CrossRef Full Text | Google Scholar

Keywords: gravimetric analysis, stoichiometry, ICP-MS, rare earth elements, lanthanum, traceability

Citation: Miura T and Wada A (2022) Precise Purity Analysis of High-Purity Lanthanum Oxide by Gravimetric Analysis Assisted With Trace Elemental Analysis by Inductively Coupled Plasma Mass Spectrometry. Front. Chem. 10:888636. doi: 10.3389/fchem.2022.888636

Received: 03 March 2022; Accepted: 20 May 2022;
Published: 15 July 2022.

Edited by:

Yong-Hyeon Yim, Korea Research Institute of Standards and Science, South Korea

Reviewed by:

Wai Hong Fung, Government of Hong Kong SAR, Hong Kong SAR, China

Copyright © 2022 Miura and Wada. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Tsutomu Miura, t.miura@aist.go.jp

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.