ORIGINAL RESEARCH article

Front. Bioeng. Biotechnol., 01 November 2024
Sec. Biomechanics

Key factors for increased tip-apex distance when treating intertrochanteric fractures with InterTAN nails

Naifeng Zhu,&#x;Naifeng Zhu1,2Lianxia Wu&#x;Lianxia Wu3Xiaofeng HanXiaofeng Han2Zhonglai Qian
Zhonglai Qian1*
  • 1Department of Orthopaedics, The First Affiliated Hospital of Soochow University, Suzhou, China
  • 2Department of Orthopaedics, Renji Hospital, School of Medicine, Shanghai Jiao Tong University, Shanghai, China
  • 3School of Social Development, Population Research Institute, East China Normal University, Shanghai, China

Background: The tip-apex distance is a key factor in predicting implant cut-out after intramedullary fixation for intertrochanteric fractures. This study aimed to evaluate the factors associated with an increased tip-apex distance when treating intertrochanteric fractures using an InterTAN nail.

Methods and Material: We retrospectively analyzed patients with intertrochanteric fractures who underwent InterTAN nail insertion between January 2017 and March 2022 at our hospital. Medical and radiological data were collected. Measurements of preoperative factors and postoperative factors were performed accordingly. Multivariate logistic regression analysis was performed to determine the statistically significant variables of the tip-apex distance.

Results: This study included 102 patients with intertrochanteric fractures. The average tip-apex distance measured 22.4 ± 7.1 mm, ranging from 9.3 to 48.0 mm. The length of the femoral neck on the non-fractured side, lag screw placement in the sagittal plane (center-inferior, superior) and coronal plane (posterior), and the angle between the line of the proximal nail axis and the femoral long axis were identified to be statistically significant factors for the tip-apex distance.

Conclusion: To obtain a shorter tip-apex distance, we recommend a medial trochanteric entry point to minimize the angle between the line of the proximal nail axis and the femoral long axis. Additionally, sufficiently deep central insertion of the lag screw was advised in both the sagittal and coronal planes.

1 Introduction

With the increase in the aging population, the incidence of intertrochanteric fractures is growing globally, and the direct costs associated with this condition are enormous (Adeyemi and Delhougne, 2019; Dong et al., 2023). Owing to the high 1-year mortality rates and other negative consequences, such as disability and depression, these fractures are commonly treated with operative management, which generally favors intramedullary nailing (Bhandari and Swiontkowski, 2017; Fischer et al., 2021; Liu P. et al., 2020). The remarkable benefits of surgical fixation are earlier mobility, accelerated recovery, and significant pain relief (Yu et al., 2015). Despite these advances, the failure of intramedullary nails is not rare and is estimated to be 2%–3% (Johnson et al., 2017). In previous literature, in patients with osteoporosis screw cut-out, cut-through, loss of reduction, or nonunion have been reported, which can result in various complications that may require reoperation (Ibrahim et al., 2019; Liu W. et al., 2020).

Compared with other conventional intramedullary nails, the InterTAN nail uniquely includes two lag screws in an integrated mechanism, which is designed to allow for linear compression and provide resistance to femoral head rotation for intertrochanteric fractures (Date et al., 2020). Additionally, the trapezoidal shape of the proximal end increases the contact surface on the lateral side of the implant for increased support and pressure resistance (Ruecker et al., 2009). Previous biomechanical studies in cadaveric models have provided evidence that the integrated dual-screw construct may provide significant stability and rotational resistance to the femoral head in older patients with osteopenia and unstable intertrochanteric fractures (Santoni et al., 2019; Santoni et al., 2016).

Although several design improvements to the InterTAN nail have been made, the importance of the correct placement of lag screws in the femoral head cannot be overemphasized. Earlier studies have concluded that the lag screws should be positioned within the central quarter of the femoral head, and an increased tip-apex distance was deemed unacceptable (Socci et al., 2017; Yam et al., 2017). Tip-apex distance was the sum of the distance in millimeters measured from the tip of the lag screw to the apex of the femoral head after controlling for magnification. If the screws are not placed optimally, the risk of ‘cut-out’ and ‘cut-through’ remains (Mayor et al., 2024; Socci et al., 2017). The purpose of this study was to investigate the reasons for an increased tip-apex distance of InterTAN nail lag screws, and how to avoid this phenomenon and reduce the risks of a cut-out, and finally establish a foundation for more effective treatment of intertrochanteric fractures in the elderly, which is crucial for enhancing their quality of life.

2 Methods

This retrospective study was conducted at our hospital between January 2017 and March 2022. Patients with intertrochanteric fractures, who underwent InterTAN nail insertion with some mobility before the fracture and had complete medical and radiological records, identified using the hospital information system were included in the study. The exclusion criteria included individuals under the age of 60 years, those with history of fracture or surgery on the contralateral hip, those with pathologic fractures, and those in whom the position of the lag screws could not be identified on radiographs.

Once a patient was extracted from our information system and included, demographic data, such as age and sex, were recorded. Preoperative radiographs (Figure 1) and computed tomography scans were recorded along with the measurement of AO foundation-Orthopedic Trauma Association (AO-OTA) classification (Marsh et al., 2007), integrity of the lateral wall and medial calcar, neck–shaft angle, and length of the femoral neck of the non-fractured side. Using radiographs obtained on the first postoperative day, the neck-shaft angle, tip-apex distance, placement of the lag screw within the femoral head, restoration of the lateral wall and medial calcar, length of the femoral neck, angle between the line of the proximal nail axis and the femoral long axis, and distance between the line of the proximal nail axis and the lateral trochanteric wall were assessed.

Figure 1
www.frontiersin.org

Figure 1. Preoperative radiographs.

Based on the integrity and comminution, the lateral wall and medial calcar were defined as either intact (or two-part and without any translation or with less than one cortical thickness of the lateral or medial cortex), a simple disruption (two-part and more than one cortical thickness of the lateral or medial cortex) or comminuted (with ≥ three fragments).

The neck-shaft angle of the non-fractured side was measured between the line of the femoral neck axis and the femoral long axis (Boese et al., 2016). The vertical length of the femoral neck was measured using a line drawn perpendicular to the femoral long axis through the top of the lesser trochanter, and the distance between the center of the femoral head and the line was considered the vertical length (Marmor et al., 2012).

Lag screws were placed in both the sagittal and coronal planes. The sagittal plane options included center-center, center-inferior, center-superior, superior, and inferior, whereas the coronal plane options included center, posterior, and anterior.

The tip-apex distance (Figure 2), namely, the sum of the distance in millimeters measured from the tip of the lag screw to the apex of the femoral head after controlling for magnification, was calculated using anteroposterior and lateral radiographs (Baumgaertner et al., 1995). Many studies have identified that tip-apex distance was one of the most important factors of cut-out.

Figure 2
www.frontiersin.org

Figure 2. Measurement of the tip apex distance on the anteroposterior and lateral radiographs (mm).

The reduction quality of the lateral wall, as previously described (Hao et al., 2019; Yoon et al., 2020) was defined as optimal reduction (no translation of the lateral cortex), acceptable reduction (translation of the lateral cortical continuity of less than one cortical thickness), or unacceptable reduction (translation of the lateral cortical continuity of more than one cortical thickness) on anteroposterior radiographs.

The reduction quality of the medial calcar was as previously described (Chang et al., 2015). Moreover, three types of medial cortex reductions including anatomical reduction, positive medial cortex support, and negative medial cortex support reduction were identified. The anatomic reduction was the complete cortex-to-cortex contact between the head-neck fragment and the femoral shaft. Positive or negative medial cortex support was defined as the displacement of the head-neck fragment medially or laterally to the superior medial edge of the shaft fragment. Anatomical and positive medial cortex support reductions were considered acceptable.

The distance between the line of the proximal nail axis and the lateral trochanteric wall and the angle between the line of the proximal nail axis and the femoral long axis were measured to evaluate the greater trochanteric entry point of the intramedullary nail (Figure 3).

Figure 3
www.frontiersin.org

Figure 3. Measurement of the great trochanteric entry point, both the angle and distance.

One reviewer completed the medical and radiological data collection and filled out the form. Another reviewer performed checks.

3 Statistical analysis

Multivariate logistic regression analysis was performed to evaluate the statistically significant variables of the tip-apex distance. By considering the variables individually, a backward likelihood ratio test was performed to carefully select the final best model. Statistical significance was set at p < 0.05, and all tests were two-sided. The R software (version 4.1.2) was used for all analyses.

4 Results

In total, 102 patients with intertrochanteric fractures were included in this study. The mean age of the patients was 81 years, with 28 men (27.5%) and 74 women (72.5%). Additionally, 47 (46.1%) patients had injuries on the left side, whereas the remaining 55 (53.9%) sustained injuries on the right side. The average tip-apex distance measured 22.4 ± 7.1 mm, ranging from 9.3 to 48.0 mm, and a mean angle of 2.9°between the line of the proximal nail axis and the femoral long axis was found.

Initially, after adjusting variables, multivariable logistic regression analyses demonstrated that comminuted lateral wall (beta −0.575; 95% confidence interval [CI] −1.02∼-0.12), optimal reduction (beta, 0.437; 95%CI, 0.030∼0.844), and unacceptable reduction of the lateral wall (beta, 0.572; 95%CI, 0.118∼1.02) were independent factors for tip-apex distance. However, these variables were removed because of multicollinearity.

Finally, the length of the femoral neck on the non-fractured side, placement of the lag screw in the sagittal plane (center-inferior, superior), the coronal plane (posterior), and the angle between the line of the proximal nail axis and the femoral long axis were statistically significant factors for the tip-apex distance (Table 1). Furthermore, the first three had a positive effect on the tip-apex distance, whereas the latter had a negative effect.

Table 1
www.frontiersin.org

Table 1. Statistical results of the multivariate model.

No statistically significant differences were observed in the AO-OTA classification or the distance between the line of the proximal nail axis and the lateral trochanteric wall.

5 Discussion

Intertrochanteric femoral fractures are associated with high morbidity and mortality rates. Various intramedullary systems and sliding hip screws have been used in the treatment of intertrochanteric femoral fractures, currently a trend exists towards increased use of intramedullary nail fixation (Grønhaug et al., 2022). Regardless of the implant chosen, poor patient outcomes such as varus collapse and cut-out continue to occur (Sisman et al., 2022). Comparison of the InterTAN nail with integrated compression screws and other available intramedullary nails in the treatment of intertrochanteric fractures displayed a reduced incidence of implant-related failure and re-intervention rates (Quartley et al., 2022). However, the risk of cut-out in unstable trochanteric fractures has been reported in recent studies (Liu P. et al., 2020). Many studies have identified that tip-apex distance, lag screw placement in the head, as well as neck-shaft angles, are important factors in predicting cut-out, with tip-apex distance being the most important factor (Oner et al., 2021; Yam et al., 2017). The tip-apex distance is also considered a radiographic index of implant placement in the head. Moreover, the importance of implant positioning has been recognized in the previous literature (Baumgaertner et al., 1995; Buyukdogan et al., 2017). This study aimed to evaluate the risk factors for high tip-apex distance after internal fixation with the InterTAN nail, which is not well studied to the best of our knowledge. We identified the placement of the lag screw in both the sagittal and coronal planes, the entry point of the InterTAN nail, and the length of the femoral neck on the non-fractured side as factors related to the issue.

Baumgaertner et al. (1995) established the concept of tip-apex distance to describe the position of a lag screw in the femoral head of a sliding hip screw device. The recommended tip-apex distance for this type of device is less than 25 mm, and this standard has been extrapolated to Cephalomedullary nail (Baumgaertner et al., 1995; Lee et al., 2022). Yam et al. (2017) studied 340 patients who underwent Proximal Femoral Nail Antirotation-II and identified that a tip-apex distance >27 mm increased the risk of the cut-out. Coviello et al. (2024) found a tip-apex distance value of 29.5 mm to be associated with a risk of cut-out in double-screw nails, when good fracture reduction is granted. Furthermore, Buyukdogan et al. (2017) study of 118 patients with hip fractures documented no reoperation when the tip-apex distance was less than <21.7 mm, and the overall mean tip-apex distance was 19.7 ± 4.8 mm. However, this value has not been extensively studied for InterTAN nails. As the InterTAN nail comprises two interlocking screws that form a worm gear, we measured the distance between the apex of the femoral head and the tip of the proximal lag screw on both anteroposterior and lateral radiographs. The overall mean tip-apex distance in our study was 22.4 mm, which is consistent with the results of previous studies.

The tip-apex distance was the sum of the distance from the tip of the lag screw to the apex of the femoral head on both anteroposterior and lateral radiographs after controlling for magnification (Baumgaertner et al., 1995). Therefore, the optimal tip-apex distance depends on the precise placement of the lag screw in both the sagittal and coronal planes. The optimal position of the lag screw in the femoral head has been suggested by numerous authors to be either center-center or center-inferior in the anteroposterior view, and centrally located in the lateral view (Konda et al., 2023; Lee et al., 2022). In this study, ideal lag screw placement was determined in 34% (35) of 102 cases, and the mean tip-apex distance was 20.3 mm. However, inadequate lag screw insertion depth can pose challenges in obtaining a suitable tip-apex distance in certain cases. Of the 35 cases, 7 had a tip-apex distance greater than 25 mm, with a maximum value of 35.7 mm. This may be the reason why the center-inferior placement of the lag screw in the sagittal plane found to be statistically significant factors for tip apex distance. 8 of the 35 cases were the center-inferior placement of the lag screw in the sagittal plane and the mean tip apex distance was 25.4 mm with a maximum value of 35.7 mm.

In the current study, we observed a significant correlation between placing the lag screw tip out of the center of the anteroposterior and lateral views and increased tip-apex distance. Other studies have also reported a connection between incorrect screw placement and an elevated risk of cut-out. Yam et al. (2017) reported a high risk of cut-out associated with posterior and superior screw positioning, whereas Şişman et al. (Sisman et al., 2022) discovered that localizations other than central or inferior on anteroposterior radiographs, as well as central on lateral radiographs, increased the risk of the cut-out. However, the aforementioned conclusions differ from those of a biomechanical study conducted by Oner et al. (2021) which identified that retroverted placement was the ideal lag screw placement for type A fractures in the sagittal plane. However, the ideal placement for type C fractures was anteverted. Although a debate is present in clinical and biomechanical studies on whether the lag screw placement should be centrally positioned on the lateral view and either centrally or inferiorly located on the anteroposterior view, it is commonly agreed that anterior placement of the lag screw on the lateral view poses an increased risk of screw cut-out (Konda et al., 2023). We identified that specifically posterior or anterior placement of the lag screws tended towards a high risk of increased tip-apex distance among cohorts in the >25 mm range.

Our study also demonstrated that the entry point of the InterTAN nail correlated with variations in the tip-apex distance. Additionally, positioning the implant correctly from the beginning to achieve sustained fracture reduction and accurate overall coronal and sagittal plane alignment of the implant is important (Lee et al., 2022). In the work done by Pan et al. (2017), they concluded that the medial posterior trochanteric entry point resulted in excellent nail and helical blade placement. In the surgical strategies recommended by Lee et al. (2022), the preferred method was to position the long axis of the implant in alignment with the femoral canal, approximately 5 mm posterior to the tip of the trochanter. In our study, we identified a mean angle of 2.9° between the line of the proximal nail axis and femoral long axis, and increasing this angle resulted in a smaller tip-apex distance.

Another finding of our study was that the length of the femoral neck on the non-fractured side was positively correlated with an increased tip-apex distance. This could also be attributed to the central insertion of the lag screw, which was not sufficiently deep in cases with a longer femoral neck.

This study has certain limitations. First, this was a retrospective study with no estimation of sample size. However, statistical principles consider a sample size of 100 to meet the most stringent rule of thumb (Norman et al., 2012), and we had 102 patients. Second, we used only one type of intramedullary nail: the InterTAN nail with angle-stable lag screws. No other types of intramedullary nails were included. In the biomechanical study by Wang C et al., they introduced a new implant and compared it with proximal femoral nail antirotation (PFNA) and InterTAN for the treatment of intertrochanteric fractures and concluded that this new implant had the best mechanical properties for the treatment of intertrochanteric fractures (Wang et al., 2024). Therefore, in future studies, we will compare factors that influence the increase in tip apex distance between more different intramedullary nails.

6 Conclusion

In conclusion, objective surgical correction using the InterTAN nail should focus on the proper positioning of the lag screw in both the sagittal and coronal planes. Moreover, utilizing a medial trochanteric entry point to minimize the angle between the proximal nail axis and the femoral long axis while ensuring sufficiently deep and central insertion of the lag screw is recommended. This approach facilitates a smaller tip apex distance and more effective treatment of intertrochanteric fractures in the elderly to enhance their quality of life and promote the rehabilitation.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Author contributions

NZ: Writing–original draft. LW: Software, Writing–review and editing. XH: Data curation, Writing–review and editing. ZQ: Supervision, Writing–review and editing.

Funding

The author(s) declare that financial support was received for the research, authorship, and/or publication of this article. National Natural Science Foundation (12171158), Fundamental Research Funds for the Central Universities in China (2020ECNU-HLYT048; 2022QKT001).

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

The reviewer YL declared a shared affiliation with the author XH to the handling editor at the time of review.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

Adeyemi, A., and Delhougne, G. (2019). Incidence and economic burden of intertrochanteric fracture: a medicare claims database analysis. JB JS Open Access 4, e0045. Epub 20190227. doi:10.2106/JBJS.OA.18.00045

PubMed Abstract | CrossRef Full Text | Google Scholar

Baumgaertner, M. R., Curtin, S. L., Lindskog, D. M., and Keggi, J. M. (1995). The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J. Bone Jt. Surg. Am. 77, 1058–1064. doi:10.2106/00004623-199507000-00012

PubMed Abstract | CrossRef Full Text | Google Scholar

Bhandari, M., and Swiontkowski, M. (2017). Management of acute hip fracture. New Engl. J. Med. 377, 2053–2062. doi:10.1056/NEJMcp1611090

PubMed Abstract | CrossRef Full Text | Google Scholar

Boese, C. K., Dargel, J., Oppermann, J., Eysel, P., Scheyerer, M. J., Bredow, J., et al. (2016). The femoral neck-shaft angle on plain radiographs: a systematic review. Skelet. Radiol. 45, 19–28. Epub 20150825. doi:10.1007/s00256-015-2236-z

PubMed Abstract | CrossRef Full Text | Google Scholar

Buyukdogan, K., Caglar, O., Isik, S., Tokgozoglu, M., and Atilla, B. (2017). Risk factors for cut-out of double lag screw fixation in proximal femoral fractures. Injury. 48, 414–418. Epub 20161119. doi:10.1016/j.injury.2016.11.018

PubMed Abstract | CrossRef Full Text | Google Scholar

Chang, S. M., Zhang, Y. Q., Ma, Z., Li, Q., Dargel, J., and Eysel, P. (2015). Fracture reduction with positive medial cortical support: a key element in stability reconstruction for the unstable pertrochanteric hip fractures. Arch. Orthop. Trauma Surg. 135, 811–818. Epub 20150404. doi:10.1007/s00402-015-2206-x

PubMed Abstract | CrossRef Full Text | Google Scholar

Coviello, M., Abate, A., Maccagnano, G., Ippolito, F., Nappi, V., Abbaticchio, A. M., et al. (2024). Tip-apex distance as a risk factor for cut-out in cephalic double-screw nailing of intertrochanteric femur fractures. Bone Jt. Open 5, 457–463. Epub 20240602. doi:10.1302/2633-1462.56.BJO-2023-0163.R1

PubMed Abstract | CrossRef Full Text | Google Scholar

Date, A., Panthula, M., and Bolina, A. (2020). Comparison of clinical and radiological outcomes in intertrochanteric fractures treated with InterTAN nail against conventional cephalomedullary nails: a systematic review. Future Sci. OA 7, Fso668. Epub 20201207. doi:10.2144/fsoa-2020-0182

PubMed Abstract | CrossRef Full Text | Google Scholar

Dong, Y., Zhang, Y., Song, K., Kang, H., Ye, D., and Li, F. (2023). What was the epidemiology and global burden of disease of hip fractures from 1990 to 2019? Results from and additional analysis of the global burden of disease study 2019. Clin. Orthop. Relat. Res. 481, 1209–1220. Epub 20221102. doi:10.1097/corr.0000000000002465

PubMed Abstract | CrossRef Full Text | Google Scholar

Fischer, H., Maleitzke, T., Eder, C., Ahmad, S., Stöckle, U., and Braun, K. F. (2021). Management of proximal femur fractures in the elderly: current concepts and treatment options. Eur. J. Med. Res. 26, 86. Epub 20210804. doi:10.1186/s40001-021-00556-0

PubMed Abstract | CrossRef Full Text | Google Scholar

Grønhaug, K. M. L., Dybvik, E., Matre, K., Östman, B., and Gjertsen, J. E. (2022). Intramedullary nail versus sliding hip screw for stable and unstable trochanteric and subtrochanteric fractures: 17,341 patients from the Norwegian Hip Fracture Register. Bone Jt. J. 104-b, 274–282. doi:10.1302/0301-620X.104B2.BJJ-2021-1078.R1

CrossRef Full Text | Google Scholar

Hao, Y., Zhang, Z., Zhou, F., Ji, H., Tian, Y., Guo, Y., et al. (2019). Risk factors for implant failure in reverse oblique and transverse intertrochanteric fractures treated with proximal femoral nail antirotation (PFNA). J. Orthop. Surg. Res. 14, 350. Epub 20191108. doi:10.1186/s13018-019-1414-4

PubMed Abstract | CrossRef Full Text | Google Scholar

Ibrahim, I., Appleton, P. T., Wixted, J. J., DeAngelis, J. P., and Rodriguez, E. K. (2019). Implant cut-out following cephalomedullary nailing of intertrochanteric femur fractures: are helical blades to blame? Injury. 50, 926–930. Epub 20190220. doi:10.1016/j.injury.2019.02.015

PubMed Abstract | CrossRef Full Text | Google Scholar

Johnson, N. A., Uzoigwe, C., Venkatesan, M., Burgula, V., Kulkarni, A., Davison, J. N., et al. (2017). Risk factors for intramedullary nail breakage in proximal femoral fractures: a 10-year retrospective review. Ann. R. Coll. Surg. Engl. 99, 145–150. Epub 20160923. doi:10.1308/rcsann.2016.0297

PubMed Abstract | CrossRef Full Text | Google Scholar

Konda, S. R., Solasz, S., Derken, M., Ganta, A., and Egol, K. A. (2023). Improving cephalad lag screw placement in the femoral head during cephalomedullary nailing using a novel augmented reality system. J. Orthop. Trauma 37, 130–134. doi:10.1097/BOT.0000000000002504

PubMed Abstract | CrossRef Full Text | Google Scholar

Lee, C., Kelley, B., Gurbani, A., and Stavrakis, A. I. (2022). Strategies for pertrochanteric fracture reduction and intramedullary nail placement: technical tips and tricks. J. Am. Acad. Orthop. Surg. 30, 867–878. Epub 20220525. doi:10.5435/JAAOS-D-21-01007

PubMed Abstract | CrossRef Full Text | Google Scholar

Liu, P., Jin, D., Zhang, C., and Gao, Y. (2020). Revision surgery due to failed internal fixation of intertrochanteric femoral fracture: current state-of-the-art. BMC Musculoskel Disord. 21, 573. Epub 20200822. doi:10.1186/s12891-020-03593-8

PubMed Abstract | CrossRef Full Text | Google Scholar

Liu W., W., Liu, J., and Ji, G. (2020). Comparison of clinical outcomes with proximal femoral nail anti-rotation versus InterTAN nail for intertrochanteric femoral fractures: a meta-analysis. J. Orthop. Surg. Res. 15, 500. Epub 20201029. doi:10.1186/s13018-020-02031-8

PubMed Abstract | CrossRef Full Text | Google Scholar

Marmor, M., Nystuen, C., Ehemer, N., McClellan, R. T., and Matityahu, A. (2012). Accuracy of in situ neck-shaft angle and shortening measurements of the anatomically reduced, varus malreduced and shortened proximal femur: can we believe what we see on the postoperative films? Injury 43, 846–849. doi:10.1016/j.injury.2011.10.010

PubMed Abstract | CrossRef Full Text | Google Scholar

Marsh, J. L., Slongo, T. F., Agel, J., Broderick, J. S., Creevey, W., DeCoster, T. A., et al. (2007). Fracture and dislocation classification compendium - 2007: orthopaedic Trauma Association classification, database and outcomes committee. J. Orthop. Trauma 21, S1–S6. doi:10.1097/00005131-200711101-00001

PubMed Abstract | CrossRef Full Text | Google Scholar

Mayor, J., Birgel, V., Clausen, J. D., Aktas, G., Sehmisch, S., Einfeldt, A. K., et al. (2024). Lessons learned from biomechanical studies on cephalomedullary nails for the management of intertrochanteric fractures. A scoping review. Injury 55, 111180. Epub 20231108. doi:10.1016/j.injury.2023.111180

PubMed Abstract | CrossRef Full Text | Google Scholar

Norman, G., Monteiro, S., and Salama, S. (2012). Sample size calculations: should the emperor's clothes be off the peg or made to measure? Bmj 345, e5278. Epub 20120823. doi:10.1136/bmj.e5278

PubMed Abstract | CrossRef Full Text | Google Scholar

Oner, K., Durusoy, S., and Ozer, A. (2021). Is the fracture morphology in the sagittal plane important in determining the ideal placement of the lag screw in intertrochanteric femoral fractures? ideal lag screw placement in intertrochanteric fractures in the sagittal plane. Injury 52, 562–568. Epub 20201006. doi:10.1016/j.injury.2020.10.023

PubMed Abstract | CrossRef Full Text | Google Scholar

Pan, S., Liu, X. H., Feng, T., Kang, H. J., Tian, Z. G., and Lou, C. G. (2017). Influence of different great trochanteric entry points on the outcome of intertrochanteric fractures: a retrospective cohort study. BMC Musculoskel Disord. 18, 107. Epub 20170314. doi:10.1186/s12891-017-1472-x

PubMed Abstract | CrossRef Full Text | Google Scholar

Quartley, M., Chloros, G., Papakostidis, K., Saunders, C., and Giannoudis, P. V. (2022). Stabilisation of AO OTA 31-A unstable proximal femoral fractures: does the choice of intramedullary nail affect the incidence of post-operative complications? A systematic literature review and meta-analysis. Injury 53, 827–840. Epub 20220203. doi:10.1016/j.injury.2022.02.002

PubMed Abstract | CrossRef Full Text | Google Scholar

Ruecker, A. H., Rupprecht, M., Gruber, M., Gebauer, M., Barvencik, F., Briem, D., et al. (2009). The treatment of intertrochanteric fractures: results using an intramedullary nail with integrated cephalocervical screws and linear compression. J. Orthop. Trauma. 23, 22–30. doi:10.1097/bot.0b013e31819211b2

PubMed Abstract | CrossRef Full Text | Google Scholar

Santoni, B. G., Diaz, M. A., Stoops, T. K., Lannon, S., Ali, A., and Sanders, R. W. (2019). Biomechanical investigation of an integrated 2-screw cephalomedullary nail versus a sliding hip screw in unstable intertrochanteric fractures. J. Orthop. Trauma. 33, 82–87. doi:10.1097/BOT.0000000000001351

PubMed Abstract | CrossRef Full Text | Google Scholar

Santoni, B. G., Nayak, A. N., Cooper, S. A., Smithson, I. R., Cox, J. L., Marberry, S. T., et al. (2016). Comparison of femoral head rotation and varus collapse between a single lag screw and integrated dual screw intertrochanteric hip fracture fixation device using a cadaveric hemi-pelvis biomechanical model. J. Orthop. Trauma. 30, 164–169. doi:10.1097/BOT.0000000000000552

PubMed Abstract | CrossRef Full Text | Google Scholar

Sisman, A., Avci, O., Cepni, S. K., Batar, S., and Polat, O. (2022). Risk factors for cut-out in intertrochanteric fractures treated with proximal femoral nail of double proximal screw design. J. Clin. Orthop. Trauma 28, 101832. Epub 20220317. doi:10.1016/j.jcot.2022.101832

PubMed Abstract | CrossRef Full Text | Google Scholar

Socci, A. R., Casemyr, N. E., Leslie, M. P., and Baumgaertner, M. R. (2017). Implant options for the treatment of intertrochanteric fractures of the hip: rationale, evidence, and recommendations. Bone Jt. J. 99-b, 128–133. doi:10.1302/0301-620X.99B1.BJJ-2016-0134.R1

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, C., Duan, N., Li, Z., Ma, T., Zhang, K., Wang, Q., et al. (2024). Biomechanical evaluation of a new intramedullary nail compared with proximal femoral nail antirotation and InterTAN for the management of femoral intertrochanteric fractures. Front. Bioeng. Biotechnol. 12, 1353677. Epub 20240223. doi:10.3389/fbioe.2024.1353677

PubMed Abstract | CrossRef Full Text | Google Scholar

Yam, M., Chawla, A., and Kwek, E. (2017). Rewriting the tip apex distance for the proximal femoral nail anti-rotation. Injury 48, 1843–1847. Epub 20170701. doi:10.1016/j.injury.2017.06.020

PubMed Abstract | CrossRef Full Text | Google Scholar

Yoon, Y. C., Oh, C. W., Sim, J. A., and Oh, J. K. (2020). Intraoperative assessment of reduction quality during nail fixation of intertrochanteric fractures. Injury 51, 400–406. Epub 20191031. doi:10.1016/j.injury.2019.10.087

PubMed Abstract | CrossRef Full Text | Google Scholar

Yu, J., Zhang, C., Li, L., Kwong, J. S., Xue, L., Zeng, X., et al. (2015). Internal fixation treatments for intertrochanteric fracture: a systematic review and meta-analysis of randomized evidence. Sci. Rep. 5, 18195. Epub 20151211. doi:10.1038/srep18195

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: tip-apex distance, cut-out, implant failure, InterTAN, intertrochanteric fracture

Citation: Zhu N, Wu L, Han X and Qian Z (2024) Key factors for increased tip-apex distance when treating intertrochanteric fractures with InterTAN nails. Front. Bioeng. Biotechnol. 12:1426307. doi: 10.3389/fbioe.2024.1426307

Received: 03 July 2024; Accepted: 16 October 2024;
Published: 01 November 2024.

Edited by:

Dejian Li, Fudan University Pudong Medical Center, China

Reviewed by:

Rongguang Ao, Seventh People’s Hospital of Shanghai, China
Ye Lu, Shanghai Jiao Tong University, China
Yinghua Li, Fudan University, China
Sun Jiao, A Suzhou Xiangcheng People’s Hospital, China, in collaboration with reviewer

Copyright © 2024 Zhu, Wu, Han and Qian. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Zhonglai Qian, MTU5Nzk5MTQyNUBxcS5jb20=

These authors have contributed equally to this work and share first authorship

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.