- 1Beijing Key Laboratory of Fundamental Research on Biomechanics in Clinical Application, Capital Medical University, Beijing, China
- 2School of Biomedical Engineering, Capital Medical University, Beijing, China
- 3Beijing Tongren Eye Center, Beijing Tongren Hospital, Beijing Ophthalmology and Visual Sciences Key Laboratory, Beijing Institute of Ophthalmology, Capital Medical University, Beijing, China
- 4Beijing Advanced Innovation Center for Big Data-Based Precision Medicine, Beijing Tongren Hospital, Beihang University and Capital Medical University, Beijing, China
- 5Beijing Tongren Eye Center, Beijing Tongren Hospital, Beijing Ophthalmology and Visual Sciences Key Laboratory, Capital Medical University, Beijing, China
Purpose: The aim of this study was to provide a method to determine corneal nonlinear viscoelastic properties based on the output data of corneal visualization Scheimpflug technology (Corvis ST).
Methods: The Corvis ST data from 18 eyes of 12 healthy humans were collected. Based on the air-puff pressure and the corneal displacement from the Corvis ST test of normal human eyes, the work done by the air-puff attaining the whole corneal displacement was obtained. By applying a visco-hyperelastic strain energy density function of the cornea, in which the first-order Prony relaxation function and the first-order Ogden strain energy were employed, the corneal strain energy during the Corvis ST test was calculated. Then the work done by the air-puff attaining the whole corneal displacement was completely regarded as the strain energy of the cornea. The identification of the nonlinear viscoelastic parameters was carried out by optimizing the sum of difference squares of the work and the strain energy using the genetic algorithm.
Results: The visco-hyperelastic model gave a good fit to the data of corneal strain energy with time during the Corvis ST test (R2 > 0.95). The determined Ogden model parameter μ ranged from 0.42 to 0.74 MPa, and α ranged from 32.76 to 55.63. The parameters A and τ in the first-order Prony function were 0.09–0.36 and 1.21–1.95 ms, respectively.
Conclusion: It is feasible to determine the corneal nonlinear viscoelastic properties based on the corneal contour information and air-puff pressure of the Corvis ST test.
Introduction
Cornea is a kind of biological soft tissue with nonlinear viscoelasticity. The biomechanical property of cornea has been considered as a helpful index in the diagnosis of keratoconus (Wang et al., 2015), selection of refractive surgery (Dupps and Wilson, 2006; Shen et al., 2014), screening before refractive surgery (Piñero and Alcón, 2014), measurement of intraocular pressure (IOP) (Brown et al., 2018), and evaluation of the curative effect of corneal cross-linking surgery (Vinciguerra et al., 2017; Sedaghat et al., 2018).
At present, corneal visualization Scheimpflug technology (Corvis ST) has been used in the diagnosis of keratoconus (Chan et al., 2018) and glaucoma (Wang et al., 2015). Corneal viscoelasticity is closely related to the progressive keratoconus (Fraldi et al., 2011; Kling et al., 2014). However, the mechanical meaning of the obtained dynamic corneal response (DCR) parameter output by Corvis ST has not been clear exactly. The relationship between DCR parameters and classical biomechanical properties has not been established either. More importantly, the DCR parameters from Corvis ST could not be directly used by numerical methods, such as finite element methods, to predict corneal deformation after refractive surgery. Therefore, many researchers have paid great attention to attempt to have a comprehensive understanding of corneal biomechanical properties based on the tests of cornea in vivo. By applying inverse finite element method and corneal deformation data from Corvis ST, investigators studied the nonlinear mechanical characteristics of the cornea (Lago et al., 2015; Sinha Roy et al., 2015; Bekesi et al., 2016), in which a complicated constitutive equation and a fine geometry model were applied. Our previous study (Qin et al., 2019a) applied a thin spherical shell model to identify corneal elastic modulus based on Corvis ST data. These studies have demonstrated the feasibility of using the output data of Corvis ST to acquire corneal biomechanical parameters, such as elastic modulus. However, a full description of corneal biomechanical properties based on in vivo tests has been still an issue of great concern to the researchers.
This study will establish a method to identify corneal nonlinear viscoelastic mechanical properties based on the output data of Corvis ST. We shall apply a visco-hyperelastic model to fit the corneal deformation data extracted from the image information during a Corvis ST test. It is expected that the method of determination of corneal nonlinear viscoelastic parameters can be used in the optimization design of corneal refractive surgery scheme and the prediction of corneal morphology after refractive surgery.
Materials and methods
Subjects
This study included 18 eyes of 12 healthy subjects. The institutional review board of Capital Medical University approved this study, and all participants signed an informed consent form in accordance with the tenets of the Declaration of Helsinki.
Subjects met the following criteria: age 18–45 years. All patients had no history of corneal, ocular surgery, trauma, or systemic diseases that might affect the eye and had abandoned soft contact lenses or rigid contact lenses at least 1 month and no contact lens utilization within 2 weeks before the examination.
Ocular Examination
All subjects underwent a thorough tomographic measurement using Corvis ST (Oculus; Optikgeräte GmbH, Wetzlar, Germany, software version: 1.6b2042) and Pentacam (Oculus; Optikgeräte GmbH, Wetzlar, Germany). The DCR parameters in the analysis mainly included deflection amplitude at the highest concavity (HCDA), central concave curvature at the highest concavity (HCR), peak distance at the highest concavity (PD), deflection amplitude ratio maximal 2 mm (DAR2), stiffness parameter at the first applanation (SPA1), corneal stress–strain index (SSI), corneal biomechanical index (CBI), and biomechanically corrected IOP (bIOP).
Method to Evaluate Corneal Nonlinear Viscoelasticity
Visco-Hyperelastic Model
Considering the time dependence of corneal stress responding to any loads, a visco-hyperelastic model based on strain energy function has been shown to describe corneal nonlinear and viscoelastic properties (Liu et al., 2020a; Su et al., 2015; Whitford et al., 2018). As one of well-known isotropic strain energy functions, the Ogden model can better describe the nonlinear stress–strain relationship of the cornea (Bao et al., 2017; Bao et al., 2018; Joda et al., 2016; Lago et al., 2015; Maklad et al., 2019). Therefore, we adopted the Ogden model in the visco-hyperelastic constitutive model. Also, the widely used Prony series model (Huang et al., 2017; Ramzanpour et al., 2020) as relaxation function was introduced in the visco-hyperelastic model. The cornea was regarded as a nonlinear incompressible viscoelastic material. The visco-hyperelastic model (Snedeker et al., 2005; Taylor et al., 2009) (Eq. 1) was applied, in which the strain energy density U was given by the Ogden model (Eq. 2), and the stress relaxation function G was given by the first-order Prony model (Eq. 3).
In Eq. 2, N, μi, and αi are material parameters, and λj (j = 1, 2, 3) is the stretch ratio of materials in three main directions. Taking into consideration that the first-order Ogden model can better describe the mechanical properties of cornea (Bao et al., 2017; Bao et al., 2018), in this study, we let N = 1. In Eq. 3, A and τ are the stress relaxation parameters,
It is assumed that the three main directions of cornea are circumferential, radial, and perpendicular (or thick) directions. It is further assumed that the circumferential elongation ratio is equal to the elongation ratio of the radial arc length under air-puff. It follows that
According to the corneal contour imagines obtained from Corvis ST, the overall corneal strain under the air-puff at each time point was calculated. The stretch ratio was set by Eq. 4.
where L1 and L0 are the arc length of the anterior corneal surface during deformation and at the initial state, respectively. L1 and L0 can be obtained from the corneal anterior surface contour by image processing.
The Work Done by Air-Puff During the Corvis ST Test
We employed cylindrical coordinates to describe the work done by air-puff attaining the whole corneal displacement during the Corvis ST test. According to literatures (Simonini et al., 2016; Eliasy et al., 2019), the pressure of the air-puff, P(r, t), at the point with polar coordinate r from the corneal apex and time t was obtained (Figure 1). We let S(r, t) be the corneal vertical displacement of the point r from the corneal apex on the anterior surface of the cornea at time t. Then for an element
where Rm is the radius of the maximum deformed area of the corneal anterior surface by the air-puff; Rm can be estimated by the data of the corneal anterior surface contour (Figure 2).
The Deformation Data of the Cornea During the Corvis ST Test
In this study, according to the corneal contour images obtained from Corvis ST, we extracted the outer contour of the cornea (Figure 3) according to the OTSU threshold method. L1 and L0 in Eq. 4 were obtained from the extracted corneal anterior surface contour, where
FIGURE 3. Original corneal image (an) (A), the corresponding processed image (bn) (B), extracted corneal anterior surface contour (cn) (C), and the position of the extracted corneal anterior surface contour in the original image (dn) (D), where n = 0 (initial state), A1 (first applanation), HC (highest concavity), A2 (second applanation), end.
FIGURE 4. Change of apex total displacement, apex pure displacement, and eye globe movement with time, indicated by blue dotted line, red solid line, and green dotted line, respectively.
In order to complete the calculation of the work done by air-puff attaining the whole corneal displacement based on Eq. 5, we need to fit these deformation data sets. We applied
to fit the vertical displacement of the anterior surface of the cornea at point (r, t), where 0 ≤ t ≤ 30 (ms), 0 ≤ r ≤ Rm (mm). According to the extracted corneal anterior surface contour, the radius of the maximum corneal deformation area is 4.29 mm, i.e., Rm is 4.29 mm. Parameter c is the time when the corneal apex in the corneal vertical displacement time curve is at the maximum position which is given from Corvis ST data.
The Determination of the Material Parameters
The cornea was regarded as a nonlinear incompressible viscoelastic material, and the Corvis ST test process was assumed to be a loading and unloading process with constant strain rate. The work done by an air-puff attaining the whole corneal displacement was completely regarded as the strain energy. Based on Eqs. 1, 2a, 3–6, the material parameters could be determined by minimizing the objective function shown in Eq. 7.
where
The material parameters needed to identify were
Statistical analysis
All analyses were performed using SPSS (version 23.0, IBM Corporation, Armonk, NY, USA). The data were tested for normality of distribution using the Shapiro–Wilk test and expressed as mean ± SD or median (interquartile range, IQR). Power of the tests was calculated using the data of the determined parameter μ.
Results
This study included 18 eyes of 12 participants (9 males, 3 females); the baseline information of participants is shown in Table 1.
DCR parameters from Corvis ST
The values of DCR parameters in the eyes are shown in Table 2. The SPA1 and SSI were normal distribution, but HCDA, HCR, PD, DAR2, and CBI were non-normal distribution.
Anterior corneal surface displacement
For each subject’s 140-frame Corvis ST image data set, we extracted the contour data of the anterior corneal surface and then obtained the data set
FIGURE 5. Vertical displacements of the anterior surface of the cornea at each position and test time point (A), spatial distribution (B), and temporal variation (C). t0, tA1, tHC, tA2, and tend represent the time at initial state, first applanation, highest concavity, second applanation, and end state, respectively;
In order to verify the effectiveness of the displacement results from image extractions, the data set
FIGURE 6. Deflection amplitude and arc length variation of corneal anterior surface vertices from image extractions with time: deflection amplitude of all eyes (A), one eye (B), and arc length variation (C). The red solid line is the image extraction result; blue dotted line is the deflection amplitude from Corvis ST.
S(r, t) (Eq. 6) was used to fit the vertical displacement of the anterior corneal surface at different distances from the corneal apex at different times, and the goodness of fit (R2) was greater than 0.97 (Table 3).
Nonlinear Viscoelastic Parameters
Descriptive statistics of the model parameters identified based on the work done by air-puff force attaining the whole corneal displacement from Corvis ST are shown in Table 4. μ and α were normal distribution, and A and τ were non-normal distribution. Table 4 shows the median, mean, standard deviation, IQR, minimum, and maximum of the material parameters for all corneas of the participants, and the median of goodness of fit (R2) was over 0.96. Furthermore, power of the tests calculated by using the data of μ was about 0.87.
A typical fitting result of the visco-hyperelastic model to the work–time curve for one eye is shown in Figure 7. It shows a good agreement and is close to the work done by air-puff force attaining the whole corneal displacement in the Corvis ST test with time.
FIGURE 7. A fitting result of work done by air-puff force attaining the whole corneal displacement during the Corvis ST test by the visco-hyperelastic model. The red solid dot is the fitting result, and the blue triangle represents the calculated work done by air-puff force attaining the whole corneal displacement during the Corvis ST test.
Discussion
This study established a method to identify the nonlinear viscoelastic properties of human corneas from the output data of Corvis ST. The results showed feasibility of the method.
The morphology of the normal cornea has almost no change under IOP. For abnormal cornea, corneal viscoelasticity plays an important role (Vellara and Patel, 2015) in the maintenance of corneal morphology under long-term IOP. Keratoconus, a common clinical disease characterized by progressive corneal dilation, is highly related to corneal viscoelasticity (Vellara and Patel, 2015; Jin et al., 2020; Rahmati et al., 2021). This study gives a more comprehensive understanding of corneal biomechanical properties from the test of cornea in vivo. Compared with other studies, the innovation of this research is the determination of the nonlinear viscoelastic properties of human corneas from the output data of Corvis ST. It is expected that the biomechanical properties of the human cornea will be widely used in the early diagnosis of keratoconus, the screening of cornea before refractive surgery, and the early diagnosis of iatrogenic corneal ectasia after refractive surgery.
IOP is always acting on the inner surface of the cornea. So the cornea has been under the tension generated by IOP, before, during, and after the Corvis ST test. In this study, the cornea under normal IOP has been regarded as an equilibrium state. It is similar in that the pre-stress is taken into account in corneal tests in vitro. For example, in the uniaxial tensile test of corneal strips, the stress–strain relationship should be
Our previous study suggested a method (Qin et al., 2019a), following the studies of Ko et al. (2013) and Reissner (1946), to determine the corneal elastic modulus from corneal apical displacements between 0.1 and 0.2 mm at the first applanation state during the Corvis ST test based on the relationship between force and displacements of the shallow spherical model. It gave a corneal elastic modulus of human which ranged from 0.16 to 0.30 MPa. Shih et al. (2017) took the average pressure of air-puff during the Corvis ST test as the external pressure, did not consider the pressure changes with time, and applied Taber’s model to obtain the elastic modulus of human which ranged from 0.053 to 0.363 MPa. These elastic moduli are given at the equilibrium of the first applanation state. The present study was initially motivated by Taber’s model (Shih et al., 2017; Taber, 1982) and applied a visco-hyperelastic model to determine corneal nonlinear viscoelastic parameters. According to Eq. 4 in the present study, the stretch ratio is about 0.995 at the first applanation state. This also demonstrates that our research was carried out under normal IOP as an equilibrium. Furthermore, in the studies of identifying corneal material parameters from Corvis ST data by the inverse finite element method, the zero stress state is also regarded as an initial state (Lago et al., 2015; Rahmati et al., 2021). Therefore, our study is different from most previous studies. We have given the in vivo corneal biomechanical parameters.
As we have known, the uniaxial stretch test is usually used to identify the tangential modulus of the cornea (Wollensak et al., 2003; Elsheikh et al., 2010; Zhang et al., 2018). If we let the stretch ratio
The researchers believe that the strain rate under loading affects the stress–strain relationship of soft tissue (Snedeker et al., 2005; Kim et al., 2012) and that the stiffness of the cornea is directly proportional to its strain rate (Rahmati et al., 2021). It should be noted that this study is based on the test data of Corvis ST to identify the corneal nonlinear viscoelastic parameters. The test time is only about 30 ms, a fast loading and unloading process with the loading rate being about 2.5/s. This may lead to the difference between the parameters of the Ogden model obtained in this study and the results of the uniaxial tensile test of the in vitro cornea (strain rate less than 0.005/s). Moreover, the parameter A (0.10–0.39) is close to the values of Ak (k = 1, 2, 3) in Prony relaxation function, according to the studies by Kling et al. (2014) and Qin et al. (2019b). However, τ (1–2 ms) is smaller than τk (k = 1, 2, 3) in Prony relaxation function compared with the reported results. Because viscoelasticity is a time-dependent property, and the strain rate (∼2.5/s) during the Corvis ST test is larger than during the stress relaxation test (less than 0.1/s), this may lead to the smaller τ in the Prony series obtained in this study.
In addition, a visco-hyperelastic model has been used to describe the stress–strain curve of the kidney capsule obtained by the uniaxial tensile test at high strain rates, and to identify the biomechanical characteristics of the tissue (Snedeker et al., 2005); this indicates that the visco-hyperelastic model has the possibility to be applied to identify the biomechanical properties of soft tissue under rapid deformation. The results of our study also confirm this possibility. The established method provides the evaluation biomechanical properties of the human cornea via in vivo test.
In addition, there are some advantages and limitations of using the visco-hyperelastic model to describe the biomechanical behavior of the human cornea. First of all, visco-hyperelastic models based on strain energy functions have been shown to reproduce both the time-dependent and large strain aspects of the response (Ahearne et al., 2007; Boyce et al., 2007). A visco-hyperelastic model can give both elastic and viscous characteristics of the cornea. Secondly, this research showed that the calculation to determine model parameters is easy to carry out and saves time compared with inverse finite element methods. Thirdly, the simple model we selected, the convenient and fast computation, and the nonlinear viscoelasticity make the method of this study a potential method adopted by clinic applications, such as corneal screening before refractive surgery and early diagnosis of keratoconus. The limitation includes that researchers have not reached a consensus on how to identify the viscoelastic properties of the cornea based on Corvis ST data. The existing literature on whether the viscoelasticity of cornea can be evaluated from Corvis ST data implemented the study based on a specific constitutive model of cornea. Due to the different methods adopted, therefore, there were some differences in the results. Francis et al. (2019) believed that corneal viscous properties cannot be determined from air-puff applanation. However, majority of studies have shown that the viscoelasticity of the cornea can be obtained according to the output data of Corvis ST; for example, based on the output data of Corvis ST, the viscoelasticity of the cornea of normal people and patients with keratoconus has been compared and analyzed by the inverse finite element method (Rahmati et al., 2021), and Abass et al. (2020) believed that the viscoelasticity of the cornea can be determined by using the output data of Corvis ST. Our research showed that Corvis ST data can be used to explore the viscoelasticity of cornea. Further, Corvis ST records the morphological changes of cornea under the action of air-puff within about 30 ms, which is a rapid loading and unloading process. In the existing studies, the period of corneal stress relaxation tests was not less than 2 min (Hammer et al., 2015; Kling et al., 2015; Zhang et al., 2018). The difference in test time between Corvis ST and classical relaxation or creep tests may result in different orders of magnitude of the model parameters. How to understand the relationship between them and corneal stress relaxation characteristics needs to be further investigated.
There are some limitations in this study. First, the number of participants included was small, considering that the population with different cornea stiffness may have different responses to Corvis ST; its clinical significance needs to be further validated with study on patients with corneal biomechanics disorder, such as keratoconus, or patients that received corneal collagen crosslinking. Second, we studied the cornea with certain tension under IOP as the initial state and failed to obtain the biomechanical parameters of the cornea under zero stress state. However, we also note that the initial state that should be paid attention to in the research of applying corneal biomechanical properties to clinical problems is the state under IOP, so the method we provide to identify corneal biomechanical properties may be more useful. Third, in this study, we did not consider the corneal anisotropic mechanical properties. It should be further studied in the future.
Conclusion
In conclusion, it is provided that the corneal nonlinear viscoelastic properties can be determined based on a visco-hyperelastic model from the corneal contour information and air-puff pressure of Corvis ST test. All in all, the present conclusion should be substantiated with more investigation, by which to provide more useful information for clinical practice.
Data availability statement
The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding authors.
Ethics statement
The studies involving human participants were reviewed and approved by the institutional review board of Capital Medical University which approved this study, and all participants signed an informed consent form in accordance with the tenets of the Declaration of Helsinki. The patients/participants provided their written informed consent to participate in this study.
Author contributions
Manuscript writing and review: DZ, HZ, and LT. Data collection, analysis, and interpretation: DZ, YZ, and CF. Design of this study and revision of the manuscript: CZ and LL. All authors read and approved the final manuscript.
Funding
This study was financially supported by the National Natural Science Foundation of China (Nos. 31370952, 31470914, 31600758, 82171101, 32171304); Beijing Nova Program (Z181100006218099); and the Open Research Fund from Beijing Advanced Innovation Center for Big Data-Based Precision Medicine, Beijing Tongren Hospital, Beihang University and Capital Medical University (BHTR-KFJJ-202001).
Conflict of Interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Publisher’s Note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors, and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
Abass, A., Roberts, C. J., Lopes, B., Eliasy, A., Vinciguerra, R., Ambrósio, R., et al. (2020). Can the Corvis St Estimate Corneal Viscoelasticity. J. Refract. Surg. 36, 346–347. doi:10.3928/1081597X-20200212-04
Ahearne, M., Yang, Y., Then, K. Y., and Liu, K.-K. (2007). An Indentation Technique to Characterize the Mechanical and Viscoelastic Properties of Human and Porcine Corneas. Ann. Biomed. Eng. 35, 1608–1616. doi:10.1007/s10439-007-9323-9
Bao, F., Deng, M., Zheng, X., Li, L., Zhao, Y., Cao, S., et al. (2017). Effects of Diabetes Mellitus on Biomechanical Properties of the Rabbit Cornea. Exp. Eye Res. 161, 82–88. doi:10.1016/j.exer.2017.05.015
Bao, F., Zheng, Y., Liu, C., Zheng, X., Zhao, Y., Wang, Y., et al. (2018). Changes in Corneal Biomechanical Properties with Different Corneal Cross-Linking Irradiances. J. Refract. Surg. 34, 51–58. doi:10.3928/1081597X-20171025-01
Bekesi, N., Kochevar, I. E., and Marcos, S. (2016). Corneal Biomechanical Response Following Collagen Cross-Linking with Rose Bengal-green Light and Riboflavin-UVA. Invest. Ophthalmol. Vis. Sci. 57, 992–1001. doi:10.1167/iovs.15-18689
Boyce, B. L., Jones, R. E., Nguyen, T. D., and Grazier, J. M. (2007). Stress-Controlled Viscoelastic Tensile Response of Bovine Cornea. J. Biomech. 40, 2367–2376. doi:10.1016/j.jbiomech.2006.12.001
Brown, L., Foulsham, W., Pronin, S., and Tatham, A. J. (2018). The Influence of Corneal Biomechanical Properties on Intraocular Pressure Measurements Using a Rebound Self-Tonometer. J. Glaucoma 27, 511–518. doi:10.1097/IJG.0000000000000948
Chan, T. C., Wang, Y. M., Yu, M., and Jhanji, V. (2018). Comparison of Corneal Dynamic Parameters and Tomographic Measurements Using Scheimpflug Imaging in Keratoconus. Br. J. Ophthalmol. 102, 42–47. doi:10.1136/bjophthalmol-2017-310355
Dupps, W. J., and Wilson, S. E. (2006). Biomechanics and Wound Healing in the Cornea. Exp. Eye Res. 83, 709–720. doi:10.1016/j.exer.2006.03.015
Eliasy, A., Chen, K.-J., Vinciguerra, R., Lopes, B. T., Abass, A., Vinciguerra, P., et al. (2019). Determination of Corneal Biomechanical Behavior In-Vivo for Healthy Eyes Using CorVis ST Tonometry: Stress-Strain index. Front. Bioeng. Biotechnol. 7, 105. doi:10.3389/fbioe.2019.00105
Elsheikh, A., Geraghty, B., Rama, P., Campanelli, M., and Meek, K. M. (2010). Characterization of Age-Related Variation in Corneal Biomechanical Properties. J. R. Soc. Interface 7, 1475–1485. doi:10.1098/rsif.2010.0108
Fraldi, M., Cutolo, A., Esposito, L., and Guarracino, F. (2011). The Role of Viscoelasticity and Stress Gradients on the Outcome of Conductive Keratoplasty. Biomech. Model. Mechanobiol. 10, 397–412. doi:10.1007/s10237-010-0242-6
Francis, M., Matalia, H., Nuijts, R. M. M. A., Haex, B., Shetty, R., and Sinha Roy, A. (2019). Corneal Viscous Properties Cannot Be Determined from Air-Puff Applanation. J. Refract. Surg. 35, 730–736. doi:10.3928/1081597x-20191010-03
Hammer, A., Kling, S., Boldi, M.-O., Richoz, O., Tabibian, D., Randleman, J. B., et al. (2015). Establishing Corneal Cross-Linking with Riboflavin and UV-A in the Mouse Cornea In Vivo: Biomechanical Analysis. Invest. Ophthalmol. Vis. Sci. 56, 6581–6590. doi:10.1167/iovs.15-17426
Huang, H., Tang, W., Tan, Q., and Yan, B. (2017). Development and Parameter Identification of a Visco-Hyperelastic Model for the Periodontal Ligament. J. Mech. Behav. Biomed. Mater 68, 210–215. doi:10.1016/j.jmbbm.2017.01.035
Jin, Z., Zhou, Y., Shen, M., Wang, Y., Lu, F., and Zhu, D. (2020). Assessment of Corneal Viscoelasticity Using Elastic Wave Optical Coherence Elastography. J. Biophotonics 13, e201960074. doi:10.1002/jbio.201960074
Joda, A. A., Shervin, M. M., Kook, D., and Elsheikh, A. (2016). Development and Validation of a Correction Equation for Corvis Tonometry. Comput. Methods Biomech. Biomed. Engin. 19, 943–953. doi:10.1080/10255842.2015.1077515
Kim, W., Argento, A., Rozsa, F. W., and Mallett, K. (2012). Constitutive Behavior of Ocular Tissues over a Range of Strain Rates. J. Biomech. Eng. 134, 061002. doi:10.1115/1.4006847
Kling, S., Bekesi, N., Dorronsoro, C., Pascual, D., and Marcos, S. (2014). Corneal Viscoelastic Properties from Finite-Element Analysis of In Vivo Air-Puff Deformation. Plos One 9, e104904. doi:10.1371/journal.pone.0104904
Kling, S., Richoz, O., Hammer, A., Tabibian, D., Jacob, S., Agarwal, A., et al. (2015). Increased Biomechanical Efficacy of Corneal Cross-Linking in Thin Corneas Due to Higher Oxygen Availability. J. Refract. Surg. 31, 840–846. doi:10.3928/1081597X-20151111-08
Ko, M. W. L., Leung, L. K. K., Lam, D. C. C., and Leung, C. K. S. (2013). Characterization of Corneal tangent Modulus in Vivo. Acta Ophthalmol. 91, e263–e269. doi:10.1111/aos.12066
Lago, M. A., Rupérez, M. J., Martínez-Martínez, F., Monserrat, C., Larra, E., Güell, J. L., et al. (2015). A New Methodology for the In Vivo Estimation of the Elastic Constants that Characterize the Patient-specific Biomechanical Behavior of the Human Cornea. J. Biomech. 48, 38–43. doi:10.1016/j.jbiomech.2014.11.009
Liu, T., Shen, M., Huang, L., Xiang, Y., Li, H., Zhang, Y., et al. (2020a). Characterization of Hyperelastic Mechanical Properties for Youth Corneal Anterior central Stroma Based on Collagen Fibril Crimping Constitutive Model. J. Mech. Behav. Biomed. Mater. 103, 103575. doi:10.1016/j.jmbbm.2019.103575
Liu, T., Shen, M., Li, H., Zhang, Y., Mu, B., Zhao, X., et al. (2020b). Changes and Quantitative Characterization of Hyper-Viscoelastic Biomechanical Properties for Young Corneal Stroma after Standard Corneal Cross-Linking Treatment with Different Ultraviolet-A Energies. Acta Biomater. 113, 438–451. doi:10.1016/j.actbio.2020.06.005
Maklad, O., Eliasy, A., Chen, K. J., Theofilis, V., and Elsheikh, A. (2019). Simulation of Air Puff Tonometry Test Using Arbitrary Lagrangian-Eulerian (ALE) Deforming Mesh for Corneal Material Characterisation. Int. J. Environ. Res. Public Health 17, 54. doi:10.3390/ijerph17010054
Piñero, D. P., and Alcón, N. (2014). In Vivo characterization of Corneal Biomechanics. J. Cataract. Refract. Surg. 40, 870–887. doi:10.1016/j.jcrs.2014.03.021
Qin, X., Tian, L., Zhang, H., Chen, X., and Li, L. (2019a). Evaluation of Corneal Elastic Modulus Based on Corneal Visualization Scheimpflug Technology. Biomed. Eng. Online 18, 42. doi:10.1186/s12938-019-0662-1
Qin, X., Yu, M., Zhang, H., Chen, X., and Li, L. (2019b). The Mechanical Interpretation of Ocular Response Analyzer Parameters. Biomed. Res. Int. 2019, 5701236. doi:10.1155/2019/5701236
Rahmati, S. M., Razaghi, R., and Karimi, A. (2021). Biomechanics of the Keratoconic Cornea: Theory, Segmentation, Pressure Distribution, and Coupled FE-Optimization Algorithm. J. Mech. Behav. Biomed. Mater. 113, 104155. doi:10.1016/j.jmbbm.2020.104155
Ramzanpour, M., Hosseini-Farid, M., McLean, J., Ziejewski, M., and Karami, G. (2020). Visco-Hyperelastic Characterization of Human Brain White Matter Micro-Level Constituents in Different Strain Rates. Med. Biol. Eng. Comput. 58, 2107–2118. doi:10.1007/s11517-020-02228-3
Reissner, E. (1946). Stresses and Small Displacements of Shallow Spherical Shells. J. Math. Phys. 25, 80–85. doi:10.1002/sapm194625180
Sedaghat, M.-R., Momeni-Moghaddam, H., Ambrósio, R., Roberts, C. J., Yekta, A.-A., Danesh, Z., et al. (2018). Long-term Evaluation of Corneal Biomechanical Properties after Corneal Cross-Linking for Keratoconus: A 4-Year Longitudinal Study. J. Refract. Surg. 34, 849–856. doi:10.3928/1081597X-20181012-02
Shen, Y., Chen, Z., Knorz, M. C., Li, M., Zhao, J., and Zhou, X. (2014). Comparison of Corneal Deformation Parameters after SMILE, LASEK, and Femtosecond Laser-Assisted LASIK. J. Refract. Surg. 30, 310–318. doi:10.3928/1081597x-20140422-01
Shih, P.-J., Huang, C.-J., Huang, T.-H., Lin, H.-C., Yen, J.-Y., Wang, I.-J., et al. (2017). Estimation of the Corneal Young's ModulusIn VivoBased on a Fluid-Filled Spherical-Shell Model with Scheimpflug Imaging. J. Ophthalmol. 2017, 5410143. doi:10.1155/2017/5410143
Simonini, I., Angelillo, M., and Pandolfi, A. (2016). Theoretical and Numerical Analysis of the Corneal Air Puff Test. J. Mech. Phys. Solids 93, 118–134. doi:10.1016/j.jmps.2016.04.012
Sinha Roy, A., Kurian, M., Matalia, H., and Shetty, R. (2015). Air-puff Associated Quantification of Non-linear Biomechanical Properties of the Human Cornea In Vivo. J. Mech. Behav. Biomed. Mater. 48, 173–182. doi:10.1016/j.jmbbm.2015.04.010
Snedeker, J. G., Niederer, P., Schmidlin, F. R., Farshad, M., Demetropoulos, C. K., Lee, J. B., et al. (2005). Strain-rate Dependent Material Properties of the Porcine and Human Kidney Capsule. J. Biomech. 38, 1011–1021. doi:10.1016/j.jbiomech.2004.05.036
Su, P., Yang, Y., Xiao, J., and Song, Y. (2015). Corneal Hyper-Viscoelastic Model: Derivations, Experiments, and Simulations. Acta Bioeng. Biomech. 17, 73–84. doi:10.5277/ABB-00142-2014-03
Taber, L. A. (1982). Large Deflection of a Fluid-Filled Spherical Shell under a point Load. J. Appl. Mech. 49, 121–128. doi:10.1115/1.3161953
Taylor, Z. A., Comas, O., Cheng, M., Passenger, J., Hawkes, D. J., Atkinson, D., et al. (2009). On Modelling of Anisotropic Viscoelasticity for Soft Tissue Simulation: Numerical Solution and GPU Execution. Med. Image Anal. 13, 234–244. doi:10.1016/j.media.2008.10.001
Vellara, H. R., and Patel, D. V. (2015). Biomechanical Properties of the Keratoconic Cornea: a Review. Clin. Exp. Optom. 98, 31–38. doi:10.1111/cxo.12211
Vinciguerra, R., Romano, V., Arbabi, E. M., Brunner, M., Willoughby, C. E., Batterbury, M., et al. (2017). In Vivo Early Corneal Biomechanical Changes after Corneal Cross-Linking in Patients with Progressive Keratoconus. J. Refract. Surg. 33, 840–846. doi:10.3928/1081597X-20170922-02
Wang, W., Du, S., and Zhang, X. (2015). Corneal Deformation Response in Patients with Primary Open-Angle Glaucoma and in Healthy Subjects Analyzed by Corvis ST. Invest. Ophthalmol. Vis. Sci. 56, 5557–5565. doi:10.1167/iovs.15-16926
Whitford, C., Movchan, N. V., Studer, H., and Elsheikh, A. (2018). A Viscoelastic Anisotropic Hyperelastic Constitutive Model of the Human Cornea. Biomech. Model. Mechanobiol. 17, 19–29. doi:10.1007/s10237-017-0942-2
Wollensak, G., Spoerl, E., and Seiler, T. (2003). Stress-strain Measurements of Human and Porcine Corneas after Riboflavin-Ultraviolet-A-Induced Cross-Linking. J. Cataract. Refract. Surg. 29, 1780–1785. doi:10.1016/s0886-3350(03)00407-3
Keywords: cornea, corneal biomechanical properties, in vivo, CorVis ST, visco-hyperelastic model
Citation: Zhang D, Zhang H, Tian L, Zheng Y, Fu C, Zhai C and Li L (2021) Exploring the Biomechanical Properties of the Human Cornea In Vivo Based on Corvis ST. Front. Bioeng. Biotechnol. 9:771763. doi: 10.3389/fbioe.2021.771763
Received: 07 September 2021; Accepted: 21 October 2021;
Published: 17 November 2021.
Edited by:
FangJun Bao, Affiliated Eye Hospital of Wenzhou Medical College, ChinaCopyright © 2021 Zhang, Zhang, Tian, Zheng, Fu, Zhai and Li. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Lin Li, lil@ccmu.edu.cn; Changbin Zhai, 17091006117@163.com