ORIGINAL RESEARCH article

Front. Bioeng. Biotechnol., 05 May 2021

Sec. Biomechanics

Volume 9 - 2021 | https://doi.org/10.3389/fbioe.2021.613616

Experimental and Modeling Analyses of Human Motion Across the Static Magnetic Field of an MRI Scanner

  • 1. Advanced Radiation Oncology Department, Cancer Care Center, Istituto di Ricovero e Cura a Carattere Scientifico (IRCCS) Sacro Cuore Don Calabria Hospital, Negrar di Valpolicella, Italy

  • 2. Dipartimento di Fisica e Chimica, Università degli Studi di Palermo, Palermo, Italy

  • 3. Istituto di Ricovero e Cura a Carattere Scientifico (IRCCS) Istituto Ortopedico Rizzoli, Movement Analysis Laboratory, Bologna, Italy

  • 4. Azienda Ospedaliera di Rilievo Nazionale e di Alta Specializzazione (A.R.N.A.S.) Civico–Di Cristina–Benfratelli, Unità Operativa Complessa (U.O.C.) Fisica Sanitaria, Palermo, Italy

  • 5. Villa Santa Teresa, Unità Operativa (U.O.) Fisica Sanitaria, Bagheria, Italy

  • 6. Dipartimento di Scienze Biomediche, Odontoiatriche e delle Immagini Morfologiche e Funzionali (BIOMORF), Università degli Studi di Messina, Messina, Italy

  • 7. Scuola di Specializzazione in Fisica Medica, Università degli Studi di Messina, Messina, Italy

Abstract

It is established that human movements in the vicinity of a permanent static magnetic field, such as those in magnetic resonance imaging (MRI) scanners induce electric fields in the human body; this raises potential severe risks of health to radiographers and cleaners exposed routinely to these fields in MRI rooms. The relevant directives and parameters, however, are based on theoretical models, and accurate studies on the simulation of the effects based on human movement data obtained in real conditions are still lacking. Two radiographers and one cleaner, familiar with MRI room activities and these directives, were gait analyzed during the execution of routine job motor tasks at different velocities. Full body motion was recorded in a gait laboratory arranged to reproduce the workspace of a room with an MRI full-body scanner. Body segments were tracked with clusters of at least three markers, from which position and velocity of the centroids were calculated. These were used as input in an established computer physical model able to map the stray field in an MRI room. The spatial peak values of the calculated electric field induced by motion of the head and of the entire body during these tasks, for both the health and sensory effects, were found smaller than the thresholds recommended by the European directives, for both 1.5 T and 3.0 T MRI. These tasks therefore seem to guarantee the safety of MRI room operators according to current professional good practice for exposure risks. Physical modeling and experimental measures of human motion can also support occupational medicine.

1. Introduction

Magnetic resonance imaging (MRI) is used largely worldwide to assess the status of tissues in patients with musculoskeletal and other diseases. There are ~50,000 MRI machines worldwide, with about 5,000 new units sold every year. Forty million scans are performed annually in the United States, and the number of examinations in 2017 has the peak of 143 per 1,000 population in Germany (Mikulic, 2019). It is also established that movements in the vicinity of a permanent static magnetic field, such as that of an MRI scanner, induce electric fields in the human body. This raises potential severe risks of health to the persons working routinely in MRI rooms, such as radiographers and cleaners. There are directives for these workers on how to move in these rooms, but today these are based on theoretical models only (inter alii, Hartwig et al., 2014, 2019; Zilberti et al., 2015, 2016; Sannino et al., 2017). In other words, accurate simulation studies based on human movement data obtained in real conditions able to provide estimations of the exact effects of these fields are still lacking. The scope of the present experimental and modeling study is to provide these estimations for the first time by enhancing with experimental data the analysis provided in a previous modeling work (Gurrera et al., 2019). The general aim is to support general good practice methods for exposure assessment of personnel moving across the permanent static magnetic field straying from MRI scanners.

This previous work in fact (Gurrera et al., 2019) suffers from some limitations. The MRI operators were supposed to move translationally and at a constant speed, with relevant parameters derived on the basis of general studies reporting allegedly normal walking speeds (Bohannon and Williams Andrews, 2011). Thus, there is still a lack of specialized literature or experiments dealing with motion analysis of MRI operators. Moreover, the exposure assessment did not include any 3.0 T facilities. What would be necessary for the enhancement of the previous work is a full body, i.e., from head to foot, kinematic characterization of real MRI operators at work. This would provide the most realistic estimation of the electric field induced by these movements, easily extendable to the scanners of 3.0 T or more.

This kinematic characterization is obtained usually by human movement analysis (HMA) using stereophotogrammetry, an established technique which allows for accurate 3D tracking in space of body segments during the execution of locomotion tasks or elementary exercises or even high-performance sport activities (Cappozzo et al., 2005). This is achieved by instrumenting the subject under analysis with small spherical markers stuck on the skin possibly according to standard protocols (Ferrari et al., 2008; Kainz et al., 2017), and in case by arranging the gait analysis laboratory with the necessary furnishing for the simulation of the required environment according to the activity under investigation (chairs, steps, stairs, obstacles, etc.).

As mentioned, given an inertial reference frame in which the magnet is at rest, electric charges movements within the body of an MRI operator may occur because of body segment voluntary movements of the subjects; other organ and tissue motion within the body (Herman, 2016) are here ignored. The motion-induced field within the body of an MRI operator is addressed in the guidelines of the International Commission on Non-Ionizing Radiation Protection (ICNIRP) (International Commission on Non-Ionizing Radiation Protection, 2014). Unfortunately, for these estimations, questionable approximations were used, and an alleged “conversion factor” should be accounted, associated to “the location within the body, the size of the body, the shape of the body, electrical properties of the tissue” as well as “the direction and distribution of the magnetic field.” Direct measurements of body segment motion of MRI operators are therefore necessary to quantify these effects.

The present work wants to contribute in this respect. In particular, the hypothesis is that the exposure of MRI personnel to static magnetic fields does comply with the current European Directive 2013/35/EU, and for both 1.5 T and 3.0 T MRI machines.

2. Materials and Methods

2.1. The Theoretical Basis

The theoretical basis, introduced and discussed in Gurrera et al. (2019), is here briefly summarized.

  • Given an inertial reference frame in which the magnet is at rest, electric charges within the body of an MRI operator may be moving because of two reasons: natural movements within the body as those related to blood flow and nervous communication system (Herman, 2016), which are ignored in the present study, and the voluntary movements of the body segments and joints associated to locomotion and upper body maneuvers according to the tasks to be performed.

  • Let γ(t) be an oriented closed conducting wire moving across a static magnetic field , the velocity of each point of the wire, Sγ(t) an arbitrary surface enclosed by γ(t).

    Then, according to the Faraday–Neumann's law:

    or equivalently:

    where and are the electric field and the overall magnetic field, respectively. The rest of the notation is assumed to be familiar to the reader.

    If the inductance of the wire is negligible, then it is possible to assume that and there is no time dependence. Therefore, according to Equation (1), the generated electric field comes out to be irrotational and, as a consequence, has to be caused only by a continuous charge redistribution within the wire. And such a redistribution is imposed by the action of the Lorentz' force. The complete cause and effect picture is as follows. Free charges within the conductor start moving across the external magnetic field with a mean velocity that, at each point, is equal to the velocity of the wire. Then the Lorentz' force accelerates them and causes a charge separation that produces an electrostatic field opposing further accumulation, i.e., opposing the electromotive field . Except for particular given arrangements of and , equilibrium is not reached and a time-varying density current flows along the wire. As far as the magnetic field produced by this induced current may be neglected, there will be only an electromotive field and a reaction electrostatic field (). Now, a human body moving inside an MRI room may be considered as a special instance of a massive extended non-ferromagnetic conductor for which all the reasoning above applies. Particularly, the motion-induced “electric field” to be tested against the exposure limit values (ELVs) of the directive will be the sum of the electromotive field and the reaction electrostatic field . If the latter, which even a computational model accounting for the electrical properties of the human body could only approximate, is neglected, then an overestimation of the induced field may be expected. But that even results in a precautionary approach.

    To summarize, an electric charge moving in a magnetic field , with overall instantaneous velocity , is subjected to the Lorentz' force and the related electromotive field may be regarded as the best approximation for the motion-induced field within the body of an MRI operator. Indeed, the induced field suggested in the guidelines of the International Commission on Non-Ionizing Radiation Protection (ICNIRP) (International Commission on Non-Ionizing Radiation Protection, 2014)

    may be considered as an unnecessary coarse approximation of it (and, possibly, an overestimation), C being a “conversion factor” that should account for “the location within the body, the size of the body, the shape of the body, electrical properties of the tissue” as well as “the direction and distribution of the magnetic field.” In fact, in a static magnetic field, the electric field induced in a moving conducting loop is irrotational (see also Bringuier, 2002). Therefore, the Faraday–Neumann's law simplifies to:

    and calculating the rate of change of the magnetic flux is tantamount to calculate the circulation of .

  • A magnetic dipole may provide (if data confirming evidence is obtained) a parsimonious, but still adequate, 3D approximation of the magnetic field straying from a closed full-body MRI scanner, its specific architecture details being unknown (see also Sannino et al., 2017).

2.2. Mapping Relevant Velocities in an MRI Room

In May 2019, three healthy MRI operators, specific job and other details provided in Table 1, volunteered to contribute the current study and to be observed and recorded during the execution of routine job tasks at the Movement Analysis Laboratory of the Istituto Ortopedico Rizzoli in Bologna (Italy). This was instrumented with an 8-TV-camera stereophotogrammetric system (Vicon Motion Systems, Oxford, UK) and related processing software and opportunely arranged, as sketched in Figure 1, to reproduce a typical workspace of an MRI closed full-body scanner. The subjects were given instructions and time to familiarize with the room environment and then were asked to execute twice each of the representative job motor tasks, involved in normal and emergency scenarios, listed in Table 2.

Table 1

OperatorJobGenderAgeHeight (cm)Weight (kg)
MRIR1RadiographerMale31189115
MRIR2RadiographerFemale2715049
MRICCleanerMale5417261

Specific job and other details of the three MRI operators who volunteered to participate the study.

Figure 1

Table 2

Task codeInvolved operatorsTask titleTask description
N1RadiographersEntering/Leaving AThe operator enters the MRI room through Door A, reaches the control panel on the magnet housing, then leaves the room through the same door.
N2RadiographersEntering/Leaving BAs in N1, but through Door B.
N3RadiographersEntering/Leaving CAs in N1, but through Door C.
N4RadiographersHead coil preparationThe operator enters the MRI room through Door A and places the head coil on the couch near the bore entrance, then leaves the room through the same door.
N5RadiographersPatient centeringThe operator is near the bore entrance, bends over the patient and positions the head coil, centers and checks the patient, then leaves the room through Door A.
N6RadiographersObject recoveringThe operator enters the MRI room through Door A, bends over the floor next to the bore entrance and picks up an object, then leaves the room through the same door.
E1RadiographersEmergency entering APatient alarm on. The operator enters the MRI room through Door A and checks the patient lying inside the bore, then leaves the room through the same door.
E2RadiographersEmergency entering BAs in E1, but through Door B.
E3RadiographersEmergency entering CAs in E1, but through Door C.
E4RadiographersEmergency patient extractionPatient alarm on. The operator enters the MRI room through Door A, checks the patient lying inside the bore, then rapidly extracts the couch manually and finally leaves the room through the same door.
C1CleanerFloor sweepingThe operator enters the MRI room through Door A and sweeps the floor (from the door toward the couch and then back), then leaves the room through the same door.
C2CleanerFloor moppingThe operator enters the MRI room through Door A and mops the floor (from the couch toward the door), then leaves the room through the same door.

Selected representative job motor tasks, involved in normal and emergency scenarios, executed by the MRI operators who volunteered to participate the study.

Refer to Figure 1 for a sketch of the measurement environment. The words “MRI room,” “door,” “control panel,” “magnet housing,” “coil,” “couch,” “bore,” and “patient” are used only evocatively. However, during the acquisition the floor was really swept and mopped by the cleaner and the patient alarm was really heard during the emergency scenarios.

State-of-the-art stereophotogrammetric HMA was performed during the execution of the tasks. Established protocols were used to track lower limbs and pelvis segments (Leardini et al., 2007), trunk and shoulders (Leardini et al., 2011), head, and upper limbs (according to Plug-in-Gait protocol, Vicon Motion Systems, Oxford, UK). All together, combining the three protocols, a total of 47 spherical reflective markers, 14 mm diameter, were stuck on the skin in correspondence of palpable anatomical landmarks, each tracked in space at 100 Hz by the stereophotogrammetric system during movement. Six of these markers served only for anatomical calibration of those landmarks necessary for body segment analysis and were removed after a single static posture acquisition in double-leg stance. Marker trajectories were smoothed by the standard software tools within the motion capture system, according to established algorithms (Woltring, 1985). Then, from each cluster of markers stuck in a single body segment, the corresponding centroid was also derived. From each marker and centroid trajectory, the corresponding velocity was computed. In particular, instantaneous position and velocity during the execution of the tasks were obtained for each of the 23 representative body points listed in Table 3.

Table 3

Point numberPoint typePoint position
1CentroidHead
2CentroidTrunk
3CentroidPelvis
4CentroidRight thigh
5CentroidLeft thigh
6CentroidRight tibia
7CentroidLeft tibia
8CentroidRight foot
9CentroidLeft foot
10CentroidRight hand
11CentroidLeft hand
12MarkerRight temple
13MarkerLeft temple
14MarkerRight side of occipital bone
15MarkerLeft side of occipital bone
16MarkerSecond thoracic vertebra
17MarkerMidpoint between the inferior angles of most caudal points of the two scapulae
18MarkerNeck in between jugular veins
19MarkerXiphoid process
20MarkerRight anterior superior iliac spine
21MarkerLeft anterior superior iliac spine
22MarkerRight posterior superior iliac spine
23MarkerLeft posterior superior iliac spine

Selected points in the body of the three operators whose instantaneous position and velocity during the execution of the tasks are used for the exposure assessment.

2.3. Mapping the Stray Field in an MRI Room

The model proposed in Gurrera et al. (2019) for the static magnetic field straying inside an MRI room was applied in the present study to a closed full-body 3.0 T scanner, the modulus of the magnetic field being mapped, in order to fit the model, as described in the paper and by the same three-axis Hall magnetometer. Particularly, the modulus of the magnetic field was measured according to the map shown in Figure 1, positioned as close as possible to the bore and the couch and whose short and long steps are 20 and 40 cm long, respectively. These measurements were recorded in each of the 70 indicated points and at each of three different heights from the floor level: 72, 119, and 156 cm. Therefore, a three-dimensional map composed of 210 measures was produced for the scanner. The modulus of the generated magnetic dipole and its height above the floor level, i.e., the preliminary estimations necessary for the fit (Gurrera et al., 2019), are 1.06 MA m2 and 100 cm, respectively.

2.4. Implementation

What follows was carried out in the R environment for statistical computing and visualization (R Core Team, 2018) by ad hoc in-house developed code.

3. Results

3.1. MRI Operators: How Fast Do They Move?

As a first result of the HMA, displayed in Figures 24 is the speed, as a function of time, of each of the three examined operators (Table 1) during the second execution of each of their tasks (Table 2). No significant difference was observed between the first and the second execution of each task. Specifically, position and velocity of the body segments here analyzed showed a good inter-trial repeatability, consistent with previous work from these authors (Manca et al., 2010; Caravaggi et al., 2011) where standard deviations of gait-analysis patterns are in the range 2÷4%.

Figure 2

Figure 3

Figure 4

The following results are drawn.

  • The feet are the part of the body that move fastest: up to 5 m s-1. Then come the hands: up to 3.5 m s-1.

  • Head marker on MRIR2 moved always slower than 2 m s-1, while this speed was occasionally exceeded by MRIR1. In the case of MRIC, head speed did not exceed 1 m s-1.

  • For the two radiographers, a U-shaped speed profile is observed: operators move fast when entering and leaving the room, while they slow down as they approach the magnet.

3.2. The Fit

Once conformed to the measured values, the dipole model in Gurrera et al. (2019) provides, in the case of the 3.0 T machine here analyzed, the response shown in Figure 5, displaying the scatter plot of all the estimated B values vs. the raw measures. Also displayed in the figure are two almost coincident straight lines, the black one being the best fit line, the other representing “perfect” modeling, i.e., y = x. As a result, despite the inherent heteroscedasticity, adherence to the model as measured by the Pearson's correlation coefficient is 0.97 (as it was in the case of the 1.5 T machines analyzed in Gurrera et al., 2019). The estimated values for the two unknowns Δx and Δy (Gurrera et al., 2019) are 69 and 8 cm, respectively.

Figure 5

3.3. Assessing the Whole-Body Exposure

By making use of the position and velocity map of each of the 23 tracked points in the body of the three volunteers (Table 3) and of the map provided by the model, in Figures 68 the peak value is shown during the (second) execution of each task. Results are given for the 3.0 T machine analyzed in this work, hereafter Machine 3.0, and for the 1.5 T Machine A analyzed in Gurrera et al. (2019).

Figure 6

Figure 7

Figure 8

For each operator and for each task, exposure results far below the 1.1 V m-1 ELV prescribed in European Union (2013) to account for possible health effects.

3.4. Assessing the Head Exposure

By making use of the position and velocity map of each of the five tracked points in the head of the three volunteers (Table 3) and of the map provided by the model, in Figures 911 the peak value is shown during the (second) execution of each task.

Figure 9

Figure 10

Figure 11

Only “object recovering” (N6) deserves being considered in detail and therefore in Figure 12 the induced pulse for MRIR1 and MRIR2 is shown in the case of Machine 3.0, along with the proper ELV adjusted by the motion-related frequency [1.43 and 1.14 Hz, respectively, as estimated by the spectral centroid of the corresponding periodogram (Massar et al., 2011)].

Figure 12

For each operator and for each task, exposure did not exceed the ELV prescribed in European Union (2013) to account for possible sensory effects, i.e., 0.7/f V m-1, where f is the motion-related frequency.

4. Discussion

In view of the obtained results, a limited risk margin (to exceed the ELV for the sensory effects) appeared only in the case of a 3.0 T machine. This raises questions about the opportunity of using lighter approaches. In fact, the “practical” reference levels in International Commission on Non-Ionizing Radiation Protection (2014) “for determining compliance with the basic restrictions for the induced internal electric field,” along with a practical but severe approximation of it (Equation 3), have proved to lead to alarming conclusions (inter alii, Acri et al., 2018 and Hartwig et al., 2019), while the estimation of the proper motion induced field introduced in Gurrera et al. (2019), once reliable measures of the components of the involved velocities and of the stray magnetic field are obtained, is definitely not less practical.

Here, in order to obtain realistic positions and velocities, a state-of-the-art HMA was used, able to track single reflective markers and relevant centroids with an accuracy smaller than 1 mm, and applied to three real MRI operators whose specific training and daily experience may be expected to replicate usual activities though in a simulated environment. Moreover, in order to span different typologies of workers, a man and a woman of very different height were recruited from a number of available radiographers. In this study, the “skin motion artifact,” due to the undesired displacements between the external markers on the skin and the underlying bone (Leardini et al., 2005), is not expected to affect the final results significantly, particularly the centroids, calculated from the external skin markers, and thus overall less sensible to this source of error.

In order to obtain the stray magnetic field, the simple dipole model already proposed in Gurrera et al. (2019) was used, which—once properly conformed to accurate static measures—has proved to yield reliable estimates of the three components of also for a 3.0 T machine.

Of course, different scenarios other than those depicted here may occur in an MRI room and other different MRI facilities should be analyzed. The obtained results appear robust and suggest that, even in the case of a 3.0 T machine, a controlled behavior in the close proximity of the magnet might easily prevent electric pulses beyond the thresholds prescribed in the Directive 2013/35/EU. In any case, the present work represents an original cross-disciplinary interaction between physical models and human motion analysis, made nowadays more necessary for the complexities implied in these approaches, and casts light on the current lack of general consensus (Stam and Yamaguchi-Sekino, 2018). In fact, this lack appears to be due to three main reasons: (i) the ICNIRP has missed the fundamental role of the Lorentz' force; (ii) a lack of specialized studies dealing with the motion analysis of MRI operators; (iii) a lack of standardized MRI personnel training throughout the European Union. Particularly, the last point appears to be fundamental since, in view of the obtained results, it would provide safe working conditions at least up to 3.0 T machines for all MRI operators. Of course, it should account for each possible task and for different statures.

5. Conclusions

The present study developed a method for introducing real human motion data into an established model now able to represent the induced electric field in the human body in the vicinity of an MRI scanner. This model, once applied to the two machines here analyzed—1.5 T one, 3.0 T the other—has resulted in a final positive compliance statement: both the health and sensory effects ELVs prescribed in the Directive 2013/35/EU are not exceeded. According to the present results, the introduction of specialized training protocols for MRI personnel throughout the European Union would prevent any possible risk to exceed the thresholds prescribed by the Directive currently in force, at least up to 3.0 T machines.

Statements

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

Ethical review and approval was not required for the study on human participants in accordance with the local legislation and institutional requirements. The patients/participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

VC, DG, and AL conceived the study and were in charge of overall direction and planning. KG and BA conceived an ad hoc precision carpet to map the static magnetic field and DG realized it. CR fabricated an ad hoc precision support for the magnetometer. DG carried out the static magnetic field measurements with support from VC, KG, BA, GI, GA, and GV. AL and MO carried out the human movement analysis with support from SD and DG. DG and AL wrote the manuscript. DG designed the theoretical model, carried out its implementation, and analyzed the data. MM provided an in-depth reading and analysis of the manuscript. All authors provided critical feedback and helped shape the research and analysis.

Funding

This work was in part supported by the Italian Ministry of Economy and Finance, programme 5 per mille, and in part by the University of Palermo.

Acknowledgments

The fundamental contribution to this study provided by the three MRI operators who volunteered their time and experience, namely, Pietro Urbani (MRIR1), Laura Federico (MRIR2), Massimo Doda (MRIC), was here deeply and gratefully acknowledged.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

References

  • 1

    AcriG.InferreraP.DenaroL.SansottaC.RuelloE.AnfusoC.et al. (2018). dB/dt Evaluation in MRI sites: is ICNIRP threshold limit (for workers) exceeded?Int. J. Environ. Res. Public Health15:1298. 10.3390/ijerph15071298

  • 2

    BohannonR. W.Williams AndrewsA. (2011). Normal walking speed: a descriptive meta-analysis. Physiotherapy97, 182189. 10.1016/j.physio.2010.12.004

  • 3

    BringuierE. (2002). Electrostatic charges in V × B fields and the phenomenon of induction. Eur. J. Phys. 24, 2129. 10.1088/0143-0807/24/1/304

  • 4

    CappozzoA.Della CroceU.LeardiniA.ChiariL. (2005). Human movement analysis using stereophotogrammetry: part 1: theoretical background. Gait Posture21, 186196. 10.1016/S0966-6362(04)00025-6

  • 5

    CaravaggiP.BenedettiM.BertiL.LeardiniA. (2011). Repeatability of a multi-segment foot protocol in adult subjects. Gait Posture33, 133135. 10.1016/j.gaitpost.2010.08.013

  • 6

    European Union (2013). Directive 2013/35/EU of the European Parliament and of the Council of 26 June 2013 on the minimum health and safety requirements regarding the exposure of workers to the risks arising from physical agents (electromagnetic fields) (20th individual Directive within the meaning of Article 16(1) of Directive 89/391/EEC) and repealing Directive 2004/40/EC. Off. J. Eur. Union56, 323. 10.3000/19770677.L_2013.179.eng

  • 7

    FerrariA.BenedettiM. G.PavanE.FrigoC.BettinelliD.RabuffettiM.et al. (2008). Quantitative comparison of five current protocols in gait analysis. Gait Posture28, 207216. 10.1016/j.gaitpost.2007.11.009

  • 8

    GurreraD.GalliasK.SpanòM.AbbateB.D'AliaF.IacovielloG.et al. (2019). Moving across the static magnetic field of a 1.5T MRI scanner: analysing compliance with Directive 2013/35/EU. Phys. Med. 57, 238244. 10.1016/j.ejmp.2018.11.004

  • 9

    HartwigV.BiaginiC.De MarchiD.FloriA.GabellieriC.VirgiliG.et al. (2019). The procedure for quantitative characterization and analysis of magnetic fields in magnetic resonance sites for protection of workers: a pilot study. Ann. Work Exposures Health63, 19. 10.1093/annweh/wxz002

  • 10

    HartwigV.VanelloN.GiovannettiG.LandiniL.SantarelliM. F. (2014). Estimation of occupational exposure to static magnetic fields due to usual movements in magnetic resonance units. Concepts Magn. Reson. B Magn. Reson. Eng. 44, 7581. 10.1002/cmr.b.21270

  • 11

    HermanI. P. (2016). “Electrical and magnetic properties,” in Physics of the Human Body, Biological and Medical Physics, Biomedical Engineering (Springer International Publishing), 819871. 10.1007/978-3-319-23932-3_12

  • 12

    International Commission on Non-Ionizing Radiation Protection (2014). Guidelines for limiting exposure to electric fields induced by movement of the human body in a static magnetic field and by time-varying magnetic fields below 1 Hz. Health Phys.106, 418425. 10.1097/HP.0b013e31829e5580

  • 13

    KainzH.GrahamD.EdwardsJ.WalshH. P.MaineS.BoydR. N.et al. (2017). Reliability of four models for clinical gait analysis. Gait Posture54, 325331. 10.1016/j.gaitpost.2017.04.001

  • 14

    LeardiniA.BiagiF.MerloA.BelvedereC.BenedettiM. G. (2011). Multi-segment trunk kinematics during locomotion and elementary exercises. Clinical Biomech. 26, 562571. 10.1016/j.clinbiomech.2011.01.015

  • 15

    LeardiniA.ChiariL.Della CroceU.CappozzoA. (2005). Human movement analysis using stereophotogrammetry: part 3. Soft tissue artifact assessment and compensation. Gait Posture21, 212225. 10.1016/j.gaitpost.2004.05.002

  • 16

    LeardiniA.SawachaZ.PaoliniG.IngrossoS.NativoR.BenedettiM. G. (2007). A new anatomically based protocol for gait analysis in children. Gait Posture26, 560571. 10.1016/j.gaitpost.2006.12.018

  • 17

    MancaM.LeardiniA.CavazzaS.FerraresiG.MarchiP.ZanagaE.et al. (2010). Repeatability of a new protocol for gait analysis in adult subjects. Gait Posture32, 282284. 10.1016/j.gaitpost.2010.05.011

  • 18

    MassarM.FickusM.BryanE.PetkieD.TerzuoliA.Jr. (2011). Fast computation of spectral centroids. Adv. Comput. Math. 35, 8397. 10.1007/s10444-010-9167-y

  • 19

    MikulicM. (2019). Number of examinations with magnetic resonance imaging (MRI) in selected countries in 2017. Available online at: https://www.statista.com/statistics/271470/mri-scanner-number-of-examinations-in-selected-countries/

  • 20

    R Core Team (2018). R: A Language and Environment for Statistical Computing. Vienna: R Foundation for Statistical Computing.

  • 21

    SanninoA.RomeoS.ScarfìM.MassaR.D'AngeloR.PetrilloA.et al. (2017). Exposure assessment and biomonitoring of workers in magnetic resonance environment: an exploratory study. Front. Public Health5:344. 10.3389/fpubh.2017.00344

  • 22

    StamR.Yamaguchi-SekinoS. (2018). Occupational exposure to electromagnetic fields from medical sources. Ind. Health56, 96105. 10.2486/indhealth.2017-0112

  • 23

    WoltringH. J. (1985). On optimal smoothing and derivative estimation from noisy displacement data in biomechanics. Hum. Mov. Sci. 4, 229245. 10.1016/0167-9457(85)90004-1

  • 24

    ZilbertiL.BottauscioO.ChiampiM. (2015). Motion-induced fields in magnetic resonance imaging: are the dielectric currents really negligible?IEEE Magn. Lett. 6, 14. 10.1109/LMAG.2015.2429641

  • 25

    ZilbertiL.BottauscioO.ChiampiM. (2016). Assessment of exposure to MRI motion-induced fields based on the International Commission on Non-Ionizing Radiation Protection (ICNIRP) guidelines. Magn. Reson. Med. 76, 12911300. 10.1002/mrm.26031

Summary

Keywords

human movement analysis, static magnetic fields, exposure limit values, MRI personnel safety, Directive 2013/35/EU

Citation

Gurrera D, Leardini A, Ortolani M, Durante S, Caputo V, Gallias KK, Abbate BF, Rinaldi C, Iacoviello G, Acri G, Vermiglio G and Marrale M (2021) Experimental and Modeling Analyses of Human Motion Across the Static Magnetic Field of an MRI Scanner. Front. Bioeng. Biotechnol. 9:613616. doi: 10.3389/fbioe.2021.613616

Received

02 October 2020

Accepted

10 March 2021

Published

05 May 2021

Volume

9 - 2021

Edited by

Simone Tassani, Pompeu Fabra University, Spain

Reviewed by

Xiaojun Chen, Shanghai Jiao Tong University, China; Michele Raggi, Turingsense EU Lab s.r.l, Italy

Updates

Copyright

*Correspondence: Davide Gurrera

This article was submitted to Biomechanics, a section of the journal Frontiers in Bioengineering and Biotechnology

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics