Skip to main content

GENERAL COMMENTARY article

Front. Cardiovasc. Med., 13 May 2019
Sec. Cardiac Rhythmology

Commentary: Interventricular Differences in Action Potential Duration Restitution Contribute to Dissimilar Ventricular Rhythms in ex vivo Perfused Hearts

  • 1BIOTRONIK SE & Co. KG, Berlin, Germany
  • 2Department of Cardiology, Ghent University Hospital, Ghent, Belgium

A Commentary on
Interventricular Differences in Action Potential Duration Restitution Contribute to Dissimilar Ventricular Rhythms in ex vivo Perfused Hearts

by Handa, B. S., Lawal, S., Wright, I. J., Li, X., Cabello-García, J., Mansfield, C., et al. (2019). Front. Cardiovasc. Med. 6:34. doi: 10.3389/fcvm.2019.00034

Since the discovery of dissimilar ventricular tachyarrhythmias using implantable cardioverter defibrillators (ICD) with cardiac resynchronization therapy (CRT-D) with the capability of left ventricular electrogram (EGM) recording (1), it was postulated that dissimilar ventricular rhythms might exist predominantly with rate disparities where the left ventricular (LV) rate is faster than the right ventricular (RV) rate.

In a recently published article in this journal entitled “Interventricular Differences in Action Potential Duration Restitution Contribute to Dissimilar Ventricular Rhythms in ex vivo Perfused Hearts”, Handa et al. used rat hearts to investigate the inducibility of dissimilar rhythms, the impact of antiarrhythmic drugs and the underlying mechanism (2). The authors correctly defined dissimilar ventricular rhythms as the occurrence of different ventricular tachyarrhythmias in the right and left ventricles, or different rates of the same arrhythmia in the two ventricles.

In the different quoted publications, dissimilar ventricular rhythms typically occur with an LV rate that is faster than the RV rate (LV>RV), which can lead to serious complications in correct ICD detection and the delay or absence of therapy. Momentarily, the detection of ventricular tachyarrhythmias (VAs) is exclusively based on the implanted electrode in the RV. A correct ICD detection cannot be expected if the RV rate is below the programmed cut-off rate, although ventricular fibrillation (VF) exists in the left ventricle.

The results in the above mentioned study (2) have shown exclusively LV>RV scenarios. A possible suggestion that was mentioned, would be to change the detection from the right to the left side.

However, in other CRT-D recordings (Figure 1) we observed episodes whereby the RV rate was faster than the LV rate (RV>LV). A fixed and predefined detection side for the ICD would therefore not be useful as no one can predict the presence of dissimilar ventricular rhythms. It is also not possible to predict in which chamber the possibly faster VA rate exists. A new dynamic detection principle could be helpful, whereby the ICD can decide itself from which side the faster heart rate originated. The technical solution requires certain special safety concepts to prevent inaccurate detection of atrial fibrillation in the event the LV electrode is dislocated into the atrium. A dislocation of one of the ventricular leads into the atrium can be safely excluded with the help of the EGM tracing in Figure 1, even in the absence of an X-ray image for two reasons: 1. The atrial channel shows sinus rhythm. Therefore, atrial fibrillation being wrongly detected by a possibly dislocated ventricular electrode can be excluded. It is genuine VF in the RV-channel; 2. The A-RV-LV signal sequence directly after the shock suggests a perfectly positioned LV electrode. The RV and LV signals are sufficiently separated from each other, which show the interventricular conduction delay caused by the left bundle branch block. A dislocated LV electrode in the atrium would present a prematurely detected LV signal at a time close to the true atrial excitation.

FIGURE 1
www.frontiersin.org

Figure 1. EGM recording of a CRT-D device with appropriate VF detection during dissimilar ventricular tachyarrhythmia. The RV-EGM shows typical morphologies of VF, whilst the LV-EGM shows RR intervals of a quasi-stable ventricular tachycardia (VT). The ICD recognized VF correctly because the ICD detection is based on the RV electrode and delivered a 40 Joule shock, restoring sinus rhythm. A, atrium; RV, right ventricle; LV, left ventricle.

Frequently, dissimilar ventricular rhythms will be triggered by anti-tachy pacing, intraoperative induction shocks or ineffective therapeutic shocks. It is rather rare that such heart rate disparities occur spontaneously. In the prehistory of the EGM recording (not shown in Figure 1), there was a previously ineffective shock that had initiated the dissimilar ventricular rhythm.

In our experience, the probability of discovering RV>LV is even higher than the incidence of LV>RV discovery, possibly because RV>LV can be better detected and treated by the ICD compared to LV>RV. A dynamic dual-side detection principle would certainly be an interesting idea. However, it can only be a possible concept in CRT-D devices. It is unfortunately not applicable in the single- or dual-chamber ICDs because of the absence of LV sensing (3, 4). Further technical investigations in favor of a functioning hemodynamic sensor are possibly more promising as this would be applicable in all types of ICDs. For the current generations of ICDs, the occurrence of RV>LV is less dramatic, as it will be recognized by the ICD.

It is remarkable that the phenomenon of RV>LV was not observed in the present study. Perhaps these cases of RV>LV could not be triggered due to the fact that this is an animal study and conditions other than those found in humans may prevail. The authors should be commended for their elegant study looking into the physiological mechanism of dissimilar rhythms, thereby lending support for clinical observations. It would be of great scientific interest if the authors can be encouraged to continue their efforts and expand their research to both observations of LV>RV as well as RV>LV, to reproduce them in the lab.

Author Contributions

All authors listed have made a substantial, direct and intellectual contribution to the work, and approved it for publication.

Conflict of Interest Statement

AK is an employee of BIOTRONIK SE & Co. KG.

The remaining author declares that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Acknowledgments

We thank Dr. Klaus Contzen of BIOTRINIK SE & Co. KG for his support.

References

1. Barold SS, Kucher A, Nägele H, Buenfil Medina JC, Brodsky M, Van Heuverswyn FE, et al. Dissimilar ventricular rhythms: implications for ICD therapy. Heart Rhythm. (2013) 10:510–6. doi: 10.1016/j.hrthm.2012.12.004

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Handa BS, Lawal S, Wright IJ, Li X, Cabello-García J, Mansfield C, et al. Interventricular differences in action potential duration restitution contribute to dissimilar ventricular rhythms in ex vivo perfused hearts. Front Cardiovasc Med. (2019) 6:34. doi: 10.3389/fcvm.2019.00034

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Kubickova M, Cervinka P, Kucher A, Nastoupil Z, Stroobandt RX. Undetected ventricular fibrillation in a single-chamber implantable cardioverter-defibrillator: when the far-field channel sees more than the intraventricular channel. Heart Rhythm Case Rep. (2016) 2:321–3. doi: 10.1016/j.hrcr.2016.03.009

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Kucher A, Barold SS. Rate disparity of near-field versus far-field ICD electrograms: a clue to the diagnosis of dissimilar ventricular rhythms. Pacing Clin Electrophysiol. (2017) 40:1047–9. doi: 10.1111/pace.13142

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: implantable cardioverter-defibrillator, cardiac resynchronization therapy (CRT), ventricular tachycardia, ventricular fibrillation, dissimilar ventricular rhythms, tachycardia detection, CRT-D, cardiac resynchronization therapy–defibrillator

Citation: Kucher A and Stroobandt RX (2019) Commentary: Interventricular Differences in Action Potential Duration Restitution Contribute to Dissimilar Ventricular Rhythms in ex vivo Perfused Hearts. Front. Cardiovasc. Med. 6:58. doi: 10.3389/fcvm.2019.00058

Received: 12 April 2019; Accepted: 23 April 2019;
Published: 13 May 2019.

Edited by:

Marina Cerrone, New York University, United States

Reviewed by:

Artur Fedorowski, Lund University, Sweden

Copyright © 2019 Kucher and Stroobandt. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Andreas Kucher, andreas.kucher@biotronik.com

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.